K103338 · New Star Lasers, Inc. · PDZ · Nov 1, 2011 · General, Plastic Surgery
Device Facts
Record ID
K103338
Device Name
COOLTOUCH
Applicant
New Star Lasers, Inc.
Product Code
PDZ · General, Plastic Surgery
Decision Date
Nov 1, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CoolTouch VariaBreeze with CoolBreeze® Handpiece is intended for: Podiatry: Ablation, vaporization, incision, excision, and coagulation of soft tissue including matrixectomy, periungal and subungal warts, plantar warts, radical nail excision, and neuromas. The CoolTouch VariaBreeze with Toenail Handpiece is intended for: For use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.).
Device Story
CoolTouch VariaBreeze is a flashlamp-pumped Nd:YAG surgical laser system (1064 nm). Device consists of a cabinet housing a power supply, cooling system, and microcontroller-based control panel. Operates via fiber optic delivery systems: CoolBreeze handpiece (with cryogen cooling and adjustable 2-10mm spot size) for soft tissue podiatry; Toenail handpiece for onychomycosis treatment. Used in clinical settings by physicians. Aiming beam (532 nm) assists targeting. Output is controlled via footswitch. Laser energy facilitates tissue ablation or temporary nail clearance. System includes remote interlock and emergency stop for safety.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Nd:YAG laser medium; 1064 nm wavelength; 532 nm diode aiming beam. Output power up to 30W; pulse duration 100-700 μsec; repetition rate 5-100 Hz. Spot size 2-10 mm (CoolBreeze) or 1 mm (Toenail). System includes cryogen cooling for CoolBreeze handpiece. Microcontroller-based control system. Console dimensions 31"H x 18"W x 21"D; weight 160 lbs. Power requirements 115/230VAC, 50/60Hz.
Indications for Use
Indicated for podiatric soft tissue procedures (ablation, vaporization, incision, excision, coagulation) including matrixectomy, warts, nail excision, and neuromas; also indicated for temporary increase of clear nail in patients with onychomycosis (dermatophytes or yeasts).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
PinPointe USA, Inc. PinPointe FootLaser Nd:YAG laser system (K093457)
Submission Summary (Full Text)
{0}------------------------------------------------
K103338
pg 1 of 3
## NOV - 1 2011
#### Premarket Notification 510(k) Summary (As Required by 21 CFR 807.93)
This 510(k) Summary of safety and effectiveness for the New Star Model CoolTouch VariaBreeze Nd:YAG Surgical Laser system is submitted in accordance with the requirements of the SMDA 1990 and following guidance concerning the organization and content of a 510(k) summary.
| Submitter: | New Star Lasers, Inc. d.b.a. CoolTouch Inc. |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 9085 Foothills Boulevard<br>Roseville, CA 95747 |
| Contact Person: | Natalie Vollrath<br>Quality and Regulatory Manager |
| Telephone/Fax/Email: | (916) 677-1912 - Phone<br>(916) 677-1901 - Fax<br>nvollrath@newstarlasers.com - Email |
| Date prepared: | September 30, 2011 |
| Device Trade Name: | CoolTouch VariaBreeze™ Nd:YAG Surgical Laser |
| Common Name: | Nd: YAG Surgical Laser |
| Classification Name: | Instrument, surgical, powered, laser<br>79-GEX<br>21 CFR §878.4810 |
| Legally Marketed Predicate Devices: | <ul><li>CoolTouch Varia Nd: YAG laser system (K092964)</li><li>PinPointe USA, Inc. PinPointe FootLaser Nd:YAG laser system (K093457)</li></ul> |
| Device Description: | The CoolTouch VariaBreeze Laser System is an Nd: YAG laser using a flashlamp-pumped solid state laser rod to produce laser emission at 1064 nm. The laser consists of a cabinet that houses the following: <ul><li>Control panel with microcontroller and emergency stop button, standby button, ready button, and key switch</li><li>Connector port for fiber optic delivery system</li><li>Remote interlock connector</li><li>630 to 680 nm diode aiming beam</li><li>Power supply</li><li>Cooling system</li><li>Connector port for footswitch and power cord</li></ul> |
{1}------------------------------------------------
K103338
pg 2 of 3
Delivery devices for the VariaBreeze include:
| | <ul><li>CoolBreeze Handpiece for CoolBreeze Mode—Contains non-replaceable fiber optic within a flexible cable connected to the laser. A standoff at the tip allows for non-contact use, and a control dial is used to set the spot size from 2mm to 10mm. The handpiece also has cryogen for topical cooling. The handpiece is cleanable and reusable.</li><li>Toenail Handpiece for Toenail Mode—The fiber optic cable attaches to the laser. The handpiece features a standoff at the tip to provide the correct placement for laser delivery to the treatment site and is reusable and cleanable.</li></ul> |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | The CoolTouch VariaBreeze with CoolBreeze® Handpiece is intended for:<br>Podiatry: Ablation, vaporization, incision, excision, and coagulation of soft tissue including matrixectomy, periungal and subungal warts, plantar warts, radical nail excision, and neuromas.<br>The CoolTouch VariaBreeze with Toenail Handpiece is intended for:<br>For use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.). |
| Comparison: | The Cooltouch VariaBreeze has the same principle of operation, the same wavelength and essentially the same pulse energy rate as the predicate devices. |
| Nonclinical Perfomance Data | None |
| Clinical Performance Data: | None |
{2}------------------------------------------------
K1.03338 pg 3 of 3
### Technological Characteristics:
| Characteristic | CoolTouch<br>Varia<br>Laser System<br>K092964 | CoolTouch<br>Varia Breeze<br>Laser System<br>K103338 |
|-----------------------|-----------------------------------------------------------------|------------------------------------------------------|
| Laser Medium | Nd:YAG | Nd:YAG |
| Wavelength | 1064 nm | 1064 nm |
| Aiming Beam | 532 nm | 532 nm |
| Output Power<br>(max) | 36W max | 30W |
| Energy per Pulse | Variable to 3.9 Joules | 20-200 mJ, 500-1000 mJ |
| Pulse Duration | 600 μsec | 100-700 μsec |
| Exposure | Single pulse or repeat for<br>continuous burst | Continuous |
| Repetition Rate | 6-200 Hz | 5-100 Hz |
| Spot Size | 2 - 10 mm | 2 - 10 mm |
| Dimensions | 31"H x 18"W x 21"D | 31"H x 18"W x 21"D |
| Weight | 160 lbs (73kg) console | 160 lbs (73 kg) console |
| Power<br>Requirements | 115VAC, 230VAC<br>50/60Hz | 115VAC, 230VAC<br>50/60Hz |
| Characteristic | Patholase<br>PinPointe<br>FootLaser<br>Laser Systems<br>K093547 | CoolTouch<br>Varia Breeze<br>Laser System<br>K103338 |
| Laser Medium | Nd:YAG | Nd:YAG |
| Wavelength | 1064 nm | 1064 nm |
| Aiming Beam | 630-680 nm | 532 nm |
| Output Power<br>(max) | 6W | 30W |
| Energy per Pulse | 20-<br>200<br>mJ | 20-200 mJ, 500-1000 mJ |
| Pulse Duration | 100-<br>700<br>µsec | 100-700 µscc |
| Exposure | Continuous | Continuous |
| Repetition Rate | 5-100 Hz | 5-100 Hz |
| Spot Size | 1 mm (published) | 1 mm |
| Dimensions | 14"H x 7"W x 16"D | 31"H x 18"W x 21"D |
| Weight | 36 lbs (16kg) console | 160 lbs (73 kg) console |
| Power<br>Requirements | 90-130VAC,<br>200-240 VAC<br>50/60 Hz | 115VAC, 230VAC<br>50/60Hz |
Conclusion:
The CoolTouch VariaBreeze Nd:YAG Laser System is substantially equivalent to the predicate devices for the indications requested.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 13, 2013
New Star Lasers, Inc. % Ms. Natalie Vollrath Quality and Regulatory Manager 9085 Foothills Boulevard Roseville, California 95747
Re: K103338
Trade/Device Name: CoolTouch VariaBreeze Nd:YAG Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: PDZ, GEX Dated: October 25, 2011 Received: October 25, 2011
Dear Mr. Vollrath:
This letter corrects our substantially equivalent letter of November 01, 2011.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
Page 2 - Ms. Natalie Vollrath
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, FOR
# Peter D. Rumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K1033.8
pg l of l
#### Indications For Use Statement
510(k) Number:
Device Name:
CoolTouch VariaBreeze Nd:YAG Laser System
Indications for Use:
#### For CoolTouch VariaBreeze™ with CoolBreeze® Handpiece:
Podiatry: Ablation, vaporization, incision, excision, and coagulation of soft tissue including matrixectomy, periungal and subungal warts, plantar warts, radical nail excision, and neuromas.
#### For CoolTouch VariaBreeze™ with Toenail Handpiece:
Indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.).
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
- 510(k) Number. K103358
11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.