K103309 · Smith & Nephew Endoscopy, Inc. · MAI · Feb 3, 2011 · Orthopedic
Device Facts
Record ID
K103309
Device Name
OSTEORAPTOR CURVED 2.3 SUTURE ANCHORS
Applicant
Smith & Nephew Endoscopy, Inc.
Product Code
MAI · Orthopedic
Decision Date
Feb 3, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Smith & Nephew OSTERAPTOR Curved Suture Anchor is intended for the reattachment of soft tissue to bone for the following indications: Hip Hip capsule repair - Acetabular labrum reattachment Shoulder Capsular stabilization - Bankart repair - Anterior shoulder instability - SLAP lesion repairs - Capsular shift or capsulolabral reconstructions Acromioclavicular separation repairs Deltoid repairs Rotator cuff tear repairs Biceps tenodesis Elbow, Wrist, and Hand Biceps tendon reattachment Ulnar or radial collateral ligament reconstructions Lateral epicondylitis repair Knee - Extra-capsular repairs: - Medial collateral ligament - Lateral collateral ligament - Posterior oblique ligament - Patellar realignment and tendon repairs – Vastus medialis obliquous advancement Iliotibial band tenodesis Foot and Ankle Hallux valgus repairs Medial or lateral instability repairs/reconstructions Achilles tendon repairs/reconstructions Midfoot reconstructions Metatarsal ligament/tendon repairs/reconstructions Bunionectomy
Device Story
Bioabsorbable 2.3 mm suture anchor; manufactured from PLLA/HA; preloaded with nonabsorbable ultra high molecular weight braided polyethylene #2 suture; preassembled to flexible stainless steel inserter. Used by surgeons in clinical settings for orthopedic soft tissue-to-bone fixation. Device provides mechanical anchor point for sutures to facilitate tissue healing. Performance characteristics include insertion, pull-out, and suture slide properties.
Clinical Evidence
Bench testing only. Performance testing demonstrated that insertion, pull-out, and suture slide properties are substantially equivalent to the predicate device.
Indicated for soft tissue-to-bone reattachment in hip, shoulder, elbow, wrist, hand, knee, foot, and ankle procedures. Includes repairs for labrum, capsule, ligaments, tendons, and bunionectomy. Prescription use only.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
K103309 p 1 of 3
Endoscopy Smith & Nephew, Inc. 150 Minuteman Road Andover, MA 01810
978 749 1000 978 749 1599 Fax www.smith-nephew.com
# >< We are smith&nephew
FEB - 3 2011
## SECTION IV 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
as required by the Safe Medical Devices Act of 1990 and codified in 21 CFR 807.92 upon which the substantial equivalence is based.
#### OSTEORAPTOR Curved 2.3 Suture Anchor
Date Prepared: November 8, 2010
#### A. Submitter's Name:
Smith & Nephew, Inc., Endoscopy Division
150 Minuteman Road
Andover, MA 01810
#### B. Company Contact
Kathleen Solomon
Regulatory Affairs Specialist II
Phone: (978) 749-1605
Fax: (978) 749-1443
#### C. Device Name
| Trade Name: | OSTEORAPTOR Curved 2.3 Suture Anchor |
|----------------------|---------------------------------------------------------------------------------|
| Common Name: | Fastener, fixation, biodegradable, soft tissue |
| Classification Name: | Single/multiple component metallic bone fixation<br>appliances and accessories. |
| Product Code: | MAI |
| Regulation Number: | 21 CFR §888.3030 |
#### D. Predicate Device
The Smith & Nephew OSTEORAPTOR Curved 2.3 Suture Anchors are substantially equivalent in Indication for Use and Fundamental Scientific Technology to the OSTEORAPTOR 2.3 Suture Anchor, K082215.
{1}------------------------------------------------
#### E. Description of Device
The Smith & Nephew OSTERAPTOR Curved 2.3 Suture Anchor is bioabsorbable 2.3 mm suture anchor manufactured from PLLA/HA and comes preloaded with nonabsorbable ultra high molecular weight braided polyethylene #2 suture preassembled to a flexible stainless steel inserter.
## F. Intended Use
The OSTERAPTOR Curved 2.3 suture anchors are intended for the fixation of soft tissue to bone in the Hip, Shoulder, Foot, Ankle, Elbow, Wrist, Hand and Knee as follows:
#### Hip
Hip capsule repair - Acetabular labrum reattachment
## Shoulder
Capsular stabilization
- Bankart repair
- Anterior shoulder instability
- SLAP lesion repairs
- Capsular shift or capsulolabral reconstructions Acromioclavicular separation repairs Deltoid repairs Rotator cuff tear repairs Biceps tenodesis
#### Elbow, Wrist, and Hand
Biceps tendon reattachment Ulnar or radial collateral ligament reconstructions Lateral epicondylitis repair
#### Knee
- Extra-capsular repairs:
- Medial collateral ligament
- Lateral collateral ligament
- Posterior oblique ligament
- Patellar realignment and tendon repairs
– Vastus medialis obliquous advancement Iliotibial band tenodesis
#### Foot and Ankle
Hallux valgus repairs Medial or lateral instability repairs/reconstructions Achilles tendon repairs/reconstructions Midfoot reconstructions Metatarsal ligament/tendon repairs/reconstructions Bunionectomy
{2}------------------------------------------------
## G. Comparison of Technological Characteristics
The Smith & Nephew OSTERAPTOR Curved 2.3 Suture Anchors are substantially equivalent in indications for use, technological characteristics, and are as safe and as effective as their currently marketed predicate device, the Smith & Nephew OSTERAPTOR 2.3 Suture Anchor (K082215).
## H. Summary Performance Data
The performance testing demonstrates that the insertion, pull out and suture slide properties of the OSTEORAPTOR Curved 2.3 anchors are substantially equivalent to the Smith & Nephew OSTEORAPTOR 2.3 anchors.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around it. Inside the circle is an emblem that resembles an abstract depiction of an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Smith & Nephew, Inc. Endoscopy Division % Ms. Kathleen Solomon Regulatory Affairs Specialist II 150 Minuteman Road Andover, Massachusetts 01810
FEB - 3 2011
Re: K103309
Trade/Device Name: OSTEORAPTOR Curved Suture Anchors Regulation Number: 21 CFR 888.3030 Regulation Name: Single / multiple component metallic bone fixation appliance and accessories Regulatory Class: Class II Product Code: MAI, JDR Dated: November 8, 2010
Received: November 10, 2010
Dear Ms. Solomon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
## Page 2 - Ms. Kathleen Solomon
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), vlease go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
AS B. R h
for.
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
K 103309
Elbow, Wrist, and Hand
reconstructions
Knee
Biceps tendon reattachment
Lateral epicondylitis repair
- Medial collateral ligament
- Lateral collateral ligament
- Posterior oblique ligament
Iliotibial band tenodesis
Patellar realignment and tendon repairs
- Vastus medialis obliquous advancement
Extra-capsular repairs:
Ulnar or radial collateral ligament
510(k) Number (if known):
Device Name: OSTEORAPTOR Curved Suture Anchors
Indications For Use:
The Smith & Nephew OSTERAPTOR Curved Suture Anchor is intended for the reattachment of soft tissue to bone for the following indications:
## Hip
Hip capsule repair - Acetabular labrum reattachment
#### Shoulder
Capsular stabilization
- Bankart repair
- Anterior shoulder instability
- SLAP lesion repairs - Capsular shift or capsulolabral
reconstructions Acromioclavicular separation repairs Deltoid repairs Rotator cuff tear repairs Biceps tenodesis
## Foot and Ankle
Hallux valgus repairs Medial or lateral instability repairs/reconstructions Achilles tendon repairs/reconstructions Midfoot reconstructions Metatarsal ligament/tendon repairs/reconstructions Bunionectomy
AND/OR
Over-The-Counter Use
(Per 21 CFR 801 Subpart D)
Prescription Use X
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Offjes of Device Evaluation (ODE)
(Division Sign Oft) Division of Surgical, Orthopedic, and Restorative Devices
22 of 126
Smith & Nephew Endoscopy OSTEORAPTOR Curved 2.3 Suture Anchor
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.