Performance of the HardyCHROM™ MRSA medium was evaluated at three geographically diverse hospitals using fresh clinical surveillance specimens to compare the device's detection accuracy against routine culture methods.
Patients undergoing MRSA screening via anterior nares swabs; Sample Size: 443 samples; Number of Sites: 3
Routine culture (TSAB) and a commercial chromogenic MRSA medium
Positive and negative percent agreement for MRSA detection at 24 and 48 hours
Indications for Use
HardyCHROM™ MRSA is a selective and differential chromogenic medium recommended for the qualitative detection of nasal colonization by methicillin-resistant Staphylococcus aureus (MRSA) to aid in the prevention and control of MRSA infections in health care settings. The test is performed on anterior nares swabs from patients and healthcare workers to screen for MRSA colonization. HardyCHROM™ MRSA is not intended to diagnose MRSA infection nor to guide or monitor therapy for MRSA infections. Concomitant cultures are necessary to recover organisms for susceptibility testing or epidemiological typing. A negative result does not preclude MRSA nasal colonization.
Device Story
HardyCHROM™ MRSA is a selective/differential chromogenic culture medium; used in clinical laboratories to screen anterior nares swabs for MRSA colonization. Input: clinical nasal swab specimens. Operation: medium contains chromogenic substrates and selective agents; MRSA strains produce enzymes that cleave substrates to release chromophores, resulting in pink/magenta colonies within 24 hours. Non-MRSA strains are inhibited or produce different colored colonies (blue/green) or remain white. Output: visual identification of colony color by laboratory personnel. Clinical utility: aids infection control programs by identifying colonized individuals; negative results do not rule out colonization. Benefits: rapid detection (24-48 hours) compared to traditional culture methods.
Clinical Evidence
Performance evaluated at three clinical sites using 443 surveillance specimens. Compared to traditional culture (TSAB with PBP2' and cefoxitin/oxacillin disk diffusion). At 24 hours, positive percent agreement (PPA) was 93.3% and negative percent agreement (NPA) was 99.7%. At 48 hours, PPA improved to 97.0%. Reproducibility testing with 20 characterized strains (15 MRSA, 5 MSSA) across three lots and sites showed 100% agreement. Interference studies confirmed no impact from human blood, transport media, or mucin; phenylephrine hydrochloride (1%) showed inhibitory effects.
Technological Characteristics
Selective and differential chromogenic culture medium. Contains chromogenic substrates for enzyme-based colorimetric detection and selective agents for mecA-mediated MRSA growth. Form factor: prepared agar plates. Shelf life: 10 weeks. Incubation: 35-37°C, aerobic, 24-48 hours. Manual interpretation.
Indications for Use
Indicated for qualitative detection of nasal colonization by methicillin-resistant Staphylococcus aureus (MRSA) in patients and healthcare workers to aid in infection control. Not for diagnosing MRSA infection or guiding therapy.
Regulatory Classification
Identification
A culture medium for antimicrobial susceptibility tests is a device intended for medical purposes that consists of any medium capable of supporting the growth of many of the bacterial pathogens that are subject to antimicrobial susceptibility tests. The medium should be free of components known to be antagonistic to the common agents for which susceptibility tests are performed in the treatment of disease.
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
K102922
B. Purpose for Submission:
To obtain a substantial equivalent determination for a Premarket notification for the HardyCHROM™ MRSA product.
C. Measurand:
Methicillin Resistant Staphylococcus aureus (MRSA)
D. Type of Test:
Detection of MRSA using a selective and differential chromogenic media
E. Applicant:
Hardy Diagnostics
F. Proprietary and Established Names:
HardyCHROM™ MRSA
G. Regulatory Information:
1. Regulation section:
21 CFR 866.1700
2. Classification:
Class II
3. Product code:
JSO: Culture media, Antimicrobial susceptibility test, excluding Mueller Hinton Agar
{1}
4. Panel:
Microbiology (83)
H. Intended Use:
1. Intended use(s):
HardyCHROM™ MRSA is a selective and differential chromogenic medium recommended for the qualitative detection of nasal colonization by methicillin-resistant Staphylococcus aureus (MRSA) to aid in the prevention and control of MRSA infections in health care settings. The test is performed on anterior nares swabs from patients and healthcare workers to screen for MRSA colonization. HardyCHROM™ MRSA is not intended to diagnose MRSA infection nor to guide or monitor therapy for MRSA infections. Concomitant cultures are necessary for susceptibility testing or epidemiological typing. A negative result does not preclude MRSA nasal colonization.
2. Indication(s) for use:
HardyCHROM™ MRSA is a selective and differential chromogenic medium recommended for the qualitative detection of nasal colonization by methicillin-resistant Staphylococcus aureus (MRSA) to aid in the prevention and control of MRSA infections in health care settings. The test is performed on anterior nares swabs from patients and healthcare workers to screen for MRSA colonization. HardyCHROM™ MRSA is not intended to diagnose MRSA infection nor to guide or monitor therapy for MRSA infections. Concomitant cultures are necessary for susceptibility testing or epidemiological typing. A negative result does not preclude MRSA nasal colonization.
3. Special conditions for use statement(s):
Prescription use only
4. Special instrument requirements:
Not applicable
I. Device Description:
HardyCHROM™ MRSA is translucent and light amber in color and contains chromogens that release chromophores when cleaved by enzymes that are produced by MRSA strains. Based on colony color, HardyCHROM™ MRSA allows for the reliable detection of most methicillin-resistant S. aureus from clinical specimens in less than 24 hours based on the appearance of a pink/magenta colored colony. Non-MRSA strains are either inhibited by the
2
{2}
addition of selective agents or utilize different chromogenic substrates in the media to produce different colored colonies. If none of the substrates are utilized, natural or white colored colonies will be present. If plates are negative for growth after 24 hours, it is recommended to re-incubate for an additional 24 hours.
# J. Substantial Equivalence Information:
1. Predicate:
BBL™ CHROMagar™ MRSA
2. Predicate K number(s):
K042812
3. Comparison with Predicate
| Product Attribute | HardyCHROM™ MRSA | BBL™ CHROMagar™ MRSA |
| --- | --- | --- |
| Intended Use | HardyCHROM™ MRSA is a selective and differential chromogenic medium recommended for the qualitative detection of nasal colonization by methicillin-resistant Staphylococcus aureus (MRSA) to aid in the prevention and control of MRSA infections in health care settings. The test is performed on anterior nares swabs from patients and healthcare workers to screen for MRSA colonization. HardyCHROM™ MRSA is not intended to diagnose MRSA infection nor to guide or monitor therapy for MRSA infections. Concomitant cultures are necessary to recover organisms for susceptibility testing or epidemiological typing. A negative result does not preclude MRSA nasal colonization. | BBL™ CHROMagar™ MRSA is a selective and differential medium for the qualitative direct detection of nasal colonization by methicillin resistant Staphylococcus aureus (MRSA) to aid in the prevention and control of MRSA infections in healthcare settings. The test is performed on anterior nares swab specimens from patients and healthcare workers to screen for MRSA colonization. BBL™ CHROMagar™ MRSA is not intended to diagnose MRSA infection nor to guide or monitor treatment for infections. |
| Specimen Type | Anterior nasal swabs | Anterior nasal swabs |
| Test Methodology | Selective and differential chromogenic prepared culture | Selective and differential chromogenic prepared culture |
{3}
| | medium | medium |
| --- | --- | --- |
| Inoculation | Direct from specimen collection device | Direct from specimen collection device |
| Incubation Temperature | Incubation at 35 -37°C | Incubation at 35 -37°C |
| Incubation Length | 24 hours. Negative plates should be re-incubated for a total of 48 hours to confirm initial findings. | 24 hours, if negative re-incubate an additional 24 hours |
| Selective Agent | Selective agents included to allow for growth of mecA mediated MRSA strains. | Selective agents included to allow for growth of mecA mediated MRSA strains. |
| Testing Method | Manual | Manual |
| Growth Detection | Pink to magenta colonies within 24 hours. Negative plates should be re-incubated to 48 hours to confirm initial findings. | Pink to magenta colonies within 24 hours. Negative plates should be re-incubated to 48 hours to confirm initial findings. |
| Organism Differentiation | Chromogenic substrates facilitate differentiation of MRSA from other organisms. | Chromogenic substrates facilitate differentiation of MRSA from other organisms. |
| Shelf Life | 10 weeks | 10 weeks |
# K. Standard/Guidance Document Referenced (if applicable):
Clinical and Laboratory Standards Institute. 2010. Performance Standards for Antimicrobial Susceptibility Testing; Twentieth Informational Supplement, M100-S20. CLSI, Wayne PA.
# L. Test Principle:
Methicillin-resistant Staphylococcus aureus strains produce pink to magenta colonies as the result of the chromogenic substrates incorporated into the HardyCHROM™ MRSA medium. The addition of specific inhibitory agents allows for the growth of mecA mediated MRSA strains while preventing growth of methicillin sensitive Staphylococcus aureus (MSSA) strains. Additional selective agents have been added to increase the sensitivity and specificity of the medium by inhibiting gram-negative organisms, yeast, and some gram-positive cocci. Bacteria other than MRSA may utilize additional chromogenic substrates present in the medium and produce blue or green colonies. HardyCHROM™ MRSA can detect most MRSA strains within 24 hours. Negative plates should be re-incubated up to 48 hours.
{4}
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Reproducibility Study with Characterized MRSA & MSSA strains
Reproducibility testing of HardyCHROM™ MRSA was conducted with twenty well-characterized test strains (fifteen MRSA strains that included Pulse-Field Gel Electrophoresis (PFGE) types USA100, USA200, USA300-0114, USA400, USA500, USA600, USA700, USA800, USA1000, USA1100, additional MRSA strains that are representative of prevalent HA(Hospital acquired)-MRSA and CA(community acquired)-MRSA (NRS100, NRS70, NRS71, BAA-43, BAA-44) and five MSSA (NRS72, SF_SNIF58, SF_SNIF104, SF_SNIF129, SF_SNIF142) that were selected from medically important S. aureus lineages. SF8300 is a wound isolate of pulse-field type USA300-0114, a subtype implicated in severe disease and in numerous outbreaks. All of the characterized test strains were obtained from the University of San Francisco and (DNA restriction patterns) were determined by pulsed-field gel electrophoresis. (J Clin Microbiol 33:2233-9)
Three different lots of HardyCHROM™ MRSA plates were tested to determine that the HardyCHROM™ MRSA plates reliably detected MRSA strains across different lots at different time intervals. A commercial chromogenic MRSA medium was also evaluated with the same strains and served as a performance standard. An acceptance reproducibility rate of 100% for both inter-lot and overall testing intervals was achieved with the HardyCHROM™ MRSA plates media. All of the characterized MRSA strains showed growth and magenta colony coloration within 24 hours following aerobic incubation at 35°C. All of the MSSA strains were inhibited and showed no growth even after 48 hours of aerobic incubation at 35°C. No discernible differences were observed between the performance testing of HardyCHROM™ MRSA in comparison with the commercial chromogenic MRSA medium.
The fifteen strains of MRSA were tested using a suspension containing approximately 10⁵ to 10⁶ CFU/ml and the five strains of MSSA were tested at a concentration of approximately 10⁶ to 10⁷. Ten μl of these suspensions were inoculated onto HardyCHROM™ MRSA. The sensitivity was 100% for the fifteen characterized MRSA strains and 100% specificity for the five characterized MSSA strains.
Staphylococcus aureus strains with other mechanisms of oxacillin resistance such as modified S. aureus (MOD-SA) strains) which have altered affinity of penicillin binding proteins for oxacillin and borderline methicillin-resistant
5
{5}
Staphylococcus aureus (BORSA) were not evaluated on HardyCHROM™ MRSA.
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Quality Control
Quality control was performed by each of the three testing sites on receipt of each shipment of the medium and on each day of use following parameters in table below. Fresh suspensions of Staphylococcus aureus (MRSA) ATCC® 43300 and Staphylococcus aureus (MSSA) ATCC® 25923 were prepared in Tryptic Soy Broth at concentrations of approximately $10^{5}$ to $10^{6}$ for the MRSA strain and $10^{6}$ to $10^{7}$ for MSSA. $10\mu l$ of these suspensions were used to inoculate HardyCHROM™ plates. Plates were incubated and examined for 24 and 48 hours for growth of the MRSA strain and inhibition of the MSSA strain. All QC testing results provided expected reactions at each of the three testing sites on each day tested.
| Test Organisms | Incubation | | | Results |
| --- | --- | --- | --- | --- |
| | Time | Temperature | Atmosphere * | |
| Staphylococcus aureus ATCC® 43300 | 24hr | 35 to 37°C | Aerobic | Growth; pink to magenta colonies |
| Staphylococcus aureus ATCC® 29213 | 24-48hr | 35 to 37°C | Aerobic | Inhibited |
* Do not incubate in $\mathrm{CO}_{2}$.
d. Detection limit:
Not applicable
e. Analytical specificity:
Cross-Reactivity Study
Common types of nasal flora (Staphylococcus and non-Staphylococcus organisms) were evaluated for growth and performance on nonselective blood agar plates, HardyCHROM™ MRSA, and BBL™ CHROMagar™ MRSA plates. Testing of other Staphylococcus and non-Staphylococcus organisms
{6}
was conducted to determine if there was any cross reactivity when tested on HardyCHROM™ MRSA medium. A total of 121 strains were tested and included the following genera: Acinetobacter, Burkholderia, Candida, Corynebacterium, Enterococcus, Enterobacter, Escherichia, Haemophilus, Klebsiella, Leuconostoc, Micrococcus, Moraxella, Neisseria, Proteus, Pseudomonas, Streptococcus and Staphylococcus. Staphylococcus strains included MRSA, MSSA, and 20 different species of coagulase negative Staphylococci (including methicillin resistant Staphylococcus epidermidis). In the cross reactivity study, only Corynebacterium jeikeium produced a purple film in the first quadrant at 24 hours. At 48 hours, very small dark purple colonies were present but were only present in the first quadrant. Staphylococcus intermedius produced gray-blue colonies. All other organisms were inhibited.
## Mixed Infection Study
Cell suspensions at a concentration of $10^{8}$ and $10^{9}$ were prepared of Staphylococcus aureus MSSA ATCC® 25923 and E. coli ATCC® 25922. Individual $10^{3}$ dilutions (LoD) of Staphylococcus aureus MRSA strains (ATCC® 33591 and 43300) were mixed with the non MRSA strains and plated in duplicate on Blood Agar and paired nasal swabs were inoculated with the same suspension. The nasal swabs were used to inoculate the HardyCHROM™ MRSA plates. On HardyCHROM™ MRSA medium, the MRSA strains tested were detected beginning at 16 hours with final counts being reported at 24 hours following aerobic incubation at $35^{\circ}\mathrm{C}$. Pink to magenta colonies were detected on all dilutions at 16 hours and colony counts were stable at 21 to 24 hours with no increase in the number of colonies counted. All dilutions plated showed typical pink to magenta colony coloration and colony morphology was consistent with the MRSA strains. No breakthrough was noted of any of the non-MRSA strains at 24 hours and all plates were incubated for a total of 48 hours. There was no change in MRSA strains colony forming units nor was there any breakthrough of non-MRSA at the 48 hour reading.
## Interference Study Results – Nasal Sprays
The strains used in this study included MRSA strains ATCC® 33591 and ATCC 43300, as well as ten clinical MRSA isolates. Commonly used nasal sprays that contain a concentration of $1\%$ Phenylephrine Hydrochloride did show an inhibitory affect for microbial growth on the nonselective Blood Agar plates, the HardyCHROM™ MRSA plates, and BBL™ CHROMagar™ MRSA plates. Nasal sprays that did not contain this ingredient showed no interference for growth or chromogenic reactions. The sensitivity and specificity criteria of $\geq 98\%$ were met for the HardyCHROM™ MRSA plates as compared to BBL™ CHROMagar™ and Blood Agar plates.
7
{7}
Interference Study Results - Human Blood
Human blood could potentially be associated with MRSA strains in a healthcare setting and therefore was evaluated for potential interference of the chromogenic reaction and growth. Isolates used in this study included ATCC® control strains of MRSA (ATCC® 33591, ATCC® 43300) and ten clinical well characterized clinical MRSA isolates. All strains were evaluated for growth and performance on HardyCHROM™ MRSA, a commercial chromogenic MRSA medium, and Blood Agar plates. Results demonstrated that none of the MRSA strains were inhibited on the three types of media evaluated, and there was a 100% recovery rate for all strains after exposure to human blood. No discernible difference was observed between the growth and performance testing of HardyCHROM™ MRSA in comparison with nonselective blood agar plates and the commercial chromogenic MRSA medium.
## Interference Study Results - Transport Media
The strains used in this study included MRSA strains ATCC® 33591 and ATCC 43300, as well as ten well-characterized clinical MRSA isolates. All of the strains were evaluated for growth and performance on HardyCHROM™ MRSA, BBL™ CHROMagar™ MRSA, and Blood Agar plates. Testing included the following commonly used types transport media: Stuart's Gel, Stuart's Liquid, Amies Liquid, Amies Gel, and Amies Charcoal. Each transport device tested contained rayon-tipped, plastic shaft swabs. Results demonstrated that none of the MRSA strains were inhibited on the three types of media evaluated, and there was a 100% recovery rate for all transport media devices tested. No discernible difference was observed between the growth and performance testing of HardyCHROM™ MRSA in comparison with nonselective blood agar plates and BBL™ CHROMagar™ MRSA plates.
## Interference Study - Mucin
Mucin from bovine submaxillary glands (type 1-S) was tested for interference using ATCC® 33591, ATCC® 43300, ten well-characterized clinical MRSA strains, and ten well-characterized clinical MSSA strains. All of the strains were exposed to suspensions of 50% and 75% mucin and after 10 minutes, were plated on blood agar and HardyCHROM™ MRSA. Results were recorded after 24 hours of incubation, with negative plates being held for an additional 24 hours (total incubation 48 hours). Suspensions prepared in mucin were held at room temperature for 24 hours and re-plated. The secondary plates were incubated at 35°C for 24 hours and plates with no growth were re-incubated for a total of 48 hours. Colonies of the MRSA strains grown on HardyCHROM MRSA™ were typical for coloration and size. Counts were stable at 18 hours with no additional growth detected at 24 or 48 hours. All MSSA strains that were tested showed growth only on the
{8}
blood agar plate and were completely inhibited at 24 and 48 hours on HardyCHROM MRSA™. Nasal swabs containing mucous or mucin were evaluated determine if there was an effect on the chromogenic reaction and the growth of MRSA strains. Isolates used in this study included ATCC® control strains of MRSA (ATCC® 33591, ATCC® 43300), ten clinical well characterized clinical MRSA isolates and ten well characterized clinical MSSA isolates.
## Incubation Study
When compared to traditional culture, 127 of 132 positive MRSA cultures (96.2%) were detected on HardyCHROM™ MRSA at 24 hours. The remaining five MRSA isolates were detected at approximately 48 hours. All testing (clinical and in-house testing) was incubated at 35 to 37 degrees C.
## f. Analytical sensitivity
### Recovery Study (Limit of Detection (LoD))
ATCC® methicillin-resistant Staphylococcus aureus strains (Staphylococcus aureus subsp. aureus ATCC® 43300 and Staphylococcus aureus subsp. aureus ATCC® 33591) were tested at seven serial dilutions in duplicate. The dilutions were plated in duplicate to HardyCHROM™ MRSA and Tryptic Soy Agar with 5% Sheep Blood (BAP).
On HardyCHROM™ MRSA medium, MRSA strains tested were detected beginning at 16 hours with final counts being reported at 24 hours following aerobic incubation at 35°C. Pink to magenta colonies were detected on all dilutions at 16 hours and colony counts were stable from 21 to 24 hours with no increase in the number of colonies counted. All dilutions plated showed typical pink to magenta colony coloration. There was no significant difference in counts noted between the two observers. Staphylococcus aureus subsp. aureus ATCC® 43300 resulted in a 67.6% recovery when present at 10² concentration and Staphylococcus aureus subsp. aureus ATCC® 33591 resulted in a 34.6% recovery at the same dilution. At 10¹, there was 80% recovery and 52.3% recovery respectively. At 10³ there was no discernable difference in recovery.
## g. Assay cut-off:
Not applicable
## 2. Comparison studies:
a. Method comparison with predicate device:
{9}
Performance of HardyCHROM™ MRSA was evaluated at three geographically diverse hospitals with fresh surveillance specimens collected from the anterior nares. The recovery of methicillin-resistant S. aureus on HardyCHROM™ MRSA was compared to routine culture, defined as isolation of staphylococci on Trypticase Soy Agar with $5\%$ blood (TSAB), with S. aureus identification confirmed by latex agglutination. All S. aureus recovered were tested for mecA mediated resistance by PBP2' latex testing and cefoxitin and oxacillin disk diffusion. Antibiotic disk susceptibility testing followed CLSI methods and interpretive criteria (Performance Standards for Antimicrobial Susceptibility Testing; Twentieth Informational Supplement, M100-S20. 2010). Performance of HardyCHROM™ MRSA was also compared to a commercially available chromogenic medium.
A total of 443 samples were tested against routine culture. A total of 131 specimens were positive on HardyCHROM™ MRSA with concordant results obtained on TSAB and confirmed by PBP2' latex testing and cefoxitin $(30\mu \mathrm{g})$ and oxacillin disk $(1\mu \mathrm{g})$ diffusion testing. An additional specimen was positive on HardyCHROM™ MRSA but did not grow on TSAB. Growth was confirmed as MRSA by PBP2' latex testing and cefoxitin $(30\mu \mathrm{g})$ and oxacillin disk $(1\mu \mathrm{g})$ diffusion testing. Agreement with PBP2', cefoxitin $(30\mu \mathrm{g})$ and oxacillin $(1\mu \mathrm{g})$ disk testing was $93.4\%$ at 24 hours and $97.0\%$ at 48 hours for mecA mediated MRSA. Percent agreement was $99.7\%$ for non-MRSA at both 24 and 48 hours.
The combined data from three clinical trial sites, demonstrated the positive and negative percent agreement at 24 and 48 hours of HardyCHROM™ MRSA with traditional culture and the commercial chromogenic MRSA medium as summarized in Table 1 and Table 2 below:
Table 1: Agreement between Traditional Culture, a Commercial Chromogenic Medium and HardyCHROM™ MRSA
| | MRSA | Non-MRSA* |
| --- | --- | --- |
| HardyCHROM™ MRSA vs. Traditional Culture 24 hours | 93.3% (126/135) (95% CI 89.6 – 98.4%) | 99.7% (307/308) (95% CI 98.9 – 100%) |
| HardyCHROM™ MRSA vs. Traditional Culture 48 hours | 97.0% (131/135) 95% CI 89.6 – 98.5% | 99.7% (307/308) 95% CI 98.9 – 100% |
| HardyCHROM™ MRSA vs. Commercial Chromogenic Medium 24 hours | 98.3% (118/120) 95% CI 97.5 – 99.3% | 97.5% (315/323) 95% CI 89.6 – 98.4% |
| HardyCHROM™ MRSA vs. a Commercial Chromogenic Medium 48 hours | 98.3% (122/124) 95% CI 97.5 – 99.3% | 96.9% (309/319) 95% CI 89.6 – 98.4% |
*Organisms producing other types of non-mecA resistance were not evaluated on this medium (e.g. MOD-SA and BORSA)
{10}
Table 2: Comparison between HardyCHROM™ MRSA and Cefoxitin 30μg Disk 24 hours
| HardyCHROM™ MRSA
24 hours | Cefoxitin 30μg
24 hours | | | Percent Agreement
Positive Percent Agreement – 93.3%
(95% CI 89.6 – 98.4%)
Negative Percent Agreement – 99.7%
(95% CI 98.9 – 100%) |
| --- | --- | --- | --- | --- |
| | MRSA | Non-MRSA^{a} | Total | |
| MRSA | 126 | 1^{b} | 127 | |
| Non-MRSA^{a,c} | 9 | 307 | 316 | |
| Total | 135 | 308 | 443 | |
a. Organisms producing other types of non-mecA resistance were not evaluated on this medium (e.g. MOD-SA and BORSA).
b. 1/1 Blood Agar Plate negative specimen was confirmed as MRSA positive by cefoxitin disk diffusion.
c. Nine MRSA isolates were detected by traditional culture methods, but did not grow on HardyCHROM™ MRSA. Discrepant results were confirmed as MRSA positive by cefoxitin disk diffusion.
The challenge clinical isolates were obtained from a private culture collection and included 10 MRSA and 10 MSSA well-characterized strains. All isolates were coagulase positive and were characterized by PBP2' and cefoxitin testing. The challenge set was tested at three testing sites for three days using three different lots of media and by three different operators performing tested in triplicate on each day of testing. The ten strains of MRSA were tested using a suspension containing approximately $10^{5}$ to $10^{6}$ CFU/ml and the ten strains of MSSA were tested at a concentration of approximately $10^{6}$ to $10^{7}$. Ten $\mu$l of these suspensions were inoculated onto HardyCHROM™ MRSA. At each clinical study site, all isolates showed expected results on the HardyCHROM™ MRSA for each operator at each individual site and across all three sites on each testing day.
b. Matrix comparison:
Not applicable
{11}
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
Not Applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.