LIFESCAN ONETOUCH SELECT / ULTRA / VITA GLUCOSE CONTROL SOLUTION
K103553 · Bionostics, Inc. · JJX · Feb 10, 2011 · Clinical Chemistry
Device Facts
Record ID
K103553
Device Name
LIFESCAN ONETOUCH SELECT / ULTRA / VITA GLUCOSE CONTROL SOLUTION
Applicant
Bionostics, Inc.
Product Code
JJX · Clinical Chemistry
Decision Date
Feb 10, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Attributes
Pediatric
Indications for Use
LifeScan OneTouch Ultra Control Solution is intended for use to verify the performance of the LifeScan OneTouch Ultra Family and the OneTouch Ping™ Meter Remote blood glucose monitoring test systems at a glucose level printed on the test strip vial. The LifeScan OneTouch Ultra Control Solution is intended for use by healthcare professionals and people with diabetes mellitus at home. LifeScan OneTouch Select Control Solution is intended for use to verify the performance of the LifeScan OneTouch Select blood glucose monitoring test system at a glucose level printed on the test strip vial. The LifeScan OneTouch Select Control Solution is intended for use by healthcare professionals and people with diabetes mellitus at home. LifeScan OneTouch Vita Control Solution is intended for use to verify the performance of the LifeScan OneTouch Vita blood glucose monitoring test system at a glucose level printed on the test strip vial. The LifeScan OneTouch Vita Control Solution is intended for use by healthcare professionals and people with diabetes mellitus at home.
Device Story
Control solution for blood glucose monitoring systems; verifies test system performance. Buffered aqueous solution containing D-glucose; viscosity modifiers; preservatives; dye. Applied to test strips by healthcare professionals or patients at home; mimics whole blood response. Output is a glucose reading compared against the target level printed on the test strip vial. Used to ensure accuracy of glucose monitoring systems; helps patients and clinicians confirm device reliability.
Clinical Evidence
No clinical data. Bench testing only, including closed bottle shelf-life stability, stability after opening (use-life), transport stability, and test response verification.
Technological Characteristics
Viscosity-adjusted, aqueous glucose control solution. Contains D-glucose, buffer, viscosity modifier, preservative, and dye. No hazardous, human, or animal-derived components. Packaged in 6 mL LDPE vials with dispensing tip and cap. Optimized to simulate whole blood response on LifeScan blood glucose test systems.
Indications for Use
Indicated for healthcare professionals and people with diabetes mellitus at home to verify the performance of specific LifeScan blood glucose monitoring test systems (OneTouch Ultra Family, OneTouch Ping Meter Remote, OneTouch Select, and OneTouch Vita) using a glucose level printed on the test strip vial.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
{0}------------------------------------------------
K103553
FEB 1 0 2011
# Quality Solut
## 510(k) Summary1
#### Submitter's name, address (a) (1)
Bionostics, Inc. 7 Jackson Road Devens, MA 01434
## Contact Person
10 February 2011
Randy Byrd VP, Chief Technical Officer (978) 772-7070 x 272
Date of preparation of this summary:
- (2) Device trade or proprietary name:
LifeScan OneTouch Select Control Solution LifeScan OneTouch Ultra Control Solution LifeScan OneTouch Vita Control solution
### Device common or usual name or classification name:
גור Single (Specified) Analyte Control, All Types, Assayed and Unassayed
| REGULATION MEDICAL<br>SPECIALTY | REGULATION<br>NUMBER | CLASS | REGULATION DESCRIPTION |
|---------------------------------|----------------------|-------|------------------------|
| Chemistry | 862.1660 | II | Glucose Control |
#### l, Substantial Equivalence
LifeScan OneTouch Select, Ultra and Vita Control Solutions are substantially equivalent in function, safety and efficacy to currently marketed devices for the same intended use as shown in the following tables:
| Characteristic | Predicate Device | Proposed Device |
|-------------------|----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Name: | OneTouch Ultra Control | OneTouch Select Control<br>OneTouch Ultra Control<br>OneTouch Vita Control |
| 510(k), Date: | K022769, 11.13.2002 | |
| Number of levels: | 1, typical fasting glucose | 1, typical fasting glucose |
| Analytes: | glucose | glucose |
| Container: | 6 mL LDPE vial with dispensing tip<br>and cap | 6 mL LDPE vial with dispensing tip<br>and cap |
| Fill volume: | 4 mL | 4 mL |
| Color: | red | red |
| Matrix: | Buffered, aqueous solution of D-<br>Glucose, viscosity modifier,<br>preservatives and other, non-<br>reactive ingredients. | Buffered, aqueous solution of D-<br>Glucose, viscosity modifier,<br>preservatives and other, non-<br>reactive ingredients. |
| Brands: | OneTouch Ultra, OneTouch<br>Select, OneTouch Vita | Unchanged: OneTouch Ultra,<br>OneTouch Select, OneTouch Vita |
LifeScan, OneTouch, OneTouch Select and OneTouch Vita are registrered trademarks of Johnson & Johnson Corporation, New Brunswick, NJ, USA.
<sup>4</sup> This summary of safety and effectiveness is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
{1}------------------------------------------------
#### ll. Description of the new device
The control solution is a buffered aqueous solution with glucose containing no ingredients of biological origin, or in concentrations qualifying as a controlled product under the Controlled Products Regulation. The primary differences in the composition of the OneTouch Select, Ultra and Vita branded Control Solutions to that of the current,
· OneTouch Ultra Control are the type and relative content of buffer system, type and relative content of the viscosity adjusting agent, type and relative content of the preservative system, a lower concentration of dye, and a relative reduction in the content of water. The OneTouch Ultra control is currently in use with all LifeScan blood glucose test systems under the OneTouch Ultra, OneTouch Select and OneTouch Vita brands. The new control solution is reformulated for optimized performance across these various brands of test strips, and will be distributed under the same brands as the current Ultra control.
#### (a) (1) Intended use of the device
LifeScan OneTouch Ultra Control Solution is intended for use to verify the performance of the LifeScan OneTouch Ultra Family and the OneTouch Ping™ Meter Remote blood glucose monitoring test systems at a glucose level printed on the test strip vial. The LifeScan OneTouch Ultra Control Solution is intended for use by healthcare professionals and people with diabetes mellitus at home.
LifeScan OneTouch Select Control Solution is intended for use to verify the performance of the LifeScan OneTouch Select blood glucose monitoring test system at a glucose level printed on the test strip vial. The LifeScan OneTouch Select Control Solution is intended for use by healthcare professionals and people with diabetes mellitus at home.
LifeScan OneTouch Vita Control Solution is intended for use to verify the performance of the LifeScan OneTouch Vita blood glucose monitoring test system at a glucose level printed on the test strip vial. The LifeScan OneTouch Vita Control Solution is intended for use by healthcare professionals and people with diabetes mellitus at home.
#### (a) (2) Technological characteristics of the device.
This material consists of viscosity-adjusted, aqueous glucose control solution prepared with a single concentration of D-glucose with recovery on the test systems in the range typically considered normal, fasting glucose for a non-diabetic person. This solution has been optimized to simulate the response of whole blood on the LifeScan OneTouch Ultra, Select and Vita branded blood glucose test systems. The solution contains no hazardous, human or animal derived components.
{2}------------------------------------------------
- (b) (1) Summary of non-clinical tests submitted with the premarket notification for the device. Tests were conducted to verify specific performance requirements:
- a) Closed bottle stability (Shelf-life)
- b) Stability after opening (Use-life)
- c) Transport Stability
- d) Test response
- Summary of clinical tests submitted with the premarket notification for the device. (b) (2) N/A
- (b) (3) Conclusions drawn from the clinical and non-clinical trials.
Comparison of technological characteristics, formulation and intended use to predicate devices listed in this summary support the claim of substantial equivalence.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird with outstretched wings.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
FEB 1 0 2011
Bionostics, Inc. c/o Randy Byrd Vice President & Chief Technical Officer 7 Jackson Road Devens, MA 01434
Re: k103553
> Trade Name: LifeScan OneTouch Select Control Solution, LifeScan OneTouch Ultra Control Solution, LifeScan OneTouch Vita Control Solution Regulation Number: 21 CFR §862.1660 Regulation Name: Quality control material Regulatory Class: Class I, reserved Product Code: JJX Dated: December 2, 2010 Received: December 3, 2010
Dear Mr. Byrd:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820).
{4}------------------------------------------------
Page 2 -
lf you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
CJC.
Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number: K103553 Device Name: Indications for Use:
LifeScan® OneTouch® Select® Control Solution The LifeScan OneTouch Select Control Solution is intended for use to verify the performance of the LifeScan OneTouch Select blood glucose monitoring test system at a glucose level printed on the test strip vial. The LifeScan OneTouch Select Control Solution is intended for use by healthcare professionals and people with diabetes mellitus at home.
For In Vitro Diagnostic Use
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrent of CDRH, Office of Device Evaluation (OIVD)
Carol Benson
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K103553
Page 1 of 3
{6}------------------------------------------------
## Indications for Use
510(k) Number: K103553 LifeScan® OneTouch® Ultra® Control Solution Device Name: Indications for Use: The LifeScan OneTouch Ultra Control Solution is intended for use to verify the performance of the LifeScan OneTouch Ultra Family and the OneTouch Ping™ Meter Remote blood glucose monitoring test systems at a glucose level printed on the test strip vial. The LifeScan OneTouch Ultra Control Solution is intended for use by healthcare professionals and
people with diabetes mellitus at home.
For In Vitro Diagnostic Use
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrent of CDRH, Office of Device Evaluation (OIVD)
Carol Benson
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K103553
{7}------------------------------------------------
## Indications for Use
K103553
510(k) Number: Device Name: Indications for Use:
LifeScan® OneTouch® Vita® Control Solution The LifeScan OneTouch Vita Control Solution is intended for use to verify the performance of the LifeScan OneTouch Vita blood glucose monitoring test system at a glucose level printed on the test strip vial. The LifeScan OneTouch Vita Control Solution is intended for use by healthcare professionals and people with diabetes mellitus at home.
For In Vitro Diagnostic Use
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrent of CDRH, Office of Device Evaluation (OIVD)
Carl Benson
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K103553
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.