K102405 · C.R. Bard, Inc. · FPA · Jul 21, 2011 · General Hospital
Device Facts
Record ID
K102405
Device Name
INFUSION SET FOR POWERPICC SOLO 2 CATHETER
Applicant
C.R. Bard, Inc.
Product Code
FPA · General Hospital
Decision Date
Jul 21, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
An extension set to be used as an extension to the intravascular administration set used for fluid delivery through Bard® proximally valved catheters. The extension set will connect the IV administration set to the catheter.
Device Story
Extension set designed for Bard proximally valved catheters; attaches to catheter hub to facilitate fluid delivery, central venous pressure monitoring, and blood sampling. Opens catheter valve upon connection. Single-use device; maximum 96-hour duration. Used in clinical settings by healthcare providers. Provides physical conduit between IV administration set and indwelling catheter; ensures secure connection via ISO-compatible luer fittings. Benefits patient by enabling reliable access for fluid therapy and diagnostic monitoring through existing valved catheter systems.
Clinical Evidence
Bench testing only. Verification activities included assembly leak testing (min 43.5 psi), gravity flow (min 750 ml/hr), assembly tensile strength (15 N for 15 seconds), and ISO 594-1/2 luer compatibility testing. Biocompatibility evaluated per ISO 10993-1. Sterilization validated per ISO 11135-1 to SAL 10^-6.
Technological Characteristics
Infusion tubing with ISO-compatible luer connections; distal cannula for valve actuation. Materials evaluated for biocompatibility per ISO 10993. Sterilization via Ethylene Oxide (EtO) per ISO 11135-1. Complies with ISO 8536-4 (infusion sets) and ISO 594-1/2 (luer fittings).
Indications for Use
Indicated for use as an extension to intravascular administration sets for fluid delivery through Bard proximally valved catheters. Connects IV administration set to catheter.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
IV Extension Set with Needleless Access Device (K980992)
Submission Summary (Full Text)
{0}------------------------------------------------
K 102405
Image /page/0/Picture/1 description: The image shows the logo for BARD ACCESS SYSTEMS. The word "BARD" is in large, bold, sans-serif font, with a unique design where the letters are connected and have a slightly angled appearance. Below the word "BARD", the words "ACCESS SYSTEMS" are printed in a smaller, sans-serif font.
:
## 510(k) Summary
21 CFR 807.92(a)
JUL 2 1 2011
## Extension Set for Bard® Proximally Valved Catheters
| General<br>Provisions | Submitter Name:<br>Address: | Bard Access Systems, Inc.<br>605 North 5600 West<br>Salt Lake City, UT 84116 |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact Person:<br>Telephone Number:<br>Fax Number:<br>Date of Preparation: | Lynn Kirchoff<br>Regulatory Affairs Manager<br>(801) 522-5636<br>(801) 522-5425<br>March 9, 2011 |
| Subject<br>Device | Trade Name:<br>Classification Name: | Extension Set for Bard Proximally Valved Catheters<br>Set, Administration, Intravascular<br>FPA, 21 CFR § 880.5440 Class II |
| Predicate<br>Device | Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer: | IV Extension Set with Needleless Access Device<br>Set, Administration, Intravascular<br>FPA, 21 CFR § 880.5440 Class II<br>K980992<br>Venetec Intl., Inc. now owned by Bard Medical<br>Division |
| Device<br>Description | The Extension Set is designed for Bard proximally valved catheters and<br>attaches to the end of the catheter hub for use in the administration of fluids.<br>The Extension Set opens the valve to facilitate central venous pressure<br>monitoring and/or blood sampling. Extension Sets are packaged individually<br>for single use for a period no longer than 96 hours. | |
| Intended Use | An extension set to be used as an extension to the intravascular<br>administration set used for fluid delivery through Bard proximally valved<br>catheters. The extension set will connect the IV administration set to the<br>catheter. | |
| Indications For<br>Use | An extension set to be used as an extension to the intravascular<br>administration set used for fluid delivery through Bard proximally valved<br>catheters. The extension set will connect the IV administration set to the<br>catheter. | |
| Technological<br>Characteristics | Technological characteristics of the subject Bard Extension Set are equivalent<br>with respect to the basic design and function to those of the predicate IV<br>Extension Set with Needleless Access Device. Both the subject and predicate<br>device have infusion tubing with ISO compatible luers connections with the<br>exception that the subject device has a cannula on the distal end of the<br>device. Differences do not raise any new questions regarding safety and<br>effectiveness. Accepted scientific methods such as, performance (bench)<br>testing do exist for assessing the effect of the differences in characteristics. | |
| | The Bard Extension Set met the bench performance testing requirements of<br>the following standards where applicable in conjunction with in-house<br>protocols which were used to determine appropriate methods for evaluating<br>the performance of the device:<br>ISO 8536-4:2007 :- Infusion equipment for medical use - Part 4:<br>Infusion sets for single use, gravity feed ISO 594-1:1986 :- Conical Fittings with a 6 % (Luer) Taper for<br>Syringes, Needles and Certain Other Medical Equipment - Part 1:<br>General Requirements ISO 594-2:1998 :- Conical Fittings with a 6 % (Luer) Taper for<br>Syringes, Needles and Certain Other Medical Equipment - Part 2:<br>Lock Fittings Guidance for Industry and FDA Review Staff - Intravascular<br>Administration Sets Premarket Notification Submissions [510(k)],<br>dated April 15, 2005 | |
| Summary of<br>Nonclincal<br>Testing | Risk management, including a failure modes and effects analysis (FMEA), of<br>the subject device was conducted in accordance with ISO 14971:2009,<br>Medical Devices - Risk Management for Medical Devices. No new types of<br>safety or efficacy questions were identified for the subject Bard Extension Set.<br><br>Biocompatibility assessment was conducted on the Bard Extension Set based<br>on the guidelines presented in blue book memorandum #G95-1 entitled "Use<br>of International Standard ISO-10993, "Biological Evaluation of Medical<br>Devices Part-1: Evaluation and Testing and ISO 10993-1:2009 - Biological<br>Evaluation of Medical Devices Part-1: Evaluation and Testing.<br><br>The sterility of the Bard Extension Set is assured by using a validated<br>sterilization method qualified in accordance with ISO 11135-1:2007 :-<br>Sterilization of Healthcare products - Ethylene Oxide- Part 1: Requirements<br>for Development, Validation and Routine Control of a Sterilization Process for<br>Medical Devices to a sterility assurance level (SAL) of 10-6. Ethylene Oxide<br>residual levels resulting from EtO Sterilization are in compliance with ISO<br>10993-7:2008 - Biological Evaluation of Medical Devices- Part 7: Ethylene<br>Oxide Sterilization Residuals. The Bard Extension is also tested for<br>pyrogenicity. | |
{1}------------------------------------------------
:
·
. : -
: .. ..
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
------------... ..
י
: :
・: : .
... . . . . . . . . .
: · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
-------- -------------------------------------------------------------------------------------------------------------------------------------------------------------------------
.. . . ...
:
・:
.
.
.
·
{2}------------------------------------------------
## Summary of Verification Activities
| Summary Results of Verification Testing | |
|-----------------------------------------|------------------------------------------------------------------------------------|
| Evaluation | Predetermined Acceptance Criteria |
| Assembly Leak | No leaks at 43.5 psi minimum (air method<br>used) |
| Gravity Flow | 750 ml/hr minimum |
| Priming Volume | NA, test and report results |
| Assembly Tensile | Must withstand an axial, static tensile force<br>of 15 N (3.37 lbs) for 15 seconds |
| ISO 594 - 1 & 2 Luer<br>Testing | Must pass acceptance criteria outlined in<br>ISO 594-1 and -2. |
| Aspiration Flow | NA, test and report results |
Summary of Substantial Equivalence
. :
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.. . .
Based on the indications for use, technological characteristics, and safety and performance testing, the subject Bard Extension Set met the requirements for its intended use/indications for use and is substantially equivalent in design, materials, sterilization, principles of operation and indications for use to current commercially available predicate infusion set.
{3}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Lynn M. Kirchoff Regulatory Affairs Manager C.R. Bard, Incorporated 605 North 5600 West Salt Lake City, Utah 84116
JUL 21 2011
Re: K102405
Trade/Device Name: Extension Set for Bard Proximally Valved Catheters Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: June 27, 2011 Received: June 29, 2011
Dear Ms. Kirchoff:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Kirchoff
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDR H/CDRHOffices/ucm1 1.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Antmony D. Orona
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number (if known):
K102405
Device Name:
Extension Set for Bard® Proximally Valved Catheters
Indications for Use:
An extension set to be used as an extension to the intravascular administration set used for fluid delivery through Bard® proximally valved catheters. The extension set will connect the IV administration set to the catheter.
> Prescription Use (Part 21 CFR §801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR §801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Richard C. Chayen 7/21/11
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K102405
32
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.