K103508 · Unimax Medical Systems, Inc. · GEI · Dec 20, 2011 · General, Plastic Surgery
Device Facts
Record ID
K103508
Device Name
UNMIAX LAPAROSCOPIC INSTRUMENT
Applicant
Unimax Medical Systems, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Dec 20, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Unimax Laparoscopic Instrument is a family of instruments which includes forceps, scissors, and probes which are intended to be used in general laparoscopic surgical procedures requiring the use of Monopolar electrosurgical cutting and/or coagulation.
Device Story
Unimax Laparoscopic Instrument family includes grasping forceps, monopolar scissors, and monopolar probe/electrodes. Devices are disposable, single-use, and individually packaged. Construction features insulated shafts with rotating tips and handles; male cautery connector attaches to standard monopolar cautery cables and generators. Probes/electrodes utilize thermally conductive metal tip electrodes. Operated by surgeons in laparoscopic surgical settings. Device functions by delivering monopolar electrosurgical energy to tissue for cutting or coagulation. Output is controlled by the surgeon via the generator interface. Benefits include precise tissue manipulation and electrosurgical control during minimally invasive procedures.
Clinical Evidence
Bench testing only. No clinical data presented. Testing included mechanical performance (drop, bending, pulling, torque, jaw clamping, blade sharpness) and electrical/thermal safety (arcing, charring, thermal spread) in accordance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2 standards. Biocompatibility testing performed per ISO 10993 series.
Technological Characteristics
Disposable, single-use laparoscopic instruments. Materials: biocompatible, medical-grade injection-molded plastic handles, insulated shafts, metal tip electrodes. Energy: Monopolar electrosurgical. Connectivity: standard monopolar cautery cable interface. Sterilization: Ethylene Oxide (EO). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, ISO 10993 (5, 7, 10, 11, 12), ISO 11135-1.
Indications for Use
Indicated for use in general laparoscopic surgical procedures requiring monopolar electrosurgical cutting and/or coagulation. Intended for prescription use.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the text "K103508" at the top, followed by a horizontal line. Below the line, the text "Pg 1 of 5" is circled. The text appears to be part of a document or page numbering system.
Unimax Laparoscopic Instrument
# 510(k) Summary
| 5.1 Type of Submission: | Traditional |
|-------------------------|-------------|
|-------------------------|-------------|
| | 5.2 Submitter: | Unimax Medical Systems Inc. |
|--|----------------|---------------------------------------------------------|
| | Address: | 8F-2, No. 127, Lane 235, Pao Chiao Rd., Hsin Tien City, |
| | | Taipei, Taiwan |
| | Phone: | 886-2-89191698 |
| | Fax: | 886-2-89191528 |
| | Contact: | Sophia Chiu |
| | | Establishment Registration Number: |
#### 5.3 Identification of the Device:
| Proprietary/Trade name: | Unimax Laparoscopic Instrument |
|-------------------------|------------------------------------------------------|
| Common Name: | Electrosurgical, Cutting & Coagulation & Accessories |
| Classification Name: | Electrosurgical, Cutting & Coagulation & Accessories |
| Device Classification: | II |
| Regulation Number: | 878.4400 |
| Panel: | General & Plastic Surgery |
| Product Code: | GEI |
### 5.4 Identification of the Predicate Device:
| Predicate Device Name: | AED Monopolar Lap Accessories |
|-----------------------------------------|-------------------------------------------|
| Manufacturer: | National Advanced Eedoscopy Devices, INC. |
| 510(k) Number or Clearance Information: | K081503 |
## 5.5 Intended Use and Indications for Use of the subject device.
The Unimax Laparoscopic Instrument is a family of instruments which includes forceps, scissors, and probes which are intended to be used in general laparoscopic surgical procedures requiring the use of Monopolar clectrosurgical cutting and/or coagulation.
{1}------------------------------------------------
K10 3508
pg 2 of 5
#### Unimax Laparoscopic Instrument
#### 5.6 Device Description
The Unimax Laparoscopic Instrument includes Grasping Forceps, Monopolar Scissors, and Monopolar Probe/Electrodes. The devices are disposable, single use, individually packaged devices that are composed of biocompatible materials. Scissors and forceps have a handle with rotating wheel attached to an insulated shaft with different tips, which allows the shaft and tip to rotate. They include a male cautery connector when attached to standard monopolar cautery cables and their generators. Probes/Electrodes have an insulated shaft with a thermally conductive metal tip electrode. The proximal end of the shaft is attached to a handle made of an injection molded, medical grade plastic.
#### 5.7 Non-clinical Testing
A series of in vitro and in vivo preclinical physical, mechanical and biocompatibility tests were performed to assess the safety and effectiveness of the Unimax Laparoscopic Instrument. The safety tests were conducted in accordance with IEC 60601-1 Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995, IEC 60601-1-2: 2001 Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic Compatibility -Requirements and Tests (Edition 2:2001 with Amendment 1:2004; Edition 2.1 (Edition 2:2001 consolidated with Amendment 1:2004)), IEC 60601-1-2:2007 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance -Collateral standard: Electromagnetic compatibility - Requirements and tests (Edition 3), ISO 10993-5 Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity, ISO 10993-7 Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals, ISO 10993-10 Biological evaluation of medical devices -Part 10: Tests for irritation and delayed-type hypersensitivity, ISO 10993-11 Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity, ISO 10993-12 Biological evaluation of medical devices -- Part 12: Sample preparation and reference materials, and ISO 11135-1 Sterilization of health care products - Ethylene oxide - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a handwritten note with the text "K10 3508" at the top, underlined. Below the line, there is a circled notation that reads "Pg 3 of 5". The writing appears to be done with a dark marker or pen, and the background is plain white.
Unimax Laparoscopic Instrument
The comparative performance testing was conducted on the subject device and the predicate device including the items listed below:
- Drop Testing
- · Bending Test
- Pulling Test
- · Torque Test
- · Jaw Clamping Test
- · Blade Sharpness Test
- Arcing Test
- · Charring Test
- · Thermal Spread Test
All the test results demonstrate the performance of Unimax Laparoscopic Instrument meets the requirements of its pre-defined acceptance criteria and intended uses.
The results of the non-clinical testing demonstrate that the Unimax Laparoscopic Instrument is as safe and effective as the predicate devices.
### 5.8 Safety and Effectiveness
The result of bench testing indicates that the new device is as safe and effective as the predicate device.
#### 5.9 Substantial Equivalence Determination
The Unimax Laparoscopic Instrument submitted in this 510(k) file is substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the cleared AED Monopolar Lap Accessories which is the subject of K081503. Differences between the devices cited in this section do not raise any new issues of safety or effectiveness.
{3}------------------------------------------------
K103508
---
Pg 4 of 5
.
### Unimax Laparoscopic Instrument
| Item | Proposed Device<br>(Unimax Medical Systems Inc.<br>Laparoscopic Instrument) | Predicate Device<br>(National Advanced Endoscopy<br>Devices, Inc. AED Monopolar Lap<br>Accessories) |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Unimax Laparoscopic Instrument is<br>a family of instruments which includes<br>forceps, scissors, and probes which are<br>intended to be used in general<br>laparoscopic surgical procedures<br>requiring the use of Monopolar<br>electrosurgical cutting and/or<br>coagulation. | AED Monopolar Lap Accessories are<br>reusable devices (forceps and<br>electrodes) intended to be used in<br>general laparoscopic surgical<br>procedures requiring the use of<br>electrosurgical cutting and/or<br>coagulation. |
| Consisted<br>Instruments | - Standard insulated monopolar handles<br>- Insulated Shafts<br>- Class I inserts (forceps, scissors)<br>- Electrodes | - Standard insulated monopolar handles<br>- Insulated Shafts<br>- Class I inserts (forceps, scissors)<br>- Probes/Electrodes |
| Models | Grasping Forceps:<br>- Straight grasping forceps<br>- Babcock grasping forceps<br>- Clinch grasping forceps<br>- Maryland dissecting forceps<br>- Rat tooth grasping forceps<br>- Duckbill grasping forceps<br>- Johan grasping forceps<br>Scissors<br>Probe/Electrode:<br>- Monopolar J hook probe<br>- Monopolar L hook probe<br>- Monopolar Spatula probe | Grasping Forceps:<br>- Graspers, with spoon<br>- Babcock Grasping Forceps<br>- Endo Clinch Forceps<br>- Maryland Forceps<br>- Cobra Style Toothed Grasper<br>- Duckbill Forceps<br>- Johan Grasping Forceps<br>Scissors<br>Probe/Electrode:<br>- J hook<br>- L hook |
| Dimension | 5mm/33cm<br>5mm/26cm<br>5mm/45cm | 5mm/33cm<br>5mm/26cm<br>5mm/45cm |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text "K103508" at the top, with a line drawn through it. Below the line, there is an oval encircling the text "pg 5 of 5". The text and markings appear to be handwritten, possibly indicating a document or page number within a larger set of materials.
Unimax Laparoscopic Instrument
| Sterilization | EO Sterility | EO Sterility |
|---------------|--------------------------|--------------------------|
| Safety | IEC 60601-1 | IEC 60601-1 |
| standards | IEC 60601-1-2 | IEC 60601-1-2 |
| | IEC 60601-2-2 | IEC 60601-2-2 |
| | ISO 10993-5 | ISO 10993-5 |
| | ISO 10993-10 | ISO 10993-10 |
| | ISO 10993-11 | ISO 10993-11 |
| | ISO 10993-12 | ISO 10993-12 |
| | ISO 10993-7 | ISO 14937 |
| | ISO 11135-1 | |
| Performance | No performance standards | No performance standards |
| standards | | |
| Compared | Drop Testing | Drop Testing |
| performance | Bending Test | Bending Test |
| testing | Pulling Test | Pulling Test |
| | Torque Test | Torque Test |
| | Jaw Clamping Test | Jaw Clamping Test |
| | Blade Sharpness Test | Blade Sharpness Test |
| | Arcing Test | Arcing Test |
| | Charring Test | Charring Test |
| | Thermal Spread Test | Thermal Spread Test |
# 5.10 Conclusion
After analyzing bench tests, electrical safety testing data, it can be concluded that Unimax Laparoscopic Instrument is as safe and effective as the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", indicating the department's name and national affiliation.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
DEC 20 2011
Unimax Medical Systems, Inc. % AcmeBiotechs Co., Ltd. Mr. Michael Lee No. 45, Minshen Rd. Danshui Town Taipei County China (Taiwan) 251
Re: K103508
Trade/Device Name: Unimax Laparoscopic Instrument Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: December 16, 2011 Received: December 16, 2011
Dear Mr. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{6}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
for Dlo. Dacus
Mark N. Mel Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
K103508
Unimax Laparoscopic Instrument
Indications for Use
510(k) Number (if known):
Device Name: Unimax Laparoscopic Instrument
Indications for Use:
The Unimax Laparoscopic Instrument is a family of instruments which includes forceps, scissors, and probes which are intended to be used in general laparoscopic surgical procedures requiring the use of Monopolar electrosurgical cutting and/or coagulation.
Prescription Use __ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use_ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
.510(k) Number K103508
Page 1 of __
5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.