AREA-CT, TZ MEDICAL MCOT ECG MONITOR AND ARRHYTHMIA DETECTOR
K102507 · Tz Medical, Inc. · DSI · Jun 13, 2011 · Cardiovascular
Device Facts
Record ID
K102507
Device Name
AREA-CT, TZ MEDICAL MCOT ECG MONITOR AND ARRHYTHMIA DETECTOR
Applicant
Tz Medical, Inc.
Product Code
DSI · Cardiovascular
Decision Date
Jun 13, 2011
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
The device is intended for use by individuals who are at risk of having cardiac disease and those that have intermittent symptoms indicative of cardiac disease and demonstrated a need to be monitored on a continuing basis. The ECG recordings can be uploaded to monitoring center in a variety of ways - transmitted via Cellular RF Modem, via RF, via TTL, or flash card to an FTP storage location or software package to be read by a healthcare professional. Patients include, but at are not limited to, those requiring for a) non-life threatening arrhythmias such as supraventricular tachycardias (e.g. atrial fibrillation, atrial flutter, PACS, PSVT) and ventricular ectopy, b) evaluation of bradyarrhthmias and intermittent bundle branch block including after cardiovascular surgery and myocardial infarction, and c) arrhythmias associated with co-morbid conditions such as hyperthyroidism or chronic lung disease. Patients with symptoms that may be due to cardiac arrhythmias. These may include, but are not limited to, symptoms such as a) syncope of unknown etiology in which arrhythmias are suspected or need to be excluded, c) dyspnea (shortness of breath). Patients with palpitations with or without known arrhythmias to obtain correlation or arrhythmias with symptoms. Patients who requiring monitoring of the effects of drugs to control ventricular rate in atrial arrhythmias (e.g. atrial fibrillation). Patients recovery from cardiac surgery who are indicated for outpatient arrhythmia monitoring. Patients with diagnosed sleep disordered breathing including sleep apnea (obstructive central) to evaluate possible nocturnal arrhythmias. Patients requiring arrhythmia evaluation of etiology of stroke or transient cerebral ischemia, possibly secondary to atrial fibrillation or flutter. Data from this device may be used by another device to analyze , measure or report QT interval . The device is not intended to sound any alarms for QT interval changes.
Device Story
Aera CT is an ambulatory ECG monitor for patients with non-life-threatening cardiac conditions. System consists of patient leads, a hand-sized monitor, and charging adapter. Device captures ECG signals, stores data, and performs automated arrhythmia analysis via on-device algorithms. Detected events or patient-triggered events are transmitted via Cellular RF modem, RF, or landline (TTL) to a remote monitoring center. Healthcare professionals or technicians review transmitted data; additional data requests can be sent to the device via SMS. Device does not provide diagnostic statements or alarms; output supports clinical decision-making by physicians regarding arrhythmia management and drug efficacy. Benefits include continuous outpatient monitoring for intermittent symptoms.
Clinical Evidence
Bench testing only. Device complies with AAMI EC 38 (Ambulatory Electrocardiographs), IEC 60601-1-1, IEC 60601-1-2, and ANSI/AAMI EC 57 (testing/reporting performance of cardiac/ST-segment algorithms). No clinical trial data provided in the summary; manufacturer committed to compiling clinical test data at a trial site for validation.
Technological Characteristics
Ambulatory ECG monitor; analog front end with MCU; flash storage; RF/Cellular/TTL transmission. Complies with AAMI EC 38, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, and ANSI/AAMI EC 57. Device-side proprietary arrhythmia detection algorithm. No alarms. Powered by battery/wall adapter.
Indications for Use
Indicated for ambulatory patients at risk of cardiac disease or with intermittent symptoms requiring continuous monitoring. Includes patients with non-life-threatening arrhythmias (SVT, AFib, flutter, PACS, PSVT, ventricular ectopy), bradyarrhythmias, bundle branch block, post-cardiac surgery/MI recovery, co-morbid conditions (hyperthyroidism, lung disease), syncope, dyspnea, palpitations, drug rate control monitoring, sleep-disordered breathing, and stroke/TIA evaluation. Contraindicated for patients with life-threatening arrhythmias requiring inpatient monitoring or those requiring hospitalization.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
CardioNet ECG Monitor with Arrhythmia Detector Model CN1005 (K072558)
Biomedical Systems Inc. Monitor with Arrhythmia Detector Model TruVue (K100155)
Submission Summary (Full Text)
{0}------------------------------------------------
### Aera CT
# 10250
Abbreviated 510(k)
JUN 1 3 2011
Summary of Safety and Effectiveness
#### 1. General
- A) Submitter Name and Address:
- TZ Medical, Inc. 7272 SW Durham Rd, Suite #800 Portland, OR 97224
- B) Contact Name and Phone Number: John Lubisich 800-944-0187 or 503-639-0282
#### 2. Device Identification
| Trade name/Proprietary: | Aera CT<br>TZ Medical MCOT ECG Monitor and Arrhythmia Detector |
|-------------------------|----------------------------------------------------------------|
| Common Name: | Arrhythmia Detector and Alarm |
| Classification Name: | Arrhythmia Detector and Alarm |
| Product Code: | DSI (21 CFR Part 870.1025) |
| Class: | Class II with Special Controls |
| Panel: | 74 Circulatory System Devices |
#### 3. Predicate Devices
K072558 CardioNet ECG Monitor with Arrhythmia Detector Model CN1005
Biomedical Systems Inc. Monitor with Arrhythmia Detector Model TruVue K100155
#### 4. Indications for Use and Contraindications
#### Indications for Use:
The device is intended for use by individuals who are at risk of having cardiac disease and those that have intermittent symptoms indicative of cardiac disease and demonstrated a need to be monitored on a continuing basis. The ECG recordings can be uploaded to monitoring center in a variety of ways - transmitted via Cellular RF Modem, via RF, via TTL, or flash card to an FTP storage location or software package to be read by a healthcare professional.
Patients include, but at are not limited to, those requiring for a) non-life threatening arrhythmias such as supraventricular tachycardias (e.g. atrial fibrillation, atrial flutter, PACS, PSVT)
{1}------------------------------------------------
and ventricular ectopy, b) evaluation of bradyarrhthmias and intermittent bundle branch block including after cardiovascular surgery and myocardial infarction, and c) arrhythmias associated with co-morbid conditions such as hyperthyroidism or chronic lung disease.
Patients with symptoms that may be due to cardiac arrhythmias. These may include, but are not limited to, symptoms such as a) syncope of unknown etiology in which arrhythmias are suspected or need to be excluded, c) dyspnea (shortness of breath). Patients with palpitations with or without known arrhythmias to obtain correlation or arrhythmias with symptoms.
Patients who requiring monitoring of the effects of drugs to control ventricular rate in atrial arrhythmias (e.g. atrial fibrillation).
Patients recovery from cardiac surgery who are indicated for outpatient arrhythmia monitoring.
Patients with diagnosed sleep disordered breathing including sleep apnea (obstructive central) to evaluate possible nocturnal arrhythmias.
Patients requiring arrhythmia evaluation of etiology of stroke or transient cerebral ischemia, possibly secondary to atrial fibrillation or flutter.
Data from this device may be used by another device to analyze , measure or report QT interval . The device is not intended to sound any alarms for QT interval changes.
#### Contraindications:
Patients with potentially life-threatening arrhythmias who require inpatient monitoring.
Patients who the attending physician thinks should be hospitalized.
Note:
The Aera CT system does not provide interpretive or diagnostic statements. Interpretation and diagnosis is the responsibility of a trained healthcare professional or physician.
#### 5. Device Description
The Aera CT is an ECG Device with auto trigger capability using MCOT compliant wireless technology for monitoring ambulatory patients with non-life threatening conditions.
The device is composed of patient (ECG) leads, monitor, and charging wall board adapter.
A sensor captures the ECG signal from the patient's body and transmits the signal to the small (hand sized) monitor where data is stored and analyzed by automated arrhythmia analysis algorithms. When events are detected by the arrhythmia analysis algorithm or when the event
{2}------------------------------------------------
processing event key is depressed, the data is transmitted by RF to a data storage location monitored by receiving center. The transmission may occur via Cellular RF modem, RF, or TTL via a landline telephone. The data is received and reviewed by the appropriately trained healthcare professionals or technicians. Request for additional data can be sent to MCT device via SMS text formatted message that will originate from the receiving software.
Image /page/2/Picture/2 description: The image shows a black and white picture of a device labeled "Aera CT". The device has a rectangular shape with rounded edges. A cable is connected to the device on the left side. There are some markings and labels on the top surface of the device.
Image /page/2/Figure/3 description: The image shows a diagram of a system with several components. A stick figure labeled 'Patient' is connected via 'Patient Cable' to a rounded rectangle labeled 'Aera-Ct'. 'Aera-Ct' is connected to a cylinder labeled 'Data Storage' via a line labeled 'Digital/RF', and to a square labeled 'Service Receiving Software' via a line labeled 'Analog'. There is also a wireless signal icon connected to 'Aera-Ct' and 'Service Receiving Software' via a line labeled 'SMS'.
#### 6. Comparison to Predicate Devices
There is no primary technological difference between TZ Medical's Aera CT device and the predicate devices (K072558) CardioNet CN1005S and (K100155) Biomedical Systems Inc TruVue System.
The following chart describes further comparison.
{3}------------------------------------------------
| | Subject Device | Predicate Device 1 | Predicate Device 2 |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|-------------------------------------------------------|
| Characteristics | Aera CT<br>TZ Medical | CN1005<br>CardioNet<br>(K072558) | TruVue<br>Biomedical Systems<br>(K100155) |
| Patient population | Patients requiring cardiac<br>monitoring who are ambulatory<br>and without life-threatening<br>arrhythmia conditions | Same | Same |
| Environment for<br>Intended Use | Ambulatory, Outpatient | Same | Same |
| Basic Technology | Analog Front end, MCU running<br>detection Algorithm, Flash Storage,<br>Modem for transmission | Same | Same |
| System<br>communication /<br>Transmission | RF Modem, Cellular RF Modem,<br>Transtelephonic Transmissions | Cellular RF Modem,<br>RF Modem to Base,<br>Transtelephonic<br>Transmission via<br>landline | Cellular RF Modem,<br>Transtelephonic<br>Transmission |
| Ambulatory ECG<br>performance<br>standards | EC 38 compliant | Same | Same |
| Algorithm | Proprietary / Device side | Proprietary / Device<br>Side | Proprietary /Server<br>Side |
| Indicators / Alarms | NONE | NONE | NONE |
#### Element 807.92 (6)(b)(1)
The TZ Medical has developed tests using the same requirements that CardioNet (K072558) and (k100155) have used and stated that their devices comply with in their respective 510k submission. The TZ Medical device and the two predicated devices are tested to and comply with all applicable tests and requirements in the relevant standards as stated in this submission.
#### Element 807.92 (6)(b)(2)
TZ Medical will compile and analyze clinical test data at a trial site prior to use. Results of data will be used as part of the validation and verification data submitted with the design file. Clinical test will be conducted at facilities that are familiar with predicate devices and will submit reports verifying substantial equivalence to competitive products and on performance to specifications.
{4}------------------------------------------------
MCT
K102507 Aera CT
page 5 OF 5
#### 7. Substantial Equivalence Conclusion
TZ Medical's Aera CT, monitor with arrhythmia detection has the same intended use, similar operating principles and technological characteristics as the predicate devices. Based on the descriptive information and the performance testing and validation, the Aera CT is as safe and effective as the predicate devices. Therefore, the Aera CT is substantially equivalent to the predicate devices.
#### 8. Referenced Standards
FDA Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm
(IEC) 60601-1-1:1998 Medical Electrical Equipment-Part 1 General Requirements for Safety
- (IEC) 60601-1-2:2001,Second edition Medical Electrical Equipment Part 1: General Requirements for Safety: Electromagnetic Compatibility
(AAMI) EC 38:2007 Ambulatory Electrocardiographs
(ANSI/ AAMI) EC 57:1998 Testing & reporting performance results of cardiac and ST-segment measurement algorithms
IEC 60601-1-4 + A1:1999 : General Requirements for safety 4. Collateral Standard: Programmable Electrical Medical Systems
IEC 14971-2007: Application of risk management to medical devices
ISO 15223-1:2007 Medical devices – Symbols to be used with medical device labels, labeling, and information to be supplied.
Note: AAMI EC 11 and EC 13 Do not apply to this device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
TZ Medical Inc. c/o Mr. John Lubisich VP of Operations 7272 SW Durham Road, Suite #800 Portland, OR 97224
JUN 1 3 2011
- Re: K102507
Device Name: Aera CT Regulation Number: 21 CFR 870.1025 Regulation Name: Patient physiological monitor (with arrhythmia detection or alarms) Regulatory Class: Class II (Two) Product Codes: DSI Dated: June 7, 201 Received: June 9, 2010
Dear Mr. Lubisich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{6}------------------------------------------------
Page 2 – Mr. John Lubisich
CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Artur Wist
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: Aera CT
Indications For Use:
The device is intended for use by individuals who are at risk of having cardiac disease and those that have intermittent symptoms indicative of cardiac disease and demonstrated a need to be monitored on a continuing basis. The ECG recordings can be uploaded to monitoring center in a variety of ways - transmitted via Cellular RF Modem, via RF, via TTL, or flash card to an FTP storage location or software package to be read by a healthcare professional.
Patients include, but at are not limited to, those requiring monitoring for
a) non-life threatening arrhythmias such as supraventricular tachycardias (e.g. atrial fibrillation, atrial flutter, PACS, PSVT) and ventricular ectopy,
b) evaluation of bradyarrhthmias and intermittent bundle branch block including after cardiovascular surgery and myocardial infarction, and
c) arrhythmias associated with co-morbid conditions such as hyperthyroidism or chronic lung disease.
(Continued on Second page)
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for B Zuckerman
6/13/2011
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K102507
Page 1 of 2
{8}------------------------------------------------
Patients with symptoms that may be due to cardiac arrhythmias.
These may include, but are not limited to, symptoms such as
a) syncope of unknown etiology in which arrhythmias are suspected or need to be excluded.
b) dyspnea (shortness of breath). Patients with palpitations with or without known arrhythmias to obtain correlation or arrhythmias with symptoms.
Patients who requiring monitoring of the effects of drugs to control ventricular rate in atrial arrhythmias (e.g. atrial fibrillation).
Patients recovery from cardiac surgery who are indicated for outpatient arrhythmia monitoring.
Patients with diagnosed sleep disordered breathing including sleep apnea (obstructive central) to evaluate possible nocturnal arrhythmias.
Patients requiring arrhythmia evaluation of etiology of stroke or transient cerebral ischemia, possibly secondary to atrial fibrillation or flutter.
Data from this device may be used by another device to analyze , measure or report QT interval . The device is not intended to sound any alarms for QT interval changes.
Contraindications:
Patients with potentially life-threatening arrhythmias who require inpatient monitoring.
Patients who the attending physician thinks should be hospitalized.
#### Note:
The Aera CT system does not provide interpretive or diagnostic statements. Interpretation and diagnosis is the responsibility of a trained healthcare professional or physician.
Arthur for B. Zuckerman
10 vision Sign-Division of C 510/ki Number
Page 2 of 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.