STOCKERT CENTRIFUGAL PUMP (SCP PLUS) (COMPLETE) OR (FOR S5/C5 SYSTEMS)
K103468 · Sorin Group Deutschland GmbH · DWA · Jan 20, 2011 · Cardiovascular
Device Facts
Record ID
K103468
Device Name
STOCKERT CENTRIFUGAL PUMP (SCP PLUS) (COMPLETE) OR (FOR S5/C5 SYSTEMS)
Applicant
Sorin Group Deutschland GmbH
Product Code
DWA · Cardiovascular
Decision Date
Jan 20, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4380
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Stöckert Centrifugal Pump System is a cardiopulmonary bypass speed control device (21CFR 870.4380) that is indicated for use with the Cobe Revolution® Pump Head/Dideco Synergy™/ECC.OTM for speed controlled pumping through a cardiopulmonary bypass circuit for typical durations of six hours or less. The SCP Plus is intended for use as a component part of or optional accessory to the S3 and S5 System (or any compatible system using the S5 firmware versions of 3.0. or greater) and the Stöckert Air Purge Control (APC) System. The SCP Plus has been qualified only for typical durations of six hours or less, appropriate to cardiopulmonary bypass procedures and has not been qualified through in vitro, in vivo, or clinical studies for long-term use as a bridge to transplant, pending recovery of the natural heart, or extracorporeal membrane oxygenation (ECMO) procedures.
Device Story
SCP Plus System acts as a speed control device for cardiopulmonary bypass centrifugal pumps. Input: electrical power and control signals from S3/S5 heart-lung machine systems. Operation: drive unit uses magnetic coupling to rotate the centrifugal pump head, recirculating blood within an extracorporeal circuit. Output: controlled pump speed. Used in surgical settings by perfusionists/clinicians. Integrates with S3/S5 systems and Air Purge Control (APC) system. Benefits: enables precise flow control during bypass procedures. Healthcare providers monitor pump performance via the integrated heart-lung machine interface to manage patient hemodynamics during surgery.
Clinical Evidence
No clinical testing performed. Bench testing only, including safety testing per IEC60601-1 and electromagnetic compatibility per IEC60601-1-2, alongside functional acceptance testing to verify performance criteria.
Technological Characteristics
Cardiopulmonary bypass speed control device. Components: drive unit, mounting receptacle, retaining key, splash protection, 3-joint mast holder. Magnetic coupling principle. Connectivity: integrates with S3/S5 systems (firmware 3.0+) and APC system. Standards: IEC60601-1, IEC60601-1-2.
Indications for Use
Indicated for speed-controlled pumping through a cardiopulmonary bypass circuit for durations of 6 hours or less. Not indicated for long-term use, bridge to transplant, or ECMO.
Regulatory Classification
Identification
A cardiopulmonary bypass pump speed control is a device used that incorporates an electrical system or a mechanical system, or both, and is used to control the speed of blood pumps used in cardiopulmonary bypass surgery.
Predicate Devices
Stöckert Centrifugal Pump (SCP) Plus System (K091008)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary for the Sorin Group Deutschland GmbH Stöckert Centrifugal Pump (SCP) Plus System
(21 CFR 807.92 and http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/default.html
## 1. SPONSOR/APPLICANT
Sorin Group Deutschland GmbH
Lindberghstrasse 25 D የብባንስ Mi
| D-80939 Munich | |
|-----------------|--------------------------|
| Germany | |
| Contact Person: | Renate Goebert |
| Telephone: | 011 +49 (0)89 323 01 153 |
Date Prepared: January 17, 2011
# 2. DEVICE NAME
| Proprietary Name: | Stöckert Centrifugal Pump (SCP) Plus System |
|----------------------|----------------------------------------------------------------------------------|
| Common/Usual Name: | Cardiopulmonary bypass centrifugal pump |
| Classification Name: | Cardiopulmonary bypass speed control device (21 CFR 870.4380: Product Code: DWA) |
# 3. PREDICATE DEVICE/S
- Stöckert Centrifugal Pump (SCP) Plus System K091008 ●
# 4. DEVICE DESCRIPTION
#### Physical description .
The SCP Plus System is intended for use as a component part of or optional accessory to the S3 and S5 System (or any compatible system using the S5 firmware versions of 3.0. or greater) and with the Stöckert Air Purge (APC) System. The System consists of the drive unit, a mounting receptacle for the pump head, a retaining key, splash protection, and a 3-joint mast holder with fast clamp connector.
#### How the device functions .
The Stöckert Centrifugal Pump (SCP) Plus System consists of hardware, firmware, and electronics that are used to drive the centrifugal pump for surgical procedures requiring cardiopulmonary bypass for typical durations of six (6) hours or less. The SCP drive unit uses magnetic coupling with the centrifugal pump.
{1}------------------------------------------------
#### Scientific concepts that form the basis for the device .
The SCP Plus System with the loaded single use and disposable centrifugal pump recirculates the contents of the extracorporeal circuit during cardiopulmonary bypass.
- Significant physical and performance characteristics of the device, such as . device design, material used, and physical properties The design, materials, and physical properties of the SCP Plus are unchanged from that already cleared by the FDA.
### 5. INTENDED USE/INDICATION FOR USE
The Stöckert Centrifugal Pump System is a cardiopulmonary bypass speed control device (21CFR 870.4380) that is indicated for use with the Cobe Revolution® Pump Head/Dideco Synergy™/ECC.OTM for speed controlled pumping through a cardiopulmonary bypass circuit for typical durations of six hours or less. The SCP Plus is intended for use as a component part of or optional accessory to the S3 and S5 System (or any compatible system using the \$5 firmware versions of 3.0. or greater) and the Stöckert Air Purge Control (APC) System.
The SCP Plus has been qualified only for typical durations of six hours or less, appropriate to cardiopulmonary bypass procedures and has not been qualified through in vitro, in vivo, or clinical studies for long-term use as a bridge to transplant, pending recovery of the natural heart, or extracorporeal membrane oxygenation (ECMO) procedures.
# 6. SUMMARY OF TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE DEVICE/S
The technological characteristics of the Stöckert Centrifugal Pump (SCP) Plus System described in this Traditional 510(k) Premarket Notification are identical to that reviewed by the FDA in K091008. They use the same hardware, firmware, and electronics as well as the same components including the drive unit, control panel, flow probe, emergency drive unit, flexible drive shaft, and connection cables.
# 7. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE
No comparative non-clinical testing served as the basis for substantial equivalence.
The Stöckert Centrifugal Pump (SCP) Plus System was tested in conjunction with the heart lung machine for safety in accordance with IEC60601-1 (with National
{2}------------------------------------------------
Deviations), for electromagnetic compatibility in accordance with IEC60601-1-2, and performance according to a formal prospectively defined functional acceptance test. Testing of the Stöckert Centrifugal Pump (SCP) Plus System (hardware, firmware, and performance) has demonstrated that the Stöckert Centrifugal Pump (SCP) Plus System fulfills prospectively defined performance criteria and that the System meets user needs.
### 8. SUMMARY OF CLINICAL TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE
Formal clinical testing of the Stöckert Centrifugal Pump (SCP) Plus System has not been performed. Therefore, this section does not apply.
#### 9. SUMMARY OF OTHER INFORMATION
No information other than that described was included in this 510(k).
### 10. CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL TESTS
Based on the non-clinical testing provided in this premarket notification, the Stöckert Centrifugal Pump (SCP) Plus System integrated with the Stöckert S5 performs in an identical manner as the System integrated with the Sorin C5 System, thus demonstrating that there are no differences and that the devices are substantially equivalent and perform in accordance with specifications and meets user needs.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract representation of a human figure or a bird in flight, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Sorin Group Deutschland GmbH c/o Ms. Rosina Robinson Medical Device Consultants, Inc. 49 Plain Street North Attleboro, MA 02760
JAN 2 0 201
Re: K103468
Trade/Device Name: Stöckert Centrifugal Pump (SCP) Plus System (SCP Plus) Regulation Number: 21 CFR 870.4380 Regulation Name: Cardiopulmonary bypass pump speed control Regulatory Class: II Product Code: DWA Dated: November 24, 2010 Received: November 24, 2010
Dear Ms. Robinson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the includions for use stated in the enclosure) to legally marketed predicate devices marked in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Dr to and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract lightlity warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 -- Ms. Rosina Robinson
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutlFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
una R. b. Auner
^Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known): K103468
Stöckert Centrifugal Pump (SCP) Plus System Device Name:
Indications for Use:
The Stockert Centrifugal Pump System is a cardiopulmonary bypass speed control device (21CFR 870.4380) that is indicated for use with the Cobe Revolution® Pump Head/Dideco Synergy™/ECC.OTM for for speed controlled pumping through ଥ cardiopulmonary bypass circuit for typical durations of six hours or less. The SCP Plus is intended for use as a component part of or optional accessory to the S3 and S5 System (or any compatible system using the SS firmware versions of 3.0. or greater) and the Stöckert Air Purge Control (APC) System.
The SCP Plus has been qualified only for typical durations of six hours or less, appropriate to cardiopulmonary bypass procedures and has not been qualified through in vitro, in vivo, or clinical studies for long-term use as a bridge to transplant, pending recovery of the natural heart, or extracorporeal membrane oxygenation (ECMO) procedures.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
uma R. V. hner
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number_K103468
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.