ANT INFLATION DEVICE/ANT INFLATION DEVICE COMPACT PACK
Applicant
Shenzhen Ant Hi-Tech Industrial Co., Ltd.
Product Code
MAV · Cardiovascular
Decision Date
Jul 1, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1650
Device Class
Class 2
Indications for Use
The ANT Inflation Device is intended for use during vascular procedures in conjunction with interventional device such as balloon catheters to create and monitor pressure in the balloon catheter. The ANT Inflation Device Compact Pack is a combination of the ANT Inflation Device and ANT Inflation Device Accessory Pack. -ANT Inflation Device: See description above. -ANT Inflation Device Accessory Pack: The Accessory Pack is recommended for use during vascular procedures in conjunction and/or diagnostic device (e.g., balloon dilatation catheters, artherector devices, sent delivery systems, intravascular ultrasound devices.}
Device Story
Single-use, sterile manual inflation device for cardiovascular procedures; used to inflate/deflate balloon catheters. Operator (physician/technician) manipulates large handle to drive internal piston; clockwise rotation provides controlled inflation; cam locking mechanism maintains pressure up to 30 ATM. Pressure monitored via integrated gauge. Compact Pack includes hemostasis valve (port for interventional system), torque device (guide wire manipulation), and guide wire introducer. Device functions as closed system; provides controlled pressure to balloon catheters to facilitate vascular interventions. Benefits include precise pressure control and monitoring during balloon dilatation and stent delivery.
Clinical Evidence
No clinical data. Bench testing only. Performance testing included pressure gauge accuracy (4, 16, 27 ATM), leak testing (1, 15, 30 ATM), and competence testing (inflation/deflation monitoring). Biocompatibility testing performed per ISO 10993-1:2003, including cytotoxicity, sensitization, irritation, systemic toxicity, hemolysis, platelet adhesion, prothrombin time, and complement activation.
Technological Characteristics
Manual, piston-driven inflation device. Materials meet ISO 10993-1:2003 biocompatibility standards. Features integrated pressure gauge and cam-locking mechanism. Rated for pressures up to 30 ATM. Single-use, sterile. No electronic components or software.
Indications for Use
Indicated for use in patients undergoing vascular procedures requiring balloon catheter inflation/deflation and pressure monitoring. Used in conjunction with interventional devices such as balloon dilatation catheters, atherectomy devices, stent delivery systems, and intravascular ultrasound devices.
Regulatory Classification
Special Controls
*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
INDEFLATIOR PLUS 30 Inflation Device and PLUS 30 PRIORITY PACK (K962495)
Submission Summary (Full Text)
{0}------------------------------------------------
K102698 p 1/2.
#### 510(k) summary
JUL - 1 2011
This 510(k) Summary is being submitted in accordance with the requirement of SMDA 1990 and 21 CFR 807.92.
#### Sponsor:
Name: Shenzhen ANT Hi-Tech Industrial Co., Ltd Building 11, Lishan Industrial Park, Xinghai Ave, Nanshan Address: District, Shenzhen, Guangdong, 518052, China.
Telephone number: 86-755-86060992
86-755-86060991 Fax Number:
| Date: | May 25, 2011 |
|-------|--------------|
|-------|--------------|
## Correspondent:
| Contact person: | Lynn Fu |
|-----------------|---------------------------------------------------------------------------------------------------------|
| Address: | Building 11, Lishan Industrial Park, Xinghai Ave, Nanshan District, Shenzhen, Guangdong, 518052, China. |
### Proposed Device:
| Trade Name: | ANT Inflation Device/ ANT Inflation Device Compact Pack |
|----------------------|---------------------------------------------------------|
| Common Name: | Angiographic injection/system |
| Classification Name: | Balloon Inflation Device |
| Device Class: | II |
| Regulation Number: | 870.1650 |
| Device Code: | MAV |
#### Predicate Device:
INDEFLATIOR PLUS 30 Inflation Device and PLUS 30 PRIORITY PACK as cleared in K962495.
#### Intended Use:
The ANT Inflation Device is intended for use during vascular procedures in conjunction with interventional device such as balloon catheters to create and monitor pressure in the balloon catheter.
The ANT Inflation Device Compact Pack is a combination of the ANT Inflation Device and ANT Inflation Device Accessory Pack.
- -ANT Inflation Device: See description above.
- -ANT Inflation Device Accessory Pack: The Accessory Pack is recommended for use during vascular procedures in conjunction and/or
{1}------------------------------------------------
# 1107648
diagnostic device (e.g., balloon dilatation catheters, artherector devices, sent delivery systems, intravascular ultrasound devices.}
### Device Description:
ANT Inflation Device is a single-use, sterile device used in cardiovascular procedures to inflate and deflate balloon catheters. Pressure can be monitored via a pressure gauge. The manually operation of the device is achieved by the manipulation of a large handle to drive a piston housed within the body of the device. Careful and controlled inflation is achieved by rotating the handle clockwise. During inflation a unique cam locking mechanism maintains pressure up to 30 ATM.
ANT Inflation Device Accessory Compact Pack contains a hemostasis valve, a torque device, and a guide wire introducer. The hemostasis valve is designed to provide a port for interventional system. The guide wire introducer is used to facilitate placement of a guide wire tip through the hemostasis valve. The torque device is designed to hold the guide wire and provide a handle for manipulating.
### Summary of the Studies
Performance test:
Accuracy of pressure gauge: Completely packaged devices are chose as the test samples. Filled them with water and connected to a calibrated pressure gauge, then pressurized to 3 settings, 4ATM, 16ATM, 27ATM, read the pressure and compared it to the calibrated gauge reading. This test showed the results comparable to other commercially device.
Leak test: Test samples were pressurized with water at 1ATM, 15ATM, 30ATM, and maintained 2 minutes. When immerse the samples into water, bubble occur indicate a leakage of the device. This test shows the connection of the device is comparable to other commercially device.
Competence test: attach the finished device to a legally marked catheter, and monitor the pressure during inflation and deflation. This test showed the results comparable to other commercially device.
## Biocompatibility test:
ANT Inflation Device, when connected to a competent device such as a balloon catheter, is a closed system and does not delivery contrast or fluids to the circulation system. ANT Inflation Device Accessory Pack may contact the blood path indirectly as an external communicating device. Both of them meet the
{2}------------------------------------------------
K102648 P73
# biocompatibility requirements of ISO 10993-1:2003.
| Test Item | Standard |
|---------------------------------------------|--------------|
| Vitro Cytotoxicity | ISO 10993-5 |
| Delayed-type hypersensitivity | ISO 10993-10 |
| irritation | ISO 10993-10 |
| System toxicity | ISO 10993-11 |
| Interactions with blood (Haemolysis test) | ISO 10993-4 |
| Interactions with blood (Platelet adhesion) | ISO 10993-4 |
| Interactions with blood (prothrombin time) | ISO 10993-4 |
| Complement Activation Test | ISO 10993-4 |
| In vitro Thrombogenicity | ISO 10993-4 |
Clinical Tests: Not Applicable
# Substantial Equivalence:
Based on the intended use, summary of the studies, the subject device, ANT Inflation Device/ ANT Inflation Device Compact Pack, meet the requirements that are considered adequate for its intended use and is substantially equivalent to the INDEFLATIOR PLUS 30 Inflation Device and PLUS 30 PRIORITY PACK.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines representing its wings and body. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Shenzhen ANT Hi-Tech Industrial Co., Ltd. c/o Ms. Lynn Fu Building 11, Lishan Industrial Park Xinghai Ave., Nanshan District Shenzhen, Guangdong China 518052
JUL - 1 2011
K102648 Re:
Trade/Device Name: ANT Inflation Device and ANT Inflation Device Compact Pack Regulation Number: 21 CFR 870.1650 Regulation Name: Syringe, Balloon Inflation Regulatory Class: Class II Product Code: MAV Dated: Not Dated Received: June 23, 2011
Dear Ms. Fu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{4}------------------------------------------------
Page 2 – Ms. Lynn Fu
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Bram D. Zuckerplan, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indication for Use
510(k) Number: K002VHB
Device Name: ATN Inflation Device ANT Inflation Device Compact Pack
Indications for Use:
The ANT Inflation Device is intended for use during vascular procedures in conjunction with interventional device such as balloon catheters to create and monitor pressure in the balloon catheter.
The ANT Inflation Device Compact Pack:
-ANT Inflation Device: See description above.
-ANT Inflation Device Accessory Kit: The Accessory Kit is recommended for use during vascular procedures in conjunction and/or diagnostic device (e.g., balloon dilatation catheters, artherectorny devices, sent delivery systems, intravascular ultrasound devices.)
Prescription Use _ V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
F.l. Q. hor BDZ
-
510(k) Number: K102648
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.