K103529 · DiaSorin, Inc. · LOL · Jan 21, 2011 · Microbiology
Device Facts
Record ID
K103529
Device Name
LIAISON XL ANALYZER MODEL 10050
Applicant
DiaSorin, Inc.
Product Code
LOL · Microbiology
Decision Date
Jan 21, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3310
Device Class
Class 2
Indications for Use
The LIAISON® Anti-HAV assay is an in vitro chemiluminescent immunoassay intended for the qualitative detection of total antibodies to hepatitis A (anti-HAV) in human serum and sodium heparin plasma samples using the LIAISON® Analyzer family. The assay is indicated as an aid in the laboratory diagnosis of current or previous HAV infections in conjunction with other serological and clinical information and to determine the presence of an antibody response to HAV in vaccine recipients. This assay is not intended for screening blood or solid or soft tissue donors. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients. The user is responsible for establishing assay performance characteristics in these populations. Caution: U.S. Federal Law restricts this device to sale by or on the order of a physician. The LIAISON® XL Analyzer is an automated discrete continuous loading chemiluminescent immunoassay (CLIA) analyzer for in vitro diagnostic analysis of CLIAs on human serum or plasma. The system menu includes infectious disease, bone and mineral, and endocrinology CLIAs. It is to be used only with FDA cleared chemilyminescence immunoassays that are marketed by DiaSorin for the LIAISON XL Analyzer. The LIAISON® Control Anti-HAV (negative and positive) is intended for use as assayed quality control samples to monitor the performance of the LIAISON® Anti-HAV assay. The performance characteristics of LIAISON® controls have not been established for any other assays or instrument platforms different from LIAISON® and LIAISON® XL.
Device Story
LIAISON® XL Analyzer is an automated, floor-standing, continuous-loading chemiluminescent immunoassay (CLIA) system for clinical laboratory use. It processes human serum or plasma samples to detect infectious disease, bone/mineral, and endocrinology markers. Input samples are processed via disposable tips; reagent levels are tracked via RF-ID technology. The system uses magnetic particle solid-phase technology and isoluminol-antibody conjugates. A competitive sandwich immunoassay principle is employed: anti-HAV in samples competes with conjugate for fixed HAV; light signal (RLU) measured by photomultiplier tube is inversely proportional to anti-HAV concentration. The system features automated pipetting, incubation, wash cycles, and data reduction. Results are displayed via touch-screen interface for physician review to aid clinical diagnosis. Benefits include increased throughput, automated maintenance, and improved resource management compared to the predicate LIAISON® Analyzer.
Clinical Evidence
No clinical data provided; substantial equivalence is based on bench testing and performance comparisons between the LIAISON® XL and the predicate LIAISON® Analyzer.
Indicated for qualitative detection of total anti-HAV antibodies in human serum and sodium heparin plasma. Used as an aid in diagnosing current or previous HAV infection and assessing vaccine response. Not for blood/tissue donor screening. Performance not established for immunocompromised/immunosuppressed patients.
Regulatory Classification
Identification
HAV serological assays are devices that consist of antigens and antisera for the detection of hepatitis A virus-specific IgM, IgG, or total antibodies (IgM and IgG), in human serum or plasma. These devices are used for testing specimens from individuals who have signs and symptoms consistent with acute hepatitis to determine if an individual has been previously infected with HAV, or as an aid to identify HAV-susceptible individuals. The detection of these antibodies aids in the clinical laboratory diagnosis of an acute or past infection by HAV in conjunction with other clinical laboratory findings. These devices are not intended for screening blood or solid or soft tissue donors.
Special Controls
*Classification.* Class II (special controls). The special control is “Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Hepatitis A Virus Serological Assays.” See § 866.1(e) for the availability of this guidance document.
{0}------------------------------------------------
K103529
DiaSorin LIAISON® XL Analyzer Premarket Notification
### 5.0 510(k) SUMMARY
### SUBMITTED BY:
JAN 2 1 2011
Mari Mever Senior Manager, Regulatory/Clinical Affairs DiaSorin Inc. 1951 Northwestern Avenue P.O. Box 285 Stillwater, MN 55082-0285 Phone (651) 351-5635 Fax (651) 351-5669 Email: mari.mever@diasorin.com
## NAME OF DEVICE:
| Trade Name: | LIAISON® Anti-HAV, LIAISON® XL Analyzer |
|----------------------------|----------------------------------------------------------------------------------------|
| Common Names/Descriptions: | Hepatitis Anti-HAV, serological assay, Automated Chemiluminescent Immunoassay Analyzer |
| Regulation Number: | 21 CFR 866.3310 |
| Regulation Name: | Hepatitis A virus (HAV) serological assays |
| Regulatory Class: | Class II |
| Product Code: | LOL, JJF |
| PREDICATE DEVICES: | LIAISON® Analyzer Reference K082050 |
### DEVICE DESCRIPTION:
### INTENDED USE:
The LIAISON® Anti-HAV assay is an in vitro chemiluminescent immunoassay intended for the qualitative detection of total antibodies to hepatitis A (anti-HAV) in human serum and sodium heparin plasma samples using the LIAISON® Analyzer family. The assay is indicated as an aid in the laboratory diagnosis of current or previous HAV infections in conjunction with other serological and clinical information and to determine the presence of an antibody response to HAV in vaccine recipients.
This assay is not intended for screening blood or solid or soft tissue donors. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients. The user is responsible for establishing assay performance characteristics in these populations. Caution: U.S. Federal Law restricts this device to sale by or on the order of a physician.
{1}------------------------------------------------
The LIAISON® XL Analyzer is an automated discrete continuous loading chemiluminescent immunoassay (CLIA) analyzer for in vitro diagnostic analysis of CLIAs on human serum or plasma. The system menu includes infectious disease, bone and mineral, and endocrinology CLIAs. It is to be used only with FDA cleared chemiluminescence immunoassays that are marketed by DiaSorin for the LIAISON XL Analyzer.
The LIAISON® Control Anti-HAV (negative and positive) is intended for use as assayed quality control samples to monitor the performance of the LIAISON® Anti-HAV assay.
The performance characteristics of LIAISON® controls have not been established for any other assays or instrument platforms different from LIAISON® and LIAISON® XL.
## DESCRIPTION:
The method for qualitative determination of anti-HAV is a competitive sandwich chemiluminescence immunoassay (CLIA) based on neutralization. Antibodies to HAV (mouse monoclonal) are used for coating magnetic particles (solid phase) and are linked to an isoluminol derivative (isoluminol-antibody conjugate). During the first incubation, anti-HAV present in calibrators, samples or controls binds to a fixed and limited amount of HAV, thus forming an HAV-anti-HAV immune complex. During the second incubation, the antibody conjugate and the solid-phase antibody compete with anti-HAV present in the specimen for HAV, that allows the conjugate to bind to the solid phase and thus form a "sandwich" of conjugate antibody-antigen-solid phase antibody. If all HAV added is sequestered in an HAVanti-HAV immune complex during the first incubation, no sandwich is formed during the second incubation. After the second incubation, the unbound material is removed with a wash cycle.
Subsequently, the starter reagents are added and a flash chemiluminescence reaction is thus induced. The light signal, and hence the amount of isoluminolantibody conjugate, is measured by a photomultiplier as relative light units (RLU) and is inversely indicative of anti-HAV present in calibrators, samples or controls
The LIAISON® XL Analyzer is an automated discrete continuous loading chemiluminescent immunoassay analyzer for in vitro diagnostic use.
{2}------------------------------------------------
.
COMPARISON TO PREDICATE DEVICE:
The following table compares the LIAISON® XL Analyzer to the LIAISON® Analyzer.
| Feature | Summary of Device Similarities and Differences:<br>LIAISON® Analyzer | LIAISON® XL Analyzer |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA K# | K032844/K082050 | N/A |
| Intended Use | Automated chemiluminescent analyzer for clinical use | Same |
| Principles of operation | Chemiluminescence using magnetic particle solid phase and<br>chemiluminescent tracer | Same |
| Optical System | High-sensitive, low-noise<br>photomultiplier tube (PMT) operating<br>as an ultra-fast photon counter.<br>Pulses are amplified by a rapid<br>electronic amplifier. | Same |
| | Circuit that suppresses PMT signal noise. | Same |
| | Linear measuring range = 300 – 650 nm | Same |
| | Light peak of chemiluminescence emitted at 450 nm | Same |
| Temperature Control:<br>• Reaction<br>Temperature | $36°C±1°C$ | $36°C±1°C$ |
| • Reagent Storage<br>Temperature | 12-19°C | 11-15°C |
| Dispense System | Automated pipetting of samples and<br>reagents. Left pipetting unit used for<br>samples; right pipetting unit used for<br>reagents | same<br>(sample pipetting: disposable tip) |
| | Precision syringes (sample and reagent) | same |
| | Sample Probe:<br>• Liquid Level Detection<br>(capacitive)<br>• Clot Detection feature<br>(software algorithm based on<br>capacitive signal) | Sample Probe (disposable tip):<br>• Liquid Level Detection and<br>• Clot Detection feature<br>(pressure) |
| | N/A | Disposable tips: 6 trays of 96 tips<br>each can be loaded on board.<br>Monitored through software<br>counter and presence sensor<br>upon tip pick-up.<br>Reloading allowed during run |
| | Reagent Probes:<br>• Liquid Level Detection<br>(capacitive), with software<br>tracking of reagent level | Reagent Probes:<br>• Liquid Level Detection<br>(capacitive), with software<br>tracking of reagent level<br>• Optical Liquid Verification<br>(real-time monitoring of<br>liquid flow inside the<br>probe) |
| Feature | LIAISON® Analyzer | LIAISON® XL Analyzer |
| Sample Handling | Capacity: Holds 12 sample racks, 12<br>places per rack<br>Tube types:<br>- primary tube<br>- aliquot tube<br>- pediatric<br>Sample presence, sample type<br>(calibrator, control, patient), tube size,<br>and processing completion tracked by<br>operating software and sample<br>barcode | Capacity: Holds 10 sample racks,<br>12 places per rack<br>same<br>same |
| Reagent Handling | Capacity: 15 Reagent Integrals (RI)<br><br>RI contains all reagents required for<br>any given assay (up to 7 vials per RI,<br>first always contains magnetic<br>particles).<br>Assay-specific processing and<br>analysis parameters, calibration, lot<br>number, expiration date, and usage<br>(number of tests run) are controlled<br>by operating software as<br>communicated by barcode. | Capacity: 25 Reagent Integrals<br>(RI), plus 4 positions for Ancillary<br>Reagents<br>same<br>Assay-specific processing and<br>analysis parameters, calibration,<br>lot number, expiration date, and<br>usage (number of tests run) are<br>controlled by operating software<br>as communicated by RF-Tag (RF-ID). |
| Additional Reagents | • Control Set (2-3 levels)<br>• LIAISON Light Check (diagnostic<br>tool only)<br>• LIAISON Starter Kit (Starter<br>Reagents 1 and 2)<br>• LIAISON Wash/System Liquid<br>• LIAISON Cleaning Kit | • Control Set (2-3 levels)<br>• LIAISON Light Check<br>(diagnostic tool only)<br>• LIAISON Starter Kit (Starter<br>Reagents 1 and 2)<br>• LIAISON Wash/System<br>Liquid (used as a wash liquid<br>only - immunometric wash<br>step)<br>• LIAISON Cleaning Kit<br>In addition:<br>• Purified water is also required<br>as System Liquid, as fluidic<br>filler and to perform:<br>- reagent needle cleaning<br>- washer needle cleaning<br>• A cleaning tank is available to<br>host a cleaning liquid suitable<br>for automated maintenance<br>purpose |
| | Level sensing by capacitive rod | same |
| Reaction Modules | Capacity: 6 individual reaction<br>compartments per Reaction Module<br>Storage capacity: maximum 120<br>strips stored in LIAISON stacker | Single-cavity Cuvettes<br>Storage capacity: >600 Cuvettes |
| Feature | LIAISON® Analyzer | LIAISON® XL Analyzer |
| Reaction Modules (cont.) | Sensors detect presence of Reaction<br>Modules, and loading and occupancy<br>of stacker. | Inventory monitoring through<br>software counter.<br>Sensors detect actual presence<br>of Cuvettes |
| | Reloading allowed during run | same |
| Test Processing | Unloading automatic into waste bag<br>Random Access and Batch | same |
| | Continuous operation | same |
| | Sample scheduling optimized for<br>throughput | same |
| Assay Protocols | 1-Step assays: 1 incubation sequence<br>/ 1 wash sequence; average<br>incubation time = 10 minutes | same |
| | 2-Step assays: 2 incubation sequence<br>/ 1 or 2 wash sequence(s); average<br>incubation time = 10 minutes | same |
| | Two-point calibration of assays | same |
| Human Interface | Computer Mouse and keyboard | Touch-screen On Screen<br>Keyboard |
| | Monitor - touch screen, color Printer Stationary barcode scanners for<br>identification of samples and<br>reagents | same Printer (optional) Stationary barcode scanner<br>for identification of samples Stationary RF-Tag reader for<br>identification of reagents<br>(Reagent Integrals and<br>Starter Reagents) Handheld barcode scanner<br>for identification of controls |
| | Computer LIS Interface | same |
| Data Analysis | Automated data reduction | same |
| | Assay-specific Master Curve with 2-<br>point recalibration | same |
| | Assay-specific data reduction | same |
| QC Software | Stored lot-specific control results | same |
| | Lot-specific Levey-Jennings plotting | same |
| | Trend identification | same |
| | Statistical analyses | same |
| Specimens | Serum or plasma | same |
| | Sampling from primary, aliquot, or<br>pediatric tubes | same |
| Disposables | Reagent Integrals Light Check (diagnostic tool) Starter Kit Wash/System Liquid Reaction Modules | same |
| | N/A | Cuvettes Disposable Tips |
| | Waste Bag Cleaning Kit | same (dedicated) |
| Feature | LIAISON® Analyzer | LIAISON® XL Analyzer |
| Hardware Improvements | LIAISON® Analyzer | LIAISON® Analyzer, with enhancements as follows:<br>floor-standing, integrated design improved average throughput, Improved data exchange for Reagent Integrals, Ancillary Reagents, Starter Reagents via RF-ID technology Disposable tip for sample pipetting Continuous loading of all liquid and solid resources or waste Increased liquid/reagent capacity Increased efficiency of washer needle cleaning Automated Maintenance tasks On-line help: User Manual and Quick Guide directly accessible on screen |
## Summary of Device Similarities and Differences:
{3}------------------------------------------------
{4}------------------------------------------------
{5}------------------------------------------------
## CONCLUSION:
The material submitted in this premarket notification is complete and supports the substantial equivalence of the LIAISON® XL Analyzer to the predicate device.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird with its wings spread, symbolizing the department's mission to protect and promote the health and well-being of the nation.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
DiaSorin, Inc. Attn: Mari Meyer 1951 Northwestern Ave. P.O. Box 285 Stillwater, MN 55082
JAN 2 1 201
Re: K103529
Trade/Device Name: LIAISON Anti-HAV, LIAISON XL Analyzer Regulation Number: 21 CFR 866.3310 Regulation Name: Hepatitis A virus (HAV) serological assays Regulatory Class: Class II Product Code: LOL, JJF Dated: November 30, 2010 Received: December 1, 2010
Dear Ms. Meyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
Commission of the
Section of the country of the county
.
{7}------------------------------------------------
### Page 2 - Mari Meyer
CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Vacayartyr
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known): K103529
LIAISON® Anti-HAV, LIAISON® XL Analyzer Device Name:
Indications For Use:
The LIAISON® Anti-HAV assay is an in vitro chemiluminescent immunoassay intended for the qualitative detection of total antibodies to hepatitis A (anti-HAV) in human serum and sodium heparin plasma samples using the LIAISON® Analyzer family. The assay is indicated as an aid in the laboratory diagnosis of current or previous HAV infections in conjunction with other serological and clinical information and to determine the presence of an antibody response to HAV in vaccine recipients.
This assay is not intended for screening blood or solid or soft tissue donors. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients. The user is responsible for establishing assay performance characteristics in these populations. Caution: U.S. Federal Law restricts this device to sale by or on the order of a physician.
The LIAISON® XL Analyzer is an automated discrete continuous loading chemiluminescent immunoassay (CLIA) analyzer for in vitro diagnostic analysis of CLIAs on human serum or plasma. The system menu includes infectious disease, bone and mineral, and endocrinology CLIAs. It is to be used only with FDA cleared chemilyminescence immunoassays that are marketed by DiaSorin for the LIAISON XL Analyzer.
The LIAISON® Control Anti-HAV (negative and positive) is intended for use as assayed quality control samples to monitor the performance of the LIAISON® Anti-HAV assay.
The performance characteristics of LIAISON® controls have not been established for any other assays or instrument platforms different from LIAISON® and LIAISON® XL.
Prescription Use _ X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Uhe Gey
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) k 103529
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.