K103310 · 3jt Enterprises, LLC · GEH · Feb 9, 2011 · General, Plastic Surgery
Device Facts
Record ID
K103310
Device Name
CRYONIZE TM
Applicant
3jt Enterprises, LLC
Product Code
GEH · General, Plastic Surgery
Decision Date
Feb 9, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
CryoNize™ is indicated, for medical professional use only, in the treatment of Verruca Vulgaris, Verruca Plantaris, Condylomà Acuminata, Verruca Plana, Molluscum Contagiosum, Skin Tags, Seborrhoeic Keratosis, Actinic Keratosis, and Lentigo.
Device Story
CryoNize™ is a portable aerosol cryosurgical system for treating various skin lesions. The device consists of an aerosol canister containing a cryogen mixture (95% dimethyl ether, 2% propane, 3% isobutane) and specialized foam tip applicators. The clinician operates the device by dispensing cryogen to saturate a foam tip, which is then applied directly to the target lesion to induce freezing. The system is intended for use by medical professionals in a clinical setting. By freezing the lesion, the device facilitates the treatment of conditions such as warts, skin tags, and keratosis. The device is provided with an instruction booklet and quick guide to assist the clinician in safe application.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Portable aerosol cryosurgical system. Cryogen mixture: 95% dimethyl ether, 2% propane, 3% isobutane (118 ml). Includes foam tip applicators (various sizes/shapes) and zone applicators. Mechanical operation via manual dispensing of cryogen onto foam tips. No electronic components, software, or energy source required.
Indications for Use
Indicated for medical professional use in the treatment of Verruca Vulgaris, Verruca Plantaris, Condyloma Acuminata, Verruca Plana, Molluscum Contagiosum, Skin Tags, Seborrhoeic Keratosis, Actinic Keratosis, and Lentigo.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
FEB - 9 2011
## PREMARKET NOTIFICATION Traditional 510(K) SUMMARY
K103310
| Submitter's Name: | 3JT Enterprises, LLC., dba CryoNize |
|-------------------|----------------------------------------------------|
| Address: | 2000 N. Racine Ave., Suite 3100, Chicago, IL 60614 |
| Telephone Number: | (312)-952-5462 |
| Fax Number: | 866-379-5579 |
| Contact Person: | Anthony K. Goldboss, M.D. |
| Date Prepared: | November 7th, 2010 |
### Device Name:
| Proprietary Name: | CryoNize™ |
|-------------------|-----------|
| Usual Name: | CryoNize™ |
#### Devices to Which Substantial Equivalence is Claimed:
Histofreezer® Device, by OraSure Technologies, Inc. (Primary Predicate including general labeling and flammability label) -- K023487, K990877, K971392, K93327, K931299, K924114, K911420, K982358.
Verruca-Freeze Cryosurgery Delivery System™, by CryoSurgery, Inc. (Secondary Predicate including labeling) - K982506, K955083, K944221, K881349.
#### Device Classification:
CryoNize™ is a cryogenic device classified under 21 CFR 878.4350 as a Class II medical device. The general and plastic surgery panel cryogenic classification name for a cryosurgical unit and accessories is product code GEH.
#### Description of Device:
CryoNize™ is a cryosurgical system used for the treatment of Verruca Vulgaris, Verruca Plantaris, Condyloma Acuminata, Verruca Plana, Molluscum Contagiosum, Skin Tags, Seborrhoeic Keratosis, Actinic Keratosis, and Lentigo.
{1}------------------------------------------------
#### lt consists of:
- An aerosol canister filled with a liquid mixture of compressed gases. This mixture, or o cryogen, is composed of 95% dimethyl ether, 2% propane, and 3% isobutane (118 ml).
- Customized applicators ০
- CryoNize™ Foam Tip Applicators-two different shapes/sizes .
- CryoNize™ Zone Applicators—six different sizes (3mm, 5mm, 7mm, 12mm, 15mm) .
- Instructions for Use Booklet O
- Quick Guide for Use o
- o Carrying Case
- Practice Mouse Pad o
#### Performance Standards:
To date, no performance standards have been finalized which affect this device.
#### Intended Use Statement:
CryoNize™ is indicated, for medical professional use only, in the treatment of Verruca Vulgaris, Verruca Plantaris, Condylomà Acuminata, Verruca Plana, Molluscum Contagiosum, Skin Tags, Seborrhoeic Keratosis, Actinic Keratosis, and Lentigo.
#### Comparision of Technological Characteristics:
CryoNize™ is for the treatment of Verruca Vulgaris, Verruca Plantaris, Condyloma Acuminata, Verruca Plana, Molluscum Contagiosum, Skin Tags, Seborrhoeic Keratosis, Actinic Keratosis, and Lentigo. It is substantially equivalent to the Histofreezer® Device (K023487) and the Verruca-Freeze™ Cryosurgery Delivery System (K982506).
All three devices are portable aerosol cryosurgical systems that contain a canister filled with cryogen and foam tip applicators. When the cryogen is dispensed, it saturates the foam tip. The foam tip is then applied directly to the desired area of treatment. Subsequently, this freezes the skin lesion. These devices are substantially equivalent.
CryoNize™ has a second type of applicator which can be used, like Verruca-Freeze™ Cryosurgery Delivery System. CryoNize™ Zone Applicators are substantially equivalent to Verruca-Freeze™ Cryosurgery Delivery System Limiting Cones.
{2}------------------------------------------------
### Labeling:
The labeling of CryoNize™ has been prepared to ensure the medical professional has adequate and clear instructions for safety and usage. It includes carrying case and canister labeling, Instructions for Use Booklet, and a Quick Guide for Use. The labeling, safety and warning statements are substantially equivalent to the predicate devices.
#### Conclusion:
Based on the information presented above, it is concluded that the proposed CryoNize™ is safe and effective for its intended use and is substantially equivalent to the primary and secondary predicate devices. Also, the labeling is substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
3JT Enterprises, LLC % Anthony K. Goldboss, M.D. 2000 North Racine Avenue, Suite 3100 Chicago, Illinois 60614
Re: K103310
FEB - 9 201
Trade/Device Name: CyroNize™ Regulation Number: 21 CFR 878.4350 Regulation Name: Cyrosurgical unit and accessories Regulatory Class: Class II Product Code: GEH Dated: January 11, 2011 Received: January 18, 2011
Dear Dr. Goldboss:
. . .
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
:
{4}------------------------------------------------
Page 2 -- Anthony K. Goldboss, M.D.
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Mark N. Melkerson Director Division of Surgical. Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Form
510(k) Number (if known): K103310
CryoNize™ Device Name:
Indication for Use: CryoNize™ is indicated for use in the treatment of the following:
Verruca Vulgaris, Verruca Plantaris, Condyloma Acuminata, Verruca Plana, Molluscum Contagiosum, Skin Tags, Seborrhoeic Keratosis, Actinic Keratosis, and Lentigo.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Nishida for anka
Division SiunOm
(Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K/033/0
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k)
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.