The DIAMOND-5A,6A,8A Digital Diagnostic X-ray System is indicated for use in generating radiographic images of human anatomy. This device is not intended for mammographic applications.
Device Story
Digital diagnostic X-ray system; consists of high voltage generator, tube support, beam limiting device, wall stand, bucky table, detector, and workstation. Operates via high-frequency inverter method; X-ray tube generates radiation passing through patient anatomy. Detector collects charges via TFT array; image processing unit performs amplification and digital conversion. Data transmitted via cable to workstation; operating software stores images in DICOM format. Used in hospital settings by clinicians for skeletal, respiratory, and urinary system diagnosis. Images transmitted to PACS for analysis. Benefits include efficient radiographic image acquisition for clinical decision-making.
Clinical Evidence
Bench testing only. Performance and safety verified through testing according to EN/IEC 60601-1, 60601-1-1, 60601-1-3, 60601-2-7, 60601-2-28, 60601-2-32, and EMC testing per EN/IEC 60601-1-2 (2001). All results were satisfactory.
Indicated for generating radiographic images of human anatomy for diagnostic purposes in skeletal, respiratory, and urinary systems. Not intended for mammographic applications.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K102408
# 510(k) Summary
DEC 2 7 2010
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: _Mar. 18, 2010
1. Company and Correspondent making the submission:
Name - DRGEM Corporation
Address - 7th FL, Dongil Techno-town 1st, 826, Guro-dong, Guro-gu,
Seoul, Republic of Korea, 152-050
Telephone - +82-2-869-8566
Fax -- +82-2-869-8567
Contact - Mr. Ki-Nam Yang
- Internet http://www.drgem.co.kr/
- 2. Device :
| Trade/proprietary name | : DIAMOND-5A,6A,8A |
|------------------------|-----------------------------------|
| Common Name | : Digital Diagnostic X-ray System |
| Classification Name | : System, x-ray, stationary |
- 3. Predicate Device :
| Manufacturer | : Sedecal, Inc. |
|---------------|-------------------------------------------|
| Device | : Sedecal X-Plus LP Plus Digital |
| 510(k) Number | : K090238 (Decision Date - Feb. 27. 2009) |
- 4. Classifications Names & Citations :
21CFR 872.1680, KPR, System, x-ray, stationary, Class2
5. Description :
- 5.1 General
The digital X-ray imaging system consists of a high voltage (HV) generator, a tube
DRGEM Corporation
{1}------------------------------------------------
support unit, an X-ray beam limiting device, a wall stand unit, a bucky table unit, a detector, operating software, and a tube, operates on a high-frequency inverter method, and is primarily used in a hospital for diagnosis of diseases in skeletal. respiratory and urinary systems.
### 5.2 Product features
Rated voltage from external power supply is introduced into the X-ray control device, and the tube voltage, tube current and time are set up. When an X-ray radiation command is given, the preset voltage is applied to the primary side of the high voltage generator and a high voltage to produce X-ray is generated from the secondary side.
When this high voltage is applied to the X-ray tube, it strikes substance named target and X-ray is generated. This X-ray passes part of a human body to be diagnosed.
Electric charges are collected by electrodes located at each pixel, and outputted to the image processing unit through the thin film transistor (TFT) array. Amplification and digital data conversion take place in the image processing unit, and the data that has been transmitted to the workstation (image processing computer) through a cable is stored in medical standard DICOM files by operating software. Stored images are used for image analysis after transmitted to the picture archiving and communication system (PACS) by operating software.
### 5 Indication for use :
The DIAMOND-5A,6A,8A Digital X-ray Imaging System is indicated for use in generating radiographic images of human anatomy. This device is not intended for mammographic applications.
#### 6 Comparison with predicate device :
DRGEM Corporation, believes that the DIAMOND-5A,6A,8A is substantially equivalent to the Sedecal X-Plus LP Plus Digital of Sedecal, Inc..
#### 7 Safety, EMC and Performance Data :
Electrical, mechanical, environmental safety and performance testing according to standard EN/IEC 60601-1, EN/IEC 60601-1-1, EN/IEC 60601-1-3, EN/IEC 60601-
DRGEM Corporation
{2}------------------------------------------------
2-7, EN/IEC 60601-2-28 and EN/IEC 60601-2-32 was performed, and EMC testing was conducted in accordance with standard EN/IEC 60601-1-2(2001). All test results were satisfactory.
- 9. Conclusions :
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification DRGEM Corporation concludes that The DIAMOND-5A,6A,8A is safe and effective and substantially equivalent to predicate devices as described herein.
- 10. DRGEM Corporation will update and include in this summary any other information deemed seasonably necessary by the FDA.
END
DRGEM Corporation
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
DRGEM Corporation % Mr. Charlie Mack International Regulatory Consultants 77325 Jovce Way ECHO OR 97826
DEC 2 7 2010
Re: K102408
R 102406
Trade/Device Name: Digital Diagnostic X-ray System / DIAMOND-5A, 6A, 8A Regulation Number: 21 CFR 892.1630 Regulation Name: Electrostatic x-ray imaging system Regulatory Class: II Product Code: KPR Dated: November 13, 2010 Received: November 22, 2010
Dear Mr. Rodenberger:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your boother be device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to conninered phay 20, 1978, in occordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmelle Act (Act) that do not requence to the general controls provisions of the Act. The 1 ou may, dictore, market the act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such If your device is classified (see acover) interesting your device can be found in Title 21, additional controls. Existing major regards on 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Prease oe advised had I DA 3 issuated or a on ot one one other requirements of the Act that I DA has made a decermination an administered by other Federal agencies. You must of any I ederal statutes and regulations and instituting, but not limited to: registration and listing (21 Comply with an the Act 31equirements, Moraning, Solanding, Solor Starting (reporting of
{4}------------------------------------------------
## Page 2
medical device-related adverse events) (21 CFR 803); and good manufacturing practice modious device reated as not the quality systems (QS) regulation (21 CFR Part 820). This letter requirences as set form in the your device as described in your Section 510(k) premarket while would to begin mading of substantial equivalence of your device to a legally marketed nontication. The I Drice results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 11 you done optice of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket 5 1501 11100, p. 1000 100 Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the I va may overni oner gearers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
signature
David G. Brown, Ph.D. Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
DEC 2 7 2010
510(k) Number(if known): という408
Device Name: Digital Diagnostic X-ray System / DIAMOND-5A,6A,8A
Indications for Use:
The DIAMOND-5A,6A,8A Digital Diagnostic X-ray System is indicated for use in generating radiographic images of human anatomy. This device is not intended for mammographic applications.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use_ (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
Page 1 of _ 1
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnosuc Device Evaluation and Safety
510K K162408
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.