← Product Code [DIH](/productcode/DIH) · K103373

# SYNCHRON SYSTEMS CHOLINESTERASE (CHEX) MODEL A53722 (K103373)

_Beckman Coulter, Inc. · DIH · Feb 7, 2011 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K103373

## Device Facts

- **Applicant:** Beckman Coulter, Inc.
- **Product Code:** [DIH](/productcode/DIH.md)
- **Decision Date:** Feb 7, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3240
- **Device Class:** Class 1
- **Review Panel:** Clinical Toxicology

## Indications for Use

System reagent for the quantitative determination of Cholinesterase activity (E.C. 3.1.1.8) in human serum and plasma on Beckman Coulter Synchron CX® Pro System(s), Synchron LX® Pro System(s) and UniCel® DxC 600/800 System(s). Measurements obtained by this device are used in the diagnosis and treatment of cholinesterase inhibition disorders (e.g., insecticide poisoning and succinylcholine poisoning).

## Device Story

In vitro diagnostic reagent for quantitative measurement of cholinesterase activity in human serum/plasma; used on Beckman Coulter Synchron CX Pro, LX Pro, and UniCel DxC 600/800 systems. Principle: enzymatic hydrolysis of butyrylthiocholine to butyrate and thiocholine; thiocholine reduces yellow hexacyanoferrate (III) to colorless hexacyanoferrate (II). Spectrophotometric measurement of absorbance decrease at 410 nm proportional to enzyme activity. Operated by laboratory technicians in clinical settings. Output provides quantitative enzyme activity levels to clinicians for diagnosing/managing cholinesterase inhibition disorders like insecticide or succinylcholine poisoning.

## Clinical Evidence

Bench testing only. Equivalence demonstrated through method comparison, stability, linearity, and imprecision experiments. No clinical data required.

## Technological Characteristics

Liquid stable, ready-to-use reagent. Principle: enzymatic hydrolysis/spectrophotometric detection. Compatible with Beckman Coulter Synchron/UniCel systems. Performance specifications include interference thresholds: Hemoglobin 1000 mg/dL, Bilirubin 60 mg/dL, Lipemia 2000 mg/dL, Ascorbic Acid 30 mg/dL.

## Regulatory Identification

A cholinesterase test system is a device intended to measure cholinesterase (an enzyme that catalyzes the hydrolysis of acetylcholine to choline) in human specimens. There are two principal types of cholinesterase in human tissues. True cholinesterase is present at nerve endings and in erythrocytes (red blood cells) but is not present in plasma. Pseudo cholinesterase is present in plasma and liver but is not present in erythrocytes. Measurements obtained by this device are used in the diagnosis and treatment of cholinesterase inhibition disorders (e.g., insecticide poisoning and succinylcholine poisoning).

## Predicate Devices

- Cholinesterase (CHE) ([K030045](/device/K030045.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K103373
FEB -7 2011

## Summary of Safety & Effectiveness SYNCHRON® Systems Cholinesterase (CHEX) Reagent

This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92.

#### 1.0 Submitted By:

Yvette Lloyd, JD Staff Regulatory Affairs Specialist Beckman Coulter, Inc. 250 S. Kraemer Blvd Mail Stop: E2.SE.08 Brea, CA 92821 Phone: (714) 961-3626 FAX: (714) 961-4234 email: vrlloyd@beckman.com

#### 2.0 Date Submitted:

November 16, 2010

#### 3.0 Device Name(s):

### 3.1 Proprietary Names SYNCHRON® Systems Cholinesterase (CHEX) Reagent

#### 3.2 Classification Name

- -- Cholinesterase test system, (Product Code -- DIH; 21 CFR & 862.3240) --

#### 4.0 Predicate Device:

····

| Candidate(s)                                          | Predicate            | Manufacturer                | Docket<br>Number |
|-------------------------------------------------------|----------------------|-----------------------------|------------------|
| SYNCHRON® Systems<br>Cholinesterase (CHEX)<br>Reagent | Cholinesterase (CHE) | OLYMPUS<br>AMERICA,<br>INC. | K030045          |

#### 5.0 Description:

System reagent for the quantitative determination of Cholinesterase activity (E.C. 3.1.1.8)1 in human serum and plasma on Beckman Coulter Synchron CX® Pro System(s), Synchron LX® Pro System(s) and UniCel® DxC 600/800 System(s).

A cholinesterase test system is a device intended to measure cholinesterase (an enzyme that catalyzes the hydrolysis of acetylcholine to choline) in human specimens. There are two principal types of cholinesterase in human tissues. True cholinesterase is present at nerve endings and in erythrocytes (red blood cells) but is not present in plasma. Pseudo cholinesterase is present in plasma and liver but is not present in erythrocytes. Measurements obtained by this device are used in the diagnosis and treatment of cholinesterase inhibition disorders (e.g., insecticide poisoning and succinylcholine poisoning).

۔ ﻧﮧ ﺑﮩﺖ ۔۔۔۔ ﺏ ، Beckman Coulter, Inc., Section 510(k) Notification SYNCHRON® Systems Cholinesterase (CHEX) Reagent SYN CHEX 510K Summary_02022011.doc____________________________________________________________________________________________________________________________________________ 2010

. . . . . . . . . . . . . . . . . . . . . . . .

D
5

{1}------------------------------------------------

The method is based on the recommendations of the German Society for Clinical Chemisty (Deutsche Gesellschaft für Klinische Chemie, DGKCh).

Cholinesterase catalyses the hydrolysis of butyrytthiocholine to butyrate and thiocholine. Thiocholine reduces yellow hexacyanoferrate (III) to colorless hexacvanoferrate (II). The decrease in absorbance at 410 nm is directly proportional to the cholinesterase activity in the sample.

#### 6.0 Intended Use:

System reagent for the quantitative determination of Cholinesterase activity (E.C. 3.1.1.8) in human serum and plasma on Beckman Coulter Synchron CX® Pro System(s), Synchron LX® Pro System(s) and UniCel® DxC 600/800 System(s).

Measurements obtained by this device are used in the diagnosis and treatment of cholinesterase inhibition disorders (e.g., insecticide poisoning and succinylcholine poisoning).

#### 7.0 Comparison to Predicate(s):

The following table shows similarities and differences between the predicate identified in Section 4.0 of this summary.

| Predicate Device: Olympus Cholinesterase""<br>(CHE) Assay                           | Current device: SYNCHRON® Systems<br>Cholinesterase (CHEX) Reagent |
|-------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| Intended Use                                                                        | same                                                               |
| Methodology                                                                         | same                                                               |
| Reagent Constituents'                                                               | same                                                               |
| Fundamental Technology: chromogenic colored<br>change measured by spectrophotometer | same                                                               |
| Analytic Interval                                                                   | same                                                               |
| Expected Values                                                                     | same                                                               |
| Sensitivity                                                                         | same                                                               |
| Within run precision                                                                | same                                                               |
| Total precision                                                                     | same                                                               |
| LIQUID stable reagents (ready to use)                                               | same                                                               |
| On board, open reagent stability                                                    | same                                                               |

| List of design inputs that are similar between the two devices |  |
|----------------------------------------------------------------|--|
|----------------------------------------------------------------|--|

Image /page/1/Figure/9 description: The image contains some text, but it is difficult to read due to the low resolution and poor contrast. The text appears to be in a small font size and is spread out across the image. It is difficult to determine the content of the text without further clarification.

D
5

16-Nov-

Beckman Coulter, Inc., Section 510(k) Notification SYNCHRON® Systems Cholinesterase (CHEX) Reagent SYN CHEX 510K Summary_02022011.doc____________________________________________________________________________________________________________________________________________ 2010

{2}------------------------------------------------

|                           | List of design inputs that are different between the two devices |                                                                    |
|---------------------------|------------------------------------------------------------------|--------------------------------------------------------------------|
|                           | Predicate Device: Olympus Cholinesterase<br>(CHE) Assay          | Current device: SYNCHRON® Systems<br>Cholinesterase (CHEX) Reagent |
| System use                | AU400/400e/480, AU600/640/640e and<br>AU2700/5400                | CX5 PRO, CX9 PRO, LX20 PRO, DxC600,<br>DxC 800                     |
| Sample type               | Serum, Plasma (Li Heparin, Na Heparin,<br>EDTA)                  | Serum, Plasma (Li Heparin, Na Heparin)                             |
| Interfering<br>substances | Hemoglobin: 500 mg/dL                                            | Hemoglobin: 1000 mg/dL                                             |
|                           | Bilirubin: 40 mg/dL                                              | Bilirubin: 60 mg/dL                                                |
|                           | Lipemia: ≤ 1000 mg/dL (Intralipid)                               | Lipemia: 2000 mg/dL (triglyceride)                                 |
|                           | Ascorbic Acid: 20 mg/dL                                          | Ascorbic Acid: 30 mg/dL                                            |

List of design inputs that are different between the two devices

... .. . . . . . . . . . . . . . . .

.

#### 8.0 Summary of Performance Data:

The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to chemistry test systems already in commercial distribution. Equivalence is demonstrated through method comparison, stability, linearity, and imprecision experiments. ,

Image /page/2/Figure/4 description: The image shows a document with some text and markings. The text includes "Beckman Coulter, Inc., Section 510(k) Notification". There are also some lines and symbols scattered throughout the document. The image appears to be a scan or photocopy of an original document.

SYNCHRON® Systems Cholinesterase (CHEX) Reagent SYN CHEX 510K Summary_02022011.doc ___________________________________________________________________________________________________________________________________________ 2010

D
5

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a bird or a human figure with outstretched arms.

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

FEB 07 2011

Beckman Coulter, Inc. C/O Ms. Y vette Lloyd 250 S. Kraemer Blvd. Brea, CA 92821

Re: k103373

Trade/Device Name: SYNCHRON Systems Cholinesterase (CHEX) assay Regulation Number: 21 CFR 862.3240 Regulation Name: Cholinesterase test system Regulatory Class: Class I Reserved Product Code: DIH Dated: November 16, 2010 Received: November 17, 2010

Dear Ms. Lloyd:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

Comments of Children

{4}------------------------------------------------

Page 2 -

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Viro Diagnostic Device Evaluation and Safety at (01) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premy aket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-fire mober (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrlv/industry/support/index.html.

Sincerely yours,

G.C.

Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use Form

510(k) Number (if known): K103373

Device Name: _________________________________________________________________________________________________________________________________________________________________

CHEX (Cholinesterase)

・

Indications for Use: of and specificance the result on your compression in the count

System reagent for the quantitative determination of Cholinesterase activity (E.C. 3.1.1.8) in human serum and plasma on Beckman Coulter Synchron CX® Pro System(s), Synchron LX® Pro System(s) and UniCel® DxC 600/800 System(s).

Measurements obtained by this device are used in the diagnosis and treatment of cholinesterase inhibition disorders (e.g., insecticide poisoning and succinylcholine poisoning).

| Prescription Use            | X      |                        |  |
|-----------------------------|--------|------------------------|--|
| (Part 21 CFR 801 Subpart D) | AND/OR | Over-The-Counter Use   |  |
|                             |        | (21 CFR 801 Subpart C) |  |

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

satilitany di be and the more and address, works were

Division Sign-Off of course and Unit of " raxCha Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K103373

Page 1 of _1_

( incle - in: 1. suith + どではない -

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

- - BCDRH, Office of In Varo Dinghostic Dessee (40)(VD

. . . . . . . . . .

---

**Source:** [https://fda-staging.innolitics.com/device/K103373](https://fda-staging.innolitics.com/device/K103373)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
