K102830 · Edan Instruments, Inc. · DPS · Dec 2, 2010 · Cardiovascular
Device Facts
Record ID
K102830
Device Name
ELECTROCARDIOGRAPH
Applicant
Edan Instruments, Inc.
Product Code
DPS · Cardiovascular
Decision Date
Dec 2, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The intended use of the SE-12 series (12-channel electrocardiograph) is to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only.
Device Story
12-channel electrocardiograph (models SE-12, SE-1200, SE-12Express, SE-1200Express) acquires, amplifies, and records ECG signals via body surface electrodes; features thermal printer, LCD display, and data storage/transmission (RS232, Ethernet, optional wireless). Used in hospitals/clinics by physicians and trained professionals. Optional stress exercise test function (SE-12Express/SE-1200Express) provides real-time ST segment analysis, trend graphs, and treadmill/bicycle control. Device provides automated measurements and interpretive statements on advisory basis to assist clinical diagnosis. Benefits include simultaneous 12-lead acquisition, real-time monitoring, and flexible reporting formats.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, including software testing, risk analysis, safety testing, and environmental testing.
Technological Characteristics
12-channel ECG; thermal printer (216mm width); LCD display; rechargeable lithium battery; connectivity via RS232, Ethernet, or wireless AP. Includes alphanumeric keyboard and optional touch screen. Frequency response ≤150Hz. Software-based measurement and interpretation functions.
Indications for Use
Indicated for adult and pediatric patients requiring ECG signal acquisition via body surface electrodes for heart disease analysis and diagnosis in hospital or clinical settings.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
1-
27
## 510(K)Summary of Safety and Effectiveness
: (
This summary of 510k safety and effectiveness is being submitted in according with 21CFR part 807.92
| Submitter: | Edan Instruments, Inc<br>3/F - B, Nanshan Medical<br>Equipments Park, Nanhai Rd 1019#,<br>shekou, Nanshan Shenzhen,<br>518067 P.R. China<br>Tel: +86755 26882220<br>Fax:+86755 26882223 | DEC 2 2010 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Contact person: | Yue Qiuhong<br>Edan Instruments, Inc. | |
| Date: | 2010-9-25 | |
| Proprietary Name: | Electrocardiograph<br>models SE-12/SE-1200/SE-12Express/SE-1200Express | |
| Classification Name: | 21 CFR 870.2340 Electrocardiograph<br>Class II | |
| Product code: | DPS | |
| Predicate Devices: | MAC 5000 ECG Analysis System K014108<br>Manufacturer: GE medical systems information technologies<br>SMART ECG SERIES ELECTROCARDIOGRAPH<br>K091513 Manufacturer: Edan Instruments, Inc | |
| Device Description: | SE-12 series is 12-channel electrocardiograph, it contains<br>four models:SE-12/SE-1200/SE-12Express/SE-1200Express.<br>These four models include the same ECG board, power<br>supply board and main controlling board. The software and<br>function of these four models are similar, except that<br>SE-12Express has the optional function which include stress<br>exercise test. SE-12 series can acquire 12 channel waveforms<br>simultaneously, which can also print out 12 channel | |
{1}------------------------------------------------
SE-12 series Special 510k SUBMISSION
28
electrocardiograph wave simultaneously by a 216mm wide thermal printer, and the waveforms also can be displayed in LCD and stored in flash memory or send to PC by RS232 or Ethernet.
## SE-12 series has the features as follows:
Supporting barcode scanner
તું.
Supporting multi-language
ECG signals of 12 leads are gathered and amplified simultaneously, 12-channel waves are displayed and recorded simultaneously
Full alphanumeric keyboard
(For SE-1200Express and SE-12Express, touch screen is available)
Real-time uploading to PC ECG
Multiple file formats
High resolution thermal recorder, recording frequency response ≤150Hz
Flexible printing formats
The auto, manual, rhythm, R-R analysis and off modes can be chosen freely
Automatic baseline adjustment for optimal printing Convenient operation of system setup and file management Measurement function and interpretation function Hint information of lead off, lack of paper, low battery
capacity etc. Built-in rechargeable lithium battery with large capacity
ECG data can be transmitted to the PC software through the serial cable, net cable, or wireless AP (optional).
The following features are only for the exercise stress test function of SE-12 Express (optional)
Real-time analysis, ST segment and trend are applied while sampling:
Real-time display and print of 12-lead simultaneous ECG waveforms:
ST segment analysis while sampling; ST position is adjustable while sampling;
Providing average templates of three rhythm leads in every stage to observe the change of ST segments between every two stages
Automatically forming elaborate reports, including Summary
{2}------------------------------------------------
.
!
/
| | Report, ST Scope Report and Trend Graph Report |
|---------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Providing classical exercise protocols; exercise protocols can be edited and created. |
| | Automatically controlling and adjusting the speed and grade of the treadmill or the power of the bicycle |
| | Supporting multi-types of treadmill or bicycle |
| Intended Use: | The intended use of the SE-12 series (12-channel electrocardiograph) is to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only. |
| Test Summary: | The following quality assurance measures were applied to the development of the SE-12 series.<br>Software testing Risk analysis Safety testing Environment test |
| Conclusion: | Verification and validation testing was done on SE-12 series. This premarket notification submission demonstrates that SE-12 series is substantially equivalent to the predicate device. |
.
: - - - - - - - -
.
ا سف ساب
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
DEC 2 2010
Edan Instruments, Inc. c/o Ms. Yue Qiuhong Registrar 3/F - B, Nanshan Medical Equipments Park, Nanhai Rd 1019# Shekou, Nanshan Shenzhen, 518067 P.R. China
Re: K102830 Trade/Device Name: SE Series Electrocardiograph (Models SE-12, SE-1200, SE12Express, SE-1200Express) Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II (two) Product Code: DPS Dated: October 28, 2010 Received: November 8, 2010
Dear Ms. Qiuhong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2 - Ms. Yue Oiuhong
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
S Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): K102830
Device Name: Electrocardiograph models SE-12/SE-1200/SE-12Express/SE-1200Express
Indication for Use
SE-12 series is intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. SE-12 series is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by SE-12 series can help users to analyze and diagnose heart disease. However the ECG with measurements and interpretive statements is offered to clinician on an advisory basis only.
Prescription Use (21 CFR Part 801 Subpart D) Ог
Over the Counter Use (21 CFR Part 801 Subpart C)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
W.M.S.
vascular Devices
510(k) Number K102830
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.