K103111 · Orthofix, Inc. · MAX · Mar 23, 2011 · Orthopedic
Device Facts
Record ID
K103111
Device Name
FORZA SPACER SYSTEM
Applicant
Orthofix, Inc.
Product Code
MAX · Orthopedic
Decision Date
Mar 23, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The FORZA™ Spacer System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels in the lumbar spine (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved levels. These patients may have had a previous non-fusion surgery at the involved level(s). The FORZA™ Spacer System is intended for use with autograft and supplemental fixation system. As an example, the supplemental fixation system that may be used is the Orthofix / Blackstone Medical, Inc. Firebird™ Pedicle Screw System. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the FORZA™ Spacer System.
Device Story
The FORZA Spacer System is an intervertebral body fusion device consisting of PEEK implants, trials, and surgical instruments. Implants are hollow cages featuring a bulleted nose for insertion and anti-migration ripples on superior/inferior surfaces to stabilize the spinal segment. Tantalum markers are embedded in the PEEK material to provide radiographic visibility for intraoperative and postoperative placement verification. The device is used by surgeons in spinal fusion procedures to restore biomechanical integrity to the lumbar spine. It is not a stand-alone device and must be used in conjunction with autograft and a supplemental fixation system (e.g., pedicle screws). The implants are provided sterile, while instruments are provided non-sterile. By maintaining disc space height and facilitating fusion, the device aims to alleviate back pain associated with discogenic degeneration.
Clinical Evidence
Bench testing only. Mechanical performance was validated per ASTM F2077-03 (static and dynamic compression) and ASTM F2267-04 (subsidence). No clinical data provided.
Indicated for spinal fusion in skeletally mature patients with lumbar (L2-S1) degenerative disc disease (DDD) at 1-2 contiguous levels. Includes patients with up to Grade 1 spondylolisthesis and prior non-fusion surgery. Requires 6 months of failed non-operative treatment. Must be used with autograft and supplemental fixation.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
### 510(k) SUMMARY
| | Orthofix Inc. FORZA Spacer System<br>MAR 2 3 2011 | | | | |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|
| Submitter Information | | | | | |
| Name:<br>Address: | Orthofix Inc.<br>3451 Plano Parkway<br>Lewisville, TX 75056 | | | | |
| Telephone Number:<br>Fax Number:<br>Email: | 214.937.2000<br>214.937.3322<br>darlachew@orthofix.com | | | | |
| Registration Number: | 3008524126 | | | | |
| Contact Person: | Darla Chew<br>Director of Regulatory Affairs | | | | |
| Date Prepared: | March 21, 2011 | | | | |
| Name of Device<br>Trade Name / Proprietary<br>Name: | き<br>FORZA™ Spacer System | | | | |
| Common Name: | Intervertebral Body Fusion Device | | | | |
| Product Code: | MAX - Intervertebral Fusion Device with Bone Graft,<br>Lumbar | | | | |
| Regulatory Classification: | Class II - 888.3080 - Intervertebral Body Fusion Device | | | | |
| Review Panel: | Orthopedic Device Panel | | | | |
| Predicate Devices: | Blackstone, PILLAR Spacer System, K081177, SE 7-23-08<br>DePuy AcroMed, Inc. Lumbar I/F Cage® with VSP<br>Spine System: P960025 (i.e., Brantigan Cage)<br>Abbott, Ardis Spacer, K073202, SE 1-30-2008<br>Advance Medical Technology, Distractable Wave Cage,<br>K083626. SE10-19-2009 | | | | |
Reason for 510(k) Submission: New product offering
{1}------------------------------------------------
## Device Description
The FORZA Spacer System consists of implants, trials, and instruments. The FORZA Spacer System is comprised of a variety of implants fabricated and manufactured from polyetheretherketone (PEEK) as described by ASTM F-2026 with Tantalum markers as described by ASTM F-560. PEEK was utilized due to its radiolucent properties, which aids the surgeon in determining if fusion in the operative site has occurred. Since PEEK is transparent in x-rays, Tantalum marker pins are inserted into the implants in order to give surgeons a visual aid in determining the location of the implants, both intra and postoperatively. The implants are offered in two geometric shapes: straight and curved, and offered in lordotic profiles to restore the natural curvature of the spine. Both the curved and straight implants feature a bulleted nose for ease of insertion and anti-migration ripples on both the inferior and superior surfaces to provide increased stability and help prevent anterior/posterior movement of the device.
The FORZA Spacer System is intended for intervertebral body fusion to aid in the surgical correction and stabilization of the spine.
The FORZA Spacer System is not intended to be used as a stand-alone device. The FORZA Spacer System must be used with supplemental fixation system. The FORZA Spacer System implants are provided sterile. The FORZATM Spacer System trials and instruments are provided non-sterile.
### Intended Use / Indications for Use
The FORZA™ Spacer System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels in the lumbar spine (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved levels. These patients may have had a previous non-fusion surgery at the involved level(s).
The FORZA™ Spacer System is intended for use with autograft and supplemental fixation system. As an example, the supplemental fixation system that may be used is the Orthofix / Blackstone Medical, Inc. Firebird™ Pedicle Screw System.
Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the FORZATM Spacer System.
{2}------------------------------------------------
| Characteristic | FORZA™ Spacer<br>System<br>(Under Review) | Orthofix Inc.,<br>PILLAR Spacer<br>System (K081177) | DePuy AcroMed, Inc.<br>Lumbar I/F Cage®<br>with VSP Spine<br>System: P960025 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Function /<br>Design | Restore Biomechanical<br>integrity of the spinal<br>column by providing<br>fusion at one or two<br>contiguous levels in the<br>lumbar spine (L2-S1).<br>The device is intended<br>to be used with<br>autograft material and<br>supplemental fixation<br>system. | Restore Biomechanical<br>integrity of the<br>anterior, middle and<br>posterior spinal<br>column even in the<br>absence of fusion for a<br>prolonged period. The<br>device is intended to<br>be used with bone<br>graft material. | The Lumbar I/F Cage<br>device indicated for an<br>open posterior approach<br>using autogenous bone<br>graft at one or two<br>spinal levels from L2-S1<br>whose condition requires<br>the use of interbody<br>fusion combined with<br>posterolateral fusion<br>and posterior pedicle<br>screw fixation. |
| Implant Size(s) | Width 9–11mm<br>Length 23–33mm | Width 9–11mm<br>Length 23–33mm | Width 9–15mm<br>Length 21–25mm |
| Implant Heights | 6–16mm | 6–16mm | 9–17mm |
| Lordotic Angles | 0, 4 and 8° | 0 and 8° | Parallel and Wedged |
| Configuration | Hollow Cage for use<br>with supplemental<br>fixation | Hollow Cage for use<br>with supplemental<br>fixation | Hollow Cage for use<br>with supplemental<br>fixation |
| Material | Polyetheretherketone<br>(PEEK, ASTM F-2026)<br>Tantalum marker<br>(ASTM F-560) | Polyetheretherketone<br>(PEEK, ASTM F-2026)<br>Titanium marker<br>(ASTM F-67) | Polyetherketone Ether<br>Ketone Ketone<br>(PEKEKK) with<br>Polyacrylonitrile (PAN)<br>carbon fibers with<br>Tantalum markers |
# Summary of the Technological Characteristics of the Device Compared to the Predicate Device
# PERFORMANCE DATA – Summary of Non-Clinical Test Conducted for Determination of Substantial Equivalence
| Characteristic | Standard / Test/ FDA Guidance |
|---------------------|-------------------------------|
| Static Compression | ASTM F 2077-03 |
| Dynamic Compression | ASTM F 2077-03 |
| Subsidence | ASTM F2267-04 |
#### Performance Data Summary
Mechanical testing of the Orthofix Inc. FORZATM Spacer System was conducted in accordance to ASTM F2077-03 standard for Static & Dynamic Compression testing and ASTM F2267-04 standards for Subsidence testing. Test results demonstrated that
> Section 4 - 510(k) Summary Section Page 3 of 4
{3}------------------------------------------------
the FORZA Space system is substantially equivalent to predicate devices that have the same intended use, similar indications, technological characteristics and principles of operation.
#### Substantial Equivalence
The FORZATM Spacer System share the same intended use, similar indications, technological characteristics and principles of operation with PILLAR Spacer System (K081177) and the DePuy Acromed, Inc. Lumbar I/F Cage® with VSP Spine System (P960025). FORZA system has similar mechanical properties to DePuy AcroMed, Inc. Lumbar I/F Cage® with VSP Spine System (P960025) and Pioneer Rotate Option implant (K073177) predicate device as stated above.
The difference between FORZA Space System implants and its predicate devices consist in technological differences i.e. the FORZA implants are distributed sterile and have minor dimensional differences. FORZA dimensional and sterility differences proposed in this submission are addressed with sterilization and mechanical verification testing. Based on test results, these changes do not present any new / additional issues of safety or effectiveness, the FORZA™ Spacer System is substantially equivalent to its predicate devices.
{4}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002
Orthofix, Inc. % Ms. Darla Chew Director of Regulatory Affairs 3451 Plano Parkway Lewisville, Texas 75056
MAR 2 3 2011
Re: K103111
Trade/Device Name: FORZA™ Spacer System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: March 18, 2011 Received: March 21, 2011
Dear Ms. Chew:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ycm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fdaggov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Aling B. Rote
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use
#### 510(k) Number (if known): K103111
#### Device Name: FORZA™ Spacer System (Intervertebral body fusion device)
#### Indications for Use:
The FORZATM Spacer System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels in the lumbar spine (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved levels. These patients may have had a previous non-fusion surgery at the involved level(s).
The FORZATM Spacer System is intended for use with autograft and supplemental fixation system. As an example, the supplemental fixation system that may be used is the Orthofix Inc. Firebird™ Pedicle Screw System.
Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the FORZATM Spacer System.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K103111
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.