K102460 · Summit Medical Products, Inc. · BSO · Jan 13, 2011 · Anesthesiology
Device Facts
Record ID
K102460
Device Name
AMBIT INTRODUCER, AMBIT NEEDLE, AMBIT SHEATH
Applicant
Summit Medical Products, Inc.
Product Code
BSO · Anesthesiology
Decision Date
Jan 13, 2011
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 868.5120
Device Class
Class 2
Indications for Use
The amblT Introducer product line is intended to allow the percutancous placement of catheters in close proximity to nerves and around or into surgical wound or non-surgical wound sites. It may be used to inject or aspirate the introduction area via the luer hub of the needle. The amblT Introducer product line is intended for use by a physician or by a trained individual under direct supervision of a physician.
Device Story
The ambIT Introducer is a sterile, single-use percutaneous catheter introducer system. It consists of a stainless steel shaft (11-17 GA, 3.25-12 inches) with a sharp or blunt tip, often paired with a plastic peelable sheath or internal plastic/metal insert. The device is used by physicians or trained personnel to create an access point for catheter placement near nerves or wound sites. The luer hub allows for fluid injection or aspiration at the site. After insertion, the stainless steel shaft is withdrawn, leaving the sheath in place for catheter introduction, or the catheter is inserted through the shaft after the insert is removed. The device facilitates regional anesthesia or wound site pain management by enabling precise catheter positioning.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through adherence to ISO 594-2 (liquid/air leakage, separation force, unscrewing torque, ease of assembly, resistance to overriding, stress cracking) and ISO 10555-1 (force to break, strength of union) standards, alongside ISO 10993 biocompatibility testing.
Technological Characteristics
Materials: 304 stainless steel needle, polystyrene needle hub, HDPE sheath shaft with 10% BaSO4, HDPE tear-away hub. Principle: Mechanical percutaneous access. Dimensions: 11-17 GA, 3.25-12 inches. Connectivity: None. Sterilization: Sterile, single-use. Standards: ISO 594-2, ISO 10555-1, ISO 10993.
Indications for Use
Indicated for percutaneous placement of catheters near nerves or into surgical/non-surgical wound sites. For use by physicians or trained individuals under physician supervision.
Regulatory Classification
Identification
An anesthesia conduction catheter is a flexible tubular device used to inject local anesthetics into a patient and to provide continuous regional anesthesia.
{0}------------------------------------------------
# 102460
## FINDING BETTER WAYS TO CARE FOR PEOPLE "
Image /page/0/Picture/2 description: The image shows three triangles of different shades. The triangle on the left is a dark shade, the triangle in the middle is a lighter shade, and the triangle on the right is a dark shade. The triangles are arranged in a row, with the middle triangle slightly taller than the other two.
SUMMIT MEDICAL PRODUCTS
JAN 1 3 2011
January 11, 2011 510(k) Summary
| Device<br>Sponsor: | Summit Medical Products, Inc.<br>2480 South Main Street Suite 212<br>Salt Lake City, UT 84115 |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact<br>Information: | Marko Van Amen<br>Summit Medical Products, Inc<br>VP of Regulatory Affairs/Quality Assurance<br>2480 South Main Street Suite 212<br>Salt Lake City, UT 84115<br>Phone: 801-352-1888<br>FAX: 801-352-1818 |
| Date<br>Submitted | January 11, 2011 |
| Trade Names: | ambIT Introducer, ambIT Sheath, ambIT Needle |
| Common<br>Name: | Catheter Introducer |
| Classification<br>Name: | Anesthesia, Conduction Catheter |
| Product Code: | BSO |
| Regulation<br>Number: | 868.5120 |
| Equivalent<br>To: | K063234 On-Q Introducers |
| Device<br>Description: | The ambIT™ Introducer is intended for the percutaneous<br>introduction of a catheter<br>1. A stainless steel shaft with a sharp tip.<br>a. Stainless steel shafts with sharp tips (various configurations such as<br>beveled or touhy) may or may not be solid (may be hollow).. The<br>diameter of the stainless steel shaft ranges from 11 to 17 GA and the<br>length is from 3.25 to 12 inches.<br>2. A plastic peelable sheath (split T-handle) on the outside of the stainless steel<br>shaft and/or a plastic/metal inserts inside hollow stainless steel shafts.<br>a. Introducers that have a plastic sheath will allow the catheter to be<br>inserted after the stainless steel shaft has been withdrawn. The tips of<br>the stainless steel shaft will be blunt or sharp.<br>b. Introducers that do not have a plastic sheath will allow the catheter to<br>be inserted through the stainless steel shaft, after the plastic/metal<br>insert has been withdrawn.<br>The ambiT Introducers may have a handle or luer hub connected to<br>the stainless steel shaft. |
Summit Medical Products, Inc. | 2480 South Main, Suite 212 | Salt Lake City, Utah 84115 Summit Medical Produces, Mor - | - 2158 Bear Han |Phone 801-352-1888 | | Fax 801-352-1818 | | Toll-Free 877-352-1888 | | www.summitmpi.com
{1}------------------------------------------------
#### FINDING BETTER WAYS TO CARE FOR PEOPLE™
ﻻﻳﺔ ﺍﻟﻤﺘ
je
Image /page/1/Picture/1 description: The image shows a logo for Summit Medical Products. The logo features three mountain peaks, with the center peak being slightly taller and textured. Below the mountain graphic, the words "SUMMIT MEDICAL PRODUCTS" are printed in a simple, sans-serif font, with "SUMMIT MEDICAL" on the first line and "PRODUCTS" on the second line.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
:
| | | The following are the configurations and names for the amblT Introducers: | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| | | 1. amblT Introducer Sheath | |
| | | a. Sterile single use product | |
| | | b. Made for use with ambIT Tunneler (without sheath) reusable product. | |
| | | 2. amblT Introducer Needle | |
| | | a. Sterile single use product | |
| | | b. Stainless steel shaft with sharp tip and insert inside stainless steel shaft. | |
| Indications<br>for Use: | The ambIT™ Introducer products are intended allow for the<br>percutaneous placement of catheters in close proximity to<br>nerves and around or into surgical wound or non-surgical<br>wound sites. It may be used to inject or aspirate the<br>introduction area via the luer hub of the needle. | | |
| Substantial<br>Equivalence<br>(SE)<br>Rational: | The ambIT™ Introducer product is considered substantially<br>equivalent based on the materials that are used, the intended<br>use, and both devices were tested to the same applicable ISO<br>standards. The product passed all applicable testing<br>requirements. The ambIT™ Introducer is intended for the<br>percutaneous introduction of a catheter. | | |
| In Section 2 of the 510(k) application for the ambIT Introducer<br>device (K102460) a comparison of the proposed and predicate<br>device is provided. A comparison of the proposed medical<br>device, ambIT™ Introducer to the predicate medical device,<br>the On-Q Introducer products (510K# K063234) is summarized<br>below in Table 1. Both products have common features, such<br>as they both use the similar materials of construction, similar<br>intended use, and similar consensus standards for the design<br>and performance for both devices | | | |
| Table 1: Substantially Equivalence Comparison Table of<br>proposed device -ambIT™ Introducer to the predicate<br>device – ON-Q Introducer™. | | | |
| | Device | Proposed Device<br>ambIT™ Introducer<br>(510(k) #K102460) | Predicate Device<br>On-Q Introducer<br>(510(k) #K063234) |
| | Intended Use | The ambIT™<br>Introducer products are<br>intended allow for the<br>percutaneous placement<br>of catheters in close<br>proximity to nerves and | The On-Q<br>Introducers™ are<br>intended for the<br>percutaneous<br>introduction and<br>placement of |
| | | | |
| | around or into surgical<br>wound or non-surgical<br>wound sites. It may be<br>used to inject or<br>aspirate the<br>introduction area via<br>the luer hub of the<br>needle. | catheters into or<br>around surgical<br>wound sites and/or<br>close proximity to<br>nerves. Introducers<br>with a luer hub may<br>be used to aspirate<br>or inject a bolus of<br>fluid or medication<br>prior to placing the<br>catheters. | |
| Trade Names | ambIT Introducer,<br>ambIT Needle, ambIT<br>Sheath | ON-Q Needle,<br>ON-Q Tunneler,<br>ON-Q Sheath | |
| Materials of<br>Construction | Needle | Needle | |
| | • Stainless steel (304)<br>needle 18 GA<br>• Needle Hub-<br>Polystyrene (clear) | • Stainless steel<br>(304) needle 18<br>GA<br>• Needle Hub-<br>Polystyrene<br>(clear) | |
| | Sheath | Sheath | |
| | • Tear away hub- High<br>density Polyethylene<br>(HDPE)<br>• Sheath Shaft (tube)-<br>High density<br>Polyethylene (HDPE)<br>with 10%BaSO4 | • Tear away hub-<br>High density<br>Polyethylene<br>(HDPE)<br>• Sheath Shaft<br>(tube)- High<br>density<br>Polyethylene<br>(HDPE) with<br>10% BaSO4 | |
| | Tunneler Sheath | Tunneler Sheath | |
| | • Sterile, single use product<br>• Plastic peelable sheath | • Sterile, single use<br>product<br>• Plastic peelable<br>sheath | |
| | | | |
| | | | |
| | Introducers Sterile, single use product Stainless steel shaft with a sharp end Plastic peelable sheath | Introducers Sterile, single use product Stainless steel shaft with a sharp end Plastic peelable sheath | |
| Standards: | The ambiT Introducers meet the requirements as set forth in the following FDA recognized consensus standards for device design and performance requirements: ISO 594-2 Liquid Leakage test ISO 594-2 Air Leakage test ISO 594-2 Separation Force test ISO 594-2 Unscrewing Torque ISO 594-2 Ease of Assembly ISO 594-2 Resistance to Overriding ISO 594-2 Stress Cracking ISO 10555-1 Force to Break ISO 10555-1 Strength of Union ISO 10993 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing | | |
| Safety and Effectiveness: | Based on the comparison to the predicate device and the conformance to the recognized standards, the ambiT Introducers are safe and effective and substantially equivalent to legally marketed devices. | | |
Summit Medical Products, Inc. | 2480 South Main, Suite 212 | Salt Lake City, Utah 84115 Summit Medical Products, Inc.
{2}------------------------------------------------
### FINDING BETTER WAYS TO CARE FOR PEOPLE "
้า และ และ ﯩﺘﺴ
్టి
Image /page/2/Picture/1 description: The image shows a logo for Summit Medical Products. The logo features three mountain peaks, with the center peak being a lighter shade than the other two. Below the mountain peaks, the text "SUMMIT MEDICAL PRODUCTS" is written in a simple, sans-serif font.
Summit Medical Products. Inc. | 2480 South Main, Suite 212 | | Salt Lake City, Utah 84115 Summit Medical Products: nel:
{3}------------------------------------------------
# FINDING BETTER WAYS TO CARE FOR PEOPLE"
Image /page/3/Picture/1 description: The image shows the logo for Summit Medical Products. The logo consists of three triangles of varying shades, arranged to resemble mountain peaks. Below the triangles, the text "SUMMIT MEDICAL PRODUCTS" is printed in a bold, sans-serif font, with "SUMMIT MEDICAL" on the first line and "PRODUCTS" on the second line.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle or bird symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Marko Van Amen Vice President of Regulatory Affairs and Quality Assurance Summit Medical Products, Incorporated 2480 South Main Street Salt Lake City, Utah 84115
JAN 1 3 201
Re: K102460
Trade/Device Name: amblT Introducers Regulation Number: 21 CFR 868.5120 Regulation Name: Anesthesia Conduction Catheter Regulatory Class: II Product Code: BSO Dated: August 28, 2010 Received: January 7, 2011
Dear Mr. Amen:
. .
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
. . . . . . . . . . .
{5}------------------------------------------------
#### Page 2- Mr. Amen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default .htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Summit Medical Products. The logo consists of three triangles of varying shades, arranged to look like mountain peaks. Below the triangles, the words "SUMMIT MEDICAL" are stacked on top of "PRODUCTS" in a sans-serif font.
Abbreviated 510(k) - ambIT Introducer
510(k) Number K 102460:
JAN 1 3 2011
法、行行中心理。以及时时并并并并建建建设的时代中心中心人民共同意见。 中国人民共和
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Partic Research - Child Child Childer of the Children
Device Name: ambIT Introducers
Indications for Use:
The amblT Introducer product line is intended to allow the percutancous placement of catheters in close proximity to nerves and around or into surgical wound or non-surgical wound sites. It may be used to inject or aspirate the introduction area via the luer hub of the needle
The amblT Introducer product line is intended for use by a physician or by a trained individual under direct supervision of a physician.
Prescription Use l (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Richard C. Chapman 1/13/11
(Division Sign-Off)
02460
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Page 16 of 27
Summit Medical Products, Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.