The ALTUM Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine. It is to be used in skeletally mature patients as an adjunct to fusion of the cervical spine (C2 to C7). The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusion in patients with: - degenerative disc disease (as defined by neck pain of discogenic origin with . degeneration of the disc confirmed by patient history and radiographic studies), - spondylolisthesis, - trauma (i.e. fractures or dislocations), - tumors. . - deformity, . - pseudarthrosis. . - failed previous fusion, - spinal stenosis. .
Device Story
ALTUM Anterior Cervical Plate System provides anterior screw fixation for cervical spine stabilization. System comprises pre-lordosed titanium plates with vision ports for graft visualization and alignment notches for midline placement/temporary pin fixation. Self-tapping bone screws provide autogenic locking to plate upon insertion; no secondary locking mechanism required. Used by surgeons in clinical settings to stabilize spine during fusion development. Device facilitates precise placement and minimizes need for intraoperative plate contouring. Benefits include temporary stabilization to support bone fusion in patients with degenerative, traumatic, or neoplastic cervical conditions.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by non-clinical bench testing, specifically static compression bending, static torsion, and dynamic compression bending tests conducted per ASTM F1717-09.
Technological Characteristics
Material: Ti-6Al-4V (ASTM F136). Design: Pre-lordosed plate with vision ports and alignment notches. Fixation: Self-tapping bone screws with autogenic locking mechanism. Dimensions: Multiple lengths; screw diameters 4.0mm and 4.5mm. Sterilization: Non-sterile, intended for hospital sterilization.
Indications for Use
Indicated for skeletally mature patients requiring temporary stabilization of the anterior cervical spine (C2-C7) as an adjunct to fusion. Conditions include degenerative disc disease, spondylolisthesis, trauma (fractures/dislocations), tumors, deformity, pseudarthrosis, failed previous fusion, and spinal stenosis.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
Predicate Devices
Envision2 Anterior Cervical Plate System (K020649)
{0}------------------------------------------------
K103505
SpineCraft, LLC
510(k) Premarket Notification ALTUM Anterior Cervical Plate
MAR 1 5 2011
# 510(k) Summarv for the ALTUM Anterior Cervical Plate
In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following 510(k) summary is submitted for the ALTUM Anterior Cervical Plate
Date Prepared: 11/24/210
Submitter: SpineCraft. LLC 2215 Enterprise Drive Westchester, IL 60154 USA Tel: 1 708-531-9700. Fax: 1708- 531-9702
Establishment Registration No: Trade name: Common Name: Classification Name:
Product Code: Classification Panel: Contact Person: Ami Akallal-Asaad Director of Regulatory Affairs. SpineCraft, LLC a.asaad@spinecraft.com
> 3004717358 ALTUM Anterior Cervical Plate System Anterior Cervical Plate Spinal intervertebral body fixation orthosis Per 21 CFR 88.3060 KWQ Class II 87
#### Predicate or legally marketed devices which are substantially equivalent:
Envision2 Anterior Cervical Plate System (K020649) / Ortho Development Spider Cervical Plating System (K052292) / X-Spine Systems
#### Description of the device:
The ALTUM Anterior Cervical Plate System is intended for anterior screw fixation of the plate to the cervical spine. The fixation construct consists of a cervical plate that is attached to the vertebral body of the cervical spine with self-tapping bone screws using an anterior approach. The ALTUM anterior cervical plate, screws and instruments were designed taking into consideration the whole procedure. Plates are available in a variety of lengths, addressing multiple levels of fixation. The ALTUM plate incorporates vision ports that allow visualization of post-operative endplate/graft incorporation. Alignment notches on the cephalad and caudal ends of the plate facilitate precise midline placement and allow for Temporary Pin fixation of the plate. To accommodate normal cervical spine lordosis and, at the same time, minimize the need for additional plate contouring, the ALTUM Anterior Cervical Plate comes with a pre-machined lordotic curve. Bone screws are available for fixed angle implantation in a variety of lengths. Each screw head forms an autogenic lock to the plate upon insertion, requiring no additional locking mechanism.
#### Materials:
{1}------------------------------------------------
K103505
SpineCraft, LLC
## Ti-6Al-4V per ASTM F136
## Function:
The ALTUM Anterior Cervical Plate system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusion.
# Substantial equivalence claimed to predicate devices
ALTUM Anterior Cervical Plate system is substantially equivalent to the Envision2 Anterior Cervical Plate System (K971730) and Spider Cervical Plating System (K052292) in terms of intended use, design, materials used, mechanical safety and performances. The table below compares the features and characteristics of the ALTUM Anterior Cervical Plate system to these predicate devices.
| Device<br>Name<br>Items | ALTUM Anterior<br>Cervical Plate | Envision2 Anterior<br>Cervical Plate<br>System | Spider Cervical<br>Plating System |
|---------------------------------------------------|------------------------------------------------------------------------------|------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| Sponsor | SpineCraft | Ortho Development | X-Spine Systems |
| 510(k)<br>Number | Current submission | K020649 | K052292 |
| Device<br>Classification<br>Name/Product<br>Codes | Spinal intervertebral<br>body fixation orthosis<br>per 21 CFR 88.3060<br>KWQ | Spinal intervertebral<br>body fixation orthosis<br>per 21 CFR 88.3060<br>KWQ | Spinal intervertebral<br>body fixation orthosis<br>per 21 CFR 88.3060<br>KWQ |
| Classification | II | II | II |
| Material | Ti-6Al-4V per ASTM<br>F136 | Ti-6Al-4V per ASTM<br>F136 | Ti-6Al-4V per ASTM<br>F136 |
| # levels | multiple | multiple | multiple |
| Graft window | Yes | Yes | Yes |
| Variable angle<br>screw | Yes | Yes | Yes |
| Fixed screw | Yes | Yes | Yes |
| Screw size | Ø4.0, Ø4.5mm<br>variety of lengths | Ø4.0, Ø4.35mm<br>variety of lengths | Ø4.0, Ø4.25mm<br>variety of lengths |
| Pre-lordosed | Yes | Yes | Yes |
| Sterility | Non-sterile, sterilized<br>at hospital | Non-sterile, sterilized<br>at hospital | Non-sterile, sterilized<br>at hospital |
Figure (5)
{2}------------------------------------------------
K103505
510(k) Premarket Notification ALTUM Anterior Cervical Plate
## Intended Use:
The ALTUM Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine. It is to be used in skeletally mature patients as an adjunct to fusion of the cervical spine (C2 to C7). The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusion in patients with:
- degenerative disc disease (as defined by neck pain of discogenic origin with . degeneration of the disc confirmed by patient history and radiographic studies),
- spondylolisthesis, �
- trauma (i.e. fractures or dislocations), �
- tumors, .
- . deformity,
- pseudarthrosis, .
- . failed previous fusion,
- spinal stenosis. .
#### Non-clinical Test Summary:
The following tests were conducted:
ASTM F1717-09, "Standard Test Methods for Spinal Implant Constructs in a Vertebrectorny Model. Testing included Static Compression Bending Tests, Static Torsion Tests and Dynamic Compression Bending Tests. The results of this testing were compared to predicate systems, with the results being equal or higher than the predicate systems.
## Clinical Test Summary
No clinical studies were performed
#### Conclusions Nonclinical and Clinical
The ALTUM Anterior Cervical Plate System is substantially equivalent to the predicate devices in terms of indications for use, design, material, and function.
P.g 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle or other bird with its wings spread, facing to the right. The bird is drawn with simple, flowing lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
SpineCraft, LLC % Ms. Ami Akallal-Asaad Director of Requlatory Affairs 2215 Enterprise Drive, Suite 1504 Westchester, Illinois 60154
MAR 1 5 2011
Re: K103505
Trade/Device Name: ALTUM Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: K WQ Dated: February 28, 2011 Received: March 01, 2011
Dear Ms. Akallal-Asaad:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 – Ms. Akallal-Asaad
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (8:00) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Aling B. n.t
for
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# A. Indication for Use Statement
KIO3505 510(k) Number (if known): __ Device Name: ALTUM Anterior Cervical Plate System
# Indication for Use:
The ALTUM Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine. It is to be used in skeletally mature patients as an adjunct to fusion of the cervical spine (C2 to C7). The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusion in patients with:
- degenerative disc disease (as defined by neck pain of discogenic origin with . degeneration of the disc confirmed by patient history and radiographic studies),
- spondvlolisthesis. .
- trauma (i.e. fractures or dislocations), .
- tumors. .
- deformity, .
- pseudarthrosis.
- failed previous fusion.
- spinal stenosis.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
KID3505 510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.