K103730 · Ge Medical Systems Ultrasound and Primary Care Dia · KGI · May 6, 2011 · Radiology
Device Facts
Record ID
K103730
Device Name
GE LUNAR VISCERAL FAT SOFTWARE
Applicant
Ge Medical Systems Ultrasound and Primary Care Dia
Product Code
KGI · Radiology
Decision Date
May 6, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1170
Device Class
Class 2
Indications for Use
The GE Lunar Visceral Fat Software option used on GE Lunar DEXA bone densitometer total body scans estimates the visceral adipose tissue (visceral fat) content within the android region in a male or female population between the age of 18 and 90 with a BMI between 18.5 and 40, excluding pregnant women. The content that is estimated is the Visceral Fat Mass and Visceral Fat Volume. The values can be displayed in user defined statistical formats and trends. The estimated visceral fat content is useful to health care professionals in their management of diseases/conditions where the disease/condition itself, or its treatment, can affect the relative amounts of visceral fat content in the android region. The GE Lunar Visceral Fat Body Composition Software option does not diagnose disease, or recommend treatment regimens, or quantify treatment effectiveness. Only the health care professional can make these judgments. Some of the diseases/conditions for which visceral fat estimation is useful include hypertension, impaired fasting glucose, impaired glucose tolerance, diabetes mellitus, dyslipidemia, and metabolic syndrome.
Device Story
Software option for GE Lunar DXA bone densitometers; utilizes total body DXA scan data to estimate visceral adipose tissue (VAT) mass and volume in the android region. Operates by measuring attenuation of dual-energy x-ray beams; combines direct fat/lean tissue measurements with modeling of fat distribution. Used in clinical settings by healthcare professionals to monitor VAT changes related to disease management (e.g., diabetes, metabolic syndrome). Output displayed as statistical formats/trends; assists clinicians in assessing patient status; does not diagnose, recommend treatment, or quantify treatment efficacy.
Clinical Evidence
Clinical trial performed comparing DXA-estimated visceral adipose tissue (VAT) against CT imaging (Advantage Windows Tissue Volume option, K963345) as the research standard. Bench testing also conducted to verify electrical, mechanical, and radiation safety per 21 CFR 820 and CDRH requirements.
Technological Characteristics
Software module for DXA bone densitometer; utilizes dual-energy x-ray attenuation principle. Operates on existing GE Lunar DXA hardware. Complies with 21 CFR 820 (Quality System Regulations) and CDRH 21 CFR subchapter J radiation safety standards.
Indications for Use
Indicated for male or female patients aged 18-90 with BMI 18.5-40 to estimate visceral adipose tissue (mass and volume) in the android region. Useful for managing conditions like hypertension, impaired fasting glucose, impaired glucose tolerance, diabetes mellitus, dyslipidemia, and metabolic syndrome. Contraindicated for pregnant women.
Regulatory Classification
Identification
A bone densitometer is a device intended for medical purposes to measure bone density and mineral content by x-ray or gamma ray transmission measurements through the bone and adjacent tissues. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
GE Lunar Body Composition Software option (K071570)
{0}------------------------------------------------
K103730 Page 1 of 3
GE Medical Systems Lunar 510(k) Premarket Notification GE Lunar Visceral Fat Software
## 510(K) Summary of Safety and Effectiveness
MAY - 6 2011
Prepared in accordance with 21 CFR Part 807.92.
| Date Prepared: | December 20, 2010 |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare, (GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC doing business as GE Medical Systems Lunar)<br>3030 Ohmeda Drive<br>Madison, WI 53718 |
| Primary Contact: | Chris Paulik<br>Regulatory Affairs Leader, GE Medical Systems Lunar<br>Telephone: 262-548-2010; Fax: 262-546-0704<br>e-mail: Christopher.Paulik@med.ge.com |
| Secondary Contact: | David Blonski<br>Regulatory Affairs Director, X-Ray<br>GE Healthcare, (GE Medical Systems, LLC)<br>Telephone: 262-513-4072; Fax: 262-546-0704<br>e-mail: David.Blonski@ge.com |
### DEVICE INDENTIFICATION
| Trade Name: | GE Lunar Visceral Fat Software |
|----------------------|--------------------------------|
| Common/Usual Name: | Bone Densitometer |
| Classification Name: | Bone Densitometer |
| | Class II, KGI, 21CFR 892.1170 |
### DEVICE DESCRIPTION:
The Visceral Fat Software option is part of a GE Lunar DXA bone densitometer. It uses data from a DXA total body scan to estimate the visceral adipose tissue (visceral fat), mass and volume within the android region in a male or female population between the age of 18 and 90 with a BMI between 18.5 and 40.
ન્ડન્ટ
Page 14-04-03
{1}------------------------------------------------
K103730 Page 2 of 3
Image /page/1/Picture/1 description: The image shows the General Electric (GE) logo, which consists of the letters 'G' and 'E' intertwined within a circular border. The logo is in black and white, with the letters and border appearing in black against a white background. The design is simple and iconic, representing the well-known multinational conglomerate.
GE Medical Systems Lunar 510(k) Premarket Notification GE Lunar Visceral Fat Software
#### INDICATIONS FOR USE:
The GE Lunar Visceral Fat Software option used on GE Lunar DEXA bone densitometer total body scans estimates the visceral adipose tissue (visceral fat) content within the android region in a male or female population between the age of 18 and 90 with a BMI between 18.5 and 40, excluding pregnant women. The content that is estimated is the Visceral Fat Mass and Visceral Fat Volume. The values can be displayed in user defined statistical formats and trends.
The estimated visceral fat content is useful to health care professionals in their management of diseases/conditions where the disease/condition itself, or its treatment, can affect the relative amounts of visceral fat content in the android region. The GE Lunar Visceral Fat Body Composition Software option does not diagnose disease, or recommend treatment regimens, or quantify treatment effectiveness. Only the health care professional can make these judgments. Some of the diseases/conditions for which visceral fat estimation is useful include hypertension, impaired fasting glucose, impaired glucose tolerance, diabetes mellitus, dyslipidemia, and metabolic syndrome.
#### TECHNOLOGY
GE Lunar Bone Densitometers use a technique called Dual-energy X-ray Absorptionetry or DXA. DXA measures the attenuation of an x-ray beam comprised of two energy levels after passing the beam through the body of a subject. When the dual-energy x-ray beam passes through soft tissue, without any bone mineral, the tissue composition for the pixel can be measured by calculating the relative attenuation by the fat and lean issue of the two energy levels in the x-ray beam.
The Visceral Fat software algorithm combines direct measurements of fat and lean tissue with modeling information about the distribution of fat in the android region.
A clinical trial was performed to compare dual-energy x-ray absorptiometry (DXA) estimated visceral fat adipose tissue (VAT) with the accepted research standard for VAT imaging, computed tomography (CT), using the Advantage Windows Tissue Volume option (K963345).
### COMPARISON WITH PREDICATE DEVICES:
The GE Lunar Visceral Fat Software option for GE Lunar DEXA Bone Densitometers is of comparable type and substantially equivalent to the GE Lunar Body Composition Software option (K071570). It has the same technological characteristics, is comparable in key safety and effectiveness features, it utilizes similar design, construction and materials, and has similar intended uses as the GE Lunar Body Composition Software option.
Image /page/1/Picture/12 description: The image shows the numbers 5 and 3 connected by a horizontal line. The number 5 is on the left and the number 3 is on the right. The numbers are written in a simple, sans-serif font. The image is black and white.
Page 12 or 95 -
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circle. The logo is in black and white, with the letters and the outline of the circle in black and the background in white.
GE Medical Systems Lunar 510(k) Premarket Notification GE Lunar Visceral Fat Software
### ADVERSE EFFECTS ON HEALTH:
The device has been evaluated for electrical, mechanical, and radiation safety, and conforms to applicable medical device safety standards, as confirmed by a Nationally Recognized Test Laboratory,
The potential hazards of electrical and mechanical are identified in a risk management summary (hazard analysis) and controlled by:
- System verification and validation to ensure performance to specifications, ● Federal Regulations, and user requirements.
- Adherence and certification to industry and international standards. .
- Compliance to applicable CDRH 21 CFR subchapter J requirements. .
The device is designed and manufactured under the Quality System Regulations of 21 CFR 820.
#### CONCLUSION:
The GE Lunar Visceral Fat Software is an extension of the currently cleared GE Lunar Body Composition Software (K071570). It does not result in any new potential safety risks, has the same technological characteristics, and performs as well as the devices currently on the market.
After analyzing performance testing on the bench and in a clinical trial it is the conclusion of GE Healthcare that the GE Lunar Visceral Fat Software is substantially equivalent to other marketed devices with similar indications for use and meeting the same standards.
5-4
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, with flowing lines above them, creating a sense of movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
MAY 1 0 2011
Mr. Chris Paulik Regulatory Affairs Leader GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC Doing business as GE Medical Systems Lunar 3030 Ohmeda Drive MADISON WI 53718
Re: K103730
Trade/Device Name: GE Lunar Visceral Fat Software Regulation Number: 21 CFR 892.1170 Regulation Name: Bone densitometer Regulatory Class: II Product Code: KGI Dated: December 20, 2010 Received: December 21, 2010
Dear Mr. Paulik:
This letter corrects our substantially equivalent letter of May 6, 2011.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Mr. Paulik
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucr1115800.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Production 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely Yours,
Mary Patel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with the letters 'GE' intertwined within a circular border. The letters are stylized and appear to be hand-drawn or calligraphic. The logo is black and white, with the letters and border in black and the background in white.
SION -510(k) Premarket Notification GE Lunar Viscoral Fat Software
# Indications for Use
510(k) Number (if Known): Device Name: GE Lunar Visceral Fat Software
Indications for Use:
The GB Lunar Visceral Pat Software option used on GE Lunar DEXA bone densitometer
s to al body scans estimates the visceral adjnose tissue (risessed friences of the lotal body scans estimates the visceral adjose tissue (visceral fat) content within the and of the on outhers are visers adjose the age of 18 and 90thin was.
between 18.5 and 40, cxcluding between the age of 18 and 90with a BMI between 18.5 and 40, excluding pregnant women. The content that is estimated is the election is estimated is the Viscently and Visceral Fat Volume. The volume. The content that is estimated is their
statistical formats and trends, statistical formats and trends.
The estimated visceral fat content is useful to health care professionals in their management of discases/conditions where the disease/oondition itself, or its treatment, oan affect the relative amounts of visce)al at content in the android region. The Greatment,
Visceral Fat Body Composition Software ontion dess not discussion. The GI Lunar Visceial Pat Body Composition Software option does not diagnose disease, CD-Library, SB-L
Visceial Pat Body Composition Software option does not diagnose disease, or recommon cannent regimens, or quantify treatment effectiveness. Only the health ears professional can make these judgments. Some of the diseases/conditions for which visceral fat estimation is useful include bypertension, impaired fasting glucose, impaired glucose, impaired glucose tolerance, diabetes mellitus, dyperiension, imparicd fasting glucose,
tolerance, diabetes mellitus, dyslipidemia, and metabolic syndrome.
Prescription Use_ X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (Part 21 CPR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
5/6/11
Division Sign-Off
Office of In Vitro Diagnostic Device Bvaluation and Safety 510(k) K103
Page 1 of 1
March 29, 2011
. . . . . .
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.