CLARITY ADVANCED CERAMIC BRACKETS

K102803 · 3M Unitek Corporation · NJM · Feb 17, 2011 · Dental

Device Facts

Record IDK102803
Device NameCLARITY ADVANCED CERAMIC BRACKETS
Applicant3M Unitek Corporation
Product CodeNJM · Dental
Decision DateFeb 17, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5470
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Clarity™ Advanced Ceramic Brackets are intended for use in orthodontic treatment. The brackets are affixed to teeth so that pressure can be exerted on the teeth.

Device Story

Clarity Advanced Ceramic Brackets are orthodontic appliances used by orthodontists to move teeth. The device consists of a translucent alumina body with a glass-grit bonding base, optionally coated with stabilized zirconia. Brackets are bonded to teeth; an orthodontic wire is inserted into the bracket slots to apply corrective pressure. The system includes a water-soluble color placement indicator to assist in positioning and identification. The device is used in a clinical setting by a physician. The primary benefit is the aesthetic correction of tooth alignment. Performance is verified through bench testing of bond strength, bracket strength, material friction, and debonding characteristics.

Clinical Evidence

No clinical performance testing was conducted. Evidence is based on non-clinical bench testing comparing the subject device to predicate devices. Evaluated properties included shear-peel bond strength, torsional bracket strength, surface material friction, and squeeze debond moments. Results confirmed performance is comparable to predicates and meets minimum requirements.

Technological Characteristics

Ceramic bracket body (microfine ceramic) with glass-grit bonding base. Optional stabilized zirconia coating. Features include true-twin tiewings, base flanges for adhesive cleanup, and vertical slot/stress concentrator for debonding. Rounded corners and edges. Water-soluble color indicators for positioning. Non-powered, passive mechanical device.

Indications for Use

Indicated for use in orthodontic treatment to exert pressure on teeth for repositioning.

Regulatory Classification

Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K102803 #### 510(k) Summary FEB 17 2011 | 510(k) Submitter | 3M Unitek Corp, 2724 S Peck Rd, Monrovia, CA 91016 | |----------------------------|---------------------------------------------------------------------| | Contact person | L. Marlyn Scheff, Regulatory Affairs<br>Phone: (626) 574-4496 | | Date Summary was Prepared | January 12, 2011 | | Device Name | Clarity™ Advanced Ceramic Brackets | | Common Name | Orthodontic ceramic bracket | | Recommended Classification | Orthodontic ceramic bracket<br>(21 CFR 872.5470, Product Code: NJM) | #### Predicate Devices K062345, Clarity™ SL Self-Ligating Ceramic Brackets K944286, Clarity™ Metal-Reinforced Ceramic Brackets K950992, InVu® Aesthetic Braces K082974, Mystique® MB Clear Braces #### Description of Device Clarity Advanced Ceramic Brackets are intended to teeth, upon which an orthodontic wire is used to move the teeth to new positions. Clarity Advanced Ceramic Brackets consist of a translucent alumina body and a glass-grit bonding base. The bracket is either uncoated or coated with a thin film of stabilized zirconia. The brackets incorporate a watersoluble color placement indicator system that marks archwire and vertical slots to aid in bracket positioning and color codes tie wing(s) to facilitate bracket identification. #### Indications for Use Clarity™ Advanced Ceramic Brackets are intended for use in orthodontic treatment. The brackets are affixed to teeth so that pressure can be exerted on the teeth. #### Substantial Equivalence Both the non-clinical data and the biocompatibility evaluation indicate that Clarity™ Advanced Ceramic Brackets are safe and effective for their intended use in orthodontic treatment and perform as well or better than predicate devices. The table on the next page compares the new device with the predicate devices. Information provided in this 510(k) submission shows that Clarity™ Advanced Ceramic Brackets are substantially equivalent to the predicate devices in terms of intended use, indications for use, composition, device design, and performance. This 510(k) also includes data from bench testing to evaluate the performance of Clarity™ Advanced Ceramic Brackets compared to the predicate devices. The properties evaluated include bond strength, bracket strength, material friction, and debond strength. {1}------------------------------------------------ ## Technological Characteristics Clarity Advanced Ceramic Brackets are substantially equivalent in design features to the predicate devices. ## Device Material Clarity Advanced, Clarity SL, Clarity, InVu and Mystique MB brackets all are have a bracket body made of ceramic, with Clarity Advanced, Clarity SL, and InVu brackets made of microfine ceramic. Clarity Advanced, Clarity SL, and Clarity have a glass-grit bonding base whereas In Vu has a molded polymer bonding base and Mystique MB have a molded alumina bonding base. Clarity Advanced has a zirconia, or no coating, on the bracket and in the archwire slot, Clarity SL has a metal liner, or no liner, in the archwire slot, Clarity has a metal liner in the archwire slot. InVu has no liner or coating in the archwire slot, and Mystique MB has a silica coating in the archwire slot. Clarity Advanced has color slot & dot indicators whereas Clarity SL. Clarity, InVu have color dot indicators and Mystique MB has no color indicators. ## Device Design Clarity Advanced, Clarity SL, Clarity, InVu and Mystique MB brackets all have tiewing undercut spaces for orthodontic ligatures. Clarity Advanced, Clarity SL, Clarity and InVu brackets have true-twin tiewings, i.e. four tiewings, for versatile use with auxiliaries. Clarity Advanced. Clarity, and In Vu brackets contain base flanges for bracket placement and adhesive flash cleanup. Clarity Advanced, Clarity SL and InVu brackets contain a molded ceramic bracket body with rounded corners and edges, which replaces the angular profile of machined ceramic brackets, and round hook on the distal-gingival tiewings. Clarity Advanced, Clarity SL and Clarity brackets contain vertical slot and stress concentrator to facilitate debonding of the bracket from the tooth. #### Nonclinical Performance Testing The nonclinical performance testing analysis shows that Clarity Advanced Ceramic Brackets perform comparably to the predicate devices as follows: - 1. The shear-peel bond strength test measures the force required to debond a bracket when a force is applied in the occlusal direction. The test results showed that the bond strengths of Clarity Advanced, Clarity SL and Clarity brackets are comparable and exceed the minimum bond strength to hold the bracket to the tooth. - 2. The bracket strength test measures the torsional force to break a bracket when a rectangular archwire is twisted in the wire slot. The test results showed that the bracket strengths of Clarity Advanced Ceramic Brackets are comparable to Clarity Metal-Reinforced Ceramic Brackets and InVu Aesthetic Braces and exceed the minimum requirements. - 3. The bracket material friction test measures the surface frictional forces of a stainless steel wire against a bracket surface. The test results showed that the zirconia-coated aluminum i oxide surface exhibited lower coefficients of friction as compared to the uncoated aluminum oxide surface. {2}------------------------------------------------ - 4. The squeeze debond test measures the forces applied to the sections left and right of the vertical slot in the Clarity Advanced, Clarity and Clarity SL brackets which cause the bracket to debond from the adhesive. The test results showed that squeeze debond moments for Clarity Advanced Ceramic Brackets are comparable to those for Clarity™ SL brackets and slightly lower for Clarity brackets. In addition, a biocompatibility assessment was developed for Clarity Advanced Ceramic Brackets using standard risk assessment techniques and consideration of FDA and internationally recognized guidelines. # Clinical Performance Testing No clinical performance testing was conducted on Clarity™ Advanced Ceramic Brackets. ## Conclusion The results from the nonclinical performance testing and the biocompatibility assessment demonstrate that Clarity Advanced Ceramic Brackets are safe and effective for their intended use and perform as well as predicate devices. {3}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three wavy lines emanating from a central point. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the caduceus. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002 Ms. Marlyn L. Scheff Regulatory Affairs 3M UNITEK Corporation 2724 South Peck Road Monrovia, California 91016 FEB 17 2011 Re: K102803 Trade/Device Name: Clarity" Advanced Ceramic Brackets Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: II Product Code: NJM Dated: January 26, 2011 Received: February 1, 2011 Dear Ms. Scheff: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2- Ms. Scheff Enclosure Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fdaggov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. Anstmen Un Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): K102803 Device Name: ClarityTM Advanced Ceramic Brackets Indications for Use: Clarity™ Advanced Ceramic Brackets are intended for use in orthodontic treatment. The brackets are affixed to teeth so that pressure can be exerted on the teeth. Prescription Use X (Per 21 CFR 801 Subpart D) OR Over-The-Counter Use (21 CFR 801 Subpart C) # PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 Susan Runse (Division Sign-Off) (Division Sign-On) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: CD 2803
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...