The Powerheart® AED G3 family of devices is intended to be used by personnel who have been trained in its operation. The user should be qualified by training in basic life support or other physician-authorized emergency medical response. The device is indicated for emergency treatment of victims exhibiting symptoms of sudden cardiac arrest who are unresponsive and not breathing. Post-resuscitation, if the victim is breathing, the AED should be left attached to allow for acquisition and detection of ECG rhythm. If a shockable ventricular tachyarrhythmia recurs, the device will change automatically and advise the operator to deliver therapy (G3) or automatically deliver the charge (G3 Automatic). When the patient is a child or infant up to 8 years of age, or up to 55 lbs (25kg), the device should be used with the Model 9730 Pediatric Attenuated Defibrillation Electrodes. The therapy should not be delayed to determine the patient's exact age or weight. Additional information exclusive to the Powerheart® AED G3 Pro, Model 9300P: At the discretion of emergency care personnel, the G3Pro with ECG display enabled can also be used with Model 5111 ECG Patient Cable to display the rhythm of a responsive or breathing patient, regardless of age. The G3Pro and ECG Patient Cable system provides a non-diagnostic display for attended patient monitoring. While connected the G3Pro ECG Patient Cable, the G3Pro evaluates the patient's ECG and disables its shock capability.
Device Story
Portable, battery-operated AED (automatic/semi-automatic) for emergency treatment of sudden cardiac arrest. Electrodes applied to patient chest; device analyzes ECG rhythm; provides voice/visual prompts. Automatic models deliver shock; semi-automatic models advise operator to press button. G3 Pro model includes ECG display for monitoring responsive patients via patient cable, which disables shock capability. Periodic automatic self-tests verify readiness. Submission focuses on software modification to enhance self-test detection capability. Used by trained BLS/emergency personnel in field/clinical settings.
Clinical Evidence
Bench testing only. Software verification and validation testing completed to demonstrate functional and performance characteristics.
Technological Characteristics
Portable, battery-operated AED. Sensing: ECG via electrodes. Energy: Defibrillation shock. Connectivity: G3 Pro supports ECG patient cable. Software: Embedded firmware with modified self-test algorithm. Sterilization: N/A.
Indications for Use
Indicated for emergency treatment of unresponsive, non-breathing victims of sudden cardiac arrest. Pediatric use (up to 8 years or 55 lbs) requires Model 9730 pediatric electrodes. G3 Pro model allows non-diagnostic ECG monitoring of responsive/breathing patients via Model 5111 cable.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
{0}------------------------------------------------
JUN - 9 2011
## 510(k) Summary
- Submitter's name, address, telephone number, contact person, and data A. summary was prepared
Submitter
Contact Person
Cardiac Science Corporation 3303 Monte Villa Parkway Bothell, WA 98021-8969
Kathleen Roberts Regulatory Affairs Manager Phone: (949) 797-3844 Fax: (949) 797-3801
Date Summary Prepared
March 24, 2011
#### B. Name of device, including trade name and classification name
Trade/Proprietary Name
Powerheart® AED G3 Semi-Automatic (Model 9300E)
Powerheart® AED G3 Automatic (Model 9300A)
Powerheart® AED G3 Pro (Model 9300P)
Powerheart® AED G3 Plus (Models 9390A and 9390E)
Automated External Defibrillator
Class III, 21CFR 870.5310
Classification Name
Classification Number
Product Code
MKJ
- C. Identification of the predicate device or legally marketed devices to which substantial equivalence is being claimed.
Company Device
Cardiac Science Corporation Powerheart® AED G3
Confidential Cardiac Science Corporation March 24, 2011
{1}------------------------------------------------
| Device<br>Identification | Model<br>Number | 510(k)<br>Number | Date Cleared |
|---------------------------------------|-----------------|------------------|--------------|
| Powerheart® AED G3 Semi-<br>Automatic | 9300E | K031987 | 7/30/2003 |
| Powerheart® AED G3 Automatic | 9300A | K040438 | 7/1/2004 |
| Powerheart® AED G3 Pro | 9300P | K040637 | 8/6/2004 |
| Powerheart® AED G3 Plus | 9390A/E | K052161 | 10/21/2005 |
#### D. Description of the device
The Powerheart® AED G3 family of devices (Model numbers 9300A/E, model numbers 9390A/E, and model number 9300P) are portable, battery operated, automatic or semi-automatic, automated external defibrillators (AED). After applying the AED electrodes (pads) to the patient's bare chest, the AED automatically analyzes the patient's electrocardiogram (EGG) and guides the operator through the rescue using a combination of voice prompts, audible alerts, and visible indicators. For the Powerheart® AED G3 Automatic, the AED automatically delivers a shock if needed. For the Powerheart® AED G3 Semi-automatic, the AED uses one button and advises the operator to press the button and deliver a shock if needed.
The Powerheart® AED G3 family of devices includes a periodic automatic self-test feature. The focus of this submission is a software modification to enhance the detection capability of the device self-test feature.
#### E. Intended use of the device
The Powerheart® AED G3 family of devices is intended to be used by personnel who have been trained in its operation. The user should be qualified by training in basic life support or other physician-authorized emergency medical response.
The device is indicated for emergency treatment of victims exhibiting symptoms of sudden cardiac arrest who are unresponsive and not breathing. Post-resuscitation, if the victim is breathing, the AED should be left attached to allow for acquisition and detection of ECG rhythm. If a shockable ventricular tachyarrhythmia recurs, the device will change automatically and advise the operator to deliver therapy (G3) or automatically deliver the charge (G3 Automatic).
When the patient is a child or infant up to 8 years of age, or up to 55 lbs (25kg), the device should be used with the Model 9730 Pediatric Attenuated Defibrillation Electrodes. The therapy should not be delayed to determine the patient's exact age or weight.
{2}------------------------------------------------
## Additional information exclusive to the Powerheart® AED G3 Pro, Model 9300P :
At the discretion of emergency care personnel, the G3Pro with ECG display enabled can also be used with Model 5111 ECG Patient Cable to display the rhythm of a responsive or breathing patient, regardless of age. The G3Pro and ECG Patient Cable system provides a non-diagnostic display for attended patient monitoring. While connected the G3Pro ECG Patient Cable, the G3Pro evaluates the patient's ECG and disables its shock capability.
### F. Functional Tests
Software verification and validation testing has been successfully completed for all models of the modified Powerheart® AED G3 family of devices to demonstrate appropriate functional and performance characteristics.
#### G. Conclusion
Based on the results of the testing, it is concluded that the modifications to the Powerheart® AED do not raise any new questions regarding the safety or effectiveness as compared with the predicate device. The Cardiac Science Corporation Powerheart® AED G3 family of devices described in this submission is substantially equivalent to the previously cleared Powerheart® AED G3 products in terms of indications for use, features and functions.
There are no technological differences between the predicate devices and the modified devices.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in capital letters.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUN - 9 2011
Cardiac Science Corporation c/o Ms. Kathleen Roberts Regulatory Affairs Manager 3303 Monte Villa Parkway Bothell, WA 98021
Re: K102496
Trade/Device Name: Powerheart AED G3 Models 9300A/E, 9390 A/E, and 9300P Regulation Number: 21 CFR 870.5310 Regulation Name: Automated External Defibrillator Regulatory Class: Class III (three) Product Code: MKJ Dated: May 10, 2011 Received: May 11, 2011
Dear Ms. Roberts:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III(PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K1024960
Device Name: Powerheart® AED G3 Model 9300A/E Powerheart® AED G3 Model 9390A/E Powerheart® AED G3 Model 9300P
Indications for Use:
The Powerheart® AED G3 family of devices is intended to be used by personnel who have been trained in its operation. The user should be qualified by training in basic life support or other physician-authorized emergency medical response.
The device is indicated for emergency treatment of victims exhibiting symptoms of sudden cardiac arrest who are unresponsive and not breathing. Post-resuscitation, if the victim is breathing, the AED should be left attached to allow for acquisition and detection of ECG rhythm. If a shockable ventricular tachyarrhythmia recurs, the device will change automatically and advise the operator to deliver therapy (G3) or automatically deliver the charge (G3 Automatic).
When the patient is a child or infant up to 8 years of age, or up to 55 lbs (25kg), the device should be used with the Model 9730 Pediatric Attenuated Defibrillation Electrodes. The therapy should not be delayed to determine the patient's exact age or weight.
Additional information exclusive to the Powerheart® AED G3 Pro, Model 9300P:
At the discretion of emergency care personnel, the G3Pro with ECG display enabled can also be used with Model 5111 ECG Patient Cable to display the rhythm of a responsive or breathing patient, regardless of age. The G3Pro and ECG Patient Cable system provides a non-diagnostic display for attended patient monitoring. While connected the G3Pro ECG Patient Cable, the G3Pro evaluates the patient's ECG and disables its shock capability.
Prescription Use X _ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
| Confidential<br>Cardiac Science Corporation | <table><tr><td>510(k) Number</td><td>K102496</td></tr><tr><td></td><td>August 27, 2010</td></tr></table> | 510(k) Number | K102496 | | August 27, 2010 | Page 1 of 1<br><br>Page 11<br>Special 510(k) |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------|---------------|---------|--|-----------------|----------------------------------------------|
| 510(k) Number | K102496 | | | | | |
| | August 27, 2010 | | | | | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.