The Prevision Hip System is intended for cementless revision hip arthroplasty in patients whose bone stock is of poor quality or inadequate for other reconstruction techniques as indicated by deficiencies of the femoral head, neck, or portions of the proximal femur. The Prevision Hip System is intended for revision only. The Recon Ring is intended to bridge the areas of acetabular bone loss in patients with acetabular bone deficiency. In addition, the Recon Ring is intended to provide support for an all polyethylene acetabular implant in a cemented application. Diagnostic indications for use of this device include acetabular dysplasia, osteoporosis, protrusion acetabuli, cystic acetabular roof, reconstruction in cases of defects after fracture, acetabular loosening, tumors or revision surgery. The general indications associated with the use of the Recon Ring in total hip arthroplasty include: Advanced joint destruction resulting from degenerative, posttraumatic or rheumatoid arthritis. Fracture or avascular necrosis of the femoral head, Failed previous surgery, e.g. osteosynthesis, joint reconstruction, arthrodesis, hemi-arthroplasty and total hip replacement.
Device Story
Modular hip prosthesis system consisting of proximal body, distal stem, and tension nut; components assembled by surgeon in OR or in situ to allow independent sizing. Designed for cementless revision hip arthroplasty; Recon Ring component bridges acetabular bone loss and supports cemented all-polyethylene acetabular implants. Manufactured from titanium alloy; proximal body features plasma-sprayed (Plasmapore) and calcium phosphate (u-CaP) coatings. Used by orthopedic surgeons in clinical settings to restore hip function in patients with significant bone deficiency or failed prior joint reconstructions.
Clinical Evidence
Bench testing only. Performance evaluated in accordance with ISO 7206 and ASTM F 2068 standards. Results demonstrated similarity to legally marketed predicate devices.
Technological Characteristics
Modular hip system; titanium alloy construction; proximal body plasma-sprayed (Plasmapore) with secondary calcium phosphate (u-CaP) coating. Recon Ring manufactured from titanium. Designed for cementless use. Testing performed per ISO 7206 and ASTM F 2068.
Indications for Use
Indicated for patients requiring cementless revision hip arthroplasty due to poor bone stock or deficiencies of the femoral head, neck, or proximal femur; also indicated for patients with acetabular bone deficiency requiring bridge support or cemented all-polyethylene acetabular implant support. Includes patients with acetabular dysplasia, osteoporosis, protrusion acetabuli, cystic acetabular roof, post-fracture defects, loosening, tumors, or failed prior surgeries (e.g., arthrodesis, hemi-arthroplasty).
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
K102424 (pg 1/2)
Page 1 of 2
#### B. 510(k) SUMMARY (as required by 21 CFR 807.92)
DEC - 9 2010
Prevision Hip System with Recon Ring December 3, 2010
| COMPANY: | Aesculap Implant Systems, LLC<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 3005673311 |
|----------|-------------------------------------------------------------------------------------------------------------------------------------|
|----------|-------------------------------------------------------------------------------------------------------------------------------------|
CONTACT: Kathy A. Racosky 610-984-9291 (phone) 610-791-6882 (fax) Kathy.racosky@aesculap.com
TRADE NAME: Prevision Hip System
COMMON NAME: Femoral Hip Prosthesis
CLASSIFICATION NAME: Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calcium-Phosphate Prosthesis, Hip, Semi-constrained, Metal/Polymer, Cemented Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non-Porous, Uncemented Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented or Uncemented
888.3353, 888.3350, 888.3360, 888.3353, 888.3390, REGULATION NUMBER:
PRODUCT CODE: MEH, JDI, LWJ, LZO, KWY
#### SUBSTANTIAL EQUIVALENCE
Aesculap Implant Systems, LLC believes that the Prevision Hip System with Recon Ring is substantially equivalent to:
- Zimmer, Inc. ZMR Hip System (K992667 & K994286) ●
- Intermedics Orthopedics Inc., Burch/Schneider Reinforcement Cage (K960678) .
- Aesculap Implant Systems Metha Hip System (K071916) .
- Aesculap Implant Systems Bicontact (K040191) .
# DEVICE DESCRIPTION
The Prevision Hip System is a modular system which consists of a proximal body, distal stem and a tension nut. The three individual components utilize a modular junction. The individual components may be assembled by the surgeon in the operating room or in situ to allow independent sizing of the proximal body and distal stem. All proximal components are able to mate with all distal components providing a comprehensive range of combinations offering optimal flexibility. It is designed for use with currently available Aesculap Implant Systems femoral heads, acetabular components and Bipolar cups.
{1}------------------------------------------------
### Page 2 of 2
The Prevision Hip System is manufactured from Titanium Alloy and intended for cementless use. The proximal body is plasma sprayed (Plasmapore®) with a secondary coating of Calcium Phosphate (u-CaP®).
The Recon Ring will be offered in three sizes with left and right configurations. The Recon Ring is manufactured from Titanium. It is designed to be used with the PE acetabular cup cleared in BiContact (K040191).
# PURPOSE FOR PREMARKET NOTIFICATION
The purpose for this submission is to gain marketing clearance for the Aesculao Implant Systems Prevision Hip System with Recon Ring.
### INDICATIONS FOR USE
The Prevision Hip System is intended for cementless revision hip arthroplasty in patients whose bone stock is of poor quality or inadequate for other reconstruction techniques as indicated by deficiencies of the femoral head, neck, or portions of the proximal femur. The Prevision Hip System is intended for revision only.
The Recon Ring is intended to bridge the areas of acetabular bone loss in patients with acetabular bone deficiency. In addition, the Recon Ring is intended to provide support for an all polyethylene acetabular implant in a cemented application.
Diagnostic indications for use of this device include acetabular dysplasia, osteoporosis, protrusion acetabuli, cvstic acetabular roof, reconstruction in cases of defects after fracture, acetabular loosening, tumors or revision surgery.
The general indications associated with the use of the Recon Ring in total hip arthroplasty include:
- Advanced joint destruction resulting from degenerative, posttraumatic or rheumatoid . arthritis.
- . Fracture or avascular necrosis of the femoral head,
- Failed previous surgery, e.g. osteosynthesis, joint reconstruction, arthrodesis, hemi-● arthroplasty and total hip replacement.
# TECHNOLIGICAL CHARACTERISTICS(compared to Predicate(s)}
The components of the Aesculap Implant Systems Prevision Hip System with Recon Ring are offered in a similar range of shapes and sizes as the predicate devices. The material used for the Aesculap Implant Systems device is the same as that used to manufacture the predicate devices.
# PERFORMANCE DATA
Testing of the Aesculap Implant Systems Prevision Hip System was performed in accordance with ISO 7206 and ASTM F 2068 and the results were found to be similar to other legally marketed predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Aesculap Implant Systems LLC % Ms. Kathy A. Racosky 3773 Corporate Parkway Center Valley, PA 18034
DEC - 9 2010
Re: K102424
:
Trade/Device Name: Prevision Hip System with Recon Ring Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: II Product Code: MEH, JDI, LWJ, LZO, KWY Dated: December 3, 2010 Received: December 6, 2010
Dear Ms. Racosky:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your Section 910(x) premained is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosule) to regarly manatinent date of the Medical Device Amendments, or to commerce province to May 26, 1976, the enasting with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a premarket approval app and Cosment Act (Act) that to not require approvinces of the general controls provisions of the Act. The You may, merelore, market the device, subject to the genirements for annual registration, listing of general controls provisions of the rict liveling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability adulteration. Prease note: CDICT does not over be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it
es an be If your device is classified (Scc above) fine viar regulations affecting your device can be
may be subject to additional controls. Existing major regulations addition, EDA m may be subject to additions: Existing mays. Existing mays found in the Code of I cucral Rogarations) This Levice in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s Issualitat equr device of a success at
that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
Page 2 - Ms. Kathy A. Racosky
I uge 2 - Ms. Ratily in Not regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm1.
Sincerely yours,
Sincerely yours,
{}
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT A.
DEC - 9 2010
**510(k) Number:** K102424(pg 1/1)
### Device Name: Prevision Hip System with Recon Ring
#### Indications for Use:
The Prevision Hip System is intended for cementless revision hip arthroplasty in patients whose bone stock is of poor quality or inadequate for other reconstruction techniques as indicated by deficiencies of the femoral head, neck, or portions of the proximal femur. The Prevision Hip System is intended for revision only.
The Recon Ring is intended to bridge the areas of acetabular bone loss in patients with acetabular bone deficiency. In addition, the Recon Ring is intended to provide support for an all polyethylene acetabular implant in a cemented application.
Diagnostic indications for use of this device include acetabular dysplasia, osteoporosis, protrusion acetabuli, cystic acetabular roof, reconstruction in cases of defects after fracture, acetabular loosening, tumors or revision surgery.
The general indications associated with the use of the Recon Ring in total hip arthroplasty include:
- Advanced joint destruction resulting from degenerative, posttraumatic or rheumatoid ● arthritis.
- Fracture or avascular necrosis of the femoral head, �
- Failed previous surgery, e.g. osteosynthesis, joint reconstruction, arthrodesis, hemi-. arthroplasty and total hip replacement.
| Prescription Use | X |
|-----------------------------|---|
| and/or Over-the-Counter Use | |
(per 21 CFR 801.109)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Apr M. Wilkerson
(Division Sign-Off (Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K102424
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.