All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for ankylosing spondylitis, congenital hip dysplasia, revision of failed previous reconstructions where sufficient bone stock is present, and to restore mobility resulting from previous fusion. Cemented femoral stems and cemented acetabular cups are intended for cemented fixation only. Press-fit femoral stems and acetabular cups are intended for press-fit fixation. Press-fit components without hydroxyapatite (HA) coating may also be used with bone cement at the discretion of the surgeon. Femoral heads and endoprostheses are intended for use in cemented and press-fit applications.
Device Story
Exactech Novation Tapered 12/14 Press-Fit Femoral Stems are orthopedic implants used in hip arthroplasty. Device serves as femoral component in total or hemi-hip replacement; implanted by orthopedic surgeons in clinical/OR settings. Stems feature tapered geometry for press-fit fixation within femoral canal; designed to restore hip joint function and mobility. Modification from predicate includes expanded sizing range/increments and updated heat treatment for new sizes. Device provides structural support for prosthetic hip joint; benefits patients by alleviating pain and restoring joint function associated with degenerative hip conditions or fractures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, specifically engineering evaluations of geometric anatomical fit and fatigue strength testing.
Technological Characteristics
Materials: Titanium alloy, commercially pure titanium plasma spray. Design: Tapered 12/14 femoral stem geometry with standard/extended offsets. Fixation: Press-fit (option for cement use if non-HA coated). Sterilization: Gamma radiation (SAL 10^-6).
Indications for Use
Indicated for skeletally mature individuals undergoing primary hip replacement for osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative hip problems, or proximal femoral fractures. Also indicated for ankylosing spondylitis, congenital hip dysplasia, revision of failed reconstructions with sufficient bone stock, and restoration of mobility after previous fusion.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
p. 1/2
## Exactech® Novation® Tapered 12/14 Press-Fit Femoral Stems Special 510(k) - 510(k) Summary of Safety and Effectiveness
NOV. - 8 2010.
Sponsor:
Exactech® Inc. 2320 N.W. 66th Court Gainesville, FL 32653
Phone: (352) 377-1140 Fax: (352) 378-2617
FDA Establishment Number 1038671
(102-487
Vladislava Zaitseva Contact: Regulatory Affairs Specialist
Date: October 12, 2010
## Trade or Proprietary or Model Name(s):
Exactech® Novation® Tapered 12/14 Press-Fit Femoral Stems
#### Common Name:
Press-Fit Femoral Stem component
#### Classification Name:
- Prosthesis, hip. semi-constrained, metal/ceramic/polymer, cemented or non-● porous, uncemented. (CFR 888.3353, Hip joint metal/ceramic/polymer semiconstrained cemented or nonporous uncemented prosthesis Class II, Product Code LZO)
- Prosthesis, hip, semi-constrained, metal/polymer, uncemented. (CFR 888.3360 . Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis Class II, Product Code LWJ)
- Prosthesis, hip, semi-constrained, metal/polymer, cemented. (CFR 888.3350. Hip joint metal/polymer semi-constrained cemented prosthesis Class II, Product Code JDI)
#### Information on devices to which substantial equivalence is claimed: 510(k) Number Trade or Proprietary or Model Name · Manufacturer
K042842 Novation 12/14 Press-Fit Femoral Stems Exactech, Inc
#### Indications for Use:
All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also
{1}------------------------------------------------
102487
p.2/2
# Exactech® Novation® Tapered 12/14 Press-Fit Femoral Stems Special 510(k) - 510(k) Summary of Safety and Effectiveness
potentially indicated for ankylosing spondylitis, congenital hip dysplasia, revision of failed previous reconstructions where sufficient bone stock is present, and to restore mobility resulting from previous fusion.
- Cemented femoral stems and cemented acetabular cups are intended for cemented . fixation only.
- Press-fit femoral stems and acetabular cups are intended for press-fit fixation. . Press-fit components without hydroxyapatite (HA) coating may also be used with bone cement at the discretion of the surgeon
- Femoral heads and endoprostheses are intended for use in cemented and press-fit � applications.
### Device Description:
The proposed Exactech Novation Tapered 12/14 Press-Fit Femoral Stems are a modification of the Exactech Novation 12/14 Press-Fit Femoral Stems cleared through premarket notification #K042842.
The predicate and proposed devices have the same intended use and the same basic fundamental scientific technology.
The modified devices share the following similarities with predicate devices:
- The same indications for use
- The same design features (e.g. neck and stem geometry, femoral head taper . design, stem insertion feature, offsets - standard and extended)
- The same materials (titanium alloy, commercially pure titanium plasma spray) .
- The same shelf life (5 years), and .
- Are packaged and sterilized using the same materials and processes (gamma ● radiation sterilization to a sterility assurance level (SAL) of 100).
This submission proposed the following design changes:
- A change to the sizing range and increments between additional femoral stem . sizes
- A heat treatment change for the additional femoral stem sizes only .
### Substantial Equivalence Conclusion:
The following engineering analyses were conducted to demonstrate substantial equivalence of the proposed device to the predicate Novation Tapered 12/14 Press-Fit Femoral Stems:
- An engineering evaluation to determine that the geometric features of the ♥ proposed device correspond to the anatomical features of the femur.
- An engineering evaluation to determine that the proposed device has fatigue . strength equivalent to other comparable legally marketed femoral stems.
The results demonstrate that the proposed device is a substantially equivalent to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA", which is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Exactech, Inc. % Ms. Vladislava Zaitseva Regulatory Affairs Specialist 2320 Northwest 66th Court Gainesville, Florida 32653
NOV - 8 2010
Re: K102487
Trade/Device Name: Exactech® Novation® Tapered 12/14 Press-Fit Femoral Stems Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: LZO, JDI, LWJ Dated: October 12, 2010 Received: October 15, 2010
Dear Ms. Zaitseva:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Or R and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act . The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contractly and warranties. We remind you, however, that device labeling must be truthful and not maching.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your davice (1 https found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Сваване вмет
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# K102487
# Exactech® Novation® Tapered 12/14 Press-Fit Femoral Stems Special 510(k) - Indications for Use
510(k) Number:
NOV: -- 8} 2010 :
ﻴﺔ
# Device Name: Exactech® Novation® Tapered 12/14 Press-Fit Femoral Stems
## INDICATIONS
All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for ankylosing spondylitis, congenital hip dysplasia, revision of failed previous reconstructions where sufficient bone stock is present, and to restore mobility resulting from previous fusion.
- Cemented femoral stems and cemented acetabular cups are intended for cemented . fixation only.
- Press-fit femoral stems and acetabular cups are intended for press-fit fixation. . Press-fit components without hydroxyapatite (HA) coating may also be used with bone cement at the discretion of the surgeon
- Femoral heads and endoprostheses are intended for use in cemented and press-fit . applications.
Prescription Use X and/or (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
Please do not write below this line - use another page if needed.
Concurrence of CDRH, Office of Device Evaluation (ODE)
K102487 Barbara Buend
(Division Sign-Off) Division of Surgical, Orthopedic. and Restorative Devices
510(k) Number K102487
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.