← Product Code [LZA](/productcode/LZA) · K103675

# CHIENMAX VIETNAM CO. LTD POWDER FREE BLUE NITRILE EXAMINATION GLOVE (K103675)

_Chienmax Vietnam Co., Ltd. · LZA · Oct 3, 2011 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K103675

## Device Facts

- **Applicant:** Chienmax Vietnam Co., Ltd.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Oct 3, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiners hand or finger to prevent contamination between patient and examiner.

## Device Story

Disposable blue nitrile synthetic rubber patient examination gloves; worn on examiner's hand or finger; intended to prevent cross-contamination between patient and examiner in medical settings. Device meets ASTM D6319-00a (2005) standards for physical dimensions and strength. Provides barrier protection; non-sterile; powder-free. Used by healthcare professionals during patient examinations.

## Clinical Evidence

Bench testing only. Testing included dermal sensitization, primary skin irritation, permeation testing (ASTM D6978-05), tensile strength, and barrier strength (pinholes). All tests met ASTM D6319-00a requirements.

## Technological Characteristics

Material: Nitrile synthetic rubber. Form factor: Disposable, powder-free, blue, patient examination glove. Standards: ASTM D6319-00a (2005). Testing: Inspection level S-2, AQL 4.0, pinholes at AQL 2.5. Non-sterile.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Sunmax Vietnam Co. Ltd Powder free Nitrile Patient Examination Gloves ([K101870](/device/K101870.md))
- ULTRAWIN SDN BHD Non-Sterile Powder Free Nitrile Examination Gloves ([K090828](/device/K090828.md))
- PT. MAHAKARYA INTI BUANA Powder Free Black Nitrile Examination gloves ([K090464](/device/K090464.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# CHIENMAX VIETNAM CO.,LTD

Image /page/0/Picture/1 description: The image shows a logo for Maslac. The logo features a stylized hand with three fingers and a thumb forming a circular shape. Below the hand is the word "Maslac" in a bold, sans-serif font. The "R" in a circle indicates that the logo is a registered trademark.

K103675

KM 8, PHAM VAN DONG ROAD, HAI THANHDOUNG KINH DISTRICT HAI PHONG, HAI PHONG, TAHNH PHO, 18671, VIETNAM

OCT - 3 2011

TEL:84-313632888 FAX: 84-31336288

EXHIBIT #3

# 510(k) Summary

This summary of 5l0(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

#### Date: August 25, 2011

#### 1. Applicant:

Chienmax Vietnam Co. Ltd Km 8, Pham Van Dong Road, Hai Thanhdoung Kinh Districthai Phong, Hai Phong, Tahnh Pho, 18671, Vietnam

#### 2. Manufacturer: Chienmax Vietnam Co. Ltd Km 8, Pham Van Dong Road,

Hai Thanhdoung Kinh Districthai Phong, Hai Phong, Tahnh Pho, 18671, Vietnam

#### 3. Submitter:

Mr. Jigar Shah Official Correspondent for Chienmax Vietnam Co. Ltd

#### 4. Address:

mdi Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, New York 11021 Tel: 516-482-9001 Fax: 516-482-0186 Jigar@mdiconsultants.com

- 5. Trade/proprietary Name: Chienmax Vietnam Co. Ltd Powder-free Blue Nitrile Patient Examination Gloves.
- Common Names: 6. POWDER-FREE Patient Examination Glove
- 7. Classification name: Device Class I, Patient examination glove (21 CFR 880.6250, product code LZA)

#### 8. Predicate Devices:

- Sunmax Vietnam Co. Ltd Powder free Nitrile Patient Examination Gloves. . (K101870)

{1}------------------------------------------------

- ULTRAWIN SDN BHD Non-Sterile Powder Free Nitrile Examination Gloves (K . 090828)
- PT. MAHAKARYA INTI BUANA Powder Free Black Nitrile Examination gloves. . (K090464)

#### 9. Device Description:

Chienmax Vietnam Co. Ltd Powder Free Blue Nitrile Patient Examination Gloves are Class I disposable device which are made up of nitrile synthetic rubber , intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner and which meets all of the requirements of ASTM standard D6319 00a (2005).

#### 10. Intended Use:

A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiners hand or finger to prevent contamination between patient and examiner.

#### 11. Substantial Equivalence Discussion:

A powder free patient examination glove is substantially equivalent to the predicate devices.

| Characteristic<br>and parameters | Chienmax<br>Vietnam Co.,<br>LTD Powder<br>Free Blue Nitrile<br>Examination<br>Glove<br>(New Device)                                                                                                                        | Sunmax Vietnam<br>Co. Ltd Powder<br>free Nitrile Patient<br>Examination<br>Gloves. (K101870)                                                                                                                            | ULTRAWIN<br>SDN BHD<br>Non-Sterile<br>Powder Free<br>Nitrile<br>Examination<br>Gloves<br>(K 090828)                                                                                                                        | PT.<br>MAHAKARY<br>A INTI<br>BUANA<br>Powder Free<br>Black Nitrile<br>Examination<br>gloves.<br>(K090464)                                                                                                                                   | SE<br>Comparison    |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Product Code                     | LZA                                                                                                                                                                                                                        | LZA                                                                                                                                                                                                                     | LZA                                                                                                                                                                                                                        | LZA                                                                                                                                                                                                                                         |                     |
| Intended Use                     | A patient<br>examination<br>gloves is a<br>disposable device<br>intended for<br>medical purpose<br>that is worn<br>on the examiner's<br>hand or fingers to<br>prevent<br>contamination<br>between patient<br>and examiner. | A patient<br>examination gloves<br>is a disposable<br>device intended for<br>medical purpose<br>that is worn<br>on the examiner's<br>hand or fingers to<br>prevent<br>contamination<br>between patient and<br>examiner. | A patient<br>examination<br>gloves is a<br>disposable device<br>intended for<br>medical purpose<br>that is worn on<br>the examiner's<br>hand or fingers to<br>prevent<br>contamination<br>between patient<br>and examiner. | A powder-free<br>patient<br>examination<br>glove is a<br>disposable<br>device intended<br>for medical<br>purposes that is<br>worn on the<br>examiner's hand<br>or finger to<br>prevent<br>contamination<br>between patient<br>and examiner. | SE                  |
| Width (size<br>medium)           | 95                                                                                                                                                                                                                         | 89                                                                                                                                                                                                                      | 93-98                                                                                                                                                                                                                      | 97.7                                                                                                                                                                                                                                        | Minor<br>Difference |

2

{2}------------------------------------------------

| verall length                           | 244       | 240       | 240       | 240.9   | Minor<br>Difference |
|-----------------------------------------|-----------|-----------|-----------|---------|---------------------|
| Palm thickness                          | 0.113     | 0.12      | Min 0.08  | 0.109   | Minor<br>Difference |
| Finger thickness                        | 0.138     | 0.12      | Min 0.08  | 0.148   | Minor<br>Difference |
| Tensile strength<br>pre aging min       | 16.78     | 22        | 15 - 21   | 18.8    | Minor<br>Difference |
| Tensile strength<br>after aging min     | 17.40     | 23.6      | 14-22     | 21.3    | Minor<br>Difference |
| Ultimate<br>elongation pre<br>aging min | 510       | 500       | 550 - 630 | 679.4   | Minor<br>Difference |
| Ultimate<br>elongation after<br>aging   | 555       | 500       | 520 - 610 | 767.4   | Minor<br>Difference |
| Meets<br>Biocompatibility               | yes       | Yes       | Yes       | Yes     | SE                  |
| Duration of bio-<br>compatibility       | Limited   | Limited   | Limited   | Limited | SE                  |
| Skin irritation test                    | Passes    | Passes    | Passes    | Passes  | SE                  |
| Dermal<br>ensitization                  | Passes    | Passes    | Passes    | Passes  | SE                  |
| Residual powder<br>test                 | Passes    | Passes    | Passes    | Passes  | SE                  |
| Labeling                                | Identical | Identical | NA        | NA      | Minor<br>Difference |

#### 12. Summary of Testing:

|    | Test                                  | Results                                              |
|----|---------------------------------------|------------------------------------------------------|
| a. | Dermal Sensitization Test             | Passes                                               |
| b. | Primary Skin irritation               | Passes                                               |
| c. | Permeation testing per ASTM D 6978-05 | Passes                                               |
| d. | Iodine Test                           | Passes                                               |
| e. | Tensile strength                      | Gloves meets the requirements of<br>ASTM D63 19-00a. |
| f. | Barrier strength                      | Gloves meets the requirements of<br>ASTM D63 19-00a. |

The standards used by Chienmax Vietnam Co. Ltd to determine substantial equivalence are based on ASTM D 631900a-2005. All testing meets requirements for physical specifications and dimensions conducted on gloves, Inspection level S-2, AQL 4.0, pinholes at AQL 2.5

There are no special labeling claims and we do not claim our gloves to be hypoallergenic.

{3}------------------------------------------------

## 13. Conclusion:

Based on the nonclinical tests performed the Chienmax Vietnam Co. Ltd Powder-free Blue Nitrile Examination Glove is as safe, as effective and performs at least as safely and effectively as the legally marketed predicate devices.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an eagle or bird in flight, with three curved lines representing the wings and a wavy line representing the body.

Food and Drug Administration 10903 New Hamoshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Chienmax Vietnam Company, Limited C/O Mr. Jigar Shah Official Correspondent MDI Consultants, Incorporated 55 Northern Boulevard, Suite 200 Great Neck, New York 11021

OCT - 3 2011

Re: K103675

Trade/Device Name: Chienmax Vietnam Company, Limited Powder Free Blue Nitrile Patient Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: September 20, 2011 Received: September 22, 2011

Dear Mr. Shah:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the enners. controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRF; does not evaluate information related to contract liability warranties. We remind you, fovever, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class iII (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 89%. In this addition, FDA may publish further announcements concerning your device in the Fight first Register.

{5}------------------------------------------------

Page 2 - Mr. Shah

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH 's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its offelire number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Cintron
Anthony D. Weiner, P.C. 160 Broadway

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

### Enclosure ·

{6}------------------------------------------------

Attachment 1

#### EXHIBIT #1

#### Indications for Use

#### 510(k) Number (if known): K103675

#### Applicant: Chienmax Vietnam Co. Ltd

Device Name: Chienmax Vietnam Co. Ltd Powder Free Blue Nitrile Patient Examination Gloves.

#### Indications for Use:

A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiners hand or finger to prevent contamination between patient and examiner.

# Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use X (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

AND/OR

Shula R. Murphy, MD, John F. Chrocien-Williams
(Division Sign-Off)

Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K:103675

---

**Source:** [https://fda-staging.innolitics.com/device/K103675](https://fda-staging.innolitics.com/device/K103675)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).