INTERLINK S / CLEARLINK SYSTEM / V-LINK LAD WITH VITALSHIELD PROTECTIVE COATING NON-DEHP CATHETER EXTENSION SET KITS,
K103551 · Baxter Healthcare Corp · FPA · Mar 25, 2011 · General Hospital
Device Facts
Record ID
K103551
Device Name
INTERLINK S / CLEARLINK SYSTEM / V-LINK LAD WITH VITALSHIELD PROTECTIVE COATING NON-DEHP CATHETER EXTENSION SET KITS,
Applicant
Baxter Healthcare Corp
Product Code
FPA · General Hospital
Decision Date
Mar 25, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The IV extension tubing kits provide the user with a "ready-to-use" set up for the administration and withdrawal of fluids avoiding the need to acquire/assemble components from various manufacturers. The kits are indicated for one-time use. The IV extension tubing set is indicated for use with a vascular access device for administration and withdrawal of fluids. The Vital-Hold catheter stabilization device and the foam tape strips are indicated to provide a method for catheter securement. The APLICARE skin prep pad is indicated to prepare the skin prior to application of a dressing or bandage.
Device Story
Kit integrates intravenous extension tubing, Vital-Hold catheter stabilization device, foam tape strips, and APLICARE skin protectant prep pad into single sterile blister package. Eliminates need for separate component assembly. Tubing connects to vascular access devices for fluid administration/withdrawal. Stabilization device and tape anchor lines to patient skin. Skin prep pad applies polymer film to skin to improve adhesion and provide barrier. Used in clinical settings; operated by healthcare providers. Benefits include convenience, reduced assembly time, and standardized securement.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Kit components include IV extension tubing (various lengths/volumes), Vital-Hold catheter stabilization device, foam tape, and APLICARE skin prep pad. Materials include non-DEHP tubing and luer-activated valves (Interlink, Clearlink, V-Link). V-Link devices feature VitalShield protective coating. Sterile, single-use, ready-to-use blister pack. No electronic or software components.
Indications for Use
Indicated for patients requiring intravenous fluid administration or withdrawal via vascular access devices. Components used for catheter securement and skin preparation prior to dressing application. For one-time use.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Modified Interlink Injection Site Septum (K925126)
Modification to Solution Administration Set with Capped Luer Activated Device (Clearlink) (K003225)
CLEARLINK Antimicrobial Luer Activated Device and Extension Sets with CLEARLINK Antimicrobial Luer Activated Device (K072576)
V-LINK Antimicrobial Luer Activated Device and Extension Sets with V-LINK Antimicrobial Luer Activated Device (K081289)
Submission Summary (Full Text)
{0}------------------------------------------------
Section 5, 510(k) Summary
MAR 2 5 2011
## 5. 510(k) SUMMARY
January 12, 2010
## OWNER:
Baxter Healthcare Corporation One Baxter Parkway Deerfield, Illinois 60015
# CONTACT PERSON:
Nanette Hedden Senior Manager, Global Regulatory Affairs 1620 Waukegan Rd. MPGR-AL McGaw Park, IL 60085 Telephone: (847) 270-4871 Fax: (847) 785-5116
#### DEVICE NAME:
Trade name: Intravenous Extension Tubing Kits
Table 5-1.
**Product Codes for Intercompany Extension Tuning & Use**
| Code number | Name |
|-------------|----------------------------------------------------------------------------------------------------------------------------------|
| 2N8220 | Clearlink System Non-DEHP Catheter Extension Set Kit, 7.7"(20.0 cm), Vol. 1.10 ml |
| 2N8221 | Interlink System Non-DEHP Catheter Extension Set Kit, 7.1"(18.0 cm), Vol. 0.85 ml |
| 2N8222 | Clearlink System Non-DEHP Catheter Extension Set Kit, 8.2"(21.0 cm), Vol. 0.57 ml |
| 2N8223 | Interlink System Non-DEHP Catheter Extension Set Kit, 7.7"(19.45 cm), Vol. 0.22 ml |
| 2N8224 | Clearlink System Non-DEHP Y-Type Catheter Extension Set Kit, 6.5"(16.45 cm), Vol. 1.2 ml |
| 2N8225 | Interlink System Non-DEHP Y-Type Catheter Extension Set Kit, 5.3"(13.4 cm), Vol. 1.21 ml |
| 2N8227 | Interlink System Non-DEHP Y-Type Catheter Extension Set Kit, 5.5"(13.9 cm), Vol. 0.85 ml |
| 6N8220 | V-Link Luer Activated Device with VitalShield Protective Coating Non-DEHP Catheter Extension Set Kit, 7.6"(19.3 cm), Vol. 1.0 ml |
{1}------------------------------------------------
| Code number | Name |
|-------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| 6N8222 | V-Link Luer Activated Device with VitalShield Protective Coating Non-DEHP<br>Catheter Extension Set Kit, 8.2"(21.0 cm), Vol. 0.5 ml |
| 6N8224 | V-Link Luer Activated Device with VitalShield Protective Coating Non-DEHP Y-<br>Type Catheter Extension Set Kit, 5.3"(13.5 cm), Vol. 1.2 ml |
Table 5-1. Intravenous Extension Tubing Kits Product Codes
Common name: Intravenous Extension Tubing Kits
Classification name: IV Administration Set (21 CFR 880.5440, Product Code FPA), Intravenous Extension Tubing Set (21 CFR 880.5440, Product Code OJA), Intravascular Catheter Securement Device (21 CFR 880.5210, Product Code KMK), Bandage, Liquid, Skin Protectant (21 CFR 880.5090, Product Code NEC).
### PREDICATE DEVICES:
| Device | Company | Previous<br>510(k) | Clearance date |
|------------------------------------------------------------------------------------------------------------------------------|----------------------|--------------------|------------------|
| APLICARE Skin Protectant Prep Pad | APLICARE,<br>Inc. | K861557 | July 11, 1986 |
| Modified Interlink Injection Site Septum | Baxter<br>Healthcare | K925126 | July 18, 1993 |
| Modification to Solution Administration Set<br>with Capped Luer Activated Device<br>(Clearlink). | Baxter<br>Healthcare | K003225 | October 19, 2000 |
| CLEARLINK Antimicrobial Luer Activated<br>Device and Extension Sets with<br>CLEARLINK Antimicrobial Luer Activated<br>Device | Baxter<br>Healthcare | K072576 | November 6, 2007 |
| V-LINK Antimicrobial Luer Activated<br>Device and Extension Sets with V-LINK<br>Antimicrobial Luer Activated Device | Baxter<br>Healthcare | K081289 | August 4, 2008 |
Table 5-2. Previous 510(k)s
{2}------------------------------------------------
### DESCRIPTION OF THE DEVICE:
The intravenous extension tubing kits will consist of an intravenous extension tubing set, a Vital-Hold catheter stabilization device, two foam tape strips and an APLICARE skin protectant prep pad. The intravenous extension sets may contain any of Baxter's cleared access ports (e.g., Interlink Injection Site, Clearlink Luer Activated Valve or V-Link Luer Activated Valve).
The IV extension tubing set is used with a vascular access device for administration and withdrawal of fluids.
The Vital-Hold catheter stabilization device is used to help anchor medical intravenous tubes and lines requiring mechanical stability while installed on the skin surface of a patient.
The two foam tape strips with release liners are used to assist with line stabilization.
The APLICARE skin protectant prep pad is a sterile pad saturated in skin prep solution. The prep solution prepares the skin prior to application of a dressing or bandage. When dry, the prep agent leaves a thin polymer film on the skin surface providing a clean surface for adhesion and a barrier to the natural body oils that can affect adhesion.
The IV extension tubing set, the Vital-Hold catheter stabilization device, the two foam tape strips and the APLICARE skin protectant prep pad will be integrated into one sterile unit blister package to provide the user with a "ready-to-use" set up avoiding the need to acquire/assemble components from various manufacturers) 4 home and comments and
#### STATEMENT OF INTENDED USE:
The IV extension tubing kits provide the user with a "ready-to-use" set up for the administration and withdrawal of fluids avoiding the need to acquire/assemble components from various manufacturers. The kits are indicated for one-time use.
The IV extension tubing set is indicated for use with a vascular access device for administration and withdrawal of fluids.
The Vital-Hold catheter stabilization device and the foam tape strips are indicated to provide a method for catheter securement.
The APLICARE skin prep pad is indicated to prepare the skin prior to application of a dressing or bandage.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized image of an eagle or bird with three lines forming its body and wings. The bird is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
C/O Ms. Nanette Hedden Baxter Healthcare Corporation 1620 Waukegan Road McGaw Park, Illinois 60085
APR - 5 2011
Re: K103551
Trade/Device Name: Intravenous Extension Tubing Kits Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: February 28, 2011 Received: March 1, 2011
Dear Ms. Hedden:
This letter corrects our substantially equivalent letter of March 25, 2011.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2- Ms. Hedden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other sequirements of the Act or any Federal statutes and regulations administered by other Federal squareins. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-fire number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Howad Bolstead
for
FD
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
#### INDICATIONS FOR USE .
510(k) Number (if known): k103551
Device Name:
Intravenous Extension Tubing Kits
Indications for Use:
The IV extension tubing kits provide the user with a "ready-to-use" set up for the administration and withdrawal of fluids avoiding the need to acquire/assemble components from various manufacturers. The kits are indicated for one-time use.
The IV extension tubing set is indicated for use with a vascular access device for administration and withdrawal of fluids.
The Vital-Hold catheter stabilization device and the foam tape strips are indicated to provide a method for catheter securement.
The APLICARE skin prep pad is indicated to prepare the skin prior to application of a dressing or bandage.
Prescription Use X AND/OR Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
1
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
3/24/4
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K103551
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.