The Trilliant Surgical Subtalar Implant is indicated for the use in the treatment of the hyperpronated foot and stabilization of the subtalar joint. It is intended to block the forward, downward, and medial displacement of the talus, thus allowing normal subtalar joint motion but limiting excessive pronation.
Device Story
Trilliant Surgical Subtalar Implant is a threaded, cannulated titanium alloy (ASTM F136) cylindrical spacer; inserted between posterior and middle facets of subtalar joint via guide wire. Used by surgeons in clinical settings to treat hyperpronated foot; device blocks forward, downward, and medial displacement of talus to limit excessive pronation while maintaining normal joint motion. Implant available in six sizes (Ø7mm to Ø12mm). Device is not provided sterile; requires hospital sterilization. Clinical decision-making relies on surgeon assessment of foot deformity; implant provides mechanical stabilization to improve patient gait and reduce arch pain.
Clinical Evidence
No clinical data; bench testing only. Compression testing was performed to compare the subject device against predicate devices.
Technological Characteristics
Material: Ti-6Al-4V alloy per ASTM F136. Design: Threaded, cannulated cylindrical implant. Dimensions: Ø7mm-12mm, length 12.75mm-19.50mm. Features: Rounded nose for insertion. Sterilization: Not sterile; requires hospital sterilization. Energy source: None (mechanical).
Indications for Use
Indicated for patients with hyperpronated foot requiring stabilization of the subtalar joint.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
## 510(k) Summary for the Trilliiant Surgical Subtalar Implant
K103183
p 1/3
In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following 510(k) summary is submitted for the Trilliant Surgical Subtalar Implant
Date Prepared: October 22, 2010
- 1. Submitter:
Trilliant Surgical LTD 602 Sawyer Street, Suite 120 Houston, TX 77007 Telephone: 214-288-1035
Contact Person: J.D. Webb The OrthoMedix Group, Inc. 1001 Oakwood Blvd Round Rock, TX 78681 Telephone: 512-388-0199
| 2. Trade name: | Trilliant Surgical Subtalar Implant |
|----------------------|-------------------------------------|
| Common Name: | Subtalar spacer |
| Classification Name: | Screw, fixation, bone |
| | 21 CFR section 888.3040 |
| | HWC |
| | Class II |
#### 3. Predicate or legally marketed devices which are substantially equivalent:
The Trilliant Surgical Subtalar Implant is substantially equivalent to the following devices:
- Subtalar MBA System, K960692 (Kinetikos Medical, Inc.) �
- HyProCure Subtalar Implant System, K042030 (GraMedical Technologies, LLC) �
- . Subtalar Arthroereisis Implant, K093820 (Memometal Technologies)
### 4. Description of the device:
The Trilliant Surgical Subtalar Implant consists of a threaded implant, designed to be inserted between the posterior and middle facets of the subtalar joint and corresponding instrumentation to facilitate insertion. The implant is cylindrical in shape and incorporates a center cannula designed for use with a guide wire to facilitate proper placement of the implant. It is available in six sizes, Ø7mm to Ø12mm in 1mm increments.
### Materials:
Ti-6Al-4V alloy per ASTM F136
### Function:
The Trilliant Surgical Subtalar Implant blocks the posterior displacement of the talus, thus allowing normal subtalar joint motion while blocking excessive pronation.
#### 5. Substantial equivalence claimed to predicate devices
The Trilliant Surgical Subtalar Implant is substantially equivalent to the Subtalar MBA System, HyProCure Subtalar Implant System and Subtalar Arthroereisis Implant devices in terms of intended use, design, and materials used. The table below compares the features and characteristics of the Trilliant Surgical Subtalar Implant to these predicate devices.
| Device<br>Name<br>Items | Trilliiant Surgical<br>Subtalar Implant | Subtalar MBA<br>System | HyProCure<br>Subtalar Implant<br>System | Subtalar<br>Arthroereisis<br>Implant |
|-------------------------|-----------------------------------------|----------------------------|-------------------------------------------|--------------------------------------|
| Sponsor | Trilliant, LTD | Kinetikos Medical,<br>Inc. | GraMedica<br>Medical<br>Technologies, LLC | Memometal<br>Technologie |
| 510(k) Number | N/A | K960692 | K042030 | K093820 |
- 8 2011
{1}------------------------------------------------
| Device<br>Name<br>Items | Trilliiant Surgical<br>Subtalar Implant | Subtalar MBA<br>System | HyProCure<br>Subtalar Implant<br>System | K103138<br>Subtalar<br>Arthroereisis<br>Implant |
|------------------------------------------|-----------------------------------------|-----------------------------------|---------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Device<br>Classification<br>Name | Screw, fixation,<br>bone | Screw, fixation,<br>bone | Screw, fixation,<br>bone | Screw, fixation,<br>bone |
| Product Code | HWC | HWC | HWC | HWC |
| Regulation # | 21 CFR 888.3040 | 21 CFR 888.3040 | 21 CFR 888.3040 | 21 CFR 888.3040 |
| Classification | II | II | II | II |
| Indications for<br>Use | [1] see at end of<br>table | [2] see at end of<br>table | [3] see at end of<br>table | [4] see at end of<br>table |
| Implant material | Titanium alloy per<br>ASTM F136 | Titanium alloy per<br>ASTM F136 | Titanium alloy per<br>ASTM F136 | Titanium alloy per<br>ASTM F136 |
| Diameter<br>Length | Ø - 7mm-12mm<br>L - 12.75mm-<br>19.50mm | Ø - 6mm-12mm<br>L - 15mm | | Ø - 6.5mm-<br>11.5mm |
| Profile /anatomy<br>location | Threaded cylinder<br>/ sinus tarsi | Threaded cylinder<br>/sinus tarsi | Threaded cylinder<br>/sinus tarsi<br>Smooth cone /<br>sinus canalis | Threaded cylinder<br>/sinus tarsi<br>Shallow thread to<br>smooth cone /<br>sinus canalis |
| Cannulated | Yes | Yes | Yes | Yes |
| Rounded nose<br>for ease of<br>insertion | Yes | Yes | Yes | No |
| Packaged sterile | No, sterilized at<br>hospital | No, sterilized at<br>hospital | Yes | Yes |
[1] The Trilliant Surgical subtalar implant is indicated for the treatment of the hyperpronated foot and stabilization of the subtalar joint. It is intended to block the forward, and medial displacement of the talus, thus allowing normal subtalar joint motion but limiting excessive pronation.
[2] The KMI Subtalar MBA System™ is indicated for use in the treatment of the hyperpronated foot and stabilization of the subtalar joint. It is design to block the anterior displacement of the talus, thus allowing normal subtalar joint motion but blocking excessive pronation and the resulting sequela:
- Severely pronated foot. .
- . Walking intemperance.
- . Calcaneal stance position greater than 5°
- . Manually correctable deformity
- . Mid-tarsal breech (arch pain)
- . Forefoot varus greater than 10°
[3] The HyProCure Subtalar Implant is indicated for use in the treatment of the hyperpronated foot by stabilization of the subtalar joint. The implant is designed to block anterior, and/or plantarflexion of the talus, while allowing normal subtalar joint motion but blocking excessive pronation and the resulting sequela.
[3] The Memometal Technologies' Subfix Arthroereisis Implant is intended to treat the hyperpronated foot and stabilize the subtalar joint. It is intended to block forward, downward and medial displacement of the talus, thus allowing subtalar joint motion but blocking excessive pronation and the resulting sequela.
{2}------------------------------------------------
#### 6. Intended Use:
K103138,p3/3
The Trilliant Surgical subtalar implant is indicated for the treatment of the hyperpronated foot and stabilization of the subtalar joint. It is intended to block the forward, and medial displacement of the talus, thus allowing normal subtalar joint motion but limiting excessive pronation.
### 7. Non-clinical Test Summary:
The following tests were conducted:
- Compression testing comparing predicate and Trilliant Surgical Subtalar Implant. ●
#### 8. Clinical Test Summary
No clinical studies were performed
#### റ്റ. Conclusions Nonclinical and Clinical
The Trilliant Surgical Subtalar Implant is substantially equivalent to the predicate devices in terms of indications for use, design, material, and function.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features an abstract eagle with three stylized wing segments, symbolizing health, services, and people. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Trilliant Surgical, Ltd. % The OrthoMedix Group, Inc. % Mr. J.D. Webb 1001 Oakwood Blvd. Round Rock, Texas 78681
FEB - 8 2", i
Re: K103183
Trade/Device Name: Trilliant Surgical Subtalar Implant Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: HWC Dated: January 28, 2011 Received: February 3, 2011
Dear Mr. Webb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
{4}------------------------------------------------
Page 2 - Mr. J.D. Webb
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Aley B. D.K
for
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K |0 3 | 8 3
Device Name: Trilliant Surgical Subtalar Implant
Indications for Use:
· The Trilliant Surgical Subtalar Implant is indicated for the use in the treatment of the hyperpronated foot and stabilization of the subtalar joint. It is intended to block the forward, downward, and medial displacement of the talus, thus allowing normal subtalar joint motion but limiting excessive pronation.
Prescription Use _ × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for M. Melkerson
(Division Sign-Off) rthopedic, Division di Surgical, Divisionative Devices
510(k) Number K/03/83
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.