OXYLOG 3000 PLUS WORKSTATION, OXYLOG 3000 PLUS BASIC DEVICE
K103625 · Draeger Medical Systems, Inc. · CBK · Oct 6, 2011 · Anesthesiology
Device Facts
Record ID
K103625
Device Name
OXYLOG 3000 PLUS WORKSTATION, OXYLOG 3000 PLUS BASIC DEVICE
Applicant
Draeger Medical Systems, Inc.
Product Code
CBK · Anesthesiology
Decision Date
Oct 6, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Oxylog 3000 plus is a time-cycled, volume controlled and pressure controlled emergency and transport ventilator for patients requiring mandatory or assisted ventilation with a tidal volume from 50 mL upwards.
Device Story
Oxylog 3000 plus is an emergency and transport ventilator; successor to Oxylog 3000. Inputs include patient airway pressure and flow; device provides time-cycled, volume-controlled, or pressure-controlled ventilation. Features include AutoFlow and CO2 measurement. Used in mobile environments (ambulances, helicopters), accident/emergency departments, and recovery rooms. Operated by trained healthcare professionals. Output displayed on front panel; assists clinicians in managing patient respiratory support. Benefits include portable, reliable ventilation for patients requiring tidal volumes ≥50 mL; supports non-intubated patients via NIV mode.
Clinical Evidence
Bench testing only. Verification and validation testing conducted per IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, ISO 10651-3, and ISO 14971. Testing included hazard analysis, system-level qualification, and verification of design requirements for mechanical, electrical, and biocompatibility performance.
Technological Characteristics
Time-cycled, volume/pressure-controlled ventilator. Features AutoFlow and CO2 measurement. Connectivity includes standalone operation or integration into a carrying system with O2 cylinder storage. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 10651-3 (emergency/transport ventilators), IEC 60601-1-6 (usability), IEC 60601-1-8 (alarms), ISO 14971 (risk management).
Indications for Use
Indicated for patients requiring mandatory or assisted ventilation with a tidal volume from 50 mL upwards. Intended for use by trained healthcare professionals (doctors, nurses, EMTs, respiratory therapists, paramedics) in emergency, transport, and hospital settings.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
Oxylog 3000 plus
ﭘ
ﻪ ﻨ
K103625
# 510(k) Summary according to 21 CFR 807.92
#### Applicant's Name and Address:
Draeger Medical GmbH Moislinger Allee 53-55 23542 Luebeck Germany
Establishment Registration Number: 9611500
#### Contact Person:
Ulrich Schröder Director Regulatory & Clinical Affairs
Tel. No .: + 49 (451) 882-3648 Fax No .: + 49 (451) 882-3018
#### Applicant's US Contact Person:
Joyce Kilroy VP Processes, Quality & Regulatory Affairs
Tel. No.: (215) 660-2626 Fax No .: (267) 721-5424
#### Date submission was prepared:
August 24, 2011
#### Device Name:
Trade Name: Oxylog 3000 plus Common Name: Ventilator
#### Classification:
Class II
Regulation No. Device Product Code Preferred Code Oxylog 3000 plus: 868.5895 Ventilator, continuous, СВК Code for predicates (Oxylog 3000, Oxylog 2000 plus and Oxalog 2000:: 868.5895 Ventilator, continuous, свк
August 2011
FDA CDRH DMC
AUG 30 2011
Received
OCT - 6 2011
{1}------------------------------------------------
ン
## Reason for the 510(k): Substantial Equivalence:
The new Oxylog 3000 plus has been compared to the following 3 predicative devices: the Oxylog 2000 plus (K082600), the Oxylog 3000 (K062267) and the Oxylog 2000 (K984577).
All 4 devices are intended to be used for patients for emergency and transport. All devices contain the equivalent set of ventilation modes and follow the same use philosophy.
The variations between the labeling are minimal due to the evolving development changes from the early 1990s until today.
The Oxylog 3000 plus and Oxylog 3000 (K062267) are intended for patients with a tidal volume from 50 mL upwards. Oxylog 2000 (K984577) plus and Oxylog 2000 plus (K082600) are intended for patients with a tidal volume from 100 mL upwards.
Drawn from all information presented in this submission it is summarized that Oxylog 3000 plus is as safe, as effective, and performs as well as or better than the predicate devices.
#### Legally Marketed Device to which Substantial Equivalence is claimed:
| 510(k) number | Trade name | Company |
|---------------|------------------|----------------------|
| K984577 | Oxylog 2000 | Draeger Medical GmbH |
| K062267 | Oxylog 3000 | |
| K082600 | Oxylog 2000 plus | |
#### Executive Summary
This documentation covers the Oxylog 3000 plus, a new device of the Oxylog family of emergency and transport ventilators.
The Oxylog 3000 plus is intended as successor of the current Oxylog 3000. The ventilator is based on the existing Oxylog 3000 and Oxylog 2000 plus with slight adaptations like different software and a different front panel to enable the new functionality AutoFlow and CO2 measurement in the device.
Safety testing was conducted per IEC60601-1, IEC60601-1-2, and other applicable standards with respect to mechanical, electrical and biocompatibility. The results of all verification and validation testing demonstrate that all system and design requirements for the Oxylog 3000 plus have been met. Qualification included hazard analysis, system level qualification and verification / validation testing.
### Device Description:
#### Indications for Use
The Oxylog 3000 plus is a time-cycled, volume controlled and pressure controlled emergency and transport ventilator for patients requiring mandatory or assisted ventilation with a tidal volume from 50 mL upwards.
August 2011
{2}------------------------------------------------
#### Intended operator
The device is intended for use by and under the supervision of trained healthcare professionals, e.g. doctors, nurses, emergency medical technicians, respiratory therapists, and paramedics.
### Enviroment of use
Intended environment of use:
- Mobile use for emergency patients, in both outdoor and indoor environments.
- During transport in ambulances or aircraft, including helicopters.
- In accident and emergency departments.
- When moving ventilated patients around the hospital.
- In the recovery room.
The following ventilation modes are offered:
·VC-CMV / VC-AC:
oVolume Controlled --- Controlled Mandatory Ventilation with or without PEEP oVolume Controlled - Assist Control with or without PEEP
·VC-SIMV: Volume Controlled -- Synchronized Intermittent Mandatory Ventilation
·PC- BIPAP: Pressure Controlled- Biphasic Positive Airway Pressure
·PC - SIMV+: Pressure Controlled- Synchronized Intermittent Mandatory Ventilation •Spn-CPAP: Spontaneous breathing- Continuous Positive Airway Pressure
Additional settings for ventilation:
·Pressure support can be supplied in the ventilation modes VC-SIMV, PC-BIPAP, Son-CPAP.
· Apnoea ventilation for spontaneously breathing patients: in the ventilation mode Spn-CPAP.
·AutoFlow (optional); in the ventilation modes VC-CMV, VC-AC and VC-SIMV. ·NIV: in the ventilation modes: Spn-CPAP (/PS), PC-BIPAP (/PS), VC-CMV / AF, VC-AC / AF and VC-SIMV / AF. Using the NIV mode, non-intubated patients can be ventilated with a mask.
The Oxylog 3000 plus can be used as a stand alone unit or in a carrying system. This system protects the device and provides space for an O2 cylinder. The bags of the carrying system can be used to store accessories like a ventilation hose or mask. The carrying system can be used in ambulances and hospitals. It can be mounted by means of an universal wall holder.
August 2011
{3}------------------------------------------------
・・・
Drägger
# List of performance testing:
Performance was tested in compliance with following standards:
| IEC 60601-1 :2006 | Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential<br>performance |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-2 : 2007 | Medical electrical equipment - Part 1-2: General<br>requirements for basic safety and essential<br>performance - Collateral standard: Electromagnetic<br>compatibility - Requirements and tests |
| ISO 10651-3: 1997 | Lung ventilators for medical use - Part 3: Particular<br>Requirements for emergency and transport ventilators |
| IEC 60601-1-6 : 2004 | Medical electrical equipment - Part 1-6: General<br>requirements for basic safety and essential<br>performance - Collateral standard - Usability |
| IEC 60601-1-8: 2006 | Medical electrical equipment - Part 1-8: General<br>requirements for basic safety and essential<br>performance - Collateral standard: General<br>requirements, tests and guidance for alarm systems in<br>medical electrical equipment and medical electrical<br>systems |
| ISO 14971: 2007 | Medical devices - Application of risk management to<br>medical devices |
August 2011
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The logo is surrounded by a circular border containing the text "U.S. Department of Health & Human Services USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Joyce Kilroy Vice President Processes. Quality & Regulatory Affairs Draeger Medical Systems, Incorporated 3135 Quarry Road Telford, Pennsylvania 18969
OCT - 6 2011
Re: K103625
Trade/Device Name: Oxylog 3000 plus Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK, BTL Dated: September 30, 2011 Received: October 3, 2011
Dear Ms. Kilroy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading,
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. Tn addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2 - Ms. Kilrov
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRHis) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Parl 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Nh for
Anthony D. Watson. B.S., M.S.. M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
## Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K103625
Device Name: Oxylog 3000 plus
Indications for Use:
The Oxylog 3000 plus is a time-cycled, volume controlled and pressure controlled emergency and transport ventilator for patients requiring mandatory or assisted ventilation with a tidal volume from 50 mL upwards.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schult
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of 1
510(k) Number: _ K 1036 25
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.