← Product Code [GCJ](/productcode/GCJ) · K103510

# UNIMAX SPECIMEN RETRIEVAL SYSTEM (K103510)

_Unimax Medical Systems, Inc. · GCJ · May 6, 2011 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K103510

## Device Facts

- **Applicant:** Unimax Medical Systems, Inc.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** May 6, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Unimax Specimen Retrieval System is indicated for use as a receptacle for the collection and extraction of tissue, organs and calculi during general and laparoscopic surgical procedures.

## Device Story

The Unimax Specimen Retrieval System is a sterile, single-use specimen container used during endoscopic surgery. It consists of a flexible polymer bag and an introducer structure designed to fit through a standard 10, 11, or 12mm trocar sheath. The device is used by surgeons to collect and extract tissue, organs, or calculi from the body cavity during laparoscopic procedures. The system is supplied in a dispensing tube for insertion. By providing a contained environment for specimen removal, the device facilitates safe extraction through the trocar port, minimizing the risk of tissue contamination or damage during retrieval.

## Clinical Evidence

Bench testing only. In vitro and in vivo preclinical physical, mechanical, and biocompatibility tests were performed to assess safety and effectiveness. Biocompatibility testing followed ISO 10993-1, 10993-10, and 10993-12 standards.

## Technological Characteristics

Sterile, single-use specimen retrieval system. Materials: Polymer bag, stainless steel components. Dimensions: Compatible with 10, 11, or 12mm trocar sheaths. Sterilization: Ethylene Oxide (EtO). Biocompatibility: Compliant with ISO 10993-1, 10993-10, and 10993-12.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Specimen Retrieval System ([K100959](/device/K100959.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K103510

MAY - 6 2011

Page 1 of 4

Unimax Medical Systems Inc. 510(k) Notification

Unimax Specimen Retrieval System

## 510(k) Summary

- Traditional Type of Submission: 5.1

| 5.2 | Submitter:                         | Unimax Medical Systems Inc.                             |
|-----|------------------------------------|---------------------------------------------------------|
|     | Address:                           | 8F-2, No. 127, Lane 235, Pao Chiao Rd., Hsin Tien City, |
|     |                                    | Taipei, Taiwan                                          |
|     | Phone:                             | 886-2-89191698                                          |
|     | Fax:                               | 886-2-89191528                                          |
|     | Contact:                           | Sophia Chiu                                             |
|     | Establishment Registration Number: | 3007791595                                              |

- Identification of the Device: 5.3 Proprietary/Trade name: Common Name: Classification Name: Device Classification: Regulation Number:
Panel:

Product Code:

Unimax Specimen Retrieval System Tissue Bags laparoscope, general & plastic surgery II 876.1500 General & Plastic Surgery GCJ

### 5.4 Identification of the Predicate Device:

| Predicate Device Name:                  | Specimen Retrieval System             |
|-----------------------------------------|---------------------------------------|
| Manufacturer:                           | Applied Medical Resources Corporation |
| 510(k) Number or Clearance Information: | K100959                               |

## 5.5 Intended Use and Indications for Use of the subject device.

The Unimax Specimen Retrieval System is indicated for use as a receptacle for the collection and extraction of tissue, organs and calculi during general and laparoscopic surgical procedures.

{1}------------------------------------------------

-K103510

Page 2 of (4)

Unimax Medical Systems Inc. 510(k) Notification

#### Unimax Specimen Retrieval System

#### 5.6 Device Description

The Unimax Specimen Retrieval System is a sterile and single-use specimen container designed for use in retrieving specimens during endoscopic surgery. The Unimax Specimen Retrieval System is supplied in a dispending tube for ease of insertion through a standard 10, 11 or 12mm trocar sheath.

#### 5.7 Non-clinical Testing

A series of in vitro and in vivo preclinical physical, mechanical and biocompatibility tests were performed to assess the safety and effectiveness of the Unimax Specimen Retrieval System. All the test results demonstrate the performance of Unimax Specimen Retrieval System meets the requirements of its pre-defined acceptance criteria and intended uses. The results of the non-clinical testing demonstrate that the Unimax Specimen Retrieval System is as safe and effective as the predicate devices.

#### 5.8 Safety and Effectiveness

The result of bench testing indicates that the new device is as safe and effective as the predicate device.

#### 5.9 Substantial Equivalence Determination

The Unimax Specimen Retrieval System submitted in this 510(k) file is substantially equivalent in intended use, design, principles of operation, materials and performance to the cleared Applied Medical Specimen Retrieval System which is the subject of K100959. Differences between the devices cited in this section do not raise any new issues of safety or effectiveness.

{2}------------------------------------------------

# K 103510

Page 3 of 4

Unimax Medical Systems Inc.
510(k) Notification

り・
 "t": " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " =

:""】

..............................................................................................................................................................................

Unimax Specimen Retrieval System

| Item             | Predicate Device<br>(Applied Medical Resources<br>Corporation Specimen Retrieval<br>System)                                                                                                                                                                                                                                                                                    | Proposed Device<br>(Unimax Medical Systems Inc<br>Specimen Retrieval System) |
|------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| Similarity       |                                                                                                                                                                                                                                                                                                                                                                                |                                                                              |
| Intended Use     | Indicated for use as a receptacle for<br>the collection and extraction of<br>tissue, organs and calculi during<br>general and laparoscopic surgical<br>procedures.                                                                                                                                                                                                             | Same                                                                         |
| Material         | Various Polymer<br>Stainless                                                                                                                                                                                                                                                                                                                                                   | Same                                                                         |
| Specification    | consists of a flexible polymer bag<br>and an introducer structure that fits<br>through a trocar port                                                                                                                                                                                                                                                                           | Same                                                                         |
| Biocompatibility | ISO 10993-1: 2003, Biological<br>evaluation of medical devices --<br>Part 1: Evaluation and testing<br>ISO 10993-10: 2002, Biological<br>evaluation of medical devices - Part<br>10: Tests for irritation and<br>delayed-type hypersensitivity<br>ISO 10993-12: 2007, Biological<br>evaluation of medical devices --<br>Part 12: Sample preparation and<br>reference materials | Same                                                                         |
| Difference       |                                                                                                                                                                                                                                                                                                                                                                                |                                                                              |
| Sterilization    | gamma irradiation, Sterility<br>Assurance Level will be 10-6                                                                                                                                                                                                                                                                                                                   | EtO Sterilization                                                            |

18

{3}------------------------------------------------

## K10 3510

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 4 96

Unimax Medical Systems Inc. 510(k) Notification

Unimax Specimen Retrieval System

#### 5.10 Conclusion

ं क

After analyzing bench tests, it can be concluded that Unimax Specimen Retrieval System is as safe and effective as the predicate device.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

MAY - 6 2011

Unimax Medical Systems, Inc. % Acmebiotechs Co., Ltd. Mr. Michael Lee No. 45, Minshen Road, Danshui Town Taipei County (Taiwan) 251 China

Re: K103510

Trade/Device Name: Unimax Specimen Retrieval System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: April 11, 2011 Received: April 11, 2011

Dear Mr. Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate · commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{5}------------------------------------------------

Page 2 - Mr. Michael Lee

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

My B. nh
for

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Unimax Medical Systems Inc. 510(k) Notification

Unimax Specimen Retrieval System

Indications for Use

510(k) Number (if known):

K103510

Device Name: Unimax Specimen Retrieval System

Indications for Use:

The Unimax Specimen Retrieval System is indicated for use as a receptacle for the collection and extraction of tissue, organs and calculi during general and laparoscopic surgical procedures.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ X (Part 21 CFR 801 Subpart D)

:

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil R.P. Ogden for mxm

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

Page 1 of 

510(k) Number K103510

7

---

**Source:** [https://fda-staging.innolitics.com/device/K103510](https://fda-staging.innolitics.com/device/K103510)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).