ANDON HEALTH CARE MANAGEMENT SYSTEM SOFTWARE, AG-608 SINGLE & AG-608 MULTI BLOOD GLUCOSE MONITORING SYSTEMS
Device Facts
| Record ID | K102678 |
|---|---|
| Device Name | ANDON HEALTH CARE MANAGEMENT SYSTEM SOFTWARE, AG-608 SINGLE & AG-608 MULTI BLOOD GLUCOSE MONITORING SYSTEMS |
| Applicant | Andon Medical Co., Ltd. |
| Product Code | NBW · Clinical Chemistry |
| Decision Date | Aug 26, 2011 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1345 |
| Device Class | Class 2 |
| Attributes | Software as a Medical Device, Pediatric |
Indications for Use
Andon Health Care Management System Software is an optional software accessory for use with the Andon blood glucose meters with data management capacities. When used with one of these meters, Andon Health Care Management System Software transfers data from the device’s memory into a computer for enhanced data management. Andon Health Care Management System Software is intended for use in home and clinical settings via the internet to assist people with diabetes and their healthcare professionals in uploading, storing, analyzing, and communicating about historical blood glucose test results and other biological statistics to support diabetes management. Andon Health Care Management System Software is not intended to provide treatment decisions or to be used as a substitute for professional healthcare judgment. All patient medical diagnoses and treatment are to be performed under the supervision and oversight of an appropriate healthcare professional. The AG-608 Single Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The AG-608 Single Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The AG-608 Single Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The AG-608 Single Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. The AGS-1000 Single Blood Glucose Test Strips are for use with the AG-608 Single Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood sample. The AG-608 Multi Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The AG-608 MULTI Blood Glucose Monitoring System is intended to be used by healthcare professionals for multiple patients in a professional healthcare setting as an aid in monitoring the effectiveness of diabetes control. The AG-608 MULTI Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. The AGS-1000 MULTI Blood Glucose Test Strips are for use with the AG-608 MULTI Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood sample. This system should only be used with single-use, auto-disabling lancing devices.
Device Story
System consists of Andon Health Care Management System Software and AG-608 blood glucose meters (Single/Multi). Meter inputs capillary whole blood via test strips; stores glucose results. Software connects to meter via USB cable; downloads stored data to PC. Software displays results in tabular/graphical formats for trend analysis; allows time setting and memory deletion. Used in home/clinical settings by patients or healthcare professionals. Output supports diabetes management communication; does not provide treatment/dosing recommendations. Infection control validated for single-patient (AG-608 Single) and multi-patient (AG-608 Multi) use via CaviWipes disinfection.
Clinical Evidence
No clinical trials performed. Evidence includes bench testing for data transfer accuracy (full memory download/rollover), usability study (n=40: 20 lay-users, 20 professionals), readability study (SMOG score ≤ 8th grade), and infection control/robustness studies (11,000 cleaning/disinfection cycles).
Technological Characteristics
Software-based data management system; Visual Basic programming; PC-based platform; USB connectivity; compatible with AG-608 series blood glucose meters; supports English and Spanish languages; generates coordinate graphs and tables.
Indications for Use
Indicated for people with diabetes and their healthcare professionals in home and clinical settings to upload, store, analyze, and communicate historical blood glucose test results and biological statistics to support diabetes management. Not for treatment decisions or as a substitute for professional judgment.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
- Clever Chek Health Care System Software (k070941)