K103274 · Sentinelle Medical, Inc. · MOS · Aug 18, 2011 · Radiology
Device Facts
Record ID
K103274
Device Name
SENTINELLE ENDO COIL ARRAY FOR PELVIC IMAGING
Applicant
Sentinelle Medical, Inc.
Product Code
MOS · Radiology
Decision Date
Aug 18, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Endo Coil Array is a receive-only MRI coil array for imaging (including spectroscopy) in the prostate, cervix and colon and surrounding pelvic tissues in adult populations with clinically present anus and rectum or vagina. The Endo Coil Array can be used with many commercial MRI scanners. This device is for exclusive use by a qualified medical professional under the order of a physician. For use with GE, Toshiba and Siemens 1.5T MR systems.
Device Story
Reusable, rigid transrectal or transvaginal receive-only MRI coil array; cylindrical housing similar to ultrasound probes. Held by articulating stabilization arm attached to patient support (transporter or tabletop versions). Used in clinical MRI settings with GE, Toshiba, or Siemens 1.5T systems. Operated by qualified medical professionals. Device captures magnetic resonance signals from pelvic tissues; signals processed by external MRI scanner to produce diagnostic images and spectroscopy data. Output used by physicians for clinical assessment of prostate, cervix, and colon. Benefits include high-resolution pelvic imaging for diagnostic purposes.
Clinical Evidence
Bench testing only. Performance evaluated via non-clinical testing including SNR comparison, spectroscopy, and image uniformity per NEMA 9:2008 standards. No clinical data provided.
Technological Characteristics
Reusable, rigid transrectal/transvaginal receive-only MRI coil array. Cylindrical form factor. Compatible with GE, Toshiba, and Siemens 1.5T MR systems. High-level disinfection required. No internal software or active electronic processing; passive receive-only coil technology.
Indications for Use
Indicated for adult patients with clinically present anus, rectum, or vagina requiring MRI imaging or spectroscopy of the prostate, cervix, colon, and surrounding pelvic tissues. Must be used by qualified medical professionals. Contraindicated for patients without the specified anatomy.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
MEDRAD 1.5T Pelvic Imaging System Interface Device (K053042)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Sentinelle Medical. The logo consists of a geometric shape on the left and the words "sentinelle MEDICAL" on the right. The geometric shape is made up of lines and curves, and it resembles a stylized flower or star. The word "sentinelle" is in a smaller font than the word "MEDICAL".
AUG 18 2011
555 Richmond Street West
Suite 800, P.O. Box 301
oronto, Ontario, Canada, M5V 3B1
phone - 416-366-4994
fax - 416-594-9696
sentinellemedical.com
# 510(k) Summary
| Submitter: | Sentinelle Medical Inc. |
|----------------------|------------------------------------------------------------------------------------------------|
| Address: | 555 Richmond Street West, Suite 800<br>Toronto, Ontario, M5V 3B1<br>Canada |
| Contact: | Cameron Piron |
| Telephone: | 647-258-3601 |
| Date: | August 17th, 2011 |
| Trade Names: | Sentinelle Endo Coil Array for Pelvic Imaging |
| Common Names: | Sentinelle Pelvic Solution, eArray, Prostate Coil Array, Female Pelvic Coil Array |
| Classification Name: | Accessory to Magnetic Resonance Diagnostic Device,<br>Class II as described in 21 CFR 892.1000 |
| Product Code: | MOS |
| Predicate Devices: | K053042 - MEDRAD 1.5T Pelvic Imaging System Interface Device |
## Device Description:
The Endo Coil Array is a reusable, rigid transrectal or transvaginal receive-only MRI coil array similar to other products. The Coil Array housing has a cylindrical shape similar to transrectal and transvaginal ultrasound probes.
The Coil Array is held by an articulating Stabilization Arm that is attached to a Patient Support. The Patient Support is available in both Transporter and Tabletop versions, which dock to the scanner or are placed on top of the scanner's existing couch respectively.
## Intended Use:
The Endo Coil Array is a receive-only MRI coil array for imaging (including spectroscopy) in the prostate, cervix and colon and surrounding pelvic tissues in adult populations with clinically present anus and rectum or vagina. The Endo Coil Array can be used with many commercial MRI scanners. This device is for exclusive use by a qualified medical professional under the order of a physician. For use with GE, Toshiba and Siemens 1.5T MR systems.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Sentinelle Medical. The logo consists of a geometric shape on the left and the words "sentinelle MEDICAL" on the right. The geometric shape is made up of lines that form a symmetrical design. The word "sentinelle" is in a larger font than the word "MEDICAL".
555 Richmond Street West
Suite 800, P.O. Box 301
Toronto, Ontario, Canada, M5V 3B1
phone - 416-366-4994
fax - 416-594-9696
sentinellemedical.com
### Indications for Use:
The Sentinelle Endo Coil Array is a receive-only MRI coil array which must be used by a qualified medical professional to obtain images of (including spectroscopy images) the prostate, cervix and colon and the surrounding tissues in the pelvis. The Coil Array is a reusable transrectal or transvaginal device that can be high level disinfected. This product is for exclusive use on adult populations with clinically present anus and rectum or vagina. For use with GE, Toshiba and Siemens 1.5T MR systems.
#### Substantial Equivalence Summary:
| Proposed Device | Predicate Device | Substantially Equivalent Aspect |
|-----------------|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Endo Coil Array | MEDRAD Interface<br>Device | • Transrectal receive-only MRI coils for<br>imaging.<br>• Regions of interest: Prostate, Colon,<br>Cervix and surrounding tissue.<br>• For use exclusively by trained medical<br>professionals. |
#### Performance Testing:
Non-clinical performance testing was done to ensure the device performed equivalently to predicates in the areas of SNR comparison, spectroscopy, and image uniformity per NEMA 9:2008 with no additional concerns of safety or effectiveness.
#### Comparison to Predicate:
In the opinion of Sentinelle, the intended use and technological characteristics of the Sentinelle Endo Coil Array are substantially equivalent to those of the predicate devices and do not pose any issues for its safety and effectiveness.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three stripes forming its body and wing. The eagle faces right and is positioned to the right of the text. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the left side of the eagle.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Sentinelle Medical. Inc. c/o Casev Conrv Senior Project Engineer Underwriters Laboratories, Inc. 1285 Walt Whitman Road MELVILLE NY 11747
AUG 1 8 2011
Re: K103274
> Trade Name: Sentinelle Endo Coil Array for Pelvic Imaging Regulation Number: 21 CFR. 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: March 25, 2011 Received: July 26, 2011
Dear Mr. Conry:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of
{3}------------------------------------------------
Page 2
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary S Postel
Mary Pastel, ScD. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number: K103274
Device Name: Sentinelle Endo Coil Array for Pelvic Imaging
Indications for Use:
The Sentinelle Endo Coil Array is a receive-only MRI coil array which must be used by a qualified medical professional to obtain images of (including spectroscopy images) the prostate, cervix and colon and the surrounding tissues in the pelvis. The Coil Array is a reusable transrectal or transvaginal device that can be high level disinfected. This product is for exclusive use on adult populations with clinically present anus and rectum or vagina. For use with GE, Toshiba and Siemens 1.5T MR systems
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
AND/OR
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mary Petty
Division Sign Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K103274
Page 1 of 1
ﮯ۔
ﺷﮩﺮ
11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.