K103355 · Boston Scientific Corp · HCG · Feb 16, 2011 · Neurology
Device Facts
Record ID
K103355
Device Name
GDC 360 DETEACHABLE COIL
Applicant
Boston Scientific Corp
Product Code
HCG · Neurology
Decision Date
Feb 16, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5950
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K103355 · Feb 16, 2011
GDC 360 DETEACHABLE COIL
Boston Scientific Corp
International Subarachnoid Aneurysm Trial (ISAT) clinical study data; Published clinical literature (The Lancet, Lancet Neurology)
The sponsor uses long-term follow-up data from the ISAT trial to demonstrate that endovascular coiling is safe and effective for ruptured intracranial aneurysms, supporting the expansion of the device's indications for use.
International Subarachnoid Aneurysm Trial (ISAT); Randomized clinical trial with long-term follow-up; Follow-up/Duration: Mean of 9 years (range 6-14 years)
Patients with ruptured intracranial aneurysms; Sample Size: 2143 patients
Neurosurgical clipping
Risk of dependency, death, and rebleeding
Indications for Use
GDC® 360° Detachable Coils are intended for the endovascular embolization of: • Intracranial aneurysms • Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae • Arterial and venous embolizations in the peripheral vasculature
Device Story
GDC 360° Detachable Coils facilitate endovascular embolization of intracranial aneurysms and vascular abnormalities. Device consists of platinum/tungsten alloy coil attached to stainless steel delivery wire. Physician delivers coils into aneurysm via microcatheter; detachment occurs via electrolytic dissolution of delivery wire using external power supply (Boston Scientific Detachable Coil Power Supply or InZone Detachment System). Multiple coils are deployed to densely pack the site. Secondary winding mandrel creates 360° shape with smaller distal loop to facilitate placement. Used in clinical settings by neurosurgeons/interventionalists. Output is physical occlusion of vascular site; benefits include reduced risk of dependency, death, or rebleeding compared to surgical clipping, as supported by ISAT clinical trial data.
Clinical Evidence
Evidence based on the International Subarachnoid Aneurysm Trial (ISAT). Long-term follow-up (mean 9 years, range 6-14) demonstrated that endovascular coiling significantly reduced the risk of death compared to surgical clipping in patients with ruptured intracranial aneurysms. Data showed low risk of rebleeding and no significant difference in rebleeding-related deaths between coiling and clipping groups.
Technological Characteristics
Platinum/tungsten alloy coil; stainless steel delivery wire. Electrolytic detachment mechanism. Secondary winding mandrel creates 360° coil shape with 75% diameter distal loop. Dimensions vary by coil subtype (GDC-10, GDC-18). Sterilized via validated process. No software or active electronic components within the implant itself; uses external power supply for detachment.
Indications for Use
Indicated for endovascular embolization of intracranial aneurysms (including ruptured aneurysms where surgical intervention is feasible), arteriovenous malformations, arteriovenous fistulae, and arterial/venous embolizations in the peripheral vasculature.
Regulatory Classification
Identification
A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).
{0}------------------------------------------------
PreMarket Notification 510(lc) Submission K103355; Boston Scientific's GDC * 360° Detachable Coils – Expanded Indications for Use
FEB 16 2011
| 510(k) SUMMARY | |
|-----------------------|-----------------------------------------------------------------------------------------------------|
| Submitter | Boston Scientific Corporation<br>47900 Bayside Parkway<br>Fremont, CA 94538-6515 |
| Correspondent | Rhoda M. Santos<br>Boston Scientific Corporation<br>47900 Bayside Parkway<br>Fremont, CA 94538-6515 |
| Contact Information | E-mail: rhoda.santos@stryker.com<br>Phone: 510-624-1262<br>Facsimile: 510-440-7752 |
| Date Summary Prepared | 21 January 2011 |
Device Name / Trade Name
The device trade names and common/classification names are as follows:
| Device Trade Name: | GDC <sup>®</sup> 360° Detachable Coil |
|-----------------------------|---------------------------------------------------------------------------------|
| Common/Classification Name: | Occlusion Coil; Vascular Embolization Device; Neurovascular embolization device |
Address and Establishment Registration Number
The address and registration numbers for the manufacturer and sterilization sites are:
| Manufacturer: | Sterilization Site: |
|--------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Boston Scientific Cork Ltd.<br>Business & Technology Park<br>Model Farm Road<br>Cork,<br>Ireland | Isotron Ireland Ltd.<br>IDA Business and Technology<br>Park<br>Tullamore,<br>County Offaly<br>Ireland |
| FDA Registration # 9616684<br>(Establishment Registration Number) | |
510(k) Submission K103355, Boston Scientific
GDC* 360° Detachable Coils
January 2011
ﺔ ﻣﺴ
{1}------------------------------------------------
# Device Classification Regulation Number and Regulatory Status
GDC® 360° Detachable Coils are vascular and neurovascular embolization devices under 21 CFR 870.3300 (KRD) and 21 CFR 882.5950 (HCG), respectively. The devices are Class II devices (special controls), the special control for which is FDA's guidance document, Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices (issued 29 Dec 2004).
# Predicate Device Information
Predicate devices are Boston Scientific's Guglielmi Detachable Coils (GDC) and GDC 360° Detachable Coils cleared under the following submissions:
- K031049, ISAT Indication for all GDC® devices (cleared 1 August 2003); .
- K042539, Guglielmi Detachable Coil 360° Coils (cleared 19 October 2004 ). .
# Device Description
Boston Scientific's GDC® 360° Detachable Coil is a device which facilitates endovascular embolization of intracranial aneurysms and other vascular abnormalities. The GDC® 360° Detachable Coil is a platinum/tungsten alloy coil attached to a stainless steel delivery wire. The GDC® 360° Detachable Coil is detached (using the Boston Scientific Detachable Coil Power Supply or InZone Detachment System) by electrolytically dissolving a small portion of the delivery wire upon its desired placement within an aneurysm or other vascular site via a microcatheter. Multiple coils can be delivered into an aneurysm or other vascular site through the same microcatheter until the aneurysm or other vascular site is densely packed.
The GDC® 360° coil is first wound into a primary or main coil and then into a secondary shape using a secondary shaping (winding) mandrel. The distal end of the main coil is formed such that there is a smaller distal loop at the end of the main coil, the diameter of which is 75% that of the main coil to facilitate placement of the coil into an aneurysm.
The GDC 360° Detachable Coil is a line extension to the GDC® family of devices and include the following coil subtypes:
- . GDC®-10 360° Coil
- GDC -10 360° Soft Coil, and .
- GDC®-18 360° Coil .
The primary differences between the GDC® and the GDC® 360° Coil include the following:
- a slightly modified shape relative to the predicate GDC® coils (accomplished . through use of a new secondary coil winding mandrel)
- an expanded range of coil sizes (i.e., in terms of coil outside, or secondary, . diameter and coil length) and,
January 2011
{2}------------------------------------------------
### PreMarket Notification 510(k) Submission K103355; Boston Scientific's GDC® 360° Detachable Coils - Expanded Indications for Use
- minor additional changes related to these two changes noted above. .
### Purpose of Submission
This Premarket Notification has been submitted to obtain clearance for a change in the indications for use of GDC® 360° Detachable Coils to include patients with ruptured intracranial aneurysms for whom surgical intervention is feasible. The proposed change in indications is based on results of the International Subarachnoid Aneurysm Trial (ISAT). There are no other changes to the GDC® 360° device as a result of the expanded indications, compared to the device described in K042539 (cleared 19 October 2004).
{3}------------------------------------------------
| Comparison to Predicate Device<br>Characteristic | Predicate Device | Predicate Device | Proposed Device |
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| Device Name | Guglielmi Detachable Coil (GDC®) | Guglielmi Detachable Coil (GDC®<br>360°) Detachable Coil | Guglielmi Detachable Coil (GDC®<br>360°) Detachable Coil |
| 510 (k) | K031049 (cleared 1 August 2003) | K042539 (cleared 19 October 2004) | CURRENT NOTIFICATION |
| Device Description<br>(Technological<br>Characteristics) | Boston Scientific's Guglielmi<br>Detachable Coil (GDC®) is a device<br>which facilitates endovascular<br>embolization of intracranial<br>aneurysms and other vascular<br>abnormalities. The GDC®<br>Detachable Coil is a<br>platinum/tungsten alloy coil attached<br>to a stainless steel delivery wire. The<br>GDC® Detachable Coil is detached<br>(using the Boston Scientific<br>Detachable Coil Power Supply or<br>InZone™ Detachment System) by<br>electrolytically dissolving a small<br>portion of the delivery wire upon its<br>desired placement within an<br>aneurysm or other vascular site via a<br>microcatheter. Multiple coils can be<br>delivered into an aneurysm or other<br>vascular site through the same<br>microcatheter until the aneurysm or<br>other vascular site is densely packed. | Boston Scientific's Guglielmi<br>Detachable Coil (GDC® 360°)<br>Detachable Coil is a device which<br>facilitates endovascular embolization<br>of intracranial aneurysms and other<br>vascular abnormalities. The GDC®<br>360° Detachable Coil is a<br>platinum/tungsten alloy coil attached<br>to a stainless steel delivery wire. The<br>GDC® 360° Detachable Coil detached<br>(using the Boston Scientific<br>Detachable Coil Power Supply or<br>InZone™ Detachment System) by<br>electrolytically dissolving a small<br>portion of the delivery wire upon its<br>desired placement within an aneurysm<br>or other vascular site via a<br>microcatheter. Multiple coils can be<br>delivered into an aneurysm or other<br>vascular site through the same<br>microcatheter until the aneurysm or<br>other vascular site is densely packed. | Same as predicate GDC 360°<br>Detachable Coil. |
| Indications for<br>Use/Intended | Guglielmi Detachable Coils are<br>intended for the endovascular | GDC 360° Detachable Coils are<br>intended for embolization of those | GDC 360° Detachable Coils are<br>intended for the endovascular |
| Use | embolization of: | | embolization of: |
| | | intracranial aneurysms that – because<br>of their morphology, their location, or<br>the patient's general medical condition<br>– are considered by the treating<br>neurosurgical team to be:<br>a) very high risk for management by<br>traditional operative techniques, or<br>b) inoperable, and for embolization of<br>other neurovascular abnormalities<br>such as arteriovenous malformations<br>and arteriovenous fistulae. GDC 360°<br>Coils are also intended for arterial and<br>venous embolizations in the peripheral<br>vasculature. | |
| | • Intracranial aneurysms | | • Intracranial aneurysms |
| | • Other neurovascular abnormalities<br>such as arteriovenous malformations<br>and arteriovenous fistula | | • Other neurovascular abnormalities<br>such as arteriovenous malformations<br>and arteriovenous fistula |
| | • Arterial and venous embolizations<br>in the peripheral vasculature | | • Arterial and venous embolizations<br>in the peripheral vasculature |
| Manufacturer | Boston Scientific Corporation | Same | Same |
| Device<br>Classification | Class II (Special Controls), HCG,<br>21 CFR §882.5950 | Same | Same |
| | Class II (Special Controls), KRD,<br>21 CFR §870.3300 | | |
eMarket Notification 510(k) Submission: Boston Scientific's GDC 360° Detachable Coils – Expanded Indications for Use
510(k) Submission K103355, Boston Scientifi
GDC 360° Detachable Coils
January 2011
:
{4}------------------------------------------------
PreMarket Notification 510(k) Submission: Boston Scientific's GDC® 360° Detachable Coils – Expanded Indications for Use
i
510(k) Submission K103355, Boston Scientific
GDC® 360° Detachable Coils
{5}------------------------------------------------
# PreMarket Notification 510(k) Submission: Boston Scientific's GDC® 360° Detachable Coils – Expanded Indications for Use
### Summary of Non-Clinical Data
A summary of the Non-clinical bench testing performed on the GDC® 360° coils is provided in Table 1. These test results were previously provided in K042539, cleared 19 October 2004.
| Table 1: Verification Test Summary |
|----------------------------------------------------------|
| Predicate GDC® Devices vs GDC®-10 and GDC®-18 360° Coils |
| Test or Point of Comparison | GDC-10 360° Coil SR<br>GDC®-10 360° Soft Coil SR<br>GDC-18 360° Coil |
|---------------------------------------------------------------------|----------------------------------------------------------------------|
| Tensile Strength<br>(Main coil to delivery wire) | Meets same acceptance criteria as predicate device (K03149) |
| Friction<br>(Coil deployment resistance through a<br>microcatheter) | Meets same acceptance criteria as predicate device (K03149) |
| Detachment Time | No change made which would affect this test. |
| Deployment / Retraction Force | Meets same acceptance criteria as predicate device (K03149) |
| Tip Ball Strength | Meets same acceptance criteria as predicate device (K03149) |
| Coil Stiffness | Meets same acceptance criteria as predicate device (K03149) |
| Heating Effect of Electrolysis | No change made which would affect this test. |
| Heating Effect of MRI | Meets same acceptance criteria as predicate device (K03149) |
| Electrostatic Discharge | No change made which would affect this test. |
| Electromagnetic Compatibility-Radiated<br>Susceptibility | No change made which would affect this test. |
| Electromagnetic Compatibility-Radiated<br>Emissions Class B | No change made which would affect this test. |
| Electromagnetic Compatibility-Magnetic<br>Immunity | No change made which would affect this test. |
| Operating System Test<br>(Assembly Source Code) | No change made which would affect this test. |
#### Conclusion
The Non-clinical testing has demonstrated that the GDC® 360° coils are substantially equivalent to the predicate GDC® coils (K03149, cleared 1 August 2003).
510(k) Submission K103355, Boston Scientific GDC 360° Detachable Coils
{6}------------------------------------------------
### PreMarket Notification 510(k) Submission: Boston Scientific's GDC® 360° Detachable Coils -Expanded Indications for Use
### Summary of Clinical Data
Boston Scientific Corporation previously submitted the 1-year results of the International Subarachnoid Aneurysm Trial (ISAT) with K031049, which demonstrated a statistically significant reduction in the risk of dependency or death at 1 year post-treatment when patients with ruptured intracranial aneurysms were treated endovascularly with GDC Detachable Coils rather than with neurosurgical clipping. Subsequent long term studies of the ISAT trial have been published in the Lancet Neurology by the investigators. The data showed that the risk of death at five years for patients with a ruptured intracranial aneurysm treated with endovascular coil embolization was significantly lower compared to patients who underwent surgical clipping. Overall, the ISAT follow-up data for a mean of 9 years (range 6-14 years) demonstrate that the risk of rebleeding from a treated aneurysm is low and although there were more rebleeds from the treated aneurysm in the coiling group than in the clipping group, there was no difference between the two groups in the number of deaths due to rebleeding. This improvement in patient outcomes compared to surgical clipping demonstrates that the GDC® 360° Coils with the proposed indications change is as safe and effective as the predicate GDC® 360° Coils with current indications.
### Conclusion
Based on the results of the clinical data of the ISAT, which demonstrate improvement in patient outcomes, and the similarities of the design, materials, and processes of the GDC® 360° Detachable Coil to the predicate GDC® Detachable Coil, Boston Scientific requests clearance for the Indications for Use on the GDC® 360° Detachable Coil to be the same as is currently referenced for our GDC devices. Since the requested change in Indications for Use does not alter the fundamental scientific technology of the underlying predicate device and the risk assessment of the modifications raises no new questions of safety and effectivess, Boston Scientific has determined that the subject device is substantially equivalent to the predicate devices cleared under K031049 and K042539.
<sup>1</sup> ISAT Collaborative Group. International Subarachnoid Aneurysm Trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised trial. The Lancet 2002; 360: 1267-74
<sup>2</sup> Molyneux, A. J., et al., Risk of recurrent subarachnoid haemorrhage, death, or dependence and standardized mortality ratios after clipping or coiling of an intracranial aneurysm in the International Subarachnoid Aneurysm Trial (ISAT): long term follow-up. Lancet Neurol, 2009. 8(5): p. 427-33.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The caduceus is depicted with three lines forming a stylized bird-like shape. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Boston Scientific Neurovascular c/o Ms. Rhoda M. Santos Senior Regulatory Affairs Specialist 47900 Bayside Parkway Fremont, CA 94538
FEB 1 6 2011
Re: K103355
Trade/Device Name: GDC® 360° Detachable Coils Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: Class II Product Code: HCG, KRD Dated: December 8, 2010 Received: December 9, 2010
Dear Ms. Santos:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
、・
{8}------------------------------------------------
Page 2 - Ms. Rhoda M. Santos
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm175809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Sincerely yours,
Kesia Alexander
for
N
I
I
O
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Device Name: GDC® 360° Detachable Coils
### Indications for Use:
GDC® 360° Detachable Coils are intended for the endovascular embolization of:
- Intracranial aneurysms ●
- Other neurovascular abnormalities such as arteriovenous malformations and ● arteriovenous fistulae
- Arterial and venous embolizations in the peripheral vasculature .
Prescription Use
(Per 21 CFR 801 Subpart D)
AND/OR Over-The-Counter Use (Per 21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
JEFFREY Toy
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number k103355
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.