K102435 · Neusoft Medical Systems Co., Ltd. · LLZ · Oct 15, 2010 · Radiology
Device Facts
Record ID
K102435
Device Name
BEYONDIMAGE WORKSTATION
Applicant
Neusoft Medical Systems Co., Ltd.
Product Code
LLZ · Radiology
Decision Date
Oct 15, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
BeyondImage Workstation is a software application that is used for viewing medical images. Beyond mage Workstation receives digital images and data from various sources (including but not limited to CT, MR, US, RF units, computed and digital radiographic devices). Images are stored, communicated, processed and displayed on the local disc of a workstation and/or across computer networks at distributed locations. Taking that users may perform when viewing images include, but are not limited to adjustment of window width and center; image stacking; annotation and measurement of regions of interest; and inversion, rotation, and flips of images, it also provides standard Multi-Planar Reconstruction (MPR) views, Curved-Planar Reconstruction (CPR) views and 3D views of Volume Rendering for digital images from CT, MR, and PET unit. In addition, using Beyondlmage Workstation, users can edit and print report. Beyondlmage Workstation cannot display and process mammograms.
Device Story
BeyondImage Workstation is a software-only PACS application for diagnostic imaging environments. It receives DICOM-formatted images from various modalities (CT, MR, US, RF, CR, DR, PET) via network or local storage. Trained medical professionals (radiologists, clinicians, technologists) operate the workstation on standard Windows hardware. The software provides tools for image manipulation, including window/level adjustment, stacking, annotation, measurement, rotation, and flipping. It performs advanced visualization, including Multi-Planar Reconstruction (MPR), Curved-Planar Reconstruction (CPR), and 3D Volume Rendering. Users can also edit and print clinical reports. The device facilitates diagnostic decision-making by enabling efficient review and analysis of medical imaging studies across distributed clinical networks.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and comparison of technological characteristics and functional specifications to predicate devices.
Technological Characteristics
Software-only application; runs on standard Windows-based computer hardware. Supports DICOM 3.0 protocol for image communication, storage, and Query/Retrieve. Features include window/level adjustment, image stacking, annotation, measurement, rotation, flipping, MPR, CPR, and 3D Volume Rendering. Not for mammography.
Indications for Use
Indicated for trained medical professionals (radiologists, clinicians, technologists) to view, process, and manipulate digital medical images (CT, MR, US, RF, CR, DR, PET) for diagnostic purposes. Contraindicated for mammography imaging.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "Neusoft" in a bold, sans-serif font. The text is black against a white background. The word is positioned on the left side of the image.
510(k)
#### Summary of Safety and Effectiveness K102438
Page 1 of 3
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h)
## General Information:
| Trade Name: | BeyondImage Workstation |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Version Number: | 2.0 |
| Common Name: | Picture Archiving and Communication System (PACS) |
| CFR Section: | 21 CFR Part 892.2050 |
| Classification Name: | Picture Archiving and Communication System (PACS). |
| Product Code: | LLZ |
| Device Class: | Class II |
| Manufacturer and<br>Distributor: | Neusoft Medical Systems Co., Ltd.<br>No.16, Shiji Road, Hunnan Industrial Area,<br>Shenyang, Liaoning, China<br>Post Code : 110179 |
| Submitter: | Contact : Tian Yuehui<br>Title : Manager of Quality Management Department<br>Tel : 86-24-83660646<br>Fax : 86-24-83660563<br>E-Mail : Tianyh@neusoft.com |
...
page 1 of 3
{1}------------------------------------------------
### Safety and Effectiveness Information
Intended Uses:
BeyondImage Workstation is a software application that is used for viewing medical images. Bevondlmage Workstation receives digital images and data from various sources (including but not limited to CT, MR, US, RF units, computed and digital radiographic devices). Images are stored, communicated, processed and displayed on the local disc of a workstation and/or across computer networks at distributed locations. Taking that users may perform when viewing images include, but are not limited to adjustment of window width and center; image stacking; annotation and measurement of regions of interest; and inversion, rotation, and flips of images, it also provides standard Multi-Planar Reconstruction (MPR) views Curved-Planar Reconstruction (CPR) views and 3D views of Volume Rendering for digital images from CT, MR, and PET unit. In addition, using BevondImage Workstation, users can edit and print report.
Beyondlmage Workstation cannot display and process mammograms.
Typical users of BeyondImage Workstation are trained medical professionals, including but not limited to radiologists, clinicians, technologists, and others.
Device Description:
BeyondImage Workstation is a software application that provides image viewing and manipulation in a diagnostic imaging setting. The functions of this application are applied to medical images that are acquired and stored on an image server in DICOM format. Beyondlmage Workstation can also transfer images in DICOM 3.0 format over a medical imaging network, as well as exporting images to applications in other proprietary formats.
Predicate Device:
K092235 : BeyondImage Workstation 1. 0 K060505 : Barco Voxar 3D 6.1
> . ::: .
Statement of Substantial Equivalence:
The BeyondImage Workstation 2.0is comparable and substantially equivalent to the Beyondlmage Workstation 1.0(K092235) and the Barco Voxar 3D 6.1 (K060505).
BeyondImage Workstation 2.0, BeyondImage Workstation 1.0and Barco Voxar 3D 6.1 are all available as "software only" application that run under Microsoft Windows operating systems on readily available computer hardware. All applications include image and report viewing and a set of imaging measurements and manipulation tools. All three are DICOM compliant systems capable of receiving and storing images and moving imaging studies using DICOM Query/Retrieve.
page 2 of 3
{2}------------------------------------------------
#### Page 3 of 3
The Beyondlmage Workstation 2.0 and the Beyondlmage Workstation 1.0 share similar technological specifications. Both of them support DICOM protocol for communication of images with other medical imaging devices, and they both provide the functions such as adjustment of window width and center; image stacking; annotation and measurement of regions of interest; and inversion, rotation, and flips of images etc. Furthermore, they both provide standard Multi-Planar Reformation (MPR) views and 3D views of Volume Rendering.
The Beyondlmage Workstation 2.0has the similar technological characteristics with the Barco Voxar 3D 6.1. Both of them support DICOM protocol for communication of images with other medical imaging devices, and they both provide the functions such as adjustment of window width and center, scout line, annotation and measurement of regions of interest; and inversion, rotation, and flips of images etc. Furthermore, they both provide function ofCurved-Planar Recontruction(CPR) ..
According to the comparison based on the requirements of 21.CFR 807.87, we state that these devices are substantially equivalent.
> : 11. 1 1 .
Page 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird in flight, with three curved lines representing its wings and tail feathers. The bird is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Tian Yuehui Manager of Q&R Department Neusoft Medical Systems, Co., Ltd. No 16, Shiji Road, Hunnan Industrial Area Shenyang, Liaoning, 110179 CHINA
OCT 1 5 2010
Re: K102435
Trade/Device Name: BeyondImage Workstation Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: August 16, 2010 Received: August 26, 2010
Dear Mr. Yuehui:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your dovice complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
signature
David G. Brown, Ph.D. Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k)
# Indications for Use Statement
OCT 1 5 2010
510(k) Number: K102435
Device Name: BeyondImage Workstation
Beyondlmage Workstation is a software application that is used for viewing medical images. Beyond mage Workstation receives digital images and data from various sources (including but not limited to CT, MR, US, RF units, computed and digital radiographic devices). Images are stored, communicated, processed and displayed on the local disc of a workstation and/or across computer networks at distributed locations. Taking that users may perform when viewing images include, but are not limited to adjustment of window width and center; image stacking; annotation and measurement of regions of interest; and inversion, rotation, and flips of images, it also provides standard Multi-Planar Reconstruction (MPR) views, Curved-Planar Reconstruction (CPR) views and 3D views of Volume Rendering for digital images from CT, MR, and PET unit. In addition, using Beyondlmage Workstation, users can edit and print report.
Beyondlmage Workstation cannot display and process mammograms.
Typical users of Beyondlmage Workstation are trained medical professionals, including but not limited to radiologists, clinicians, technologists, and others.
Prescription Use YES (Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
Over-The-Counter Use NO
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF DEEDED)
Concurrence of CDRH, @ffice of Bevice Evan
(Division Sign Off)
510K K102435
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.