# TDS ABUTMENT FOR FRIADENT XIVE (K103339)

_Pou YU Biotechnology Co., Ltd. · NHA · Mar 31, 2011 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K103339

## Device Facts

- **Applicant:** Pou YU Biotechnology Co., Ltd.
- **Product Code:** [NHA](/productcode/NHA.md)
- **Decision Date:** Mar 31, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

TDS Abutment for Friadent Xive is intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient. This device is compatible with the following implant systems which have an internal hex with flat-to-flat dimensions of 1.78mm or greater: Firadent: FRIALIT Implant, XiVA Implant; 3i: Internal Connect Type; Astra: Osseospeed Implant, Osseospeed TX Implant; BioHorizons: Internal Implant System, Tapered Internal Implant System, Single-Stage Implant System, Laser-lok 3.0 implant system; Lifecore: Lifecore RENOVA™ Internal Hex Implant System; Zimmer: Tapered Screw-Vent Implant System, Screw-Vent Implant System, AdVent Implant System; Osstem: GS System; Nobel Biocare: Active Implant.

## Device Story

TDS Abutment for Friadent Xive consists of titanium and ceramic-titanium abutments used with a Ti-6Al-4V ELI titanium screw to secure the abutment to an endosseous dental implant. The device supports cement-retained single or multi-unit restorations. It is compatible with various third-party implant systems featuring an internal hex with flat-to-flat dimensions of 1.78mm or greater. Used in dental clinics by clinicians, the abutment serves as an interface between the implant and the prosthetic crown or bridge. The provider selects the appropriate abutment based on the implant system, secures it with the provided screw, and cements the restoration. This system allows for the restoration of function and aesthetics in edentulous patients.

## Clinical Evidence

Bench testing only. Mechanical fatigue testing was performed per ISO 14801 on a worst-case scenario (angled abutment) to verify structural integrity. Compatibility testing evaluated dimensions, tolerances, and rotation parameters to ensure proper fit with designated dental implant systems.

## Technological Characteristics

Materials: Ti-6Al-4V ELI titanium and Y-TZP Zirconia. Design: Implant-level abutment for cement-retained restorations. Compatibility: Internal hex interface (≥1.78mm flat-to-flat). Mechanical testing: ISO 14801 fatigue testing. Connectivity: None (mechanical device).

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- Xive TG Abutments ([K032302](/device/K032302.md))
- PrimaConnex® CAD/CAM Abutment System ([K072241](/device/K072241.md))
- Atlantis M Abutment in Zirconia for 3i Certain Interface ([K063734](/device/K063734.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K103339

# 510(k) Summary Pou Yu Biotechnology Co., Ltd TDS Abutment for Friadent Xive

# APPLICANT'S NAME AND ADDRESS

Applicant' Name: Address:

Telephone: Fax: Official Contact: Date Prepared:

Pou Yu Biotechnology Co., Ltd. No. 6 Fugong Rd. Fusing Township Changhua County 506, Taiwan +886-(0)4 768 5660 x5122 +886-(0)4 768 9032 Daniel Tsao November 5, 2010

# DEVICE NAME AND CLASSIFICATION

Trade/Proprietary Name: Common Name: Classification Regulations:

Product Code: Classification Panel: Reviewing Branch:

TDS Abutment for Friadent Xive, Dental implant abutment Endosseous dental implant abutment Class II, 21 CFR 872.3630 NHA Dental Products Panel Dental Devices Branch

### LEGALLY MARKETED DEVICES

FRIADENT GmbH. – Xive TG Abutments –K032302 Lifecore Biomedical Inc. - PrimaConnex® CAD/CAM Abutment System- K072241 Atlantis Components, Inc. - Atlantis M Abutment in Zirconia for 3i Certain Interface- K063734.

## DEVICE DESCRIPTION

TDS Abutment for Friadent Xive are titanium and ceramic-titanium abutments designed to be used in conjunction with specific dental implants utilizing the TDS Abutment screw, which is made of Ti-6A1-4V ELI titanium and is used to secure the abutment to the implant. In combination with

Confidential

MAR 3 1 2011

{1}------------------------------------------------

the implant, the abutments support single or multi-unit cement-retained restorations in the maxillary and/or mandibular arch. TDS Abutment for Friadent Xive is compatible with the following implant systems which have an internal hex with flat-to-flat dimensions of 1.78mm or greater: Firadent: FRIALIT Implant, XiVA Implant; 3i: Internal Connect Type; Astra: Osseospeed Implant, Osseospeed TX Implant; BirHorizons: Internal Implant System, Tapered Internal Implant System, Single-Stage Implant System, Laser-lok 3.0 implant system; Lifecore: Lifecore RENOVA™ Internal Hex Implant System; Zimmer: Tapered Screw-Vent Implant System, Screw-Vent Implant System, AdVent Implant System; Osstem: GS System; Nobel Biocare: Active Impalnt.

# INTENDED USE OF THE DEVICE

ﺑ

TDS Abutment for Friadent Xive is intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.

This device is compatible with the following implant systems which have an internal hex with flat-to-flat dimensions of 1.78mm or greater: Firadent: FRIALJT Implant, XiVA Implant; 3i: Internal Connect Type: Astra: Osseospeed Implant, Osseospeed TX Implant; BioHorizons: Internal Implant System, Tapered Internal Implant System, Single-Stage Implant System, Laser-lok 3.0 implant system: Lifecore: Lifecore RENOVA™ Internal Hex Implant System; Zimmer: Tapered Screw-Vent Implant System, Screw-Vent Implant System, AdVent Implant System; Osstem: GS System; Nobel Biocare: Active Implant.

## TECHNOLOGICAL CHARACTERISTICS

TDS Abutment for Friadent Xive has the following similarities to the predicate devices which have been determined by FDA:

- · has the same intended use,
- · uses the same operating principle,
- · incorporates the same basic design,
- · incorporates the same materials, and
- · is produced using the same processes.

The basis for Pou Yu Biotechnology Co. Ltd.'s belief that TDS Abutment for Friadent Xive is substantially equivalent to the predicate devices is summarized in the following table.

{2}------------------------------------------------

510(k) Premarket Notification

:

|                       | Subject Device<br>Pou Yu Biotechnology Co.<br>TDS Abutment for Friadent Xive                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | FRIADENT GmbH.<br>XIVE TG Abutments<br>K032302                                                                            | Predicate Devices<br>Lifecore Biomedical Inc.<br>PrimaConnex® CAD/CAM<br>Abutment System<br>K072241                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Atlantis Components, Inc.<br>Atlantis™ Abutment in Zirconia for 3i<br>Certain Interface<br>K063734                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use          | TDS Abutment for Friadent Xive is<br>intended for use with dental implants<br>as a support for single or multiple<br>tooth prostheses in the maxilla or<br>mandible of a partially or fully<br>edentulous patient.<br><br>TDS Abutment for Friadent Xive is<br>compatible with the following implant<br>systems which have an internal hex<br>with flat-to-flat dimensions of<br>1.78mm or greater: Firadent:<br>FRIALIT Implant, XiVA Implant; 3i:<br>Internal Connect Type; Astra:<br>Osseospeed Implant, Osseospeed TX<br>Implant; BioHorizons: Internal<br>Implant System, Tapered Internal<br>Implant System, Single-Stage Implant<br>System, Laser-lok® 3.0 implant<br>system; Lifecore: Lifecore<br>RENOVAT™ Internal Hex Implant<br>System; Zimmer: Tapered<br>Screw-Vent Implant System,<br>Screw-Vent Implant System, AdVent<br>Implant System; Osstem: GS System;<br>Nobel Biocare: Active Impalnt. | The XIVE TG Abutment is<br>intended for use in the fabrication<br>of screw-retained and cementable<br>crowns and bridges. | The Lifecore PrimaConnex®<br>CAD/CAM Abutment System is<br>intended for use as an accessory to a<br>Lifecore PrimaConnex endosseous<br>implant to support a prosthetic<br>device in a partially or completely<br>edentulous patient. It is intended for<br>use to support single and multiple<br>tooth prostheses in the mandible or<br>maxilla. The prosthesis can be<br>cement retained to the abutment.<br>The abutment screw is intended to<br>secure the abutment to the<br>endosseous implant. The copings are<br>intended for use as a core structure<br>for a prosthetic restoration in<br>partially or fully edentulous<br>mandibles or maxillae in the<br>construction of single-unit cement<br>retained restorations on Lifecore<br>PrimaConnex CAD/CAM<br>Abutments. | The devices covered by this submission are<br>abutments which are placed into a dental<br>implant to provide support for a prosthetic<br>reconstruction. The Atlantis Abutment is<br>intended for use as an accessory to an<br>endosseous implant to support a prosthetic<br>device in a partially or completely edentulous<br>patient. It is intended for use to support<br>single and multiple tooth prosthesis, in the<br>mandible or maxilla. The prosthesis can be<br>cement retained to the abutment. The<br>abutment screw is intended to secure the<br>abutment to the endosseous implant. Please<br>note: Highly angled abutments (i.e. 30<br>degrees) on implants with diameters less than<br>4 mm are intended for the anterior region of<br>the mouth and are not intended for the<br>posterior region due to the limited strength of<br>the implant fixture. |
| Custom Design         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Attachment            | Implant level                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Implant level                                                                                                             | Implant level                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Implant level                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Restoration           | Cement-retained                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Cement-retained                                                                                                           | Cement-retained                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Cement-retained                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| CAD/CAM<br>processing | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Material              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Abutment              | Ti-6Al-4V ELI;<br>Y-TZP Zirconia                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Ti-6Al-4V ELI                                                                                                             | Y-TZP Zirconia                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Y-TZP Zirconia                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Screw                 | Provided by Pou Yu Biotechnology<br>Co. Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Provided by individual<br>manufacturers for specified<br>implants                                                         | Provided by individual<br>manufacturers for specified implants                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Provided by Atlantis Components, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |

{3}------------------------------------------------

# NON-CLINICAL TESTING DATA

Mechanical testing, according to ISO 14801 Dentistry - Fatigue test for endosseous dental implants, was conducted on a worst-case scenario with construction of an angle abutment to ensure that the strength of TDS Abutment for Friadent Xive is appropriate for its intended use.

Compatibility testing was conducted on the abutments and corresponding dental implants with designated screws, the dimensions, tolerances and rotation parameters were evaluated in determining appropriate fit.

These testing results show that TDS Abutment for Friadent Xive made of titanium, zirconia-titanium and zirconia materials for their respective dental implant systems have sufficient mechanical strength for their intended clinical application and are compatible with the implant systems for which they are indicated for use.

### CONCLUSION

Pou Yu Biotechnology Co. Ltd demonstrated that, for the purposes of FDA's regulation of medical devices, TDS Abutment for Friadent Xive is substantially equivalent in indications and design principles to predicate devices, each of which has been determined by FDA to be substantially equivalent to pre-amendment devices.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three wing-like shapes, symbolizing health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Daniel Tsao Pou Yu Biotechnology Company, Limited No. 6, Fugong Road Fusing Township, Changhua City China Taiwan 506

MAR 3 1 2011

Re: K103339

Trade/Device Name: TDS Abutment for Friadent Xive Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment · Regulatory Class: II Product Code: NHA Dated: March 17, 2011 Received: March 17, 2011

Dear Mr. Tsao:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2- Mr. Tsao

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

hh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{6}------------------------------------------------

#### 510(k) Premarket Notification

# Indications for Use

103339 510(k) Number (if known):

Device Name: TDS Abutment for Friadent Xive

Indications for Use:

TDS Abutment for Friadent Xive is intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.

This device is compatible with the following implant systems which have an internal hex with flat-to-flat dimensions of 1.78mm or greater: Firadent: FRIALIT Implant, XiVA Implant; 3i: Internal Connect Type; Astra: Osseospeed Implant, Osseospeed TX Implant; BioHorizons: Internal Implant System, Tapered Internal Implant System, Single-Stage Implant System, Laser-lok® 3.0 implant system; Lifecore: Lifecore RENOVA™ Internal Hex Implant System; Zimmer: Tapered Screw-Vent Implant System, Screw-Vent Implant System, AdVent Implant System; Osstem: GS System; Nobel Biocare: Active Implant.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Qffice of Device Evaluation (ODE)

Susan Kurpe

(Division Sign-Off) Division of Anesthesiology, General Hospital infection Control, Dental Devices

Page 1 of 1

510(k) Number: K103339

---

**Source:** [https://fda-staging.innolitics.com/device/K103339](https://fda-staging.innolitics.com/device/K103339)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).