← Product Code [MHI](/productcode/MHI) · K103673

# GIARDIA/ CRYPTOSPORIDIUM QUIK CHEK (K103673)

_Techlab Inc., Corporate Research Center · MHI · Aug 18, 2011 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K103673

## Device Facts

- **Applicant:** Techlab Inc., Corporate Research Center
- **Product Code:** [MHI](/productcode/MHI.md)
- **Decision Date:** Aug 18, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3220
- **Device Class:** Class 2
- **Review Panel:** Microbiology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K103673 · Aug 18, 2011 | GIARDIA/ CRYPTOSPORIDIUM QUIK CHEK | Techlab Inc., Corporate Research Center | Retrospective human fecal specimens (fresh, frozen, and preserved) | Clinical performance evaluation comparing the device to Microscopy (IFA) and commercial ELISA predicate devices using retrospective patient fecal specimens. | Clinical performance; Retrospective specimens; Diagnostic accuracy |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Clinical Performance Evaluation; Retrospective clinical performance study | Patients with gastrointestinal symptoms; Sample Size: 791 (Microscopy comparison); 849 (ELISA comparison); Number of Sites: 3 | Microscopy (IFA) and commercially available ELISAs | Sensitivity, specificity, and correlation/agreement |

## Indications for Use

The GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test is a rapid membrane enzyme immunoassay for the simultaneous qualitative detection and differentiation of Giardia cyst antigen and Cryptosporidium oocyst antigen in a single test device. It is intended for use with human fecal specimens from patients with gastrointestinal symptoms to aid in the diagnosis of Giardia and/or Cryptosporidium gastrointestinal infection. The test results should be considered in conjunction with the patient history.

## Device Story

Rapid membrane enzyme immunoassay for simultaneous qualitative detection/differentiation of Giardia cyst antigen and Cryptosporidium oocyst antigen in human fecal specimens. Device utilizes monoclonal and polyclonal antibodies immobilized on a membrane within a test cassette. Sample mixed with diluent and conjugate (polyclonal antibodies coupled to horseradish peroxidase) is added to the device; antigen-antibody-conjugate complexes migrate to test lines where they are captured. After washing and substrate addition, results are read visually by the presence of blue lines. Used in clinical laboratory settings by trained personnel. Provides rapid diagnostic aid for gastrointestinal infection; results interpreted alongside patient history to guide clinical management.

## Clinical Evidence

Clinical performance evaluated at three sites (N=791) comparing to reference microscopy/IFA. Giardia sensitivity 98.9%, specificity 100%. Cryptosporidium sensitivity 100%, specificity 99.8%. Method comparison (N=849) against predicate ELISA showed 99.5% overall agreement for both analytes. Analytical sensitivity: 6000 cysts/oocysts per mL. No cross-reactivity observed with 40+ common enteric pathogens/parasites. Interference testing showed no effect from common medications or fecal components.

## Technological Characteristics

Rapid membrane enzyme immunoassay. Employs mouse monoclonal antibodies against Giardia and Cryptosporidium, and anti-HRP antibodies for control. Conjugate consists of polyclonal antibodies coupled to horseradish peroxidase. Visual readout via blue line appearance. Manual test procedure with 25-30 minute incubation. No electronic components or software.

## Regulatory Identification

Entamoeba histolytica serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Entamoeba histolytica in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identify Entamoeba histolytica directly from clinical specimens. The identification aids in the diagnosis of amebiasis caused by the microscopic protozoan parasite Entamoeba histolytica and provides epidemiological information on diseases caused by this parasite. The parasite may invade the skin, liver, intestines, lungs, and diaphragm, causing disease conditions such as indolent ulcers, an amebic hepatitis, amebic dysentery, and pulmonary lesions.

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.

## Predicate Devices

- Merifluor™ Cryptosporidium / Giardia Kit ([K912408](/device/K912408.md))
- GIARDIA II ([K033274](/device/K033274.md))
- CRYPTOSPORIDIUM II ([K052932](/device/K052932.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
K103673

B. Purpose for Submission:
To obtain a substantial equivalence determination for the detection of Giardia and Cryptosporidium antigens in human stool

C. Measurand:
Giardia and Cryptosporidium antigens

D. Type of Test:
Qualitative membrane enzyme immunoassay

E. Applicant:
TechLab® Inc.

F. Proprietary and Established Names:
The GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test

G. Regulatory Information:
1. Regulation section:
21 CFR Part 866.3220 Entamoeba histolytica Serological Reagents
2. Classification:
II
3. Product code:
MHJ - Cryptosporidium spp.
MHI - Giardia spp.

{1}

4. Panel:
83, Microbiology

H. Intended Use:

1. Intended use(s):

The GIARDIA/CRYPTOSPORIDIUM QUIK CHEK test is a rapid membrane enzyme immunoassay for the simultaneous qualitative detection and differentiation of Giardia cyst antigen and Cryptosporidium oocyst antigen in a single test device. It is intended for use with human fecal specimens from patients with gastrointestinal symptoms to aid in the diagnosis of Giardia and/or Cryptosporidium gastrointestinal infection. As with other Giardia and/or Cryptosporidium tests, results should be considered in conjunction with the patient history.

FOR IN VITRO DIAGNOSTIC USE.

2. Indication(s) for use:

The GIARDIA/CRYPTOSPORIDIUM QUIK CHEK test is a rapid membrane enzyme immunoassay for the simultaneous qualitative detection and differentiation of Giardia cyst antigen and Cryptosporidium oocyst antigen in a single test device. It is intended for use with human fecal specimens from patients with gastrointestinal symptoms to aid in the diagnosis of Giardia and/or Cryptosporidium gastrointestinal infection. As with other Giardia and/or Cryptosporidium tests, results should be considered in conjunction with the patient history.

3. Special conditions for use statement(s):

For prescription use only

4. Special instrument requirements:

None

I. Device Description:

The GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test device contains 25 pouches, each containing a membrane filter pad. Each device has a Reaction Window with three vertical lines of immobilized antibodies. The Giardia test line ("Giar") contains mouse monoclonal antibodies against Giardia. The Crypto test line ("Cryp") contains mouse monoclonal antibodies against Cryptosporidium. The control line ("C") is a dotted line that contains anti-horseradish peroxidase (HRP) antibodies. The

2

{2}

Conjugate consists of polyclonal antibodies coupled to horseradish peroxidase. The antigen-antibody-conjugate complexes migrate through a filter pad to a membrane where they are captured by the immobilized Giardia and/or Cryptosporidium-specific antibodies in the test lines. After addition of substrate and proper incubation, the reaction is examined visually for the appearance of a vertical blue line on either side of the Reaction Window. A blue line indicates a positive test. A positive "control" reaction, indicated by a vertical dotted blue line under the "C" portion of the Reaction Window, confirms that the test is working properly and the results are valid.

## J. Substantial Equivalence Information:

1. Predicate device name(s):

GIARDIA II

CRYPTOSPORIDIUM II

2. Predicate 510(k) number(s):

K033274

K052932

3. Comparison with predicate:

|  Similarities  |   |   |   |
| --- | --- | --- | --- |
|  Item | Device | Predicate 1 | Predicate 2  |
|  Intended Use | Detection of Giardia cyst and Cryptosporidium oocyst antigen | Detection of Giardia cyst antigen | Detection of Cryptosporidium oocyst antigen  |
|  Technology | Enzyme Immunoassay | Enzyme Immunoassay | Enzyme Immunoassay  |
|  Antibody Format | Monoclonal/Polyclonal | Monoclonal/Polyclonal | Monoclonal/Polyclonal  |
|  Specimen Types | Human Fecal Specimens- fresh, frozen, or preserved in 10% Formalin or SAF | Human Fecal Specimens- fresh, frozen, or preserved in 10% Formalin or SAF | Human Fecal Specimens- fresh, frozen, or preserved in 10% Formalin or SAF  |

{3}

|  Differences  |   |   |   |
| --- | --- | --- | --- |
|  Item | Device | Predicate 1 | Predicate 2  |
|  Intended Use | Detection and differentiation of *Giardia* cyst antigen and *Cryptosporidium* oocyst antigen | Detection *Giardia* cyst antigen only | Detection *Cryptosporidium* oocyst antigen only  |
|  Technology | Membrane Enzyme Immunoassay | Enzyme Immunoassay – Microassay Plate ELISA | Enzyme Immunoassay – Microassay Plate ELISA  |
|  Amount of Specimen required | 25 μL – fresh
100 μL - preserved | 100 μL – fresh or preserved | 100 μL – fresh or preserved  |
|  Incubation time | 15 minutes | 1 hour | 1 hour  |
|  Time to Result | 30 minutes | 1 hour 45 minutes | 1 hour 45 minutes  |

## K. Standard/Guidance Document Referenced (if applicable):

Not applicable

## L. Test Principle:

The GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test is a membrane enzyme immunoassay that uses monoclonal and polyclonal antibodies to cell-surface antigens of the organisms. The polyclonal antibodies are coupled to horseradish peroxidase which are contained in the diluent. Specific mouse monoclonal antibodies against Giardia and Cryptosporidium antigens are bound on different location of the membrane. A fecal specimen is added to the diluent and transferred to the sample well and incubated. During the incubation, cyst and/or oocyst antigens in the sample bind the antibody-peroxidase conjugates. The antigen-antibody-conjugate complexes migrate through a filter pad to a membrane where they are captured by the immobilized Giardia and/or Cryptosporidium-specific antibodies in the test lines.

## M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

A total of 22 fecal specimens were pre-characterized by commercially available predicate devices. The samples included six Giardia-positive specimens (three mid-range positives), six Cryptosporidium-positive

{4}

specimens (three low positives), and four Giardia/Cryptosporidium-positive specimens (two of which were low Giardia-positives and two of which were low Cryptosporidium-positives), and six specimens negative for both parasites. All specimens were coded to prevent their identification during testing. Testing was performed at three sites. The samples were tested, twice a day over a 5-day period by multiple technicians at each site using two different kit lots. A positive and negative control was run with each panel of the masked samples. The results were consistent among the different locations, and exhibited a correlation of 100%.

b. Linearity/assay reportable range:

Not applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Not applicable.

d. Detection limit:

The analytical sensitivity of the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK test was determined by spiking various dilutions of purified Giardia cysts or Cryptosporidium oocysts into negative human stool. The cyst/oocyst preparations were quantified using the reference assay of IFA microscopy. Test results determined the limit of detection to be 6000 Giardia cysts per mL (133 cysts/test) and 6000 Cryptosporidium oocysts/mL (133 cysts/test).

e. Analytical specificity:

Cross- Reactivity Study

No cross reactions were seen with the following organisms: Aeromonas hydrophila, Bacillus cereus, Bacillus subtilis, Bacteroides fragilis, Campylobacter coli, Campylobacter fetus, Campylobacter jejuni, Candida albicans, Clostridium bifermentans, Clostridium difficile (strain 630), Enterococcus faecalis, Escherichia coli, Escherichia coli 0157:H7, Escherichia coli ETEC (enterotoxic), Escherichia coli EPEC (enteropathogenic), Escherichia coli EIEC (enteroinvasive), Klebsiella pneumoniae, Salmonella typhimurium, Shigella dysenteriae, Shigella flexneri, Shigella sonnei, Staphylococcus aureus, Staphylococcus aureus (Cowan's), Staphylococcus epidermidis, Vibrio parahaemolyticus, Yersinia enterocolitica

Human Adenovirus 1 and 3, Adenovirus, Type 2, 5, 40 and 41, Human Coxsackievirus B2, B3, and B4, Coxsackievirus B5, Human Coronavirus, Echovirus 11, 18, 33, Human Echovirus 9, Human paraechovirus 1 (Echovirus 22), Enterovirus 68, 69, Human Enterovirus 70, 71, Human rotavirus

5

{5}

Additionally, the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test was tested using fecal specimens documented to be positive for other parasites by microscopy; *Ascaris lumbricoides* eggs, *Dientamoeba fragilis*, *Entamoeba hartmanni*, *Blastocystis hominis*, *Diphyllobothrium latum* eggs, *Entamoeba histolytica/E. dispar*, *Chilomastix mesnili*, *Endolimax nana*, Hookworm eggs, *Cyclospora cayetanensis*, *Entamoeba coli*, *Iodamoeba butschlii*, and *Trichuris trichiura* eggs. No cross-reactivity was seen with the following organisms for either the *Giardia*-portion or the *Cryptosporidium*-portion of the test.

Cross-reactivity to Astrovirus and Caliciviruses has not been evaluated.

The results obtained from the cross reactive study is acceptable.

## Interference Study

The following substances had no effect on positive or negative test results analyzed at the concentrations indicated: Hog gastric mucin (3.5% w/v), Human blood (40% v/v), Barium sulfate (5% w/v), Imodium (5% v/v), Kaopectate (5% v/v), Pepto-Bismol (5% v/v), Steric/Palmitic Acid (40% w/v), Metronidazole (0.25% w/v), Vancomycin (0.25% w/v).

The results obtained from the interference study is acceptable.

### f. Assay cut-off:

Not Applicable

### 2. Comparison studies:

#### a. Method comparison with predicate device:

The performance of the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK test was evaluated at three geographically diverse sites. Two sites compared the performance to commercially available ELISA (separate devices for *Giardia* and *Cryptosporidium*). A total of 849 specimens were evaluated and included 349 fresh, 322 frozen, 36 preserved (formalin) and 142 preserved (SAF).

Two hundred and fifteen (215) specimens were positive for *Giardia* by ELISA. The GIARDIA/CRYPTOSPORIDIUM QUIK CHEK test exhibited a 99.1% agreement for *Giardia* positive specimens, a 99.7% agreement for *Giardia* negative specimens, with an overall agreement of 99.5%.

One hundred thirty one (131) specimens were positive for *Cryptosporidium* by ELISA. The GIARDIA/CRYPTOSPORIDIUM QUIK CHEK test exhibited a 99.2% agreement for *Cryptosporidium* positive specimens, 99.6% agreement for *Cryptosporidium* negative specimens, and an overall

6

{6}

agreement of 99.5%.

The results obtained from the method comparison study are acceptable.

b. Matrix comparison:

Not Applicable

3. Clinical studies:

a. Clinical Sensitivity:

The performance of the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK test was evaluated at three geographically diverse sites. Two sites compared performance to the reference microscopy/immunofluorescence (IFA). A total of 791 were evaluated and included 220 fresh, 140 frozen, 216 preserved (formalin) and 215 preserved (SAF).

One hundred eighty three (183) specimens were positive for Giardia by IFA. The GIARDIA/CRYPTOSPORIDIUM QUIK CHEK test exhibited a sensitivity of 98.9%, a specificity of 100%, and an overall correlation of 99.7% with microscopy.

One hundred forty (140) specimens were positive for Cryptosporidium by IFA. The GIARDIA/CRYPTOSPORIDIUM QUIK CHEK test exhibited a sensitivity of 100%, a specificity of 99.8%, and an overall correlation of 99.9% with Microscopy.

The results obtained from the clinical study appear to be acceptable.

Combined Results - Clinical Performance Comparing the Giardia Line of the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK Test to Microscopy - IFA

|  N = 791 | Microscopy - IFA
Giardia positive | Microscopy - IFA
Giardia negative  |
| --- | --- | --- |
|  GIARDIA/CRYPTOSPORIDIUM
QUIK CHEK
Giardia Line Positive | 181 | 0  |
|  GIARDIA/CRYPTOSPORIDIUM
QUIK CHEK
Giardia Line Negative | 2 | 608  |
|   |   | 95% Confidence Limits  |
| --- | --- | --- |
|  Sensitivity | 98.9% | 95.7 – 99.8%  |
|  Specificity | 100% | 99.2 – 100%  |
|  Correlation | 99.7% | 99.7 – 99.7%  |

{7}

8

# Combined Results - Clinical Performance Comparing the Cryptosporidium Line of the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK Test to Microscopy – IFA

|  N = 791 | Microscopy - IFA Crypto. positive | Microscopy - IFA Crypto. negative  |
| --- | --- | --- |
|  GIARDIA/CRYPTOSPORIDIUM QUIK CHEK
Cryptosporidium Line Positive | 140 | 1  |
|  GIARDIA/CRYPTOSPORIDIUM QUIK CHEK
Cryptosporidium Line Negative | 0 | 650  |
|   |   | 95% Confidence Limits  |
| --- | --- | --- |
|  Sensitivity | 100% | 96.7 – 100%  |
|  Specificity | 99.8% | 99.0 – 100%  |
|  Correlation | 99.9% | 100 – 100%  |

b. Clinical specificity:

See section M.3.a

c. Other clinical supportive data (when a. and b. are not applicable):

Not Applicable

4. Clinical cut-off:

Not Applicable

5. Expected values/Reference range:

Normal healthy individuals should not be infected with Giardia or Cryptosporidium and should test negative in the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK test. A positive test result in the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK test indicates that the person is shedding detectable amounts of Giardia and/or Cryptosporidium antigen. The incidence of Giardia and Cryptosporidium infection varies significantly between populations and geographic regions. Children in daycare settings have exhibited higher rates of infection with Giardia than the normal population. In addition, homosexual men have shown higher rates of infection. In general, laboratory-confirmed incidence of cryptosporidiosis in developed countries ranges from 1 to 2% overall with a higher incidence in children.

Prospective studies to evaluate the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK test were conducted at an Asian hospital specializing in diarrheal diseases. The studies included 129 freshly collected fecal specimens. A positive test result in the

{8}

GIARDIA/CRYPTOSPORIDIUM QUIK CHEK test was obtained in 22.5% for Giardia and in 5.4% for Cryptosporidium antigen.

N. Proposed Labeling:

The labeling is sufficient and satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

9

---

**Source:** [https://fda-staging.innolitics.com/device/K103673](https://fda-staging.innolitics.com/device/K103673)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
