← Product Code [LZO](/productcode/LZO) · K103352

# VERSAFITCUP CC TRIO (K103352)

_Medacta International · LZO · Feb 11, 2011 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K103352

## Device Facts

- **Applicant:** Medacta International
- **Product Code:** [LZO](/productcode/LZO.md)
- **Decision Date:** Feb 11, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3353
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Versafitcup® CC Trio is intended for cementless use in total hip arthroplasty and in primary or revision surgery. The patient should be skeletally mature. The patient's condition should be due to one or more of: - . Severely painful and/or disabled joint: as a result of osteoarthritis, posttraumatic arthritis, rheumatoid arthritis or psoriactic arthritis, congenital hip dysplasia, or ankylosing spondylitis. - Avascular necrosis of the femoral head. . - Acute traumatic fracture of the femoral head or neck. - Failure of previous hip surgery: joint reconstruction, internal fixation, . arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.

## Device Story

Versafitcup® CC Trio is a cementless acetabular component for total hip arthroplasty; used in primary or revision surgery. System comprises a titanium alloy (Ti-6Al-4V) shell with Ti plasma spray and hydroxyapatite coating; fixed liner made of UHMWPE or HighCross® highly crosslinked UHMWPE. Optional titanium alloy cancellous bone screws and screw hole covers available. Device implanted by orthopedic surgeons in clinical/hospital settings. Shell features macrostructures for stability; accepts 28, 32, and 36 mm CoCrMo or ceramic ball heads. Provides stable acetabular surface for hip joint articulation; restores joint function; reduces pain in patients with degenerative or traumatic hip conditions.

## Clinical Evidence

Bench testing only. Design verification testing conducted per international standards and FDA guidance. Testing included push-out, lever-out, and rotational stability of the modular liner-to-shell connection, coating validation, metal shell deformation resistance during impaction, range of motion, and bone screw testing. All tests met pre-defined acceptance criteria.

## Technological Characteristics

Materials: Titanium alloy (Ti-6Al-4V) per ISO 5832-3; UHMWPE/HXUHMWPE per ISO 5834. Coatings: Ti plasma spray and Hydroxyapatite. Features: Macrostructures on equatorial region. Dimensions: 10 shell sizes; 28/32/36 mm head compatibility; liner thickness ≥5mm. Sterilization: Sterile, single-use.

## Regulatory Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

## Predicate Devices

- Exactech AcuMatch™ Integrated Hip System A-Series Porous Coated Acetabular Component ([K993082](/device/K993082.md))
- U2 Acetabular Component ([K050262](/device/K050262.md))
- Versafitcup® Double Mobility Acetabular Family ([K083116](/device/K083116.md))
- Versafitcup® Double Mobility HighCross® UHMWPE Liners ([K092265](/device/K092265.md))
- Medacta Bone Screws ([K091069](/device/K091069.md))
- Allofit Acetabular Cup ([K003758](/device/K003758.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K 103352

p. 1/4

Image /page/0/Picture/2 description: The image shows the logo for Medacta International. The logo consists of two black triangles pointing upwards, followed by the word "Medacta" in black lowercase letters. Below the word "Medacta" is the word "International" in a smaller font size, also in black. To the right of the word "International" is a small square with a plus sign inside.

FEB 1 1 20ri

# 510(k) Summary

Manufacturer:

Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 FAX (+41) 91 696 60 66

| Contact Person: | Ms. Natalie J. Kennel     |
|-----------------|---------------------------|
|                 | Consultant                |
|                 | NJK & Associates, Inc.    |
|                 | 13721 Via Tres Vista      |
|                 | San Diego, CA 92129 USA   |
| Phone:          | (858) 705-0350            |
| Fax:            | (858) 764-9739            |
| email:          | nkennel@njkconsulting.com |

Date Prepared: February 8, 2011

# DEVICE INFORMATION

| Trade/Proprietary Name: | Versafitcup® CC Trio                                                                                              |
|-------------------------|-------------------------------------------------------------------------------------------------------------------|
| Common Name:            | Total Hip Acetabular Components                                                                                   |
| Classification Name:    | LZO, 21 CFR 888.3353 Hip joint, metal/ ceramic/<br>polymer semi-constrained or nonporous uncemented<br>prosthesis |
| Classification Name:    | MEH, 21 CFR 888.3353 Hip joint, metal/ ceramic/<br>polymer semi-constrained or nonporous uncemented<br>prosthesis |

{1}------------------------------------------------

Predicate Devices:

K993082 Exactech AcuMatch™ Integrated Hin System A-Series Porous Coated Acetabular Component K050262 U2 Acetabular Component K083116 Versafitcup® Double Mobility Acetabular Family, Medacta International K092265 Versafitcup® Double Mobility HighCross® UHMWPE Liners, Medacta International K091069 Medacta Bone Screws. Medacta International K003758 Allofit Acetabular Cup. Zimmer

# Product Description:

The Versafitcup® CC Trio family of acetabular components is designed to be used with the Medacta Total Hip Prosthesis System. The Medacta Total Hip Prosthesis system includes the Quadra S, H, R, and C Stems and CoCrMo and ceramic ball heads (K072857, K073337, K080885, K082792, and K083558). The AMIStem femoral stems also work with the Medacta Total Hip Prosthesis System (K093944). The Medacta Total Hip Prosthesis System is a total hip replacement system consisting of the femoral stem made of metal, a modular femoral head made of metal or ceramic, and acetabular components. The Versafitcup® CC Trio acetabular components that are the subject of this 510(k) consist of a metal cup made of Titanium alloy and a fixed liner that is made of ultra-high molecular weight polyethylene (UHMWPE) or HighCross® highly crosslinked ultra-high molecular weight polyethylene (HXUHMWPE).

K103352

The metal acetabular component or shell is machined from titanium alloy (Ti-6Al-4V) conforming to ISO 5832-3. The outside of the metal component has macrostructures in the equatorial region. The outer surface of the metallic cup has a dual layer of coatings: Ti plasma spray and Hydroxyapatite. The polyethylene liner is a fixed liner made of UHMWPE conforming to ISO 5834. The liner has a minimum thickness of at least 5mm.

If supplemental bone screw fixation is needed, cancellous screws made of titanium allov (Ti-6Al-4V) can be inserted through screw holes. Screw hole covers made of titanium alloy (Ti-6Al-4V) are also provided.

The metal acetabular component is available is 10 sizes which accept both CoCrMo and MectaCer BIOLOX® forte Ceramic ball heads with diameters of 28, 32 and 36 mm. The fixed liners are available in both flat and hooded options in a range of sizes.

All the Versafitcup® CC Trio components are supplied sterile in single-use individual packages.

{2}------------------------------------------------

# Indications for Use:

The hip prosthesis is designed for cementless use in total hip arthroplasty in primary or revision surgery.

03352

The patient should be skeletally mature.

The patient's condition should be due to one or more of:

- . Severely painful and/or disabled joint: as a result of osteoarthritis, posttraumatic arthritis, rheumatoid arthritis or psoriactic arthritis, Congenital hip dysplasia, ankylosing spondylitis.
- Avascular necrosis of the femoral head. .
- Acute traumatic fracture of the femoral head or neck. .
- Failure of previous hip surgery: joint reconstruction, internal fixation, . arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.

# Comparison to Predicate Devices

The Versafitcup® CC Trio family is similar to the Exactech® AcuMatch™ Integrated Hip System A-Series Porous Coated Acetabular Components and U2 Acetabular components in indications for use, design, materials, and technological characteristics. The bone screws are similar to the Medacta Bone Screws in materials and design. The design and coating of the metal acetabular shell of the Versafitcup® CC Trio is very similar to the Versafitcup® Double Mobility acetabular shells. The HighCross® highly crosslinked UHMWPE liners of the Versafitcup® CC Trio is the same HighCross® highly crosslinked UHMWPE of which the Versafitcup® Double Mobility HighCross® UHMWPE liners are made. The design of the macrostructures on the outside of metal acetabular shells is identical to the Versafitcup® Double Mobility and similar to the Zimmer Allofit acetabular component.

### Performance Testing

No performance standards applicable to this device have been adopted under Section 514 of the Food, Drug and Cosmetic Act. Performance testing of the Versafitcup® CC Trio was conducted in accordance with various international standards and FDA guidance documents.

The Versafitcup® CC Trio was tested as part of design verification to written protocols with pre-defined acceptance criteria. The testing was conducted on the worst case component size and option/design. The design verification testing included push out, lever out and rotational stability of the modular connection of the fixed liner to the metal shell, coating validation, metal shell deformation resistance during impaction, range of motion, and bone screw testing. The testing met all

Versafitcup® CC Trio 510(k) February 8, 2011

Section 5 - Page 3 of 4

{3}------------------------------------------------

# 103352

p. 4/4

acceptance criteria and verifies that the performance of the Versafitcup CC Trio is substantially equivalent to the predicate devices.

# Conclusion:

The data and information provided in this submission support the conclusion that the Versafitcup® CC Trio is substantially equivalent to its main predicates devices: Exactech® AcuMatch™ Integrated Hip System A-Series Porous Coated Acetabular Component and U2 Acetabular Component with respect to indications for use and technological characteristics. The Versafitcup® CC Trio is also substantially equivalent to the Medacta Bone Screws, Versafitcup® Double Mobility Family, Versafitcup® DM HighCross® Liners and Zimmer Allofit Acetabular Cup for various technological characteristics.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three lines forming the snake and staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Medacta International SA % NJK & Associates, Inc. Ms. Natalie J. Kennel Consultant 13721 Via Tres Vista San Diego, California 92129

FEB 1 1 201

Re: K103352

Trade/Device Name: Versafitcup CC Trio Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: LZO, MEH Dated: November 5, 2010 Received: November 16, 2010

Dear Ms. Kennel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Ms. Natalie J. Kennel

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

A. B. Mh
ha

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

K103352

## Indications for Use Statement

510(k) Number (if known):

Device Name: Medacta Versafitcup® CC Trio

Indications for Use:

The Versafitcup® CC Trio is intended for cementless use in total hip arthroplasty and in primary or revision surgery. The patient should be skeletally mature. The patient's condition should be due to one or more of:

- . Severely painful and/or disabled joint: as a result of osteoarthritis, posttraumatic arthritis, rheumatoid arthritis or psoriactic arthritis, congenital hip dysplasia, or ankylosing spondylitis.
- Avascular necrosis of the femoral head. .
- Acute traumatic fracture of the femoral head or neck.
- Failure of previous hip surgery: joint reconstruction, internal fixation, . arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.

Prescription Use × Over-The-Counter Use Prescription Ose -------------------------------------------------------------------------------------------------------------------------------------------------------------(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

---

(Division/Sign-Off) or M.metkerson

Division of Surgical, Chopedic, and Restorative Devices

Page of

510(k) Number K103352

---

**Source:** [https://fda-staging.innolitics.com/device/K103352](https://fda-staging.innolitics.com/device/K103352)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
