K102766 · C.R. Bard, Inc. · FTL · Dec 16, 2010 · General, Plastic Surgery
Device Facts
Record ID
K102766
Device Name
COMPOSIX L/P MESH WITH ECHO PS POSITIONING SYSTEM
Applicant
C.R. Bard, Inc.
Product Code
FTL · General, Plastic Surgery
Decision Date
Dec 16, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Composix™ L/P Mesh is indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias and chest wall defects. The Echo PSTM Positioning System is intended to be used to facilitate the delivery of soft tissue prosthetics during laparoscopic hernia repair.
Device Story
Composix™ L/P Mesh with Echo PS™ Positioning System combines a nonabsorbable hernia repair mesh with a removable mesh deployment balloon. Mesh consists of large pore polypropylene and ePTFE layers stitched with PTFE monofilament. Positioning system features an inflatable channel to facilitate laparoscopic unrolling, positioning, and placement of mesh. System includes an introducer tool (metal tines/T-cap assembly) to roll the device for laparoscopic introduction. Inflation assembly updated to a syringe-based apparatus. Used in clinical/surgical settings by surgeons. Device provides structural reinforcement for soft tissue defects; positioning system aids delivery, potentially reducing procedural complexity and improving mesh placement accuracy.
Clinical Evidence
Bench testing and in vivo simulated use experiments were conducted. Biocompatibility testing performed per ISO 10993-1. No clinical data provided.
Indicated for patients requiring reconstruction of soft tissue deficiencies, including hernia repair and chest wall defects. Used by clinicians during laparoscopic hernia repair procedures.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
K102766
pg lot 3
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) Summary is provided per the requirements of section 807.92(c).
DEC 1 6 2010
# Submitter Information:
| Submitter's Name:<br>Address: | Davol, Inc., Subsidiary of C. R. Bard, Inc.<br>100 Crossings Boulevard<br>Warwick, RI 02886 |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Telephone: | (401) 825-8575 |
| Fax: | (401) 825-8763 |
| Contact Person: | Keti Sino |
| Device Name: | |
| Trade Namc: | Composix™ L/P Mesh with Echo PSTM<br>Positioning System |
| Common/Usual Name: | - Surgical Mesh<br>- Endoscope & Accessories<br>- Mesh Deployment Balloon |
| Classification Name: | - Mesh, Surgical, Polymeric<br>- Mesh Deployment Balloon<br>- Laparoscope, General & Plastic Surgery |
| Classification Code: | - Class II, § 878.3300, Product Code FTL<br>- Class II, § 878.3300, Product Code OQL<br>- Class II, § 876.1500, Product Code GCJ |
# Predicate Device Names:
- Bard Composix® L/P Mesh, K061754 (Davol Inc.), FDA cleared on . 10/23/2006
- Surgical Structures Mesh GPS™ Deployment Balloon, K092726, FDA . cleared on 03/15/2010
# PREMARKET NOTIFICATION FOR THE COMPOSIX™ L/P MESH WITH ECHO PSTM POSITIONING SYSTEM
-CONFIDENTIAL-SECTION 7
{1}------------------------------------------------
#### Device Description:
The proposed Composix™ L/P Mesh with Echo PSTM Positioning System is comprised of a hernia repair mesh with a pre-attached mesh positioning system (mesh deployment balloon).
The mesh, Composix™ L/P, is a low profile, nonabsorbable, sterile prosthesis designed for the reconstruction of soft tissue deficiencies, previously cleared under K061754. It is constructed of one layer of large pore polypropylene mesh and one layer of expanded polytetrafluoroethylene (ePTFE) stitched together with PTFE monofilament.
The Composix™ L/P Mesh described above will be preassembled with the Echo PSTM Positioning System, a removable mesh deployment balloon, previously cleared under K092726. The positioning system is designed to help facilitate laparoscopic deployment, including unrolling, positioning, and placement of the Composix™ L/P Mesh. Additionally, all sizes of the proposed product will be packaged with an Introducer Tool. The Introducer Tool consists of a metal tines/T-cap assembly which is used to roll the proposed device in order to facilitate laparoscopic introduction. The Introducer Tool is identical to that included and cleared under the Composix® L/P Mesh K061754.
#### Intended Use:
The Composix™ L/P Mesh is indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias and chest wall defects. The Echo PSTM Positioning System is intended to be used to facilitate the delivery of soft tissue prosthetics during laparoscopic hernia repair.
# Summary of Similarities and Differences in Technological Characteristics, Performance and Intended Use:
The Composix™ L/P Mesh and the predicate mesh have the same indication: reinforce soft tissue, i.e., repair of hernias and chest wall defects. The proposed product has the same intended use, physical attributes, performance characteristics, and materials as the predicate Bard Composix™ L/P Mesh.
The proposed Composix™ L/P Mesh with Echo PSTM Positioning System simply combines two previously cleared devices: K061754 (which included the Introducer Tool) and K092726, the predicate mesh positioning system. However, for optimal functionality in combination with the Composix™ Mesh that is part of the proposed product, two changes have been made to the positioning system. These modifications include a change to the shape of the positioning system and a change to the type of inflation assembly that is to be used with the positioning system. Similar to its predicate device, the Echo PSTM
# PREMARKET NOTIFICATION FOR THE COMPOSIX™ L/P MESH WITH ECHO PSTM POSITIONING SYSTEM
#### -CONFIDENTIAL-SECTION 7
{2}------------------------------------------------
K102766
pg 3 of 3
Positioning System has an inflatable channel or tubular structure. However, the predicate device shape was based on a universal design concept in order for it to be compatible with currently marketed mesh devices from various manufacturers, whereas the proposed product is provided in an oval shaped to better match the shape and configuration of the Composix™ L/P Mesh. The second change addresses the positioning system inflation assembly which has been updated to a syringe inflation apparatus, replacing the original bulb type inflation apparatus, as used for the predicate device positioning system. The Echo PSTM Positioning System maintains the same intended use, physical attributes, performance characteristics, and materials as the Mesh GPSTM Deployment Balloon (K092726).
The Introducer Tool that will be packaged with all size configurations of the proposed device is the same as the Introducer Tool currently provided in the predicate Composix™ L/P Mesh device, as cleared in K061754.
#### Performance Data:
Biocompatibility testing in accordance to ISO 10993-1 standards was conducted on the proposed device (mesh and positioning system) and the results indicate that the device is biocompatible per these standards.
No biocompatibility testing was conducted on the Introducer Tool to be packaged with the proposed device as the 304 Stainless Steel material used for the rolling tines is a recognized biocompatible material as per ASTM F899 - 09e1 Standard Specification for Wrought Stainless Steels for Surgical Instruments. In addition. the polymer handle attached to the rolling tines and the T-cap are non patient contacting material, and therefore, not subject to biocompatibility testing requirements.
Bench testing results and in vivo simulated use experiments demonstrate that the proposed device design meets product specifications and intended uses.
All test results provided in this submission support the safety and effectiveness of the device for its intended use and demonstrate that the proposed device is substantially equivalent to its predicate devices.
### PREMARKET NOTIFICATION FOR THE COMPOSIX™ L/P MESH WITH ECHO PSTM POSITIONING SYSTEM
-CONFIDENTIAL-SECTION 7
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is an abstract image of an eagle.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
# DEC I 6 2010
C. R. Bard, Inc. Davol. Inc. % Keti Sino 100 Crossing Boulevard Warwick, Rhode Island 02886
Re: K102766
Trade/Device Name: Composix" L/P Mesh with Echo PS™ Positioning System Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTL, OQL, GCJ Dated: September 23, 2010 Received: September 24, 2010
Dear Keti Sino:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration: listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
Page 2 – Keti Sino
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm11.78001 html far the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.hum for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Sincerely yours,
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K102766
#### INDICATION FOR USE STATEMENT
510(k) Number (if known):
Device Name:
# Composix™ L/P Mesh with Echo PSTM Positioning System
Composix™ L/P Mesh is indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias and chest wall defects. The Echo PSTM Positioning System is intended to be used to facilitate the delivery of soft tissue prosthetics during laparoscopic hernia repair.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Usc (21 CFR 801 Subpart C)
(PLËASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Kumefa MXM
(Division Sign Off)
(Division Sign-Oil) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K102766
20
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.