K103562 · Centurion Medical Products Corporation · FPA · Jun 23, 2011 · General Hospital
Device Facts
Record ID
K103562
Device Name
CENTURION PRESSURE INJECTABLE EXTENSION SET
Applicant
Centurion Medical Products Corporation
Product Code
FPA · General Hospital
Decision Date
Jun 23, 2011
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
Centurion® Pressure Injectable Extension Sets are prescription devices intended to allow the aspiration, injection, or gravity/pump flow of fluids and may be used with power injectors having a maximum pressure setting of 325 psi. When used with a power injector, the Centurion® Pressure Injectable Extension Set must be secured to other devices, via a luer lock connection. These devices must also be rated for power injection applications.
Device Story
Centurion® Pressure Injectable Extension Set is a sterile, single-use intravascular administration set; consists of non-DEHP PVC tubing with female luer locks; available in straight, Y, or T configurations; includes optional slide/snap clamps and needlefree valves. Used in clinical settings by healthcare professionals for fluid delivery/aspiration or power injection (max 325 psi, 10 mL/s). Device connects between power injector and IV catheter; requires luer-lock security and compatibility with other power-rated components. Benefits include safe, high-pressure fluid administration during diagnostic or therapeutic procedures.
Clinical Evidence
Bench testing only. Biocompatibility testing per ANSI/AAMI/ISO 10993-1:2009 confirmed non-cytotoxic, non-sensitizing, non-hemolytic, and low complement activation. Sterilization validated per ANSI/AAMI/ISO 11135-1:2007 (SAL 1x10^-6). Packaging validated per ISO 11607-1.
Technological Characteristics
Materials: non-DEHP PVC tubing. Configuration: straight, Y, or T with female luer locks. Sterilization: Ethylene oxide (ANSI/AAMI/ISO 11135-1:2007). Pressure rating: max 325 psi. Flow rate: max 10 mL/s. Connectivity: luer-lock mechanical interface.
Indications for Use
Indicated for delivery and/or aspiration of fluids in patients requiring IV access. Compatible with power injectors up to 325 psi and 10 mL/second flow rate. Contraindicated for use with blunt cannulae systems.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "K103562 page 1 of 2" written in cursive. The text appears to be handwritten, with a slightly uneven stroke. The text indicates that the document is page 1 of a 2-page document and has the identifier K103562.
JUN 2 3 2011
# 510(k) Summary
Centurion Medical Products Corporation 100 Centurion Way Williamston, MI 48895
Mr. Matthew K. Price Director of Quality Assurance and Regulatory Affairs (517) 546-5400, x1135 Phone: Facsimile: (517) 546-3356
| Date Summary Prepared: | November 29, 2010 |
|--------------------------------|------------------------------------------------------------|
| Trade Name (Proprietary Name): | Centurion® Pressure Injectable Extension Set |
| Common Name of Device: | Intravascular Administration Set |
| Classification Name: | Set, Administration, Intravascular |
| Device Classification: | II |
| Regulation: | 880.5440 |
| Product Code: | FPA |
| Review Panel: | General Hospital |
| Reason for 510(k): | Marketing of Centurion® Pressure Injectable Extension Sets |
#### Predicate SE Device(s):
Manufacturer:
Contact:
This product is similar in design, composition, function, and method of use to the following products:
Cardinal Health, Alaris ® Products, SmartSite® Needle Free Valve Administration Sets (K061285) .
for Physician and Hospital use.
# Description:
The Centurion® Pressure Injectable Extension set intended for the delivery and/or aspiration of fluids and may be used in conjunction with power injectors having a maximum pressure setting of 325 psi and a maximum flow rate of 10 mL/second.
Centurion® Pressure Injectable Extension Sets consist of polyvinyl chloride (PVC) tubing with non-DEHP plasticizer. They are configured with female luer locks on the proximal and distal ends, respectively, and are available in straight, Y, and T configurations. A slide or snap clamp may be positioned over the tubing to obstruct the fluid flow when needed.
The Centurion® Pressure Injectable Extension Set is available in configurations with and without a needlefree valve. These valves are separately marketed devices and the applicable 510(k)'s are maintained by the respective manufacturers.
The Centurion® Pressure Injectable Extension Sets are prescription devices provided sterile and nonpyrogenic for single use only and may be packaged individually or in medical convenience kits.
Centurion® Pressure Injectable Extension Sets 5-1 510(k) Summary, Rev. 0 510(k) Submission
> 100 CENTURION WAY | F.O. BOX 510 | WILLIAMSTON | MICHIGAN | 48895 phone 517.546 5400 | toll free 800 248 4059 | www.centurionmp.com
{1}------------------------------------------------
ION K.103562
### Intended Use:
Centurion® Pressure Injectable Extension Sets are prescription devices intended to allow the aspiration, injection, or gravity/pump flow of fluids and may be used with power injectors having a maximum pressure setting of 325 psi. When used with a power injector, the Centurion® Pressure Injectable Extension Set must be secured to other devices, via a luer lock connection. These devices must also be rated for power injection applications.
# Summary of Technological Characteristics between Subject and Predicate Device:
The differences in technological characteristics (e.g., design specifications) between the subject and predicate devices raise no new questions of safety or effectiveness. Slight differences in materials, dimensions, sterilization method, and warnings exist; however, these differences do not affect the safety and effectiveness of the subject device.
### Performance Data:
The results of performance testing demonstrate that the functionality, integrity, and safety and effectiveness of the Centurion® Pressure Injectable Extension Set are sufficient for its intended use and support a determination of substantial equivalence.
### Summary of Testing:
Biocompatibility testing was performed on Centurion® Pressure Injectable Extension Sets in accordance with ANSI/AAMI/ISO 10993-1:2009: Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing and AAMI Standards and Recommended Practices, Required Biocompatibility Training and Toxicology Profiles for Evaluation of Medical Devices. Results of testing validate Centurion® Pressure Injectable Extension Sets are non-cytotoxic, non-sensitizing, a negligible irritant, non-hemolytic, and a low potential activator of the complement system.
Centurion® Pressure Injectable Extension Sets will be available only in sterile packaged form. The sterile product will be sterilized using ethylene oxide. The sterilization method was validated and performed in accordance with ANSI/AAMI/ISO 11135-1:2007, Sterilization of health care products – Ethylene oxide- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices. A sterility assurance level of 1 x 10 ° has been validated for this product.
Centurion's packaging has been validated for performance and shelf life in accordance with ISO 11607-1 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems.
Centurion® Pressure Injectable Extension Sets 510(k) Submission
5-2
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The logo is simple and uses a black and white color scheme.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Matthew K. Price Director of Quality Assurance & Regulatory Affairs Centurion Medical Products Corporation 100 Centurion Way Williamston, Michigan 48895
JUN 2 3 2011
Re: K103562
Trade/Device Name: Centurion® Pressure Injectable Extension Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: June 15, 2011 Received: June 16, 2011
Dear Mr. Price:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2- Mr. Price
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.html for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Susan Runner
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K103542
# Indications for Use Statement
# 510(k) Number (if known):
Device Name: Centurion® Pressure Injectable Extension Set
### Indications For Use:
The Centurion® Pressure Injectable Extension set intended for the delivery and /or aspiration of fluids and may be used in conjunction with power injectors having a maximum pressure setting of 325 psi and a maximum flow rate of 10 mL/second.
### Contraindications:
Contraindicated for use with blunt cannulae systems.
### Cautions:
1. For unattended medication delivery use only a luer-locking 1.V. set or syringe.
2. Refer to the Centurion pressure information sheet for applicable valve pressure rating and valve instructions.
3. When used with a power injector, the Centurion Pressure Injectable Extension Set must be secured to other devices via a luer lock connection. These devices must also be rated for power injection applications.
4. Centurion Pressure Injectable Extension Sets are intended for use with power injectors under the following conditions:
a. Before attempting power injection, ensure patency and that any unused access ports are securely capped.
- b. All components (from injector to catheter) must be rated for use with a power injector at the selected pressure setting.
- c. A 20-gauge or larger IV catheter should be used.
- d. A minimum 60″ coiled connector tubing must be used between the injector syringe and Centurion set.
- e. The injector pressure limit must be set to 325psi or below and the flow rate set to a maximum of 10ml/second.
# Warning:
Do not use this product with pressure infusion systems at a pressure setting above 325 psi.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | x |
|-------------------------------------------------|---|
|-------------------------------------------------|---|
AND/OR
| Over-The-Counter Use<br>(21 CFR 801 Subpart C) | |
|------------------------------------------------|--|
|------------------------------------------------|--|
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K103562
Centurion® Pressure Injectable Extension Sets 4-1 510(k) Submission
Indications for Use, Rev. 0
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.