SUNSET HEALTHCARE SOLUTIONS BACTERIAL/VIRAL HME FILTER, MODEL BHF 104
Applicant
Sunset Healthcare Solutions, LLC
Product Code
CAH · Anesthesiology
Decision Date
Sep 13, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5260
Device Class
Class 2
Indications for Use
The BHF 104 is a disposable single-use filter for patients who require humidification during the delivery of ventilator gases and provide filtration for reducing possible cross-contamination between patient and equipment. The BHF 104 is for use in hospital, ICU, anesthesia, and respiratory therapy.
Device Story
BHF 104 is a disposable, single-use, barrier-type heat and moisture exchanger (HME) and bacterial/viral filter. Device consists of a translucent plastic housing (K-Resin) with 15mm/22mm connector ports and a gas sampling luer port. It is placed in-line within a breathing circuit to provide humidification of inspired gases and filtration of pathogens. Operated by clinicians in hospital, ICU, anesthesia, and respiratory therapy environments. Device captures moisture from exhaled breath to humidify subsequent inhaled gases; filter media traps bacteria and viruses to prevent cross-contamination between patient and ventilator equipment. Output is passive; clinical decision-making relies on standard ventilator management protocols. Benefits include reduced risk of cross-contamination and maintenance of airway humidity for ventilated patients.
Clinical Evidence
Bench testing only. Performance data includes Bacterial Filtration Efficiency (>99.9999%), Viral Filtration Efficiency (99.999%), flow resistance at 30/60/90 L/min, HME moisture output (37.5 mg H2O/L at 500ml VT), dead space (53ml), and mechanical integrity of conical connectors.
Technological Characteristics
Housing: K-Resin KR03 (Styrene Butadiene Block Copolymer). Filter: 3M Filtrete Air Media Filter Type G. Passive HME/filtration principle. Dimensions: 90mm length, 55mm diameter. Connectors: 15mm ID x 22mm OD and 22mm ID x 15mm OD. Gas sampling luer port. Non-sterile delivery.
Indications for Use
Indicated for patients requiring humidification during ventilator gas delivery and filtration to reduce cross-contamination between patient and equipment. For use in hospital, ICU, anesthesia, and respiratory therapy settings.
Regulatory Classification
Identification
A breathing circuit bacterial filter is a device that is intended to remove microbiological and particulate matter from the gases in the breathing circuit.
{0}------------------------------------------------
K102483
ال
# 510(k) Summary for the Bacterial/Viral HME Filter (BHF 104)
#### Name and Address of Sponsor
Sunset Healthcare Solutions, Inc 2201 S. Halsted St. Suite 1344 Chicago, IL 60608
#### Name and Address of Manufacturer
Altera Tibbi Malzeme San Ve Tic As. Turan Mah Tire Organize Sanayi Bolgesi. Tire/Izmir Aegean Region, 35900 Turkey Phone: 90-232-2375949
#### Establishment Registration Number
3006534321
### Name and Address of Official Correspondent
Regulatory Insight, Inc. 5401 S. Cottonwood Ct. . Greenwood Village, Colorado 80121 Contact: Mr. Kevin Walls, RAC Tel: 720-962-5412 Fax: 720-962-5413 Email: kevin@reginsight.com
#### Date Summary was Prepared
December 17, 2010
#### Device Name
Trade Name: Bacterial/Viral HME Filter (BHF 104) Common Name: Disposable Heat and Moisture Exchanger and Bacteria/Viral Filter Classification Name: Breathing circuit bacterial filter
#### Classification, Panel and Product Code
Class 2, Anesthesiology, Regulation Number: 868.5260 Product Code: CAH
{1}------------------------------------------------
### Indications for Use
The BHF 104 is a disposable single-use filter for patients who require humidification during the delivery of ventilator gases and provide filtration for reducing possible crosscontamination between patient and equipment. The BHF 104 is for use in hospital, ICU, anesthesia, and respiratory therapy.
## Legally Marketed Device to Which Substantial Equivalence is Claimed
DATEX-OHMEDA HMEF 500 (K021265)
## Device Description
The BHF 104 filter is a single-use, barrier type filter, fabricated from a plastic resin, and delivered to the customer in a sealed, non-sterile package. The translucent housing has aligned connector ports (15mm ID x 22mm OD, 22mm ID x 15mm OD), a perpendicular filter media, and a gas sampler luer port, protruding at 45 degrees from the base of the filter housing. The entire filter is 90mm (3.54 inches) in length, with a maximum cross section diameter of 55mm (2.16 inches), and a mass of 24g (.85 ounces). The surface area is 26.4cm2.
The BHF 104 has 15mm ID x 22mm OD and 22mm ID x 15mm OD male and female connectors, which allows compatibility with various compressors, ventilators, masks, and aerosol tubes.
### Specifications
| Tidal Volume: | 150-1500ml |
|--------------------------|------------------------------------------------------------|
| Flow rate resistance: | 30lt/min 5mm/H20<br>60lt/min 13mm/H20<br>90lt/min 28mm/H20 |
| Bacterial filtration: | >99.9999% |
| Viral filtration: | 99.999% |
| Dead Space: | 53ml |
| HME Output per ISO 9360: | 37.5mg/L H2O at 500ml VT |
### Performance Testing
Performance was based on the following tests, which are contained in the body of the 510(k) application:
- . Bacterial Filtration Efficiency (BFE)
- Viral Filtration Efficiency (VFE)
- Dead Space 트
- I Resistance to Flow
- Heat and Moisture Exchange (HME) l
- . Conical Connectors
{2}------------------------------------------------
## Substantial Equivalence
| | DATEX-OHMEDA HMEF 500 | Bacterial/Viral HME Filter (BHF 104) |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) # | K021265 | N/A |
| Indications for use | The HMEF500 is a disposable single-use device indicated for patients who require humidification during the delivery of ventilator gases and provide filtration for reducing possible cross contamination between patient and equipment. The HMEF 500 is for use in hospital, ICU, anesthesia, respiratory therapy, during transport and with resuscitators. The device can be used on adult and pediatric patients. The device is indicated for use by qualified medical personnel only. | The BHF 104 is a disposable single-use filter for patients who require humidification during the delivery of ventilator gases and provide filtration for reducing possible cross-contamination between patient and equipment. The BHF 104 is for use in hospital, ICU, anesthesia, and respiratory therapy. |
| Diameter | 45mm | 55mm |
| Length | 74mm | 90mm |
| BFE | 99.999% | 99.9999% |
| VFE | 99.98% | 99.999% |
| Housing Material | PP Polypropylene | K-Resin KR03 - Styrene Butadiene Block Copolymer |
| Filter Material | PP and Acrylic Fibers | 3M Filtrete Air Media Filter - Type G |
| Fittings | 15mm ID x 22mm OD<br>15mm OD x 22mm ID | 15mm ID x 22mm OD<br>15mm OD x 22mm ID |
| Moisture Output | 30mg H2O/L @ 500ml VT | 37.5 mg H2O/L @ 500ml VT |
| Tidal Volume Range | 120-500 | 150-1500 |
| Dead Space | 30ml | 53ml |
| Weight | 15g | 24g |
| Sampling Port | Luer lock | Luer lock |
None of the differences listed in the table above have any affect on safety or effectiveness.
. .
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake entwined around it, and three horizontal bars above it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Sunset Healthcare Solutions. LLC C/O Mr. Kevin Walls, RAC Principal Consultant Regulatory Insight, Incorporated 5401 S. Cottonwood Court Greenwood Village, Colorado 80121
SEP 13 2011
Re: K102483
Trade/Device Name: Sunset Healthcare Solutions Bacterial/Viral HME Filter (BHF 104) Regulation Number: 21 CFR 868.5260 Regulation Name: Breathing Circuit Bacterial Filter Regulatory Class: II Product Code: CAH Dated: August 19, 2011 Received: August 22, 2011
Dear Mr. Walls:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Walls
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem /default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm.
Sincerely yours.
hr for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
Indications for Use
510(k) Number (if known): K102483
Device Name: Sunset Healthcare Solutions Bacterial/Viral HME Filter (BHF 104)
Indications for Use: The BHF 104 is a disposable single-use filter for patients who require humidification during the delivery of ventilator gases and provide filtration for reducing possible cross-contamination between patient and equipment. The BHF 104 is for use in hospital, ICU, anesthesia, and respiratory therapy.
Prescription Use × AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L Schutter
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of 1
K 102483 510(k) Number: _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.