K102584 · International Medsurg Connection · FMI · Jan 14, 2011 · General Hospital
Device Facts
Record ID
K102584
Device Name
IMC HYPODERMIC NEEDLE
Applicant
International Medsurg Connection
Product Code
FMI · General Hospital
Decision Date
Jan 14, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This device is intended for use to inject fluids into or withdraw fluids from parts of the body below the surface of the skin.
Device Story
IMC Hypodermic Needle consists of a hub, cannula, and protector. Used to inject or withdraw fluids from body tissues below the skin surface. Device features color-coded hubs corresponding to specific gauges (16G-30G). Manufactured using polypropylene hubs/protectors, SUS 304 stainless steel cannulas, and epoxy resin adhesive. Available in sterile and non-sterile configurations. Operates via manual insertion by healthcare professionals. Performance verified through hub/needle bond strength testing per ISO 7894:1993.
Clinical Evidence
Bench testing only. Hub/needle bond strength was tested according to ISO 7894:1993. Results demonstrated that the IMC Hypodermic Needle meets standard acceptance criteria for all gauge sizes (16G-30G), consistent with the predicate device.
Technological Characteristics
Materials: Polypropylene (hub/protector), SUS 304 (cannula), Epoxy Resin (adhesive). Design: Single lumen needle with bevel tip. Sizes: 16G-30G, lengths 1/2" to 1 1/2". Sterilization: Sterile and non-sterile options. Testing: ISO 7894:1993 for bond strength.
Indications for Use
Indicated for injection of fluids into or withdrawal of fluids from body tissues below the skin surface. Available in 16G-30G sizes.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
# JAN 1 4 2011
## 8. 510(K) SUMMARY
| 8. 510(K) SUMMARY | |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary Name: | IMC Hypodermic Needle |
| Common Name: | HypodermicNeedle |
| Classification Name: | Hypodermic Single Lumen Needle (21 CFR 880.5570) |
| Device Clarification: | Class II |
| Panel Code: | 80 |
| Product Code: | FMI |
| Submitter Information: | International Medsurg Connection<br>935 N Plum Grove Rd, STE F<br>Schaumburg, Illinois 60173 |
| Summary Prepared By: | Peter Kim<br>Director of Quality Assurance<br>International Medsurg Connection<br>935 N Plum Grove Rd, STE F<br>Schaumburg, Illinois 60173<br>Telephone: 847-619-9926<br>Fax: 847-619-9927<br>e-mail: peterkim@intlmedsurg.com |
| Date Prepared: | January 4, 2010 |
| Predicate Devices: | • Nipro Medical Corporation (K052474) |
#### Device Name(s):
IMC Hypodermic Needle (non-sterile and sterile)
### Classification Panel:
General Hospital
### Legally Marketed Device Under Which Substantial Equivalence is Being Claimed:
International Medsurg Connections, Inc is claiming substantial equivalence of the IMC Hypodermic Needle with the currently marketed:
| Description | 510(k) Number |
|-------------------------|---------------|
| Nipro Hypodermic Needle | K052474 |
### Device Description
This device is intended for use to inject fluids into or withdraw fluids from parts of the body below the surface of the skin.
{1}------------------------------------------------
This device is consist of hub, cannula and protector and has difference hub color per each of gauges.
# Statement of Intended Use
Indications For Use: This device is intended for use to inject fluids into or withdraw fluids from parts of the body below the surface of the skin.
| Description | Size | Sterility |
|----------------------------|---------------------------------------|-----------------------|
| 16 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 17 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 18 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 19 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 20 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 21 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 22 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 23 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 24 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 25 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 26 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 27 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 28 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 29 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 30 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
# Name/Descrintion
# New Devices as Compared to Marketed Device(s)
The IMC Hypodermic needle and the predicate device (Nipro Hypodermic needle) are intended to be used to inject fluids into or withdraw fluids from parts of the body below the surfince of the skin.
| Feature/ Characteristic | IMC Hypodermic Needle | Nipro Hypodermic Needle: K052474<br>(Predicate) |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| Intended Use | is device is intended for use to inject<br>fluids into or withdraw fluids from<br>parts of the body below the surface<br>of the skin | ipro Hypodermic Needles are intended<br>for use to inject fluids into or withdraw<br>fluids from the parts of the body |
| Material | | |
| Hub of Needle | Polypropylene (PP) | Polypropylene (PP) |
| Protector | Polypropylene (PP) | Polypropylene (PP) |
| Feature/ Characteristic | IMC Hypodermic Needle | Nipro Hypodermic Needle: K052474<br>(Predicate) |
| Cannula | SUS 304 | SUS 304 |
| Adhesive. | Epoxy Resin | Epoxy Resin |
| colors | | |
| 16G | White color | White color |
| 17G | Red-violet color | Mauve color |
| 18G | Pink color | Pink color |
| 19G | Cream color | Cream color |
| 20G | Yellow color | Yellow color |
| 21G | Deep green color | Green color |
| 22G | Black color | Black color |
| 23G | Deep Blue color | Blue color |
| 24G | Medium purple color | Purple color |
| 25G | Orange color | Orange color |
| 26G | Brown color | Brown color |
| 27G | Medium gray color | Gray color |
| 28G | Blue-green color | Blue-green color |
| 29G | Red color | Red color |
| 30G | Yellow color | Yellow color |
| Length | ½", 5/8", ¾", 7/8", 1", 1 ¼" and<br>1 ½" | ½", 5/8", ¾", 7/8", 1", 1 ¼" and<br>1 ½" |
| Gauge | 16G, 17G, 18G, 20G, 21G, 22G, 23G,<br>24G, 25G, 26G, 27G, 28G, 29G & 30G | 16G, 17G, 18G, 20G, 21G, 22G, 23G,<br>24G, 25G, 26G, 27G, 28G, 29G & 30G |
| Cover dimension | ½" size: 40mm | ½" size: 42mm |
| | 5/8" size: 40mm | 5/8" size: 42mm |
| | ¾" size: 40mm | ¾" size: 42mm |
| | 7/8" size: 40mm | 7/8" size: 42mm |
| | 1" size: 40mm | 1" size: 42mm |
| | 1 ¼" size: 56mm | 1 ¼" size: 55mm |
| | 1 ½" size: 56mm | 1 ½" size: 55mm |
| | All gauges have the same dimension. | |
| Cover color | Clear (for all gauges) | Clear (for all gauges) |
| Tip configuration | Bevel | Bevel |
{2}------------------------------------------------
.
.
·
{3}------------------------------------------------
### Performance Data:
| Performance Characteristics | Test Method | Acceptance<br>Criteria | IMC Hypodermic<br>Needle | Nipro Hypodermic<br>Needle<br>K052474 |
|-----------------------------|-----------------|----------------------------------------------------------------------------------------|----------------------------|---------------------------------------|
| Hub/needle bond strength | ISO 7894 : 1993 | 16G-19G: >69N<br>20G: >54N<br>21G: >44N<br>22G: >40N<br>23G-24G: >34N<br>25G-30G: >22N | Meets Standard<br>Criteria | Meets Standard<br>Criteria |
### Conclusions:
The indications for use, technology, specification, safety of the IMC Hypodermic needle and the predicate devices K052474 are essentially the same. The differences between the hypodermic needle are minor and do not raise new issues of safety or effectiveness. Hence, the IMC Hypodermic needles are substantially equivalent to the predicate device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Peter Kim Director of Quality Assurance International Medsurg Connection 935 North Plum Grove Road, Suite F Schaumburg, Illinois 60173
#### JAN 1 4 2011
Re: K102584
Trade/Device Name: IMC Hypodermic Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: December 10, 2010 Received: December 20, 2010
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2- Mr. Kim
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours
Jum J. Robert
for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital,
Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{6}------------------------------------------------
# Indications for Use
Kio2584
510(k) Number :
Device Name: IMC Hypodermic needle
Indications For Use: This device is intended for use to inject fluids into or withdraw fluids from parts of the body below the surface of the skin.
| Gauge | Size | Sterility |
|----------------------------|---------------------------------------|-----------------------|
| 16 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 17 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 18 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 19 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 20 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 21 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 22 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 23 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 24 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 25 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 26 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 27 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 28 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 29 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
| 30 Gauge Hypodermic needle | ½", 5/8", ¾", 7/8", 1", 1 ¼" and 1 ½" | Sterile & Non-sterile |
Name/Description
X Prescription Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
filter for the River Richard as Appines of (Division Sign-Off) Division of Anesthesiology, General Hospital Page 13 Infection Control, Dental Devices
510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.