← Product Code [GXZ](/productcode/GXZ) · K103103

# PRESSON ELECTRODE (K103103)

_Persyst Development Corp. · GXZ · Dec 8, 2011 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K103103

## Device Facts

- **Applicant:** Persyst Development Corp.
- **Product Code:** [GXZ](/productcode/GXZ.md)
- **Decision Date:** Dec 8, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.1350
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The PressOn™ electrode is intended for use in the recording of the electroencephalogram (EEG), the evoked potential (EP), or as ground and reference in an EEG or EP recording.

## Device Story

PressOn electrode is a single-use, disposable EEG electrode; features 0.6 cm footprint and 0.004 inch height profile. Constructed from super-elastic Nitinol; tri-pod configuration with three legs terminating in micro-teeth. Micro-teeth penetrate top layers of scalp dermis for subdermal placement. Used with reusable PressOn lead (48" length) to connect to external EEG monitoring equipment. Operated by clinicians in clinical settings. Device facilitates signal acquisition for EEG/EP recordings; provides stable electrical contact via subdermal penetration. Benefits include ease of placement and secure attachment during neurological monitoring.

## Clinical Evidence

Bench testing only. Biocompatibility and performance testing demonstrated acceptable results.

## Technological Characteristics

Material: Super-elastic Nitinol. Design: Tri-pod configuration with micro-teeth for subdermal penetration. Dimensions: 0.6 cm footprint, 0.004 inch height. Connectivity: Wired connection via accessory lead to external EEG monitor. Single-use, disposable electrode; reusable lead.

## Regulatory Identification

A needle electrode is a device which is placed subcutaneously to stimulate or to record electrical signals.

## Predicate Devices

- Rhythmlink subdermal needle electrodes ([K022914](/device/K022914.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K103103

DEC - 8 2011

#### 510(k) Summary of Safety and Effectiveness II.

# PressOn™ electrode

# General Information

| Criteria             | Information                                                                                                                                    |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name           | PressOn™ electrode                                                                                                                             |
| Catalog/Model Number | PR01                                                                                                                                           |
| Common Name          | Needle (electroencephalograph) electrode                                                                                                       |
| Classification       | 21 CFR 882.1350 - Needle electrode<br>Class II; product code: GXZ                                                                              |
| 510(k) Submitter     | Persyst Development Corporation<br>3177 Clearwater Drive<br>Prescott, AZ 86305                                                                 |
| Contact Person       | Scott B. Wilson<br>Chief Executive Officer<br>Persyst Development Corporation<br>scottw@persyst.com<br>858.755.4568 (phone) 858.755.4565 (fax) |

# Summary of Substantial Equivalence

The PressOn electrode does not raise any new safety or effectiveness issues and is substantially equivalent to legally marketed needle electrodes that are in commercial distribution, and have been determined to be substantially equivalent to devices in commercial distribution, prior to May 28, 1976.

Date: October 5, 2010

{1}------------------------------------------------

### Substantially Equivalent Devices

| Manufacturer                                               | Substantially equivalent device           | 510(k)  |
|------------------------------------------------------------|-------------------------------------------|---------|
| Rhythmlink International, LLC<br>Cayce, South Carolina USA | Rhythmlink subdermal<br>needle electrodes | K022914 |

### Device Description

The PressOn electrode is a single-use, disposable EEG electrode. The electrode is very small, with a footprint of approximately 0.6 cm and a very thin (~0.004 inch) height profile. It is made from a super-elastic Nitinol and is flexible. The electrode has 3 legs (a tri-pod configuration) that each terminate in micro-teeth that penetrate the top layers of the scalp dermis during subdermal electrode placement.

The PressOn electrode is used in conjunction with an accessory PressOn lead that is provided non-sterile and is reusable. The lead is approximately 48" in length and is used to connect one of the PressOn electrodes to a commercially available EEG monitor. The EEG monitor/ equipment is not part of the PressOn device offering.

### Indications for Use

The PressOn™ electrode is intended for use in the recording of the electroencephalogram (EEG), the evoked potential (EP), or as ground and reference in an EEG or EP recording.

### Testing

The PressOn electrode underwent biocompatibility, bench and performance testing and demonstrated acceptable results in all testing.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Persyst Development Corporation c/o Mr. Scott B. Wilson Chief Executive Officer 3177 Clearwater Drive Prescott, Arizona 86305

Re: K103103

Trade/Device Name: PressOn Electrode Regulation Number: 21 CFR 882.1350 Regulation Name: Needle Electrode Regulatory Class: Class II Product Code: GXZ Dated: November 18, 2011 Received: November 21, 2011

Dear Mr. Wilson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

DEC - 8 2011

{3}------------------------------------------------

Page 2 - Mr. Scott B. Wilson

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of madical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm1118001.html 01 the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

/MedicalDevices/ResourceSion/Fou/Industry/default.ill

Sincerely yours,

Kesia Alexander

for

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### 1. Indications for Use Statement

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: PressOn™ electrode

Indications for Use:

The PressOn™ electrode is intended for use in the recording of the electroencephalogram (EEG), the evoked potential (EP), or as ground and reference in an EEG or EP recording.

| Prescription Use            | X |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) |   |

AND/OR

| Over-The-Counter Use   |  |
|------------------------|--|
| (21 CFR 801 Subpart C) |  |

Concurrence of CDRH, Office of Device Evaluation (ODE)

| page 1 of 1 | (Division Sign-Off)                           |
|-------------|-----------------------------------------------|
|             | Division of Ophthalmic, Neurological and Ear, |
|             | Nose and Throat Devices                       |

| 510(k) Number | K103103 |
|---------------|---------|
|---------------|---------|

| PressOn Electrode 510(k)        | page 3 of 44 |
|---------------------------------|--------------|
| Persyst Development Corporation | Confidential |

---

**Source:** [https://fda-staging.innolitics.com/device/K103103](https://fda-staging.innolitics.com/device/K103103)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).