K103719 · Everyway Medical Instruments Co.,Ltd · GXY · Aug 24, 2011 · Neurology
Device Facts
Record ID
K103719
Device Name
LIFECARE CONDUCTIVE GARMENTS
Applicant
Everyway Medical Instruments Co.,Ltd
Product Code
GXY · Neurology
Decision Date
Aug 24, 2011
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 882.1320
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Lifecare HC-88 Series Conductive Garments are used in conjunction with electrical stimulators both in the home and physical therapy setting, and can stimulate large or multiple. They are knitted in different shapes, as gloves, socks, kneecaps and wristlets/elbow guards, and in different sizes.
Device Story
Lifecare HC-88 Series Conductive Garments are single-patient reusable, non-sterile, flexible fabric electrodes; interface between electrical stimulators (TENS, powered muscle stimulators, galvanic devices) and patient skin. Constructed from metal snap terminals and cloth knitted with Spandex and stainless-steel/nylon fibers. Used in home or physical therapy settings; operated by patients or clinicians. Device transmits electrical signals from stimulator through conductive garment to skin surface. Available in various shapes (gloves, socks, kneecaps, wristlets/elbow guards). Not MRI compatible. Benefits include delivery of electrical stimulation to large or multiple body areas simultaneously.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (skin irritation and sensitization). Impedance testing conducted using resistance meters and oscilloscopes; results showed impedance <5 ohms/cm and total impedance <50 ohms, maintained after 20 wash cycles.
Indicated for use as cutaneous electrodes with legally marketed electrical stimulators (e.g., powered muscle stimulators, TENS) to deliver stimulation signals to body surfaces including hands, feet, wrists, elbows, and knees.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
{0}------------------------------------------------
### 510(k) Summary As Required by 21 section 807.92 ( c )
Everyway Medical Instruments Co., Ltd. 1-Submitter Name:
2-Address:
3 FL., No.5, LANE 155, Sec. 3, Peishen Rd Shen Keng Hsiang. 222 Taipei Hsien. Taiwan (ROC)
3-Phone: (886) 2 2662 0038
4-Fax: (886) 2 2664 5566
5-Contact Person: Mr. Robert Tu (General Manager)
6-Date summary prepared: July 15", 2011
7 - Official Correspondent: Mansour Consulting LLC
8- Address: 845 Aronson Lake Court. Roswell, GA 30075 USA
9- Phone: 678-908-8180
10- Fax: 678-623-3765
11- Contact Person: Jav Mansour, President
12-Device Trade or Proprietary Name: Lifecare HC-88 Series Conductive Garments
13-Device Common or usual name: Lifecare HC-88 Series Conductive Garments
14-Device Classification Name: Cutaneous electrode
15-Substantial Equivalency is claimed against the following device:
TheraKnit Garment Cutaneous Electrodes from Neurotron Medical, Inc. 510k# K053214 16-Description of the Device:
The single-patient reusable and non-sterile Lifecare HC-88 Series Conductive Garments are cutaneous. flexible, conductive garment/fabric electrodes.
They are passive devices acting as interface between electrical stimulators (such as powered muscle stimulators, interferential devices, galvanic device and transcutaneous electrical nerve stimulators) and a patient's skin. The signal is transmitted from the conductive knob through the cloth of conductive garment to the surface of the patient's skin.
The Lifecare HC-88 Series Conductive Garments are used in conjunction with electrical stimulators both in the home and physical therapy setting, and can stimulate large or multiple. They are knitted in different shapes, as gloves, socks, kneecaps and wristlets/elbow guards, and in different sizes.
The Lifecare HC-88 Series Conductive Garments are constructed with four basic components (materials): a metal conductive snap terminal; cloth knitted together with Spandex and Stainless-steel yarns that twisted and composed by Stainless-steel and Nylon fibers. The garments are not packaged with lead wire.
Impedance testing evaluated by the "Resistance meter of electronic multimeter" to measure the impedances of all conductive garments are below 5 ohms between each centimeter and evaluated by the "oscilloscope measuring circuit" to measure the total impedance of all conductive garments are below 50 ohms" . Even after over 20 times washing, the impedance of Lifecare HC-88 Series Conductive Garments are still within specification.
The garments are not compatible with MRI spectrometry.
#### 17-Intended use of the device: (refer to FDA form attached)
The Lifecare HC-88 Series Conductive Garments are used in conjunction with electrical stimulators both in the home and physical therapy setting, and can stimulate large or multiple. They are knitted in different shapes, as gloves, socks, kneecaps and wristlets/elbow guards, and in different sizes.
## 18-Safety and Effectiveness of the device:
This device is safe and effective as the predicate devices cited above as detailed within this submission. This type of device has been marketed and used in US for several years as Prescription use and proven to be safe and effective without side effects or complications generated from the use of the device itself.
#### 19-Biocompatibility of material:
The material of Lifecare HC-88 Series Conductive Garments were evaluated and passed the evaluation of biocompatibility testing by SGS Taiwan Ltd.
- Cytotoxicity test (ISO 10993-5, SGS report No.UB/2010/50533), -
- । Skin irritation test (ISO 10993-10, SGS report No. UB/2010/50533A-01) and
- Skin sensitization test (ISO 10993-10, SGS report No. UB/2010/50533A-02)
AUG 2 4 2011
{1}------------------------------------------------
# 20-Approximate surface area of each garment model:
| REF.NO | SIZE | SURFACE AREA | Type |
|-----------|--------------|--------------|----------------|
| HC-884A | M(L21cm) | 450cm² | Glove |
| HC-885A | L(L24cm) | 580cm² | Glove |
| HC-886A | S(L18cm) | 330cm² | Glove |
| REF.NO | SIZE | SURFACE AREA | Type |
| HC-883 | (L33xW9cm) | 600cm² | Sock |
| REF.NO | SIZE | SURFACE AREA | Type |
| HC-881A | M(L18xW11cm) | 400cm² | Wristlet/Elbow |
| HC-881-1A | S(L18xW10cm) | 360cm² | Wristlet/Elbow |
| REF.NO | SIZE | SURFACE AREA | Type |
| HC-882A | M(L27xW14cm) | 760cm² | Kneecaps |
719
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with three stripes forming its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Everyway Medical Instruments Co., Ltd. c/o Mr. Jay Mansour Mansour Consulting, LLC 845 Aronson Lake Court Roswell, GA 30075
AUG 2 4. 2011
Re: K103719
Trade/Device Name: Lifecare HC 88-Series Conductive Garments Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous electrode Regulatory Class: Class II Product Code: GXY Dated: August 1, 2011 Received: August 5, 2011
Dear Mr. Mạnsour:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section 910(x) promation is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosule) to tegally manoves personal perice Ameridae, or to commerce prof to way 20, 1970, the classified in accordance with the provisions of the Federal Food, Drug, devices that have been recursed in accordance was a premarket approval approval application (PMA). and Cosment Act (Act) that do not roques approvinces in controls provisions of the Act. The
You may, therefore, market the device, subject to the general controls provision You may, merelote, market the device, saljers to the gents for annual registration, listing of the general Controls provisions of the vice is a prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability adulteration. Please note. CDITI docs how verse labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or classica on be II your device is classifica (see above) Existing major regulations affecting your device can be may be subject to additional controlly "Linziting Collections, Title 21, Parts 800 to 898. In addition, FDA may found in the Code of I occar 105g ming your device in the Federal Register,
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act of the Act one of our must oomn that FDA has made a determination that Jour cered by other Federal agencies. You must comply of any Pederal statutes and regulations administered of the registration and listing (2) CFR Patt With an the Act s requiremons; areasing, a with the reporting of medical device-related
{3}------------------------------------------------
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the adverse events) (21 CFR 805); good manufacture production in electronic product
quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product quality systems (QS) regulation (21 CFR 100 SE6), 21 CFR 1000-1050.
radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
r (1580), please and 15 (1) (15) (1)PRIVOR (CDRUCCDRUCC) If you desire specific advice for your device of our laborary copanese (
go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucml (5809.html for go to http://www.tda.gov/AboutFDACEmers/crices/Office of Compliance. Also, please
the Center for Devices and Radiological Health's (CDRH) s (CDRH) stopperket notification" ( the Center for Devices and Radiological ricallis systemarks notification" (21CFR Patt
note the regulation entitled, "Misbranding by reference sunder the MDR regulation (2 note the regulation entitled, "Misoraliung of relected to promation
807.97). For questions regarding the reporting of adverse events under the MDR regulation (21
CFR Part 803), please go to CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.html for the CDRH's Office ntlp://www.ida.gov.nice.etrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the You may obtain other general inormational and Consumer Assistance at its toll-free number
Division of Small Manufacturers, International and Consumer Assistance at its toll-f (800) 638-2041 or (301) 796-7100 or at its Internet address (800) 638-2041 of (501) 790-7100 of at its career)
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Alexander
Image /page/3/Picture/7 description: The image shows the handwritten word "for" in cursive. The "f" has a large loop extending below the baseline, and the "o" and "r" are connected with a smooth curve. The writing style is fluid and elegant, with a slight slant to the right.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Lifecare HC-88 Series Conductive Garments
Indications For Use:
The Lifecare HC-88 Series Conductive Garments are cutaneous electrodes to be used with legally marketed electrical stimulators (such as powered muscle stimulators and transcutaneous electrical nerve stimulators).
The knitted garment electrodes will deliver the stimulation signals generated by the stimulator to the body surface with which they are in contact.
These body parts include hands (gloves), feet (socks), wrists / elbows (wristlets / elbow guards) and knees (kneecaps).
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
B
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
Page 1 of
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.