← Product Code [BZR](/productcode/BZR) · K103317

# RESUSCITATION UNIT 105 (K103317)

_Atom Medical Corporation · BZR · Apr 20, 2011 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K103317

## Device Facts

- **Applicant:** Atom Medical Corporation
- **Product Code:** [BZR](/productcode/BZR.md)
- **Decision Date:** Apr 20, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5330
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Resuscitation Unit 105 PV and Resuscitation Unit 105 EV are intended to remove secretions from the neonatal / infant airways and to provide blended air and oxygen at a desired oxygen concentration in the range of 21-100% to a respiratory assist device connected to the patient such as a ventilator (including a manually operated one) and an oxygen tent.

## Device Story

Resuscitation Unit 105 (models 105PV and 105EV) provides blended air/oxygen and suction for neonates/infants. Inputs: compressed air and oxygen; 105PV uses wall vacuum, 105EV uses internal venturi. Operation: manual blending of gases; pressure regulation via relief valve; monitoring via integral pressure gauge. Output: blended gas mixture to respiratory devices (tents, masks, ventilators) and suction for airway clearance. Used in NICU/delivery suites by trained clinicians. Output allows clinicians to manage oxygen concentration and airway pressure; benefits include improved respiratory support and secretion management.

## Clinical Evidence

Bench testing only. Testing included flow accuracy, pressure gauge accuracy, O2 concentration (ISO 10651-5), and suction performance (ISO 10079-3: overfill, seal, aspiration, pressure, noise). Results demonstrated device performance met all pass/fail criteria and was equivalent to predicate devices.

## Technological Characteristics

Gas-powered blender and suction unit. Features: pressure relief valve, pressure gauge, suction pressure gauge, dual outlets. 105PV: wall vacuum; 105EV: venturi vacuum. Gas sources: compressed air/oxygen (206-517 kPa). Blender range: 21-100% O2 (±3%). Flow rate: <15 lpm. Standards: ISO 10651-5 (resuscitators), ISO 10079-3 (suction). Mechanical/pneumatic operation; no software/electronics described.

## Regulatory Identification

A breathing gas mixer is a device intended for use in conjunction with a respiratory support apparatus to control the mixing of gases that are to be breathed by a patient.

## Predicate Devices

- Atom V-505 Infant Warmer ([K060461](/device/K060461.md))
- Bio-Med Devices BMD air/oxygen blender ([K925982](/device/K925982.md))
- Precision Medical Precision Blender ([K053232](/device/K053232.md))
- GE Giraffe and Panda system ([K070210](/device/K070210.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

.K103317

## 510(k) Summary Page 1 of 8 5-Apr-11

APR 2 0 2011

| Atom Medical Corporation<br>3-18-15, Hongo, Bunkyo-ku<br>Tokyo, Japan 113-0033 | Tel - 011-81-3-3815-3632<br>Fax - 011-81-3-3812-3199                                                                                                                                       |
|--------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact:                                                              | Tsuyoshi Sugino - Regulatory Affairs Manager                                                                                                                                               |
| Proprietary or Trade Name:                                                     | Resuscitation Unit 105                                                                                                                                                                     |
| Common/Usual Name:                                                             | Air / oxygen blender                                                                                                                                                                       |
| Classification Name/Code:                                                      | BZR - Breathing gas mixer<br>CFR 868.5330                                                                                                                                                  |
| Device:                                                                        | Resuscitation Unit 105EV<br>Resuscitation Unit 105PV                                                                                                                                       |
| Predicate Devices:                                                             | Bio-Med Devices_BMD_air/oxygen_blender_K925982<br>Precision Medical - Precision Blender - K053232<br>Atom Medical - V-505 Infant Warmer - K060461<br>GE Giraffe and Panda system - K070210 |

## Device Description:

There are two (2) models of the Resuscitation Unit 105 air / oxygen blender with suction.

- Resuscitation Unit 105PV .
- . Resuscitation Unit 105EV

They have the same basic design. Table 1 outlines the basic features and differences of each model.

## Indications for Use:

The Resuscitation Unit 105 EV and Resuscitation Unit 105 PV are intended to remove secretions from the neonatal / infant airways and / or to provide blended air and oxygen at a desired oxygen concentration in the range of 21-100% to a respiratory assist device connected to the patient such as a ventilator (including a manually operated one) and an oxygen tent.

Environment of Use: Hospitals, delivery suites, NICU

{1}------------------------------------------------

## 510(k) Summary Page 2 of 8 5-Apr-11

## Table 1 Resuscitation Models 105PV and 105EV

| Feature                | Resuscitation Unit 105PV                                                                                                                                                                                                | Resuscitation Unit 105EV       |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| Suction(Piping Type)   | Yes                                                                                                                                                                                                                     | No                             |
| Suction(Venturi Type)  | No                                                                                                                                                                                                                      | Yes                            |
| Oxygen / Air Blender   | Yes                                                                                                                                                                                                                     | Yes                            |
| Inlet Gas Sources      | Oxygen compressed gas                                                                                                                                                                                                   | Oxygen compressed gas          |
|                        | Air compressed gas                                                                                                                                                                                                      | Air compressed gas             |
|                        | Suction compressed gas                                                                                                                                                                                                  |                                |
| Pressure Relief Valve  | Yes                                                                                                                                                                                                                     | Yes                            |
| Pressure Gauge         | Yes                                                                                                                                                                                                                     | Yes                            |
| Suction Pressure Gauge | Yes                                                                                                                                                                                                                     | Yes                            |
| Outlet                 | Double outlets<br>(Outlet A/B)                                                                                                                                                                                          | Double outlets<br>(Outlet A/B) |
|                        | Outlet A / B is for delivery of the air/oxygen gas mixture to a<br>variety of devices, i.e., oxygen tent, oxygen mask, nasal cannula,<br>manual resuscitator or manual ventilator, in-line humidifier jar,<br>etc.      |                                |
|                        | Airway Pressure tube and adapter can be connected at the patient<br>connected to measure and monitor the airway pressure which is<br>displayed on the integral pressure gauge as well as a pressure<br>regulating valve |                                |

As outlined in Table 1, the Resuscitation Unit 105 EV and Resuscitation Unit 105 PV are identical except for the source for vacuum. Model 105PV is from a wall vacuum source and Model 105EV is from an internal venturi.

Table 2 compares the Resuscitation Unit 105 EV and Resuscitation Unit 105 PV to the predicates:

- Atom V-505 Infant warmer with optional oxygen / suction module (K060461) .
- . Bio-Med Devices – BMD air/oxygen blender (K925982) a blender which is incorporated into the proposed devices
- . Precision Medical – Precision Blender (K053232) which has the indications for use and product classification we are seeking.

Table 3 compares the Resuscitation Unit 105 EV and Resuscitation Unit 105 PV to the predicates Atom V-505 Infant warmer with optional oxygen / suction module (K060461) and the GE Giraffe and Panda system (K070210) which incorporates suction function and features for the intended patient population and environment.

{2}------------------------------------------------

510((k) Summary

S-Apr-11

# able of the Similarities and Differences to the Predicates for Air / Oxygen Blender feature

| General Attributes                                               | Resuscitation Unit<br>Unit 105PV<br>Unit 105EV                                                                                                                                                                                                                                                                                                                                                                 | Atom V-505 Infant<br>Warmer with oxygen /<br>suction module K060461                                                                                                                                                                                                                                                                                                                                                        | Bio-Med Devices<br>BMD air/oxygen<br>blender K925982                                                                                                                                                                         | Precision Medical<br>Precision Blender<br>K053232                                                                                                                                                          |                                                                                                    |
|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| Indications for Use                                              | The Resuscitation Unit<br>105 EV and<br>Resuscitation Unit 105<br>PV are intended to remove<br>secretions from the<br>neonatal / infant airways or<br>to provide blended air and<br>oxygen at a desired oxygen<br>concentration in the range<br>of 21-100% to a respiratory<br>assist device connected to<br>the patient such as a<br>ventilator (including a<br>manually operated one)<br>and an oxygen tent. | The blender and suction<br>modules incorporated in the<br>V-505 Infant Warmer are<br>intended to remove<br>secretions from the neonatal<br>/ infant airways or to<br>provide blended air and<br>oxygen at a desired oxygen<br>concentration in the range of<br>21-100% to a respiratory<br>assist device connected to<br>the patient such as a<br>ventilator (including a<br>manually operated one) and<br>an oxygen tent. | There is no indications<br>for sue statement in the<br>FDA database but the<br>BMD air/oxygen<br>blender is intended to<br>provide blended air and<br>oxygen at a desired<br>oxygen concentration in<br>the range of 21-100% | Is intended to deliver<br>blended air and oxygen in<br>a hospital setting.<br>Oxygen concentrations<br>can be dialed in from 21<br>to 100%. The blender is<br>not intended as a life<br>supporting device. |                                                                                                    |
| Patient Population                                               | Neonates and Infants                                                                                                                                                                                                                                                                                                                                                                                           | Neonates and Infants                                                                                                                                                                                                                                                                                                                                                                                                       | Not specified                                                                                                                                                                                                                | Not specified                                                                                                                                                                                              |                                                                                                    |
| Environment of Use                                               | Hospital, delivery suites,<br>NICU                                                                                                                                                                                                                                                                                                                                                                             | Hospital, delivery suites,<br>NICU                                                                                                                                                                                                                                                                                                                                                                                         | Hospital, delivery<br>suites, NICU                                                                                                                                                                                           | Hospital setting                                                                                                                                                                                           |                                                                                                    |
| Prescriptive                                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                        | Yes                                                                                                                                                                                                                          | Yes                                                                                                                                                                                                        |                                                                                                    |
|                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                | Page 4 of 8<br>5-Apr-11                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                              |                                                                                                                                                                                                            |                                                                                                    |
|                                                                  | Resuscitation Unit<br>Unit 105PV<br>Unit 105EV                                                                                                                                                                                                                                                                                                                                                                 | Atom V-505 Infant<br>Warmer with oxygen /<br>suction module K0604618                                                                                                                                                                                                                                                                                                                                                       | Bio-Med Devices<br>BMD air/oxygen<br>blender K925982                                                                                                                                                                         | Precision Medical<br>Precision Blender<br>K053232                                                                                                                                                          |                                                                                                    |
| General Attributes (continued)                                   | Outlet connections to                                                                                                                                                                                                                                                                                                                                                                                          | Oxygen tent<br>Oxygen mask<br>Cannula<br>Manual ventilator<br>Manual resuscitator                                                                                                                                                                                                                                                                                                                                          | Oxygen tent<br>Oxygen mask<br>Cannula<br>Manual ventilator<br>Manual resuscitator                                                                                                                                            | Oxygen tent<br>Oxygen mask<br>Cannula<br>Manual ventilator<br>Manual resuscitator<br>None supplied                                                                                                         | Oxygen tent<br>Oxygen mask<br>Cannula<br>Manual ventilator<br>Manual resuscitator<br>None supplied |
|                                                                  | Accessories/Components                                                                                                                                                                                                                                                                                                                                                                                         | Airway pressure tube<br>Hoses and tubing<br>Humidifier jar<br>Suction jar                                                                                                                                                                                                                                                                                                                                                  | Hoses and tubing<br>Humidifier jar<br>Suction jar                                                                                                                                                                            |                                                                                                                                                                                                            |                                                                                                    |
| Technical specifications                                         | Oxygen / air supply module                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                              |                                                                                                                                                                                                            |                                                                                                    |
|                                                                  | Gas sources                                                                                                                                                                                                                                                                                                                                                                                                    | Air/O₂<br>300-500kPa                                                                                                                                                                                                                                                                                                                                                                                                       | Air/O₂<br>206-517kPa                                                                                                                                                                                                         | Air/O₂<br>206-517kPa                                                                                                                                                                                       | Air/O₂<br>206-517kPa                                                                               |
|                                                                  | input pressure                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                              |                                                                                                                                                                                                            |                                                                                                    |
|                                                                  | Airway pressure<br>manometer<br>Range Accuracy                                                                                                                                                                                                                                                                                                                                                                 | -20 to 80cmH₂O<br>±1% of full scale                                                                                                                                                                                                                                                                                                                                                                                        | N/A                                                                                                                                                                                                                          | N/A                                                                                                                                                                                                        | N/A                                                                                                |
|                                                                  | Air/O₂ blender<br>Range<br>Accuracy                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                            | 21-100%<br>±3%O₂                                                                                                                                                                                                             | 21-100%<br>±3%O₂                                                                                                                                                                                           | 21-100%<br>±3%O₂                                                                                   |
|                                                                  | Alarm                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                            | Differential pressure<br>between oxygen and air<br>supply                                                                                                                                                                    | Differential pressure<br>between oxygen and air<br>supply                                                                                                                                                  | Differential pressure<br>between oxygen and air<br>supply                                          |
|                                                                  | Bleed flow rate                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                            | 2.5-3.5 lpm                                                                                                                                                                                                                  | N/A                                                                                                                                                                                                        | N/A                                                                                                |
|                                                                  | Gas flow rate range                                                                                                                                                                                                                                                                                                                                                                                            | < 15 lpm                                                                                                                                                                                                                                                                                                                                                                                                                   | < 15 lpm                                                                                                                                                                                                                     | ≥ 30 lpm                                                                                                                                                                                                   | ≥ 30 lpm                                                                                           |
|                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                | 510(k) Summary<br>5 of 8<br>5-Apr-11                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                              |                                                                                                                                                                                                            |                                                                                                    |
|                                                                  | Resuscitation Unit<br>Unit 105PV<br>Unit 105EV                                                                                                                                                                                                                                                                                                                                                                 | Atom V-505 Infant<br>Warmer with oxygen /<br>suction module K060461                                                                                                                                                                                                                                                                                                                                                        | Bio-Med Devices<br>BMD air/oxygen<br>blender K925982                                                                                                                                                                         | Precision Medical<br>Precision Blender<br>K053232                                                                                                                                                          |                                                                                                    |
| Technical specifications (continued)                             |                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                              |                                                                                                                                                                                                            |                                                                                                    |
| Relieving valve<br>(for preventing excessive<br>airway pressure) | (Default setting : 5.9kPa<br>(60cmH2O))                                                                                                                                                                                                                                                                                                                                                                        | N/A (no reliving valve)                                                                                                                                                                                                                                                                                                                                                                                                    | N/A(no relieving valve)                                                                                                                                                                                                      | N/A(no relieving valve)                                                                                                                                                                                    |                                                                                                    |
| Flow outlet                                                      | Dual flow outlet to connect to<br>a device requiring a gas<br>supply<br>Oxygen density of both is the<br>same.                                                                                                                                                                                                                                                                                                 | Single flow outlet to connect<br>to device requiring a gas<br>supply                                                                                                                                                                                                                                                                                                                                                       | Flow outlet for<br>connecting to device<br>requiring a gas supply                                                                                                                                                            | Flow outlet for connecting<br>to device requiring a gas<br>supply                                                                                                                                          |                                                                                                    |
| Standards                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                              |                                                                                                                                                                                                            |                                                                                                    |
| ISO 10651-5 Gas<br>powered resuscitators                         | Yes, applicable sections                                                                                                                                                                                                                                                                                                                                                                                       | N/A                                                                                                                                                                                                                                                                                                                                                                                                                        | N/A                                                                                                                                                                                                                          | N/A                                                                                                                                                                                                        |                                                                                                    |
| ISO 10079-3 suction<br>equipment                                 | Yes, applicable sections                                                                                                                                                                                                                                                                                                                                                                                       | N/A                                                                                                                                                                                                                                                                                                                                                                                                                        | N/A                                                                                                                                                                                                                          | N/A                                                                                                                                                                                                        |                                                                                                    |
| Section 514                                                      | None                                                                                                                                                                                                                                                                                                                                                                                                           | None                                                                                                                                                                                                                                                                                                                                                                                                                       | None                                                                                                                                                                                                                         | None                                                                                                                                                                                                       |                                                                                                    |

Page 52

---

{3}------------------------------------------------

510(k) Summary

.

.

Page 53

-. r

.

{4}------------------------------------------------

and the contraction of the comments of the control of

.

{5}------------------------------------------------

510(k) Summary 

## able of the Similarities and Differences to the Predicates for Suction features

| General Attributes                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                         |                                 |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|
| Indications for Use                                                                                                                                                                                                                                 | Resuscitation Unit<br>Unit 105PV<br>Unit 105EV                                                                                                                                                                                                                                                                                                                                                      | Atom V-505 Infant Warmer with<br>oxygen / suction module<br>K060461                                                                                                                                                                                                                                                                                                                                              | GE<br>Giraffe and Panda<br>K070210                                                                                                                                                                                                                                                                                      |                                 |
|                                                                                                                                                                                                                                                     | The Resuscitation Unit 105 EV<br>and Resuscitation Unit 105 PV are<br>intended to remove secretions from<br>the neonatal / infant airways and to<br>provide blended air and oxygen at a<br>desired oxygen concentration in the<br>range of 21-100% to a respiratory<br>assist device connected to the patient<br>such as a ventilator (including a<br>manually operated one) and an<br>oxygen tent. | The blender and suction modules<br>incorporated in the V-505 Infant<br>Warmer are intended to remove<br>secretions from the neonatal / infant<br>airways and to provide blended air<br>and oxygen at a desired oxygen<br>concentration in the range of 21-<br>100% to a respiratory assist device<br>connected to the patient such as a<br>ventilator (including a manually<br>operated one) and an oxygen tent. | Provides the basic equipment<br>required for pulmonary<br>resuscitation of infants.<br>Pulmonary resuscitation includes<br>practices necessary to establish a<br>clear airway and provide oxygen<br>and air/oxygen mixtures and/or<br>manual ventilation to the infant.<br>(Includes a suction function and<br>feature) |                                 |
| Patient Population                                                                                                                                                                                                                                  | Neonates and Infants                                                                                                                                                                                                                                                                                                                                                                                | Neonates and Infants                                                                                                                                                                                                                                                                                                                                                                                             | Infant                                                                                                                                                                                                                                                                                                                  |                                 |
| Environment of Use                                                                                                                                                                                                                                  | Hospital, delivery suites, NICU                                                                                                                                                                                                                                                                                                                                                                     | Hospital, delivery suites, NICU                                                                                                                                                                                                                                                                                                                                                                                  | Hospital, delivery suites, NICU                                                                                                                                                                                                                                                                                         |                                 |
| Prescriptive                                                                                                                                                                                                                                        | Persons trained in infant / neonate<br>resuscitation                                                                                                                                                                                                                                                                                                                                                | Persons trained in infant / neonate<br>resuscitation                                                                                                                                                                                                                                                                                                                                                             | Persons trained in infant / neonate<br>resuscitation                                                                                                                                                                                                                                                                    |                                 |
| Technical specifications                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                         |                                 |
| Suction bottle capacity                                                                                                                                                                                                                             | Up to 1200 ml                                                                                                                                                                                                                                                                                                                                                                                       | Not specified                                                                                                                                                                                                                                                                                                                                                                                                    | Not specified other than various<br>sizes                                                                                                                                                                                                                                                                               |                                 |
| Dimensions                                                                                                                                                                                                                                          | 200mm(W)x180 mm(D)x370mm (H)                                                                                                                                                                                                                                                                                                                                                                        | Built-in to warmer, not standalone                                                                                                                                                                                                                                                                                                                                                                               | 340mm(W)x250mm(D)x250mm(H)                                                                                                                                                                                                                                                                                              |                                 |
| Weight                                                                                                                                                                                                                                              | 7kg                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                  | 9 kg                                                                                                                                                                                                                                                                                                                    |                                 |
| Operating and Storage<br>limits                                                                                                                                                                                                                     | -20°C to 65°C<br>Up to 97% RH                                                                                                                                                                                                                                                                                                                                                                       | -20°C to 65°C<br>Up to 97% RH                                                                                                                                                                                                                                                                                                                                                                                    | -20°C to 60°C<br>Up to 95% RH                                                                                                                                                                                                                                                                                           |                                 |
| 510(k) Summary<br>Page 7 of 8<br>5-Apr-11                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                         |                                 |
|                                                                                                                                                                                                                                                     | Resuscitation Unit<br>Unit 105PV<br>Unit 105EV                                                                                                                                                                                                                                                                                                                                                      | Atom V-505 Infant Warmer with<br>oxygen / suction module K060461                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                         | GE<br>Giraffe and Panda K070210 |
| Technical specifications (continued)                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                         |                                 |
|                                                                                                                                                                                                                                                     | Vacuum source                                                                                                                                                                                                                                                                                                                                                                                       | Model 105PV - wall suction<br>vacuum                                                                                                                                                                                                                                                                                                                                                                             | Model 105PV - wall suction<br>vacuum                                                                                                                                                                                                                                                                                    | Wall suction<br>vacuum          |
|                                                                                                                                                                                                                                                     | Vacuum range                                                                                                                                                                                                                                                                                                                                                                                        | Model 105EV - venturi method<br>0-200 mmHg                                                                                                                                                                                                                                                                                                                                                                       | 0-200 mmHg                                                                                                                                                                                                                                                                                                              | 0-150 mmHg                      |
|                                                                                                                                                                                                                                                     | Vacuum Flow range                                                                                                                                                                                                                                                                                                                                                                                   | ~ 20 lpm                                                                                                                                                                                                                                                                                                                                                                                                         | N/A                                                                                                                                                                                                                                                                                                                     | 15 lpm                          |
|                                                                                                                                                                                                                                                     | Standards                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                         |                                 |
|                                                                                                                                                                                                                                                     | ISO 10651-5 Gas<br>powered resuscitators                                                                                                                                                                                                                                                                                                                                                            | Yes, applicable sections                                                                                                                                                                                                                                                                                                                                                                                         | N/A                                                                                                                                                                                                                                                                                                                     | Yes, applicable sections        |
|                                                                                                                                                                                                                                                     | ISO 10079-3 - Suction<br>equipment                                                                                                                                                                                                                                                                                                                                                                  | Yes, applicable sections                                                                                                                                                                                                                                                                                                                                                                                         | N/A                                                                                                                                                                                                                                                                                                                     | Yes, applicable sections        |
|                                                                                                                                                                                                                                                     | Section 514                                                                                                                                                                                                                                                                                                                                                                                         | None                                                                                                                                                                                                                                                                                                                                                                                                             | None                                                                                                                                                                                                                                                                                                                    | None                            |
| Summary of substantial equivalence - Resuscitation Unit 105PV and Resuscitation Unit 105EV                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                         |                                 |
| Indications for Use - The Resuscitation Unit 105 EV and Resuscitation Unit 105 PV have equivalent indications for use to the predicates. That<br>is as a breathing gas mixer (blender) with a suction feature.                                      |                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                         |…

---

**Source:** [https://fda-staging.innolitics.com/device/K103317](https://fda-staging.innolitics.com/device/K103317)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com