DESARA MINI

K103418 · Caldera Medical, Inc. · PAH · Nov 30, 2011 · General, Plastic Surgery

Device Facts

Record IDK103418
Device NameDESARA MINI
ApplicantCaldera Medical, Inc.
Product CodePAH · General, Plastic Surgery
Decision DateNov 30, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Desara Mini is intended to be used in females to position a suburethral mesh sling for treatment of Genuine Stress Urinary Incontinence (SUI) and mixed incontinence resulting from urethral hypermobility. The mesh is placed transvaginally using a reusable introducer (sold separately)

Device Story

Desara Mini is a sterile, single-use pubourethral sling designed for transvaginal placement to treat female urinary incontinence. The device consists of a monofilament polypropylene mesh, two self-fixating tips, and two guiding suture loops; a visible midline marker assists in surgical placement. It is used in a clinical setting by a surgeon, requiring a separate reusable introducer for deployment. The device functions by positioning the mesh to support the urethra, addressing urethral hypermobility. It provides a mechanical support structure to restore continence. The device is a modification of existing sling technologies, combining the mesh material of the Desara Mesh with the self-fixating tip design of the MiniArc.

Clinical Evidence

No clinical data. Evidence consists of bench testing, cadaver lab testing, and validation testing. Bench testing assessed mesh thickness, weave characteristics, pore size, density, tensile strength, stiffness, flexural rigidity, tear resistance, burst strength, and suture pullout. Biocompatibility testing followed ISO-10993 standards. Sterility and shelf-life testing followed ASTM F-1980-07. Results demonstrated performance equivalent to predicate devices.

Technological Characteristics

Monofilament polypropylene mesh; self-fixating tips; guiding suture loops; midline marker. Sterile, single-use. Mechanical fixation principle. Biocompatibility per ISO-10993. Accelerated aging per ASTM F-1980-07.

Indications for Use

Indicated for female patients with Genuine Stress Urinary Incontinence (SUI) or mixed incontinence caused by urethral hypermobility.

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for Caldera Medical. The logo consists of a circular graphic on the left, followed by the text "CALDERA MEDICAL" in bold, sans-serif font. The graphic appears to be an abstract representation of a biological or medical element. ## 510(k) Summary Date of Summary: July 12, 2011 Submitted by: Submitter: Address: Contact: Phone: Caldera Medical, Inc. 5171 Clareton Drive Agoura Hills, CA 91301 Vicki Gail, Manager QA/RA (818) 879-6555 x 102 Device Name: | Device Name: | Surgical Mesh (878.3300) | |--------------------|--------------------------------------------------------------------------------------------| | Trade Name: | Desara Mini | | Common Name: | Surgical Mesh | | Device Class: | Class II, Product Code FTL, 21 CFR 878.3300, Surgical Mesh,<br>Plastic and General Surgery | | Predicate Devices: | Desara mesh, #K101169<br>MiniArc, #K073703 | ### Description of Device: The Desara Mini is a sterile, single-use pubourethral sling. The device is comprised of monofilament polypropylene yarn, which is knitted into a mesh, two self-fixating tips and two guiding suture loops at each end of the device. The device contains a visible midline marker to assist the surgeon in placement of the device. The Desara Mini is available in one configuration and a reusable introducer is available separately. ### Intended Use of Device: Desara Mini is intended to be used in females to position a mesh for treatment of Genuine Stress Urinary Incontinence (SUI) and mixed incontinence resulting from urethral hypermobility. The mesh is placed transvaginally using a reusable introducer (sold separately) Technological Characteristics The Desara Mini utilizes a similar concept, mesh with self-fixating tips, as that of the predicate device, MiniArc. The Desara Mini is comprised of the same mesh and suture as the predicate device, Desara Mesh. Self-fixating tips are not contained in the predicate device, Desara Mesh. The Desara Mini is a modification of the predicate devices, MiniArc, and Desara Mesh with the same intended use and does not change the fundamental scientific technology of the predicate device. Confidential {1}------------------------------------------------ K103418 PAGE 2/2. # CALDERA MEDICAL ### Performance Data Summary In accordance with the FDA's Guidance for Industry and FDA Staff "Formal for Traditional and Abbreviated 510/k)'s the results of bench, cadaver lab and validation testing has shown the Desara Mini to be substantially equivalent to the predicate devices in its intended use, technological characteristics and performance. The mesh component of the Desara Mini is comprised of the same mesh as the predicate device. Desara mesh. FDA 510(k) #K110169, also a product of Caldera Medical. In accordance with the FDA's Guidance for the Preparation of a Premarket Notification Application for Surgical Mesh, the following mesh characteristics were assessed: mesh thickness, mesh weave characteristics, pore size, mesh density, tensile strength, mesh stiffness, flexural rigidity, tear resistance, burst strength, suture pullout and pyrogen levels. The Desara Mini mesh demonstrates substantial equivalence to the predicate device, Desara Mesh. Desara Mini has passed all biocompatibility testing as indicated per the FDA quidance documents, FDA Device Advice, Guidance Documents (Medical Device and Radiation Emitting Product), 3. Biocompatibility and FDA Blue Book Memorandum #G95-1 Entitled "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing". The Desara Mini mesh demonstrates substantial equivalence to the predicate device, Desara Mesh. In accordance with the FDA Guidance, Updated 510(k) Sterility Review Guidance K90-1; Final Guidance for Industry, FDA, FDA Device Advice, Guidance Documents (Medical Device and Radiation Emitting Product), 4. Labeling and FDA Consensus standard, ASTM F-1980-07, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices. Desara Mini has passed all testing requirements in terms of aging. shelf life, transportation and sterilization and has demonstrated substantial equivalence to the predicate device. Desara Mesh. Results of mechanical bench and functional cadaver testing demonstrate equivalent junction strength, device function and device performance (fixation tips, polypropylene mesh, suture loops and introducer) based upon its intended use to the predicate device, MiniArc. The performance of the Desara Mini demonstrates that the device is as safe, and as effective, and performs at least as safely and effectively as the predicate devices, Desara Mesh and MiniArc. Summary of Substantial Equivalence The Desara Mini sling is safe and effective for its intended use and is substantially equivalent to the predicate devices Desara Mesh, also a product of Caldera Medical, Inc. and MiniArc, a product of American Medical Systems, Inc. Confidential {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo features a stylized eagle with its wings spread, symbolizing strength and freedom. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle, indicating the organization's name and location. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Ms. Vicki Gail Manager, QA/RA Caldera Medical, Inc. 5171 Clareton Drive AGOURA HILLS CA 91301 SEP 2 8 2012 Re: K103418 Trade/Device Name: Desara Mini Regulation Number: 21 CFR& 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: PAH Dated: November 14, 2011 Received: November 18, 2011 Dear Ms. Gail: This letter corrects our substantially equivalent letter of November 30, 2011. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {3}------------------------------------------------ Page 2 - comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Benjamin R. Perkins Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Caldera Medical. The logo consists of a stylized image to the left of the company name. The company name is written in all capital letters. ## Indications For Use 510 (k) Number (if known): K103418 Device Name: Desara Mini Indications for Use: Desara Mini is intended to be used in females to position a suburethral mesh sling for treatment of Genuine Stress Urinary Incontinence (SUI) and mixed incontinence resulting from urethral hypermobility. The mesh is placed transvaginally using a reusable introducer (sold separately) | Prescription Use | --X-- | AND/OR | Over the Counter | |-----------------------------|-------|--------|------------------| | Use | | | | | (Part 21 CFR 801 Subpart D) | | | (Part 21 CFR 807 | | Subpart C) | | | | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices 510(k) Number: K103418 Desara® Mini Caldera Medical, Inc. Confidential
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...