← Product Code [IYE](/productcode/IYE) · K102650

# CYBERKNIFE ROBOTIC RADIOSURGERY SYSTEM (K102650)

_Accuray Incorporated · IYE · Nov 17, 2010 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K102650

## Device Facts

- **Applicant:** Accuray Incorporated
- **Product Code:** [IYE](/productcode/IYE.md)
- **Decision Date:** Nov 17, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic

## Indications for Use

The CyberKnife and CyberKnife VSI Systems are intended to provide treatment planning and image-quided stereotactic radiosurgery and precision radiotherapy for tumors, lesions and conditions anywhere in the body when radiation treatment is indicated.

## Device Story

CyberKnife and CyberKnife VSI are computer-controlled robotic radiosurgery systems for stereotactic radiosurgery and precision radiotherapy. System inputs include diagnostic X-ray images of the treatment area. The system utilizes a 6 MV linear accelerator mounted on a robotic manipulator to deliver radiation. Dynamic positioning and pointing are achieved via target locating subsystems, including skull tracking, fiducial tracking, Xsight Spine Tracking, Xsight Lung Tracking, and Synchrony Respiratory Tracking. Used in clinical settings by physicians and trained staff to treat tumors and lesions throughout the body. The system provides image-guided precision, allowing for real-time tracking of target motion during treatment. Output is high-energy radiation delivered to the target, enabling minimally invasive treatment of various oncological and neurological conditions. Benefits include precise dose delivery to targets while sparing surrounding healthy tissue.

## Clinical Evidence

No clinical trials were conducted. Evidence consists of bench testing and performance verification of the tracking software. Tracking performance was validated using archived patient lung treatment images, demonstrating equivalence to the predicate system.

## Technological Characteristics

Robotic radiosurgery system featuring a 6 MV linear accelerator, robotic manipulator, and X-ray-based target locating subsystem. Supports multiple tracking modalities: skull, fiducial, Xsight Spine, Xsight Lung, and Synchrony Respiratory. Dose rate up to 1000 MU/minute. System is computer-controlled for treatment planning and delivery.

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- CyberKnife Robotic Radiosurgery System ([K102650](/device/K102650.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92.

### Name. Address, Phone and Fax number of the Applicant

Accuray Incorporated 1310 Chesapeake Terrace Sunnyvale, California 94089 Ph: (408) 716-4600 Fax: (408) 789-4260

#### Contact Person

Anne Schlagenhaft

Date Prepared September 30, 2010

#### Device Name

Trade Names: CyberKnife VSI™ Robotic Radiosurgery System CyberKnife® Robotic Radiosurgery System Common Name: Radiosurgery/radiotherapy treatment planning and delivery system Classification Name: Medical charged particle radiotherapy device

## Device Description

The CyberKnife Robotic Radiosurgery System and CyberKnife VSI Robotic Radiosurgery System are computer controlled medical systems for planning and performing minimally invasive stereotactic radiosurgery and precision radiotherapy using a treatment radiation generator, linear accelerator, manipulator (robot), and a target locating subsystem to accurately deliver radiation to the treatment target. The target locating system provides X-rays of the treatment area taken by a diagnostic X-ray system that lets the user know the position of the target. The CyberKnife and CyberKnife VSI Systems use skull tracking, fiducial tracking, Xsight® Spine Tracking, Xsight® Lung Tracking, and Synchrony® Respiratory Tracking for dynamic positioning and pointing of the linear accelerator. The system uses a 6 MV linear accelerator to provide a dose rate up to 1000 MU/minute.

#### Intended Use

The CyberKnife and CyberKnife VSI Systems are intended to provide treatment planning and image-quided stereotactic radiosurgery and precision radiotherapy for tumors, lesions and conditions anywhere in the body when radiation treatment is indicated.

NOV 1 7 2010

{1}------------------------------------------------

The CyberKnife Systems may be used to treat astrocytoma, glioma, skull base tumors, metastases (brain and bony), nasopharyngeal carcinoma, meningioma, acoustic neuroma, schwannoma, pituitary adenoma, hemangioblastoma, craniopharngioma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, and tumors of the neck, spine, pancreas, liver, lungs, ovary, prostate, and bladder. Patients should be examined by a team of physicians to determine if they are candidates for CyberKnife treatment.

## Substantial Equivalence

The CyberKnife and CyberKnife VSI Systems are substantially equivalent to the predicate CyberKnife System. The source energy, beam properties, design, materials and other physical properties are the same or equivalent to the predicate CyberKnife Testing included in the premarket notification demonstrated that the System. performance characteristics of the device are equivalent to the predicate CyberKnife System.

Testing was done to verify that the new tracking software, performs as designed, as well as regression testing to verify integrity of existing features. Archived images from CyberKnife patient lung treatments were used to test tracking performance which proved to be equivalent to the predicate CyberKnife tracking systems.

The intended use, principles of operation, technological characteristics and labeling are the same or equivalent to the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three stripes forming its wing, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" encircling the eagle. The text is arranged in a circular fashion around the eagle, with "DEPARTMENT OF HEALTH & HUMAN SERVICES" on the left side and "USA" on the right.

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Ms. Anne Schlagenhaft Senior Regulatory Affairs Associate Accuray Incorporated 1301 Chesapeake Terrace SUNNYVALE CA 94089

NOV 1 7 2010

Re: K102650

Trade/Device Name: CyberKnife® Robotic Radiosurgery System and CyberKnife VSI™ Robotic Radiosurgery System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: September 13, 2010 Received: September 23, 2010

Dear Ms. Schlagenhaft:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{3}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Signature

David G. Brown, Ph.D. Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): K102650 Device Name: CyberKnife® Robotic Radiosurgery System and CyberKnife VSI™ Robotic Radiosurgery System Indications For Use:

The CyberKnife® Robotic Radiosurgery System and CyberKnife VSI™ Robotic Radiosurgery System are indicated for treatment planning and image guided stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|

*(Division Sign-Off)* 
Division of Radiological Devices

Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety

510K 5102650

Page 1 of

---

**Source:** [https://fda-staging.innolitics.com/device/K102650](https://fda-staging.innolitics.com/device/K102650)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com