K103354 · Natec Medical , Ltd. · LIT · Feb 23, 2011 · Cardiovascular
Device Facts
Record ID
K103354
Device Name
EBONY PTA .035 PERIPHERAL DILATATION CATHETER
Applicant
Natec Medical , Ltd.
Product Code
LIT · Cardiovascular
Decision Date
Feb 23, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ebony® .035 PTA Catheter is intended for the dilation of lesions in the femoral, iliac, popliteal, infrapopliteal and renal arteries. The Ebony® .035 PTA Catheter is not for use in coronary arteries or neuro-vasculature.
Device Story
Over-the-wire (OTW) peripheral transluminal angioplasty (PTA) catheter; features semi-compliant inflatable balloon at distal end; atraumatic tapered tip for crossing stenoses; double lumen design (one for guidewire/inflation, one for distal access). Radiopaque markers on balloon provide visual positioning reference. Hydrophilic coating on distal shaft improves trackability/pushability. Luer lock fitting connects to external inflation device. Used by clinicians in clinical settings to dilate arterial lesions; balloon diameter/length selected based on vessel anatomy. Inflation device expands balloon to known diameter at specific pressures per compliance chart. Benefits patient by restoring blood flow in peripheral arteries.
Clinical Evidence
Bench testing only. No clinical data. Testing included balloon compliance, burst pressure, fatigue, shaft resistance, bond strength, dimensions, deflation time, and compatibility with guidewires/introducers. Results met all predetermined acceptance criteria.
Technological Characteristics
Double lumen OTW catheter; semi-compliant balloon; hydrophilic coating; radiopaque markers. Materials biocompatible per ISO 10993 (Parts 1, 2, 4, 5, 10, 11, 12) and ASTM F756-00. Available in diameters 5-12mm, lengths 20-80mm, catheter lengths 80-150cm. Sterile, single-use.
Indications for Use
Indicated for dilation of lesions in femoral, iliac, popliteal, infrapopliteal, and renal arteries. Contraindicated for use in coronary arteries or neuro-vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
K103354
# 510(k) Summary 807.92(c)
# FEB 2 3 2011
### SPONSOR
#### 807.92(a)(1)
Company Name: Company Address
Natec Medical Ltd. Maeva Centre Building - Ebene Business Park Reduit, Mauritius
Telephone: Fax:
+230 466 30 54 +230 466 67 70
Contact Person:
Alain Valorge Quality manager info@natec-medical.com
Summary Preparation Date: October 28, 2010
| DEVICE NAME | 807.92(a)(2) | |
|----------------------|------------------------------------------------|--|
| Trade Name: | Ebony® PTA .035 Peripheral Dilatation Catheter | |
| Common/Usual Name: | PTA Catheter | |
| Classification Name: | Percutaneous Catheter | |
| Regulation Number: | 21 CFR 870.1250 | |
| Product Code: | LIT | |
| Device Class: | Class II | |
#### PREDICATE DEVICE
#### 807.92(a)(3)
Legally Marketed Equivalent Device
| K Number | Trade Name | Manufacturer |
|----------|---------------------------------|------------------------------------|
| K062809 | ADMIRAL XTREME™ PTA<br>Catheter | Invatec Innovative<br>Technologies |
#### DEVICE DESCRIPTION
#### 807.92(a)(4)
The Ebony® PTA .035 Peripheral Dilatation Catheter is a standard Over the Wire (OTW) PTA Catheter with a semi-compliant inflatable balloon at the distal part and an atraumatic, tapered tip to aid in crossing tight stenoses.
It is a double lumen catheter, one lumen for the guide wire used for inflation/deflation of the balloon and one lumen for the guide wire used to access to the distal tip of the catheter (max/0.035″). A luer lock fitting allows the connection with an inflation device (Y hub). The balloon has two radiopaque markers to provide visual reference points for balloon positioning relative to the stenosis within the vessel.
{1}------------------------------------------------
The distal catheter, proximal to the balloon, is covered with a hydrophilic coating to improve the trackability and pushability characteristics.
The balloon material expands to a known diameter at specific pressure defined in a compliance chart supplied with the catheter. The device is available in balloon. diameters of 5, 6, 7, 8, 9, 10 or 12 mm, balloon lengths of 20, 40, 60, or 80 mm and catheter lengths of 80, 130, or 150 cm.
lt will be supplied sterile and is intended for one time use.
## DEVICE INTENDED USE
#### 807.92(a)(5)
The Ebony® .035 PTA Catheter is intended for the dilation of lesions in the femoral, iliac, popliteal, infrapopliteal and renal arteries.
The Ebony® .035 PTA Catheter is not for use in coronary arteries or neuro-vasculature.
## COMPARISON OF TECHNICAL CHARACTERISTICS 807.92(a)(6)
The Ebony® .035 PTA Catheter and the predicate device, Admiral Xtreme™ PTA Catheter have the same intended use for the dilatation of lesion in the femoral, iliac, infra popliteal and renal arteries. The technological characteristics such as material, biocompatibility, mode of operation, performance properties, sterilization and packaging are substantially equivalent to the predicate device.
## NONCLINICAL TESTS
807.92(b)
## SAFETY and EFFECTIVENESS
#### BIOCOMPATIBILITY
All materials used in the Ebony® .035 PTA Catheter are biocompatible based on biocompatibility testing results. The device has been tested according to ISO 10993 Part 1, 2, 4, 5, 10, 11, 12, ASTM F756-00 and 21 CFR 58 (GLP regulations)
## PERFORMANCE DATA
The safety and effectiveness of the Ebony® .035 PTA Catheter has been demonstrated through data collected from in vitro bench tests and analyses. Testing results demonstrated equivalent performance of the Ebony® .035 PTA Catheter with the predicate device.
The testing included balloon compliance, balloon burst pressure, balloon fatigue, shaft resistance, bond strength, catheter dimensions, deflation time and guide wire and introducer compatibility.
### CONCLUSION
## 807.92(b)(3)
The subject device, the Ebony® .035 PTA Catheter, met all the predetermined acceptance criteria of design verification and validation as specified by applicable
{2}------------------------------------------------
:
:
:
. . . . .
. .
.
1. 1. 1. 1.
standards, guidance, and test protocols. It is the conclusion of Natec Medical, Ltd that the Ebony® .035 PTA Catheter is as safe and effective as the predicate device.
and the state of the state of the states of the states
:
:
:
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes wrapped around a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Natec Medical Ltd c/o Mr. E.J. Smith Smith Associates 1468 Harwell Ave. Crofton, MD 21114
FEB 2 3 2011
ﻟﻤﻤﻠﻜﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘ
Re: K103354
Trade/Device Name: Ebony .035 PTA Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Regulatory Class: Class II (two) Product Code: LIT Dated: February 8, 2011 Received: February 9, 2011
Dear Mr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2 - Mr. E.J. Smith
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
er D. V. hner
Image /page/4/Picture/8 description: The image shows a black and white drawing of a symbol. The symbol appears to be a stylized letter or character, with curved lines and a loop at the bottom. The lines are thick and bold, and the overall impression is abstract and calligraphic. The symbol is isolated against a white background.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Form
Indications for Use
510(k) Number (if known): K103354
Device Name: Ebony .035 PTA Catheter
Indications for Use:
The Ebony® .035 PTA Catheter is intended for the dilation of lesions in the femoral, illac, popliteal, infrapopliteal and renal arteries.
The Ebony® .035 PTA Catheter is not for use in coronary arteries or neuro-vasculature.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
R. Vihnes
(Division Sign-Off) Division of Cardiovascular Devices
Page _1_of ।
510(k) Number_KIO 3354
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.