K102771 · Somatex Medical Technologies GmbH · KNW · Jun 24, 2011 · Gastroenterology, Urology
Device Facts
Record ID
K102771
Device Name
BIOPSY HANDY, MRI BIOPSY HANDY
Applicant
Somatex Medical Technologies GmbH
Product Code
KNW · Gastroenterology, Urology
Decision Date
Jun 24, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The Biopsy Handy and MRI Biopsy Handy are intended to obtain diagnostic samples of soft tissues for histological examination during a percutaneaous biopsy procedure. Both devices are indicated to be used with ultrasound or CT-Guidance techniques. Only the MRI Biopsy Handy is indicated to be used with magnetic resonance imaging (MRI) techniques.
Device Story
Biopsy Handy and MRI Biopsy Handy are sterile, single-use, semiautomatic percutaneous soft tissue biopsy systems. Components include handle, cannula, stylet button, stylet with specimen chamber, hub, guide bush, slide, and pressure spring. Device operates by manual activation of spring-loaded mechanism to capture tissue sample in chamber. Used in clinical settings under ultrasound, CT, or MRI guidance (MRI version only) by healthcare professionals. Provides histological samples for diagnostic evaluation; assists in disease diagnosis through tissue analysis.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including biocompatibility, sterility, packaging, and MR-compatibility assessments.
Technological Characteristics
Semiautomatic percutaneous biopsy system. Components: handle, cannula, stylet with specimen chamber, pressure spring. Single-use, sterile. MRI Biopsy Handy compatible with MRI environments. No software or electronic components.
Indications for Use
Indicated for patients requiring percutaneous soft tissue biopsy for histological examination. Compatible with ultrasound, CT-guidance, and (for MRI Biopsy Handy) MRI techniques.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for SOMATEX Medical Technologies GmbH. Above the company name is the number k/02771. The SOMATEX logo has a registered trademark symbol next to it.
Rheinstraße 7d | D-14513 Teltow
JUN 2 4 2011
## 510(k) SUMMARY
### Biopsy Handy / MRI Biopsy Handy
Submitter:
Somatex Medical Technologies GmbH Rheinstrasse 7d 14513 Teltow Germany Phone: +49 3328 3076 13 +49 3328 3076 99 Fax:
Official Correspondent: Susanne Raab
Regulatory Affairs Consultant 1480 Cambridge Street Cambridge, MA 02139 Phone: 617 547 0628 Fax: 617 520 2136 e-mail: sbraab@comcast.net
Biopsy Handy / MRI Biopsy Handy Trade Name:
Common Name: Semiautomatic Biopsy System
II
Biopsy Instrument, 21 C.F.R. 876.1075 Classification Name:
Regulatory Class:
Product Code: KNW
Device Description: The Biopsy Handy and the MRI Biopsy Handy are sterile, single use, semiautomatic percutaneous soft tissue biopsy systems each consisting of the following major components: a device handle, a cannula, a stylet button, a stylet with a specimen chamber, a hub of the cannula, a guide bush, a slide and a pressure spring. In addition, the cannulas of the
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for SOMATEX Medical Technologies GmbH. The logo is in black and white and features the word "SOMATEX" in large, bold letters. Below the word "SOMATEX" is the phrase "MEDICAL TECHNOLOGIES GMBH" in smaller letters.
einstraße 7d | D-1451 3328 30 76-0 | Fax +49 (0) 3328 30 76 @somatex.com | www.somatex.com
proposed devices will be delivered with a protective tube.
The Biopsy Handy and MRI Biopsy Handy are intended to Intended Use: obtain diagnostic samples of soft tissues for histological examination during a percutaneaous biopsy procedure. Both devices are indicated to be used with ultrasound or CT-Guidance techniques. Only the MRI Biopsy Handy is indicated to be used with magnetic resonance imaging (MRI) techniques.
> The Biopsy Handy and MRI Biopsy Handy are substantially equivalent to the Speedybell manufactured by Biopsybell S.A.S., which has been cleared by FDA on August 17, 2001 (K010735), and the Daum Biopsygun manufactured by Daum Corp., which has been cleared by FDA on March 6, 1998 (K974576). Specifically, the proposed devices and the predicate devices are all intended to obtain diagnostic samples of soft tissues for histological examination during a percutaneaous biopsy procedure. In addition, the proposed devices and the predicate device Speedybell are indicated to be used with ultrasound or CT-Guidance techniques. Finally, both the MRI Biopsy Handy and the predicate device Daum Biopsygun are indicated to be used with magnetic resonance imaging (MRI) techniques.
Furthermore, the Biopsy Handy and MRI Biopsy Handy are identical or similar in technology, design and material to the
#### Substantial Equivalence:
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for SOMATEX Medical Technologies GmbH. The word "SOMATEX" is in large, bold, sans-serif font, with a registered trademark symbol to the right of the "X". Below the main wordmark, "MEDICAL TECHNOLOGIES GMBH" is printed in a smaller, sans-serif font.
MADE IN GERM.
Rheinstraße 7d | 0-14513 Teltow Tel +49 [0] 3328 30 76-0 | Fax +49 [0} 3328 30 76-95 info@somatex.com | www.somatex.co
predicate devices. Both the proposed and the predicate devices consist of the same primary components and the component materials of the proposed device and the predicate devices are substantially equivalent.
Based on the same intended use and the similiarities in technolgy, design and materials, the Biopsy Handy and MRI Biopsy Handy are substantially equivalent to their predicate devices. The minor technological differences between the proposed devices and the predicate devices do not raise new questions of safety and effectiveness.
Finally, biocompatibility, sterility, packaging and MRtesting demonstrate the safety and effectiveness of the proposed devices.
Date Prepared:
February 24, 2011
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its wing. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circle around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Somatex Medical Technologies GmbH % Ms. Susanne Raab Regulatory Affairs Consultant 1480 Cambridge Street Cambridge, Massachusetts 02139
# JUN 2 4 2011
Re: K102771
Trade/Device Name: Biopsy Handy/MRD Biopsy Handy Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: Class II Product Code: KNW Dated: June 23, 2011 Received: June 24, 2011
Dear Ms. Raab:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
Page 2 - Ms. Susanne Raab
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark A. Millman
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indication for Use
510(k) Number (if known):
Device Name: Biopsy Handy / MRI Biopsy Handy
Indications for Use:
The Biopsy Handy and MRI Biopsy Handy are intended to obtain diagnostic samples of soft tissues for histological examination during a percutaneaous biopsy procedure. Both devices are indicated to be used with ultrasound or CT-Guidance techniques. Only the MRI Biopsy Handy is indicated to be used with magnetic resonance imaging (MRI) techniques.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (OL
Nil RP Qyle fortkn
ivision Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K102771
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.