K102750 · Geto2, Inc. · CAW · Feb 11, 2011 · Anesthesiology
Device Facts
Record ID
K102750
Device Name
O2-IN-A-BOX, O2-IN-A-BOX
Applicant
Geto2, Inc.
Product Code
CAW · Anesthesiology
Decision Date
Feb 11, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Emergency Use - OTC: Intended to produce oxygen for emergency use. There is a minimum 6 LPM flow rate for at least 15 minutes established a minimum total oxygen capacity of 90 liters. Prescriptive Use: A portable oxygen generator intended to produce oxygen to provide supplemental oxygen as prescribed by a physician.
Device Story
Portable oxygen generator; utilizes chemical reaction to produce oxygen; single-use, disposable device. Emergency version provides minimum 6 LPM for 15 minutes (90L capacity). Rx version provides variable flow (0.5 LPM for 60 min or 3.0 LPM for 30 min). Used in emergency settings (OTC) or home/institutional/travel environments (Rx). Operated by patients or caregivers. Output is supplemental oxygen delivered via standard mask or nasal cannula. Benefits include immediate access to oxygen in emergency or prescribed supplemental settings.
Clinical Evidence
Bench testing only. Testing included VOC/PM2.5, residual hydrogen peroxide, bacteria/mold, environmental (temperature), mechanical (vibration), flow rate, flow duration, and oxygen concentration percentage. All tests passed.
Technological Characteristics
Portable oxygen generator; chemical reaction-based oxygen generation. Gas pathway materials tested for biocompatibility per EPA TO-15 and NIOSH 0500. Single-use, disposable form factor. No electronic components or software.
Indications for Use
Indicated for individuals requiring emergency oxygen (OTC) or supplemental oxygen as prescribed by a physician (Rx). No specific age or gender limitations. No contraindications.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}------------------------------------------------
# FEB 1 1 2011
## 510(k) Summary Page 1 of 5 10-Feb-11
| Get-O2, Inc.<br>27 Bayberry Lane<br>Westport, CT 06880 | Tel - 603-570-2579 |
|--------------------------------------------------------|----------------------------------------------------------------------------------------|
| Official Contact: | Richard P. Imbruce, Ph.D, President |
| Proprietary or Trade Name: | O2-in-a-boxTM - Emergency Use - OTC<br>O2-in-a-boxTM - Rx use |
| Common/Usual Name: | Portable oxygen generator |
| Classification Name: | Portable Oxygen Generator<br>CAW - 868.5440 |
| Predicate Devices: | OxySure - Model 615 - K052396 - Emergency Use<br>Oxytec - Model 900 - K043615 - Rx use |
# Device Description
Indications for Use
#### Emergency Use - OTC
Intended to produce oxygen for emergency use.
There is a minimum 6 LPM flow rate for at least 15 minutes established a minimum total oxygen capacity of 90 liters.
## Prescriptive Use -
A portable oxygen generator intended to produce oxygen to provide supplemental oxygen as prescribed by a physician.
| Flow rate | Delivery Time | Total Oxygen Capacity |
|-----------|---------------|-----------------------|
| 0.5 LPM | 60 minutes | 30 liters |
| 3 LPM | 30 minutes | 90 liters |
#### Environment of Use
Emergency - OTC - locations where emergency oxygen maybe needed Rx use - Home, institutional, travel / mobile environments
## Performance Testing
Performance testing was performed:
- VOC / PM 2.5 and Residual Hydrogen Peroxide .
- Bacteria / Mold Testing .
- Environmental testing high and low temperature conditions .
{1}------------------------------------------------
## 510(k) Summary Page 2 of 5 10-Feb-11
- Mechanical testing vibration ●
- . Flow rate and flow duration for Emergency OTC and Rx Use including external container temperature and temperature of gas at outlet
- . % oxygen generated
The proposed devices met and passed all the performance testing as outlined above.
## Comparative Table - OTC - Emergency Use
| Features | Predicate<br>OxySure K052396 | Proposed Device<br>O₂-in-a-box™<br>OTC - Emergency Use |
|------------------------------|---------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| Indications for use | To produce oxygen for emergency<br>use | To produce oxygen for<br>emergency use |
| Environment of Use | Home, emergency locations | emergency locations |
| Patient Population | Not discussed | No limitations |
| Contraindications | None | None |
| Method for oxygen generation | Chemical reaction | Chemical reaction |
| Patient interface | Standard oxygen mask or nasal<br>cannula | Standard oxygen mask or nasal<br>cannula |
| Specifications | % Oxygen - 99%<br>Flow rate minimum - 6 LPM<br>Duration at least 15 minutes<br>90 liter oxygen capacity | % Oxygen - 99%<br>Flow rate minimum - 6 LPM<br>Duration at least 15 minutes<br>90 liter oxygen capacity |
| Single use, disposable | Yes | Yes |
| OTC designation | Yes | Yes |
| Performance testing | | |
| Flow rate | > 6 LPM | > 6 LPM |
| Duration of flow | At least 15 minutes | At least 15 minutes |
| % oxygen | >99% | > 99% |
| VOC / PM₂.₅ | Yes | Yes |
| Housing temperature | Yes | Yes |
| Temperature of gas at outlet | Yes | Yes |
| Accessories | Mask<br>Oxygen cannula | Mask<br>Oxygen cannula |
## OTC - Emergency Use Substantial Equivalence
The O2-in-a-box™ for Emergency Use - OTC is viewed as substantially equivalent to the predicate device because:
## Indications for Use -
Intended to produce oxygen for emergency use. There is a minimum 6 LPM flow rate for at least 15 minutes established a minimum total oxygen capacity of 90 liters.
{2}------------------------------------------------
#### 510(k) Summary Page 3 of 5 10-Feb-11
## Environment of use -
Locations where emergency oxygen may be needed.
## Technology -
The O2-in-a-box™ for Emergency Use - OTC utilizes chemicals to be mixed when the device is ready for use. There chemical reaction generates oxygen. The used a chemical reaction to generate oxygen is identical technology to predicate OxySure (K052396).
## Materials -
The materials in the gas pathway have been tested via VOC and PM2,5 as is expected for such devices. The test results demonstrate that the O2-in-a-box™ for Emergency Use - OTC does not generate unwanted particulate matter or undesirable gases. This is similar to the predicate OxySure (K052396).
## Environment of Use -
The O2-in-a-box™ for Emergency Use - OTC has the identical environments of use, i.e. emergency settings, as does the predicate OxySure (K052396).
## Patient Population -
The patient population is not specified by the predicate in the 510(k) Summary but the intended use is for individuals require oxygen in an emergency setting. This would be identical to the predicate OxySure (K052396).
## Performance Specifications and Testing -
The O--in-a-box™ for Emergency Use - OTC has the minimum performance specifications required to meet emergency use, which are - at least a flow rate of 6 LPM for at least 15 minutes. This is identical to the predicate OxySure (K052396).
The O2-in-a-box™ for Emergency Use - OTC is equivalent to the predicate, OxySure (K052396), for performance specifications and testing.
The rationale for choosing the predicate is summarized as follows:
## RX Use Substantial Equivalence
The O2-in-a-box™ for Rx Use is viewed as substantially equivalent to the predicate devices because:
{3}------------------------------------------------
## 510(k) Summary Page 4 of 5 10-Feb-11
Comparative Table - Rx Use
| Features | Predicate | Proposed Device |
|---------------------------------|------------------------------------|--------------------------------|
| | Oxytec Model 900 | O2-in-a-boxTM |
| | K043615 | OTC - Emergency Use |
| Indications for use | For prescription use by patients | For prescription use by |
| | requiring high concentration of | patients requiring oxygen on a |
| | oxygen on a supplemental basis | supplemental basis |
| Environment of Use | Home, institutional, travel / | Home, institutional, travel / |
| | mobile environments | mobile environments |
| Patient Population | Require supplemental oxygen | Require supplemental oxygen |
| Contraindications | None | None |
| Method for oxygen generation | Molecular sieve | Chemical reaction |
| Patient interface | Standard oxygen mask or nasal | Standard oxygen mask or nasal |
| | cannula | cannula |
| Single use, disposable | No | Yes |
| Rx designation | Yes | Yes |
| Who designates use and sets | Clinician prescribes and designate | Clinician prescribes and |
| flow rate | flow rate | designate flow rate |
| Who administers oxygen | Caregivers, self directed | Caregivers, self directed |
| Performance testing | | |
| Flow rate range | > 1-5 1pm | 0.5 LPM and 3.0 LPM |
| Duration of flow / O2 capacity | Limited by battery life | 0.5 LPM - 60 min - 30 l O2 |
| | | 3.0 LPM - 30 LPM - 90 l O2 |
| % oxygen | 89% +/- 3% | > 99% |
| VOC / PM25 | Yes | Yes |
| Housing temperature | < 45°C | < 45°C |
| Temperature of gas at outlet | Ambient at patient interface | < 45°C at outlet |
| | | Ambient at patient interface |
| Accessories / Patient interface | Mask | Mask |
| | Oxygen cannula | Oxygen cannula |
## Indications -
The O2-in-a-box™ for Rx Use is intended to provide supplemental oxygen to individuals as prescribed by a clinician. These indications are identical to predicate - Oxytec (K043615).
## Technology -
The O2-in-a-box™ for Rx Use utilizes chemical reaction technology which is equivalent to the predicate OxySure (K052396) for flow rates < 6 LPM.
{4}------------------------------------------------
#### 510(k) Summarv Page 5 of 5 10-Feb-11
#### Environment of Use -
The O2-in-a-box™ for Rx Use is intended to be used in environments which are equivalent to the predicate Oxytec (K043615). These can include home, institutional, travel / mobile environments.
#### Patient Population -
The O4-in-a-box™ for Rx Use has the identical patient population, those requiring supplemental oxygen, as the predicate Oxytec (K043615). In addition, as defined in the FDA guidelines for devices delivery < 6 LPM or 90 liters oxygen capacity they are to be considered prescriptive.
## Performance Specifications -
The O2-in-a-box™ for Rx Use has similar performance specifications, variable flow rates, as the predicate Oxytec (K043615).
Since this is a prescriptive device, the prescriber is to instruct the patient as to the prescribed flow rates of oxygen needed and the proposed device will be matched to those prescribed requirements. This is similar to the predicate Oxytec (K043615) device.
## Performance Testing -
To demonstrate safety we have performed the following bench tests:
- VOC / PM > < and Residual Hydrogen Peroxide .
- Bacteria / Mold Testing .
- Environmental testing high and low temperature conditions .
- Mechanical testing vibration .
- Flow rate and flow duration for Emergency OTC and Rx Use including external . container temperature and temperature of gas at outlet
- % oxygen generated .
## Materials and Biocompatibility -
To demonstrate biocompatibility, all the materials in the gas pathway have been tested via VOC and PM2 < and found to be within the acceptable limits as outlined in Volatile Organic Compounds (VOC) test method - EPA TO-15 VOC and Particulate Material (PM) test method -NIOSH 0500 PM2 5.
## Conclusion -
The O2-in-a-box™ for Rx Use is equivalent to the predicates for indications for use, environments of use, patient population, technology, and performance specifications.
{5}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
# FEB 1 1 2011
Geto2. Incorporated C/O Mr. Paul Dryden Promedic, Incorporated 24301 Woodsage Drive Bonita Springs, Florida 34134
Re: K102750
Trade/Device Name: O²-in-a-box™ portable oxygen generator –Emergency – OTC
O²-in-a box" - Rxx use Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: II Product Code: CAW Dated: February 3, 2011 Received: February 4, 2011
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2- Mr. Dryden
Enclosure
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices.and Radiological Health
{7}------------------------------------------------
## Indications for Use Statement
Page 1 of 2
510(k) Number: K102750
O2-in-a-box™ portable oxygen generator -Device Name: Emergency - OTC
Indications for Use:
Intended to produce oxygen for emergency use.
There is a minimum 6 LPM flow rate for at least 15 minutes established a minimum total oxygen capacity of 90 liters.
Environment of use - locations where emergency oxygen maybe needed
Prescription Use (Part 21 CFR 801 Subpart D) or
Over-the-counter use XX (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) (Division Sign-off)
Division of Anesthesiology, General Hospital
Division of Anest Partist Davises Division of Anoot, Dental Devices
K102750 510(k) Number: _
{8}------------------------------------------------
## Indications for Use Statement
Page 2 of 2
| 510(k) Number: | K102750 |
|----------------|---------------------------------------------------------------|
| Device Name: | O <sub>2</sub> -in-a-box™ portable oxygen generator -- Rx Use |
Indications for Use:
A portable oxygen generator intended to produce oxygen to provide supplemental oxygen as prescribed by a physician.
| | Flow rate | Delivery Time | Total Oxygen Capacity |
|--|-----------|---------------|-----------------------|
| | 0.5 LPM. | 60 minutes | 30 liters |
| | 3 LPM | 30 minutes | 90 liters |
Environment of use – Home, institutional, travel / mobile environments
1 100
.....
.
| Prescription Use XX_ | or | Over-the-counter use _ |
|-----------------------------|----|------------------------|
| (Part 21 CFR 801 Subpart D) | | (21 CFR 807 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L Schultheis
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
11/02750 510(k) Number: __
> . .
. . .
100 100 100
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.