K103478 · Olympus Medical Systems Corporation · FGE · Feb 24, 2011 · Gastroenterology, Urology
Device Facts
Record ID
K103478
Device Name
SINGLE USE BILIARY DRAINAGE STENT V
Applicant
Olympus Medical Systems Corporation
Product Code
FGE · Gastroenterology, Urology
Decision Date
Feb 24, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This instrument has been designed to be used with Olympus Endoscopes for endoscopic retrograde biliary drainage (ERBD). The stent is not intended to be permanently implanted in the patient. The stent is intended only for short-term implantation.
Device Story
Single-use biliary drainage stent system used during endoscopic retrograde biliary drainage (ERBD) procedures. System includes stent, insertion kit, and preloaded stent configurations. Operated by physicians using Olympus endoscopes; guidewire and pusher catheter used to place stent into duodenal papilla to maintain bile flow. Stent features side flap at proximal end for positioning. Device is for short-term implantation only. Benefits include maintenance of biliary drainage. No electronic or software components.
Clinical Evidence
Bench testing only. Testing included biocompatibility (cytotoxicity, sensitization, intracutaneous reaction, implantation, subacute/systemic toxicity, chemistry), stability (accelerated and real-time aging), and bench performance testing (flexibility, tensile strength, patency).
Technological Characteristics
Biliary catheter/stent. Materials: unspecified polymer/pigment. Dimensions/shape modified from predicate. Accessories: protection sleeve, straightener. Standards: ISO 14971:2007, ISO 11135-1:2007, ISO 10993-1:2003, ISO 10993-5:1999, ISO 10993-6:1994, ISO 10993-10:2002, ISO 10993-11:2006, ASTM F1980-07. Sterilization: Ethylene oxide (implied by ISO 11135-1).
Indications for Use
Indicated for endoscopic retrograde biliary drainage (ERBD) in patients requiring short-term biliary stenting. Not for permanent implantation.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
K103478
pg 1 of 3
FEB 2 4 2011
## 510(k) SUMMARY
PBD-1030/1031/1032/1033 Series, MAJ-1818/1819/1820/1821/1822 Series, PBD-V630P/V631P/V632P Series Single Use Biliary Drainage Stent V
February 23, 2011
# 1 General Information
| Applicant: | OLYMPUS MEDICAL SYSTEMS CORP.<br>2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507<br>Establishment Registration No: 8010047 |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Correspondent: | Stacy Abbatiello Kluesner, M.S., RAC<br>Regulatory Affairs & Quality Assurance<br>Olympus America Inc.<br>3500 Corporate Parkway<br>PO Box 610<br>Center Valley, PA 18034-0610, USA<br>Phone: 484-896-5405<br>FAX: 484-896-7128<br>Email: stacy.kluesner@olympus.com |
| Manufacturer: | Aomori Olympus Co., Ltd.<br>248-1 Okkonoki 2-chome Kuroishi-shi,<br>Aomori, Japan 036-0367<br>Establishment Registration No.: 9614641 |
**2 Device Identification**
| Device Trade Name: | PBD-1030/1031/1032/1033 Stent Series,<br>MAJ-1818/1819/1820/1821/1822 Insertion Kit Series,<br>PBD-V630P/V631P/V632P Preloaded Stent Series<br>Single Use Biliary Drainage Stent V |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Biliary catheter |
| Regulation Number: | 876.5010 |
| Regulation Name: | Biliary catheter and accessories |
| Regulatory Class: | II |
| Classification Panel: | Gastroenterology/Urology |
| Product Code: | FGE |
{1}------------------------------------------------
K103478
pg 2 of 3
### 3 Predicate Device Information
- Device Name: PBD-3Z through 7Z
- 배 OLYMPUS PBD STENTS Common Name:
- I Aomori Olympus Co., Ltd. Manufacturer: 248-1 Okkonoki 2-chome Kuroishi-shi. Aomori, Japan 036-0367 Establishment Registration No .: 9614641
- 510(k) No. #K933200
#### ব Device Description
The subject Single Use Billary Drainage Stent V PBD-1030/ 1031/ 1032/ 1033 Stent Series, MAJ-1818/ 1819/ 1820/ 1821/ 1822 Insertion Kit Series and PBD-V630P/ V631P/ V632P Preloaded Stent Series are designed to be used with Olympus endoscopes for Endoscopic Biliary Drainage(ERBD) procedure.
The subject Biliary Stent is placed into the duodenal papilla by using a Guidewire and Pusher Catheter to maintain the flow of bile by indwelling in the duodenal papilla. It is introduced to the biliary duct as follows:
- The user pushes the distal end of the guidewire into the papilla of Vater and 1) advance the guidewire to the target area.
- The user inserts the guide catheter, stent and pusher catheter over the guidewire. 2) The distal portion of the stent is inserted first.
- The user advances the stent until the side flap at its proximal end contacts the 3) papilla of Vater.
- The user withdraw the guide catheter from the pusher catheter to release the 4) stent.
- The user withdraw the pusher catheter from the endoscope. 5)
### Indications for Use റ
This instrument has been designed to be used with Olympus Endoscopes for endoscopic retrograde biliary drainage (ERBD). The stent is not intended to be permanently implanted in the patient. The stent is intended only for short-term implantation.
#### Comparison of Technological Characteristics 6
The subject Single Use Biliary Drainage Stent V PBD-1030/1031/1032/1033 Stent Series, MAJ-18181819/1820/1821/1822 Insertion Kit Series and PBD-V630P/V631P/V632P Preloaded Stent Series are basically identical to the predicate biliary stent PBD-62, included in MD-274 except as follows:
1) The stent material and a pigment have been changed.
- 2) The dimention and shape have been changed.
- 3) An accessory device, the Protection Sleeve and straightener have been added.
{2}------------------------------------------------
K103478
pg 3 of 3
### Summary of Non-Clinical Testing 7
The following non-clinical tests were performed:
- 1) Biocompatibility
- a) Cytotoxicity Test
- b) Sensitization Test
- c) Intracutaneous Reaction Test
- d) Implantation Test
- e) Subacute toxicity Test
- Systemic toxicity Test f)
- g) Chemistry special Test (60 day)
- h) Bile solution PH
## 2) Stability
Representative samples of devices were subjected to accelerated and real time ageing. The results of the accelerated age testing demonstrates that the device will be stable for the stated shelf-life. In addition, real time age testing will confirm the results of the accelerated age testing. The test devices were evaluated to ensure they continued to meet their specifications, including flexibility, tensile strength, tensile strength after stent placement, patency and compatibility with the endoscope.
3) Bench Testing
Bench testing was performed to compare the flexibility of the stent
## 4) Use of Standards
The following standards were used during the design and validation of the subject devices:
- a) ISO 14971, 2007
- b) ISO 11135-1, 2007
- C) ISO 10993-1, 2003
- d) ISO 10993-5, 1999
- ISO 10993-6, 1994 e)
- f) ISO 10993-10, 2002
- ISO 10993-11, 2006 g)
- h) ASTM F1980-07
The risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verifications tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment.
#### Conclusion 8
When compared to the predicate device, the PBD-1030/1031/1032/1033 Stent Series MAJ-1818/1819/1820/1821/1822 Insertion Kit Series and PBD-V630P/V631P/V632P Preloaded Stent Series do not incorporate any significant changes in intended use, method of operation, material, or design that could affect the safety or effectiveness of the device.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Olympus Medical Systems Corp. c/o Stacy Abbatiello Kluesner, M.S., RAC Regulatory Affairs Project Manager · Regulatory Affairs & Ouality Assurance Olympus America. Inc. 3500 Corporate Parkway, P.O. Box 610 CENTER VALLEY PA 18034-0610
FFR 2 4 2011
Re: K103478
Trade/Device Name: PBD-1030/1031/1032/1033 Stent Series, MAJ-1818/1819/1820/1821/1822 Insertion Kit Series, and PBD-V630P/V631P/V632P Preloaded Stent Series Single Use Biliary Drainage Stent V
Regulation Number: 21 CFR \$876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: FGE Dated: November 23, 2010 Received: November 26, 2010
Dear Ms. Kluesner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading,
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{4}------------------------------------------------
## Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Hubert H. Humphrey
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive. Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
K103478
Device Name:
PBD-1030/1031/1032/1033 Series, MAJ-1818/1819/1820/1821/1822 Series PBD-V630P/V631P/V632P Series
Indications For Use:
This instrument has been designed to be used with Olympus Endoscopes for endoscopic retrograde biliary drainage (ERBD). The stent is not intended to be permanently implanted in the patient. The stent is intended only for short-term implantation.
Prescription Use_ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use_ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
| 510(k) Number | K103478 |
|---------------|---------|
|---------------|---------|
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.