K102917 · Exacta Dental Direct, Inc. · EBG · Jun 17, 2011 · Dental
Device Facts
Record ID
K102917
Device Name
FIXTEMP C&B
Applicant
Exacta Dental Direct, Inc.
Product Code
EBG · Dental
Decision Date
Jun 17, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3770
Device Class
Class 2
Indications for Use
FixTemp C&B ™ is a chemically catalyzed, highly filled, resin based restorative material used to fabricate temporary crowns and bridges. U S.A. Federal Law restricts the dispensing of device except on the order of a licensed practitioner.
Device Story
FixTemp C&B is a two-component (catalyst and base) resin-based dental material used to fabricate temporary crowns and bridges. It is dispensed via an industry-standard cartridge and static mixing tip. The material consists of ethoxylated Bis-GMA (multifunctional acrylate) and inorganic fillers (silanated barium glass and silanated silicic acid). It is radiopaque for radiographic identification and compatible with light-cured composites for repair. Used in dental clinics by licensed practitioners to create provisional prosthetics for patients. The material provides a structural form for temporary restorations, aiding in patient dental management during the interim period before permanent restoration placement.
Clinical Evidence
Bench testing only. Biocompatibility testing conducted per DIN EN ISO 10993: Agar Diffusion Test (cytotoxicity) showed non-cytotoxic results using L929 mouse fibroblasts; modified Guinea Pig Maximization Test (sensitization) per ASTM F720-81 showed no sensitization; Ames test (genotoxicity) showed non-mutagenic results. Carcinogenicity, implantation, and inhalation toxicity testing were deemed unnecessary based on low exposure levels and negative results in other biocompatibility assays.
Technological Characteristics
Two-component resin-based material; contains ethoxylated Bis-GMA, silanated barium glass, and silanated silicic acid fillers. Chemically catalyzed polymerization. Radiopaque. Delivered via cartridge/static mixing tip. Biocompatibility tested per ISO 10993 standards.
Indications for Use
Indicated for fabrication of short and long term dental provisional prosthetics (crowns and bridges) in patients requiring temporary dental restorations. Prescription use only.
Regulatory Classification
Identification
A temporary crown and bridge resin is a device composed of a material, such as polymethylmethacrylate, intended to make a temporary prosthesis, such as a crown or bridge, for use until a permanent restoration is fabricated.
{0}------------------------------------------------
5. 510(k) Summary (per 21 CFR 807.92)
102917
### 1) Applicant Information
10 May 2011 Sponsor Mr. Chase Wade EXACTA Dental Direct, Inc. . 44780 Macomb Industrial Dr. Clinton Township MI 48036 800-474-7665 / f- 416 322 7735 chase@exactadental.com
Establishment registration No.
#### 2) General Device Information
Proprietary Name: Trade Name: Common Name Device Classification Name Classification Number: Product Code Review Advisory Committee Classification Advisory Committee Device Classification
#### Consultant
Mr. Richard Keen Compliance Consultants 1151 Hope Street Stamford, CT 06907-1659 203 329 2700 F 203 329 2345 rkeen@fda-complianceconsultants.com
1836392
FixTemp C&B 1m FixTemp C&B ™ FixTemp C&B ™ Crown and Bridge Temporary resin. 21 CFR 872.3770 EBG Dental Dental Class II (nonexempt)
### 3) Predicate Device
FixTemp C&B "" is substantially equivalent to other legally marketed devices in the United States. FixTemp C&B "" functions in a manner similar to and is intended for the same use as the original Temphase "" formulation that was manufactured by Kerr Dental Materials Center.
| Predicate Device | Temphase tm, Temporary Dental Restorative Material.<br>See K020092..<br>Kerr Dental Materials Center<br>1717 West Collins Avenue<br>Orange, California 92867<br>Colleen Boswell |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Device for K020092 | Kerr Corporation, TempFil C&B |
Trademark Notice: All Trademarks used other than those of EXACTA Dental Direct, Inc. are registered to their respective owners.
#### 4) Device Description
The FixTemp C&B *** is a Class II (nonexempt) device that is ideal suited as a dental acrylic to fabricate temporary crown and bridges.
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | Comments of the controlled to the control of the consisted to the controllable of the comments of the comments of the comments of the comments of the comments of the comments | CHANAGE AND A CALL A CHANNEL CONTRACT COLLEGIAL CONTRACT COLLEGIAL CONTRACT COLLEGION |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| EXACTA Dental Direct, Inc.<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | Page . | FixTemp C&B um |
{1}------------------------------------------------
# 5. 510(k) Summary
## (per 21 CFR 807.92)
FixTemp C&B "" provisional material is a two component, catalyst and base, material dispensed and mixed by a industry standard cartridge/static mixing tip combination. FixTemp C&B "" is intended for . use in both short and long term crown and bridge temporaries. The material is compatible with light cured composites for repair and characterization. FixTemp C&B "" contains methacrylate components and is radiopaque for easy radiographic identification.
The scientific concept on which this device is based is the principle that ethoxylated Bis-GMA is a multifunctional acrylates and inorganic fillers as silinated barium glass and silinated silicic acid to form a suitable structure for a short and long term crown and bridge temporaries. This device functions by providing a dental material to short and long term dental provisional prosthetics.
### 5a) Intended Use
The intended use of the FixTemp C&B "" is to form short and long term dental provisional prosthetics.
# 5b) Comparison to Predicate
EXACTA Dental Direct, Inc. has determined that the FixTemp C&B "" device is substantially equivalent to the performance of the predicate device. The differences between our device and the predicate is incidental and not significant. Both our device and the predicate have the same technological and chemical characteristics.
The FixTemp C&B"" device has the same technological characteristics when compared to the predicate device:
- . The Indications for Use for the FixTemp C&B *** device is identical to the Temphase 100, Temporary Dental Restorative Material.
- The Operational principals are the same. ●
- Manufacturing materials and processes are similar. .
Any differences do not affect the safety and effectiveness of the device when used as labeled.
### 5c) Bench Tests
The following bench tests were conducted.
### Cytotoxicity.
Cytotoxicity test was performed as Agar Diffusion Test with cured material, according to Din E N ISO 10993-5 (DIN/EN 30993-5). The recommended cell strain L929 (mouse fibroblasts) was used. Fixtemp C&B met the criteria of the test and was found to be non-cylotoxic.
# Sensitization.
The potential for sensitization of Fixtemp C&B was tested with a modified Guinea Pig Maximization Test (GMPT) according to Magnusson and Kligman (ASTM standards, Standard Practice F720-81), Under the test conditions, Fixtemp C&B did not induce any sensitization,
| | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
| EXACTA Dental Direct, Inc. | Page 2 | FixTemp C&B "m |
{2}------------------------------------------------
# 5. 510(k) Summary
(per 21 CFR 807.92)
## Genotoxicity (Arnes-Test)
Fixtemp C&B was tested with an Ames test for its genotoxic potential.
Fixtemp C&B (extract) was non mutagenic in the Ames test either with or without metabolic activation,
### Carcinogenicity.
A test for carcinogenicity was not conducted, as this test is only sensible according to DIN EN ISO 10993, if there are hints for such a suspect. in view of the extraordinarily low exposition levels and the negative genotoxicity and cytoxicity tests a test for carcinogenicity was not thought to be necessary. This conforms with procedures, that are generally applied for the testing of plastics.
#### Implantation.
Implantation of materials into the muscular tissue of rabbits is done to assess negative reactions of tissue towards biomaterials. If Fixtemp C&B is used as prescribed, the material only comes into contact with oral hard tissues and the surface of oral soft tissues. Therefore no implantation test for Fixtemp C&B was conducted,
#### Acute Systemic Toxicity.
The acute systemic toxicity is not explicitly obligatory according to DIN EN ISO 10993 and therefore was not specifically evaluated. Assessment of components In Fixtemp C&B that amount to more than 3%, have all a very low acute oral toxicity. They are generally used in dental materials in similar or even higher closes, which have already been reviewed and approved by the FDA,
### Inhalation Toxicity.
Fixtemp C&B contains practically no volatile compounds and is not used as an aerosol. Therefore a test for inhalation toxicity, as suggested by the FDA G95-1 memorandum, was not thought useful and therefore was not conducted
#### Test Conclusion
Different in vivo and in vitro tests of Fixtemp C&B according to DN EN ISO 10993 have proven Fixtemp C&B to be not cytotoxic, not genotoxic and not sensitizing.
### Summary
Based on the information, FixTemp C&B "" is substantially equivalent to Temphase "".
{3}------------------------------------------------
# 102917 4. Indications for Use Statement
510(K) Number (If known): no 510(K) number assigned Device Name: FixTemp C&B 'm
# Indications for Use
FixTemp C&B ™ is a chemically catalyzed, highly filled, resin based restorative material used to fabricate temporary crowns and bridges.
U S.A. Federal Law restricts the dispensing of device except on the order of a licensed practitioner.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | XXX |
|------------------|--------------------------|
| | or |
| | Over - The - Counter Use |
| | (Per 21 CFR 801.109) |
| | (Optional Format 1-2-96) |
| | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EXACTA Dental Direct, Inc. | Page 7 | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
| | | FixTemp C&B 'm |
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | And Charles of Children Comments of Children Comments of Children Comments of Children Comments of | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three lines representing the head, body, and legs. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the figure. The logo is black and white.
Food.and Drug_Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUN 1 7 2011
Exacta Dental Direct, Incorporated C/O Mr. Richard Keen Vice President Operations Compliance Consultants 1151 Hope Street Stamford, Connecticut 06907-1659
Re: K102917
Trade/Device Name: Temphase™ Temporary Dental Regulation Number: 21 CFR 872.3770 Regulation Name: Temporary Crown and Bridge Resin Regulatory Class: II Product Code: EBG Dated: May 23, 2011 Received: June 1, 2011
Dear Mr. Keen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2- Mr. Keent
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. nater
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{6}------------------------------------------------
# 102415 4. Indications for Use Statement
510(K) Number (If known): ____________________________________________________________________________________________________________________________________________________ no 510(K) number assigned Device Name: FixTemp C&B ""
# Indications for Use
FixTemp C&B "" is a chemically catalyzed, highly filled, resin based restorative material used to fabricate temporary crowns and bridges.
U S.A. Federal Law restricts the dispensing of device except on the order of a licensed practitioner.
Susan Runnes
(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K102
K102917
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)<br>Concurrence of CDRH, Office of Device Evaluation (ODE) | | |
|------------------------------------------------------------------------------------------------------------------------------------|-----------|--------------------------------------------------|
| Prescription Use<br>XXX | <b>Or</b> | Over - The - Counter Use<br>(Per 21 CFR 801.109) |
| | | (Optional Format 1-2-96 |
| EXACTA Dental Direct, Inc. | Page 7 |
|----------------------------|----------------|
| | FixTemp C&B "" |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.