The AOI Cranial Helmet is intended for medical purposes to apply pressure to prominent regions of an infant's cranium in order to improve cranial symmetry and/or shape in infants from age 3 to 18 months of age, with moderate to severe nonsynotic positional plagiocephaly, including infants with plagiocephalic- and brachycephalic- shaped heads.
Device Story
Custom-fabricated cranial orthosis; designed to apply mild pressure to prominent cranial regions to restrict growth while providing voids for flattened areas to expand. Fabrication involves creating a plaster mold of the infant's head; positive model modified by adding plaster to flattened areas to create growth voids. Device consists of a rigid thermoformable copolymer outer shell and polyethylene foam inner liner; secured with a Velcro strap. Used in clinical settings under supervision of a Certified Orthotist; requires physician prescription. Orthotist monitors fit and alignment at bi-monthly re-evaluation appointments. Output is the physical shaping of the skull over time; benefits patient by improving cranial symmetry. No electronic components or automated processing.
Clinical Evidence
Bench testing only. No clinical trials performed. Biocompatibility supported by ISO 10993-1 compliance and historical use of materials (copolymer plastics and polyethylene foams) in similar pediatric medical devices. Safety further supported by custom-fit design and ongoing monitoring by certified orthotists.
Technological Characteristics
Custom-fabricated cranial orthosis. Materials: 5/32 inch copolymer plastic outer shell, 1/2 inch polyethylene foam inner liner, Dacron strap, nylon washer. Sensing/actuation: Passive mechanical pressure via custom-molded voids. No energy source. Non-sterile. No software or connectivity.
Indications for Use
Indicated for infants aged 3 to 18 months with moderate to severe nonsynostotic positional plagiocephaly, including plagiocephalic and brachycephalic head shapes. Contraindicated for infants with craniosynostosis or hydrocephalus.
Regulatory Classification
Identification
A cranial orthosis is a device that is intended for medical purposes to apply pressure to prominent regions of an infant's cranium in order to improve cranial symmetry and/or shape in infants from 3 to 18 months of age, with moderate to severe nonsynostotic positional plagiocephaly, including infants with plagiocephalic-, brachycephalic-, and scaphocephalic-shaped heads.
Special Controls
*Classification.* Class II (special controls) (prescription use in accordance with § 801.109 of this chapter, biocompatibility testing, and labeling (contraindications, warnings, precautions, adverse events, instructions for physicians and parents)).
In addition to the general controls of the act, the Dynamic Orthotic Cranioplasty - DOC™ Band is subject to the following special controls in order to provide reasonable assurance of the safety and effectiveness: (1) The sale, distribution and use of this device are restricted to prescription use in accordance with 21 CFR 801.109. (2) The labeling must include (a) contraindications for the use of the device on infants with synostosis or with hydrocephalus; (b) warnings indicating the need: (i) to evaluate head circumference measurements and neurological status at intervals appropriate to the infant’s age and rate of head growth, and to describe steps that should be taken in order to reduce the potential for restriction of cranial growth and possible impairment of brain growth and development; (ii) to evaluate the skin at frequent intervals, e.g., every three to four hours, and to describe steps that should be taken if skin irritation or breakdown occurs; (c) precautions indicating the need: (i) to additionally treat torticollis, if the positional plagiocephaly is associated with torticollis; (ii) to evaluate device fit and to describe the steps that should be taken in order to reduce the potential for restriction of cranial growth, possible impairment of brain growth and development and skin irritation and/or breakdown; (iii) to evaluate the structural integrity of the device and to describe the steps that should be taken to reduce the potential for the device to slip out of place and cause asphyxiation or trauma to the eyes or skin; (d) adverse events, i.e., skin irritation and breakdown that have occurred with the use of this device; (e) clinician’s instructions for casting the infant, for fitting the device, and for care and use of the device; and (f) parents’ instructions for care and use of the device. (3) The materials must be assessed for biocompatibility with testing appropriate for long term direct skin contact.
{0}------------------------------------------------
## K103362
Image /page/0/Picture/1 description: The image shows the letters "AOI" in white against a black, hexagonal background. The letters are in a bold, sans-serif font and are evenly spaced. The hexagon is oriented with two flat sides at the top and bottom, and pointed corners on the sides.
APR 1 8 2011
## AOI 510(K) Summary of Effectiveness and Safety
The summary contained hereinafter is provided pursuant to Section 513(1)(3)(A) of the Federal Food, Drug and Cosmetic Act.
A. Applicant Information
1. Submitter Name: Advanced OrthoPro Inc. (AOI)
- 2. Address: 1820 N. Illinois St., Indianapolis, IN 46202
- 3. Telephone: 317-924-4444
- 4. Contact Person: Keon Mansoori
- 5. Date Prepared: February 8, 2011
- 6. Registration Number:
B. Device Name and Classification
- 1. Name: Cranial Helmet
- 2. Trade Name: AOI Cranial Remolding Helmet
- 3. Common Name: Cranial Orthosis
- 4. Classification Name: Cranial Orthosis
- 5. Product Code: MVA
- 6. Class: II
- 7. Regulation Number: 21 CFR § 882.5970
- C. Identification of Legally Marketed Devices
- 1. Name: OPIBand by Orthomerica
- 2. K Number: K001167
- 3. Date Cleared: July 7, 2000
D. Device Description
The AOI Cranial Helmet is a cranial orthosis which is made through a custom fabrication process to apply mild pressure to prominent regions of a young child's cranium to slow further growth in that area while allowing flattened portions of the cranium to grow into voids built into the helmet in order to improve overall cranial symmetry.
The helmet itself, once finished and delivered, covers the majority of the head and is trimmed as follows. The anterior inferior edge is trimmed just above the eyebrows, and the sides are trimmed above the ears. The posterior inferior is trimmed low to capture the occiput. A Velcro strap is attached to one side of the helmet to keep it
{1}------------------------------------------------
securely in place. The details of the materials used in this helmet and process are found below.
The AOI Cranial Helmet is not packaged for sale due to the fact that it is a custom fabricated device. The device is fitted to the child's head at the time of delivery. The AOI device has no accessories, and no options are available for the infant or the caregiver.
The AOI Cranial Helmet is custom made by first having a Certified Orthotist take a plaster mold of the patient's head. Since this device is custom fabricated for each patient there is no such thing as a standard size, model variance or otherwise. This individual's mold is then filled with plaster and a positive model of the child's head is made. The prominent areas are left alone on the model while the flattened areas of the model (which correspond to flat areas of the child's head) are built up with plaster to create rounded and symmetrical voids. As the child grows and the head fills the voids, which are made for just that purpose, the skull is slowly shaped into a more symmetrical shape. All steps of fabrication and production are done under the supervision of a Certified Orthotist.
The AOI Cranial Helmet is made of substantially the same materials as the predicate. The finished AOI Cranial Helmet is composed of a hard outer shell, made of thermoform-able plastic (Copolymer 5/32 inch) and an inner layer of polyethylene foam ( 1/2 inch). These materials are the exact same materials used on many types of medical devices and come into protracted contact with the skin of adults and children every day. Throughout the duration of use, the fit and alignment of the device are monitored by the Certified Orthotist on at least a bi-monthly basis at re-evaluation appointments. Instructions on the care and use of the device, including wearing schedule and helpful hints are provided to caregiver at time of delivery of the product. In addition to the evaluations by the Orthotist the AOI Cranial Helmet is only provided to person's with a prescription from their physician for the device, and therefore this device is also only provided to persons whom are under the care of a physician who has decided to apply this product. Please see the Figures below which show the AOI Cranial Helmet.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image contains three figures of a medical helmet. The helmet is white with a pattern of black and gray designs. The helmet has a strap on the side and is sitting on a white surface. The figures are labeled as Figure 1, Figure 2, and Figure 3.
E. Intended Use and Indications of Use
The AOI Cranial Helmet is intended for medical purposes to apply pressure to prominent regions of an infant's cranium in order to improve cranial symmetry and/or shape in infants from age 3 to 18 months of age, with moderate to severe nonsynotic positional plagiocephaly, including infants with plagiocephalic- and brachycephalicshaped heads.
F. Comparison to Predicate Device
{3}------------------------------------------------
The AOI Cranial Helmet and the predicate device it is based on are very similar with respect to the production processes, the instructions for use which are provided to the patient, the materials used to produce the items, safety and effectiveness and special controls used in production. In fact, the AOI Cranial Helmet and its predicate device have exactly the same purposes, are made from the same materials and work in exactly the same way, as will be shown below.
The material in the production of the AQI Cranial Helmet is handled in the exact same manner as the polymer used in the predicate, incorporating all of the safety and standards of practice. The proposed indications of use are analogous to those presented by the predicate device, and biocompatibility, function and effectiveness also closely shadow those of the predicate device.
| Comparison<br>Criteria | Subject Device | Predicate Device<br>(OPIBand K001167) |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product<br>Code | MVA | MVA |
| Prescription<br>Device | YES | YES |
| Materials | 1. Outer rigid shell made of 5/32<br>inch copolymer plastic.<br>2. Inner liner of 1/2 inch<br>polyethelene foam.<br>3. 1 1/2 inch Dacron Strap.<br>4. 1 1/2 inch Chafe with roller loop.<br>5. Nylon Washer<br>6. Gap Block 1/2 inch firm pelite<br>polyethelene foam. | 1. Outer shell of 5/32 inch<br>copolymer plastic.<br>2. An inner liner liner of 1/2 inch<br>pelite polyethelene foam or<br>Aliplast foam.<br>3. 1 1/2 in. Dacron Strap<br>4. Gap block 1/2 inch firm pelite<br>polyethelene foam.<br>5. Nylon Washer |
| Product<br>Design | Custom made cranial orthosis.<br>approx. 6 oz. | Custom made cranial orthosis,<br>approx 6 oz. |
| Production | 1. Form orthosis from positive<br>mold of infant's head. | 1. Form orthosis from a positive<br>mold of infant's head. |
{4}------------------------------------------------
| Comparison<br>Criteria | Subject Device | Predicate Device<br>(OPIBand K001167) |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for use | The AOI Cranial Helmet is<br>intended for medical purposes to<br>apply pressure to prominent<br>regions of an infant's cranium in<br>order to improve cranial symmetry<br>and/or shape in infants from age 3<br>to 18 months of age, with<br>moderate to severe nonsynostotic<br>positional plagiocephaly, including<br>infants with plagiocephalic- and<br>brachycephalic- shaped heads. | "The OPIBand is intended for<br>medical purposes to apply<br>pressure to prominent regions of<br>an infant's cranium in order to<br>improve cranial symmetry and/or<br>shape in infants from 3 to 18<br>months of age, with moderate to<br>severe nonsynostotic positional<br>plagiocephaly, including infants<br>with plagiocephalic- and<br>brachycephalic- shaped heads." |
| Contraindicat<br>ions | The AOI Cranial Helmet is not for<br>use on infants with<br>craniosynostosis or<br>hydrocephalus. | "This device is not for use on<br>infants with pre-surgical<br>craniosynostosis or<br>hydrocephalus." |
See Table (supra)
## G. Performance Data
The safety of cranial helmets has also been shown through past biocompatibility assessments which reveal that this type of device is not expected to adversely affect children under intended conditions of wear.1 The AOI Cranial Helmet has not undergone its own individual clinical trials with respect to biocompatibility. This is in compliance with ISO 10993-1 in that after review of the production methods and materials involved "no testing is needed if the material has a demonstrable history of use in a specified role that is equivalent to that of the device under design." In the case of the AOI Cranial helmet the materials used in the manufacture and completed product are copolymer plastics and polvethelene foams that are used in devices which are used directly against the skin of children and even infants. Devices such as custom night AFOs or custom made Cervical devices are fabricated using the same methods and materials and contact the skin for protracted wear periods. They are prepared using the same vacuum forming and glueing method. The only surface contacting the skin is the foam. These materials are provided to AOI by companies which certify that their
<sup>1</sup> Littlefield TR. Beals SP. Manwaring KH. Pomatto JK. Joganic EF, Golden KA, Ripley CE. Treatment of Craniofacial Asymmetry with Dynamic Orthotic Cranioplasty. Journal of Craniofacial Surgery. 1988; 11-17.
{5}------------------------------------------------
products are manufactured and delivered in a manor which is compliant with ISO standards.2
Importantly, the only part of the device which contacts the patient's skin is the foam liner which is made of polvethelene foam. This foam is not reported to cause skin irritation or abrasion and is used widely in the medical device field as a safe interface between derma and device. Also, the safety of these helmets is furthered by the fact the each device is custom made to the individual patient, this prevents excessive slipping or migration of the helmet or excessive pressure on the child's cranium. Most importantly all patients wearing helmets are closely monitored by AOI personnel who are certified and have had years of experience as orthotists, including bi-monthly evaluations of the helmet and patient and an open dialoque with the treating physicians and caregivers. Also. the proposed device is not in any quantifiable way different from the predicate device with regard to manufacture or materials used, thus when looking at the predicate device we can see that the helmet has been effectively used for years without complications of this sort.
## H. Summarv
As can be seen from the 510(K) Summary supra the proposed AOI Cranial Helmet does not differ substantially from the predicate device. The predicate is made of the same materials and in the same fashion as the AOI proposed device. Based on a comparison of the technological specifications of the predicate device and the AOI Cranial Helmet it is believed that the AOI Helmet has been shown to be substantially equivalent to the predicate device.
<sup>2</sup> O&P Enterprises, Inc. and American Plastics Inc.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Advanced OrthoPro, Inc. c/o Keon Mansoori B.A., J.D. Counsel 1820 N. Illinois St. Indianapolis, IN 46202
Re: K103362
Trade/Device Name: AOI Cranial Helmet Regulation Number: 21 CFR 882.5970 Regulation Name: Cranial Orthosis Regulatory Class: Class II Product Code: MVA Dated: February 15, 2011 Received: February 15, 2011
APR 1 8 2011
Dear Mr. Mansoori:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - Keon Mansoori B.A., J.D
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Alexander
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
510(k) Number (if known): K103362 Device Name: AOI Cranial Helmet Indications for Use:
The AOI Cranial Helmet is intended for medical purposes to apply pressure to prominent regions of an infant's cranium in order to improve cranial symmetry and/or shape in infants from age 3 to 18 months of age, with moderate to severe nonsynotic positional plagiocephaly, including infants with plagiocephalic- and brachycephalic- shaped heads.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
OR
JEFFREY Toy
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
K103362 510(k)
Page I of --
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.