CHAD THERAPEUTIC EVOLUTION ELECTRONIC OXYGEN CONSERVER

K103392 · Inovo, Inc. · NFB · Dec 17, 2010 · Anesthesiology

Device Facts

Record IDK103392
Device NameCHAD THERAPEUTIC EVOLUTION ELECTRONIC OXYGEN CONSERVER
ApplicantInovo, Inc.
Product CodeNFB · Anesthesiology
Decision DateDec 17, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 868.5905
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Inovo Evolution is intended for prescription use only, to be used as part of a portable oxygen delivery system for patients that require supplemental oxygen up to 7 liters per minute, in their home and for ambulatory use.

Device Story

Inovo Evolution OM-900 is a microprocessor-controlled oxygen conserver and pressure regulator for portable oxygen systems. Device connects to oxygen cylinders; internal regulator reduces pressure for conserver operation. Breath detection circuitry and inhalation sensors monitor breathing cycle to trigger delivery of precise oxygen boluses. Output is equivalent to 1 to 7 liters per minute flow rate. Used in home and ambulatory settings by patients requiring supplemental oxygen. Healthcare providers use device to manage oxygen delivery; bolus delivery optimizes oxygen usage compared to continuous flow. Benefits include extended oxygen supply duration and improved patient mobility.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Microprocessor-controlled oxygen conserver and pressure regulator. Sensing via breath detection circuitry and inhalation sensors. Connectivity: None. Power: Battery-operated. Form factor: Portable oxygen delivery system component.

Indications for Use

Indicated for patients requiring supplemental oxygen up to 7 liters per minute for home and ambulatory use. Prescription use only.

Regulatory Classification

Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K103392 DEC 1 7 2010 ### 510(k) Summary Inovo, Inc. Date Prepared: November 15, 2010 Submitter Information: Official Contact: Phone: FAX: E-mail: Proprietary Names: Common Name: Inovo Model Number Classification Name: lnovo, Inc. 2975 S. Horseshoe Dr. Naples Fl. 34104 Michael T. Dildine Director, Quality Assurance (239) 643-6577 (239) 643-6530 mdildine@inovoinc.com Chad Therapeutic Evolution Electronic Oxygen Conserver Oxygen Conserver OM-900 Series Class II, 21 CFR 868.5905 Non Continuous Ventilator Product Code: ﮯ ﺑﮩ NFB Predicate Device Equivalence: K042142 - Chad Therapeutic Lotus Models OM-700 & OM-700S {1}------------------------------------------------ ### Device Description: The Inovo Evolution is a microprocessor-controlled device, which is a combination of a oxygen pressure regulator and a oxygen conserver, designed for use with ambulatory oxygen systems. The built in oxygen regulator reduces the oxygen pressure from the oxygen cylinder to ensure proper operation of the oxygen conserving device. The low pressure oxygen enters the conserver portion of the device where the breath detection circuitry and inhalation sensors convert the low pressure oxygen to deliver a precise amount of supplemental oxygen at a specific point in the breathing cycle. It delivers boluses of oxygen that is equivalent to 1 to 7 liters per minute depending on the flow rate setting ### Intended Use: The Inovo Evolution is intended for prescription use only, to be used as part of a portable oxygen delivery system for patients that require supplemental oxygen up to 7 liters per minute, in their home and for ambulatory use. # Comparison of Device Technological Characteristics to Predicate Devices: The submitted Inovo Evolution OM-900 Series has the following similarities to the predicate OM-700 Series - Has the same intended use . - Incorporates the same basic modes and settings . - Incorporates similar materials ● - Oxygen delivery method is fundamentally equivalent . There is a technological characteristic difference between the Evolution and the predicate device. The breath detection circuitry and sensors utilized in the Evolution are new but are based on the same operating principles of the Lotus breath detection circuitry and sensor. The bolus sizes for the Evolution are the same with the exception of one additional setting No. 7. The software and electronics have been modified to extend battery life {2}------------------------------------------------ ## Statement of Safety and Effectiveness Analysis of comparison of design, function and features of the Inovo Evolution OM-900 Series to the (K042142) Lotus OM-700, OM700S Series Electronic Conserver, together with the results of testing demonstrates the device to be substantially equivalent to the predicate device in terms of meeting performances criteria and functioning as intended. ## Conclusion Based on the above, we conclude that the Inovo Evolution OM-900 Series Electronic Conserver is substantially equivalent to the predicate device listed and does not raise any new issues of safety and effectiveness. . {3}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows a logo for the Department of Health & Human Services - USA. The logo features a stylized symbol to the right of the text. The symbol consists of three curved lines that resemble a stylized human figure. The text is arranged in a circular pattern around the symbol. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 # DEC 1 7 2010 Mr. Michael T. Dildine Director, Quality Assurance Inovo. Incorporated 2975 S. Horseshoe Drive, Suite 600 Naples, Florida 34104 Re: K103392 Trade/Device Name: Chad Therapeutics Evolution Model OM-900 Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: NFB Dated: November 15, 2010 Received: November 19, 2010 Dear Mr. Dildine: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ ### Page 2- Mr. Dildine Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours for Anthony D. Watson, B.S., M.S., M.B.A. Director . Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health ### Enclosure {5}------------------------------------------------ # Indications for Use ### 510(k) Number (if known): Device Name: Chad Therapeutics Evolution Model OM-900 ### Indications for Use: The Chad Therapeutics Evolution Model OM-900 is intended for prescription use only, to be used as part of a portable oxygen delivery system for patients that require supplemental oxygen up to 7 liters per minute, in their home and for ambulatory use × Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of OORH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices Page 1 of *_*_ 510(k) Number: K103392
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...